K253874 · Shenzhen Hanhua Opto Co., Ltd. · OHS · Mar 3, 2026 · General, Plastic Surgery
Device Facts
Record ID
K253874
Device Name
LED Therapy Light (COB1400, HP1800PRO)
Applicant
Shenzhen Hanhua Opto Co., Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Mar 3, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LED Therapy Light is an over-the-counter (OTC) device intended for the following uses: For Red Mode: The device emits energy in the red spectrum and is intended for the treatment of full-face wrinkles. For NIR Mode: The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. For Red + NIR Mode: The device emits energy in the red and infrared spectrum and is intended for the treatment of full-face wrinkles.
Device Story
LED Therapy Light (models COB1400, HP1800PRO) uses dual-spectrum LEDs to emit red (660 nm) and near-infrared (850 nm) light; device operates in Red, NIR, or combined modes; intended for home use by consumers; provides topical heating and wrinkle reduction; output intensity ranges from 1.425 to 7.687 mW/cm² depending on mode and model; user-operated via simple controls; benefits include temporary pain relief and aesthetic improvement of facial wrinkles; device functions as a non-laser light source for therapeutic and cosmetic applications.
Clinical Evidence
No clinical or animal testing performed. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), photobiological safety (IEC 62471), and performance requirements for non-laser light sources (IEC 60601-2-57, IEC 60601-2-83).
Technological Characteristics
Dual-spectrum LED light source; Red (660 nm) and NIR (850 nm) wavelengths; irradiance 1.425–7.687 mW/cm²; home-use form factor; electrical power source; software-controlled; complies with IEC 60601-1, 60601-1-2, 60601-1-11, 60601-1-6, 62471, 60601-2-57, 60601-2-83, and IEC 62304.
Indications for Use
Indicated for OTC use by adults for treatment of full-face wrinkles (Red/Red+NIR modes) and for temporary relief of minor muscle/joint pain, arthritis, muscle spasm, stiffness, and promotion of local blood circulation/muscle relaxation (NIR mode).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
LED Light Therapy Mask (FM-01, FM-02, FM-03) (K223544)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 3, 2026
Shenzhen Hanhua Opto Co., Ltd.
% Jun Da Huang
Project Manager
Registrar Corp.
Rm. 502, Bldg. A, Phase I, Qianhai Economic And Trade Center
# 151 Free Trade W. St., Nanshan Sub-District, Qianhai
Shenzhen, Guangdong Province 518066
China
Re: K253874
Trade/Device Name: LED Therapy Light (COB1400, HP1800PRO)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: Class II
Product Code: OHS, ILY
Dated: December 1, 2025
Received: December 4, 2025
Dear Jun Da Huang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253874 - Jun Da Huang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253874 - Jun Da Huang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L.
HITHE -S
Digtally signed by TANISHA L. HITHE -S
Date: 2020.03.03 15:45:56 -03'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253874 | |
| Device Name LED Therapy Light (COB1400, HP1800PRO) | |
| Indications for Use (Describe) The LED Therapy Light is an over-the-counter (OTC) device intended for the following uses: For Red Mode: The device emits energy in the red spectrum and is intended for the treatment of full-face wrinkles. For NIR Mode: The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. For Red + NIR Mode: The device emits energy in the red and infrared spectrum and is intended for the treatment of full- face wrinkles. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
Shenzhen Hanhua Opto Co., Ltd.
# 510(k) Summary - K253874
LED Therapy Light
Date of 510(k) Summary Preparation: 2025-12-01
## 1. Applicant Information
Shenzhen Hanhua Opto Co., Ltd.
Building A, Yishida Baoyicheng Science and Technology Park, Dahe Community,
Guanlan Street, Longhua District, Shenzhen, Guangdong, 518000, China
Contact Person: Mr. Wu Taishan
Telephone Number: +86 135 1056 0868
Email: wutaishan@hanhualed.com
## 2. Submission Correspondent
Registrar Corp
Room 502, Bldg A, Phase I, Qianhai Economic and Trade Center, No. 151 Free Trade
West Street, Nanshan Sub-district, Qianhai Shenzhen-Hong Kong Cooperation Zone,
Shenzhen, Guangdong Province, 518066, People's Republic of China
Contact Person: HUANG Jun Da Ken
Telephone Number: +86 137 6074 9097
Email: adnil1330@vip.qq.com
## 3. Proposed Device
Trade Name: LED Therapy Light
Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology.
Infrared lamp.
Model(s): COB1400, HP1800PRO
Classification: 21 CFR 878.4810
Regulation Number: 21 CFR 890.5500
Product Code: OHS (Light Based Over The Counter Wrinkle Reduction)
ILY (Lamp, Infrared, Therapeutic Heating)
Device Class: II
FDA CDRH Review: General & Plastic Surgery
Panel: Physical Medicine
## 4. Predicate Devices
{5}
510(k) Summary
Shenzhen Hanhua Opto Co., Ltd.
## Predicate Device #1:
510(k) Number: K223544
Applicant: Guangdong Newdermo Biotech Co.,Ltd
Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology. Infrared lamp.
Trade Name: LED Light Therapy Mask (FM-01, FM-02, FM-03)
This predicate device has not been subject to a design-related recall.
## Predicate Device #2:
510(k) Number: K212275
Applicant: Aesthetic Technology, Ltd.
Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology. Infrared lamp.
Trade Name: Dermalux® Flex MD
This predicate device has not been subject to a design-related recall.
## Predicate Device #3:
510(k) Number: K171323
Applicant: Biophotas, Inc.
Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology.
Trade Name: Biophotas Celluma³
This predicate device has not been subject to a design-related recall.
## 5. Device Description
The LED Therapy Light employs dual-spectrum LEDs that emit red and near-infrared light, allowing simultaneous delivery of both wavelengths for synergistic treatment effects.
## 6. Indications for Use
The LED Therapy Light is an over-the-counter (OTC) device intended for the following uses:
For Red Mode: The device emits energy in the red spectrum and is intended for the treatment of full-face wrinkles.
For NIR Mode: The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
{6}
510(k) Summary
Shenzhen Hanhua Opto Co., Ltd.
For Red + NIR Mode: The device emits energy in the red and infrared spectrum and is intended for the treatment of full-face wrinkles.
7. Comparison to the Predicate/Reference Device(s)
Indications for Use/Intended Use:
The subject LED Therapy Light is an Over-the-Counter (OTC) device mainly intended for treatment of wrinkles, and providing topical heating for multiple purposes. All devices share the similar anatomical location and intended use for treatment. The proposed device can be used at home and healthcare settings.
Technological Characteristics:
The subject LED Therapy Light has the similar principle of operation and characteristics as the predicate devices. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is as safe and effective as the predicate devices for its intended use.
The subject LED Therapy Light is substantially equivalent to the predicate devices in that it has the same/similar intended use and technological characteristics, as demonstrated in the following Table 1 Substantial Equivalence Comparison.
Page 3 of 6
{7}
510(k) Summary
Shenzhen Hanhua Opto Co., Ltd.
Table 1 Substantial Equivalence Comparison
| Feature | Proposed Device | Predicate Device #1 (K223544) | Predicate Device #2 (K212275) | Predicate Device #3 (K171323) | Remarks |
| --- | --- | --- | --- | --- | --- |
| Manufacturer | Shenzhen Hanhua Opto Co., Ltd. | Guangdong Newdermo Biotech Co.,Ltd | Aesthetic Technology, Ltd. | Biophotas, Inc. | / |
| Device Name | LED Therapy Light | LED Light Therapy Mask | Dermalux® Flex MD | Biophotas Celluma³ | / |
| Model(s) | COB1400, HP1800PRO | FM-01, FM-02, FM-03 | Not Specified | Not Specified | / |
| Intended Use | The LED Therapy Light is an over-the-counter (OTC) device intended for the following uses:For Red Mode: The device emits energy in the red spectrum and is intended for the treatment of full-face wrinkles.For NIR Mode: The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.For Red + NIR Mode: The device emits energy in the red and infrared spectrum and is intended for the treatment of full-face wrinkles. | Red Light: Treatment of full-face wrinkles.Blue Light: Treatment of mild to moderate inflammatory acne.Infrared Light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.Mixed Light: Treatment of mild to moderate inflammatory acne. | Intended to deliver heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.The blue spectrum light is intended to reduce mild to moderate inflammatory acne vulgaris.Intended to emit energy in the red and infrared spectrum for use in dermatology for the treatment of periorbital and full face wrinkles. | The BioPhotas Celluma³ is intended to emit energy in the visible and infrared region of the spectrum for use in the treatment of full face wrinkles. | Same |
| Rx or OTC | OTC | OTC | OTC | OTC | Same |
| Product Code | OHS, ILY | OHS, OLP, ILY | OHS, OLP, ILY | OHS | Same |
| Light Source | LED | LED | LED | LED | Same |
| Anatomical Location | Face and Body | Face | Face and Body | Face | Same |
| Wavelength | Red Light Mode: 660 nmNIR Mode: 850 nmRed + NIR Mode: 660 nm + 850 nm | Red Light: 620 nmInfrared Light: 850 nmMixed Light: 460 nm, 620 nm, 850 nm | Red Light: 633 nm ± 5 nmNIR Light: 830 nm ± 5 nm | Red: 640 nm ± 25 nmNIR: 880 nm ± 50 nm | Same |
| Intensity / Irradiance | COB1400 (~20 inches / 50 cm Distance):Red Light Mode: 6.147 mW/cm²NIR Mode: 1.425 mW/cm²Red + NIR Mode: 7.687 mW/cm²HP1800PRO (~20 inches / 50 cm Distance):Red Light Mode: 4.116 mW/cm²NIR Mode: 1.56 mW/cm²Red + NIR Mode: 5.675 mW/cm² | Red: 2.0~3.0 mW/cm²Infrared: 2.0~4.0 mW/cm²Mixed: 9.0~12.0 mW/cm² | Red Light: 11.5 mW/cm²NIR Light: 5.5 mW/cm² | 6.5 mW/cm² | SameWithin the range of the predicate devices |
| Treatment Time | Maximum 30 min per session | Manual Mode: 15 minutes each timeAutomatic Mode: 10 minutes each time.3~4 treatment a week, reduce to 1~2 treatment a week once the results shown. | Up to 30 Minutes | 3 treatments per week(1800 Seconds)4 weeks | SameWithin the range of the predicate devices |
{8}
510(k) Summary
Shenzhen Hanhua Opto Co., Ltd.
## 8. Performance Data
The following performance data are provided in support of the substantial equivalence determination:
The proposed device performs testing in accordance with the following recognized consensus standards:
IEC 60601-1:2005+AMD1:2012+AMD2:2020 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
IEC 60601-1-2:2014+AMD1:2020 Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests
IEC 60601-1-11:2015+AMD1:2020 Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
IEC 60601-1-6:2010+AMD1:2013+AMD2:2020 Medical Electrical Equipment - Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability
IEC 62471:2006 Photobiological safety of lamps and lamp systems
IEC 60601-2-57:2023 Medical electrical equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring, cosmetic and aesthetic use
IEC 60601-2-83: 2019+AMD1:2022 Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
IEC 62304:2006+AMD1:2015 Medical Device Software - Software Life Cycle Processes
Software verification and validation testing were conducted and basic level of documentation was provided as recommended by FDA’S Guidance for Industry and FDA Staff, “Content of Premarket Submissions for Device Software Functions (2023).”
## 9. Clinical/Animal Testing
Clinical/animal testing was not performed for the proposed device as part of the submission.
{9}
510(k) Summary
Shenzhen Hanhua Opto Co., Ltd.
## 10. Conclusions
Based on the above performance as documented in this application, the LED Therapy Light is found to have a safety and effectiveness profile that is similar to the predicate devices.
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.