K223544 · Guangdong Newdermo Biotech Co., Ltd. · OHS · Feb 23, 2023 · General, Plastic Surgery
Device Facts
Record ID
K223544
Device Name
LED light therapy mask (FM-01, FM-02, FM-03)
Applicant
Guangdong Newdermo Biotech Co., Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Feb 23, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Red light: Treatment of full-face wrinkles. Blue light: Treatment of mild to moderate inflammatory acne. Infrared light: Provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. Mixed light: Treatment of mild to moderate inflammatory acne.
Device Story
Wearable LED phototherapy mask (models FM-01, FM-02, FM-03) for home use. Device consists of mask, controller, adapter, USB cable, and straps. Emits red (620nm), blue (460nm), infrared (850nm), or mixed light. User operates device via controller to select light mode for skin treatment or topical heating. Output is light energy delivered to face/body. Healthcare provider does not operate; patient self-administers. Benefits include wrinkle reduction, acne treatment, and muscle pain relief. Device transforms electrical input into therapeutic light output. Safety verified via electrical and biocompatibility testing.
Clinical Evidence
No clinical data. Bench testing only. Compliance with ANSI AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 60601-2-57. Biocompatibility testing per ISO 10993-5 and ISO 10993-10.
Technological Characteristics
Wearable LED mask. Light source: LED. Wavelengths: Red (620nm), Blue (460nm), Infrared (850nm), Mixed. Intensity: 2.0-12.0 mW/cm². Power: 100-240V AC input, 5V DC output. Biocompatible materials per ISO 10993. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57.
Indications for Use
Indicated for treatment of full-face wrinkles (red light), mild to moderate inflammatory acne (blue/mixed light), and topical heating for temporary relief of minor muscle/joint pain, arthritis, muscle spasm, stiffness, and increased local blood circulation (infrared light). Intended for OTC home use.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
LED FACIAL LIGHT THERAPY MASK, FLEXIBLE LED LIGHT THERAPY (K200983)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.
February 23, 2023
Guangdong Newdermo Biotech Co.,Ltd Annie Cai Certificate Engineer Building C28. Huachuang Industrial Park, Jinshan Avenue Shiji Town, Panyu Guangzhou. 511450 China
Re: K223544
Trade/Device Name: LED light therapy mask (FM-01, FM-02, FM-03) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS, OLP, ILY Dated: November 24, 2022 Received: November 25, 2022
Dear Annie Cai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jianting Wang -S
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K223544
Device Name LED light therapy mask (FM-01, FM-02, FM-03)
Indications for Use (Describe)
Red light: Treatment of full-face wrinkles.
Blue light: Treatment of mild to moderate inflammatory acne.
Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. Mixed light: Treatment of mild to moderate inflammatory acne.
Type of Use (Select one or both, as applicable)
| For activities identified in 37.44, Subpart D | <div> <input type="checkbox"/> </div> |
|---------------------------------------------------|--------------------------------------------------|
| For Streamlined permitting under 37.44, Subpart C | <div> <input checked="" type="checkbox"/> </div> |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) Summary K223544
{4}------------------------------------------------
# 510 (K) Summary
## Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92
#### 1. Submitter:
| 510(k) owner's name: | Guangdong Newdermo Biotech Co.,Ltd |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Address: | Building C28. Huachuang Industrial Park<br>Jinshan Avenue, Shiji Town,<br>Panyu, Guangzhou,<br>511450 P.R. China |
| Tel: | +86(020) 31105688 |
| Contact person: | Annie Cai |
| Email: | caidejiao@qq.com |
| Preparing date: | February 22, 2023 |
| 2. Device name and classification: | |
| Device Name: | LED light therapy mask |
| Model: | FM-01, FM-02, FM-03 |
| Classification Name/<br>Product code: | 21 CFR 878.4810 OHS<br>21 CFR 878.4810 OLP<br>21 CFR 890.5500 ILY |
| Regulatory Class: | Class II |
## 3.Premarket Notification Class III Certification and Summary
Not applicable, the subject device is Class II.
#### 4. Predicate Device(s):
#### 1) Predicate device1
| Sponsor: | NINGBO HESI ELECTRIC CO., LTD |
|----------------------|--------------------------------------------------------------|
| Device name: | LED FACIAL LIGHT THERAPY MASK, FLEXIBLE LED LIGHT<br>THERAPY |
| 510(k) Number: | K200983 |
| Product Code: | OHS, OLP, ILY |
| 2) Predicate device2 | |
| Sponsor: | Hunan Guangye Biotechnology Co., Ltd. |
| Device name: | Beauty LED Mask |
{5}------------------------------------------------
Traditional 510(k) Submission of LED light therapy mask
| 510(k) Number: | K221151 |
|----------------------|------------------------------------------------------|
| Product Code: | OHS, OLP |
| 3) Predicate device3 | |
| Sponsor: | Zhongshan Bisen Plastic Electronic Products Co.,Ltd. |
| Device name: | RED Light Device |
| 510(k) Number: | K162489 |
| Product Code: | OHS |
| 4) Predicate device4 | |
| Sponsor: | Theragun, Inc. |
| Device name: | TheraFace LED |
| 510(k) Number: | K212155 |
| Product Code: | OHS, OLP |
| 5) Reference device | |
| Sponsor: | Xuzhou Kernel Medical Equipment Co., Ltd. |
| Device name: | LED Light Therapy Device, KN-7000L |
| 510(k) Number: | K222751 |
| Product Code: | GEX |
#### 5. Reason for Submission
New device, there were no prior submissions for the device.
#### 6. Pre-Submission, IDE
Not applicable, there is no prior submission.
#### 7. Device Description:
LED light therapy mask is a home use wearable LED phototherapy device which can help solve various skin problems. LED light therapy mask is consisting of mask, controller, adapter, USB cable and straps.
There are 4 kinds of light which include Red light (wavelength 620mm), Blue light (wavelength 460nm), Infrared light (wavelength 850mm), Mixed light (wavelength 620mm and 850nm and 460nm).
#### 8. Intended Use:
Red light: Treatment of full-face wrinkles.
Blue light: Treatment of mild to moderate inflammatory acne.
{6}------------------------------------------------
Infrared light: Provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
Mixed light: Treatment of mild to moderate inflammatory acne.
| Item | Subject Device | Predicate<br>device1(K200983) | Predicate<br>Device2(K221151) | Comparison<br>Result |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Trade name | LED light therapy<br>mask | LED FACIAL<br>LIGHT<br>THERAPY<br>MASK,<br>FLEXIBLE LED<br>LIGHT<br>THERAPY | Beauty LED Mask | / |
| 510 (k) number | K223544 | K200983 | K221151 | / |
| Manufacturer | Guangdong<br>Newdermo Biotech<br>Co.,Ltd | NINGBO HESI<br>ELECTRIC CO.,<br>LTD | Hunan Guangye<br>Biotechnology Co.,<br>Ltd. | / |
| Regulation<br>number | 21 CFR 878.4810<br>21 CFR 890.5500 | 21 CFR 878.4810,<br>21 CFR 890.5500 | 21 CFR 878.4810 | Same |
| Regulation<br>Name | Light Based Over<br>The Counter<br>Wrinkle<br>Reduction(OHS),<br>Over-The-Counter<br>Powered Light<br>Based Laser For<br>Acne(OLP),<br>Infrared,<br>Therapeutic<br>Heating(ILY) | Light Based Over<br>The Counter<br>Wrinkle<br>Reduction(OHS),<br>Over-The-Counter<br>Powered Light<br>Based Laser For<br>Acne(OLP),<br>Lamp, Infrared,<br>Therapeutic<br>Heating(ILY) | Light Based Over<br>The Counter<br>Wrinkle<br>Reduction(OHS),<br>Over-The-Counter<br>Powered Light<br>Based Laser For<br>Acne(OLP) | Same |
| Product code | OHS, OLP, ILY | OHS, OLP, ILY | OHS, OLP | Same |
| Classification | II | II | II | Same |
| Indications for<br>use/ Intended<br>use | Red light:<br>Treatment of<br>full-face wrinkles.<br>Blue light:<br>Treatment of mild<br>to moderate<br>inflammatory acne. | The LED FACIAL<br>LIGHT<br>THERAPY<br>MASK (Model:<br>HK207) is<br>intended to:<br>The device | The device is<br>intended to use<br>LED light for the<br>treatment of<br>wrinkles and mild<br>to moderate acne. | Similar<br>Note2 |
#### 9. Predicate Device Comparison
{7}------------------------------------------------
| Infrared light:<br>Provide topical<br>heating for the<br>purpose of<br>elevating tissue<br>temperature;<br>arthritis and muscle<br>spasm; relieving<br>stiffness; promoting<br>the relaxation of<br>muscle tissue; and<br>to temporarily<br>increase local blood<br>circulation.<br>Mixed light:<br>Treatment of mild<br>to moderate<br>inflammatory acne. | emitting energy in<br>the blue is<br>intended to reduce<br>the mild to<br>moderate<br>inflammatory acne<br>vulgaris.<br>- The device<br>emitting energy in<br>the red and<br>infrared spectrum<br>is intended for the<br>treatment of<br>full-face wrinkles.<br>The FLEXIBLE<br>LED LIGHT<br>THERAPY<br>(Model: HK209) is<br>intended to:<br>- The device<br>emitting energy in<br>the blue is<br>intended to reduce<br>the mild to<br>moderate<br>inflammatory acne<br>vulgaris.<br>- The device<br>emitting energy in<br>the red and<br>infrared spectrum<br>is intended for the<br>treatment of<br>full-face<br>wrinkles.<br>- The device is<br>intended to deliver<br>heat in the IR<br>spectrum to<br>provide topical<br>heating for the<br>purpose of<br>elevating tissue | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
{8}------------------------------------------------
| | | the temporary<br>relief of minor<br>muscle and joint<br>pain, arthritis and<br>muscle spasm;<br>relieving stiffness;<br>promoting the<br>relaxation of<br>muscle tissue; and<br>to temporarily<br>increase local<br>blood circulation. | | |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| Location for use | Face and body | Entire Face and<br>body | Face | Same |
| OTC or<br>prescription | OTC | OTC | OTC | Same |
| Power supply | Input:<br>100-240 V~, 50/60<br>Hz, 0,25 A<br>Output: DC 5 V,<br>500 mA | Input: 100 –<br>240Vac,<br>2.0 A, 50/60Hz | Input: 100 -240 V<br>~ 50 /60 Hz<br>Output: 5V 1A | Similar<br>Note4 |
| Light source | LED | LED | LED | same |
| Wavelength | Red: 620nm<br>Blue: 460nm<br>Infrared: 850nm<br>Mixed: 620nm and<br>850nm and 460nm | 465nm,<br>640nm,<br>880nm | Red (637nm±5nm)<br>and<br>IR (854nm±5nm);<br>Blue (465±5nm) | Compare<br>with<br>predicate<br>device 4<br>Similar<br>Note 5 |
| LED Intensity | Red light: 2.0~3.0<br>mW/cm²<br>Blue light:2.0~4.0<br>mW/cm²<br>Infrared light:<br>2.0~4.0 mW/cm²<br>Mixed light:<br>9.0~12.0 mW/cm² | 6.5 mW/cm² | Red+IR:<br>25.5mW/cm²<br>Blue: 1.36mW/cm² | Similar<br>Note6 |
| Treatment time | Manual Mode:<br>15minutes each<br>time,<br>Automatic Mode:<br>10minutes each<br>time.<br>3-4 treatment a<br>week, reduce to 1-2 | 3 times a week for<br>30<br>min. 4 weeks | 10min each time | Compare<br>with<br>predicate<br>device 3<br>Similar<br>Note7 |
| | treatment a week<br>once the results<br>shown. | | | |
| Dimensions<br>(mm) | FM-01:<br>207X277X43mm,<br>FM-02:<br>198X383X33.5mm,<br>FM-03:<br>237.5X108X8.1mm | Not public | LED Mask:<br>Approximately 183 mm x 238 mm x 98 mm<br>Controller: 100mm x 50 mm x 21.5 mm | / |
| Electrical safety | IEC 60601-1;<br>IEC 60601-1-2 | IEC 60601-1;<br>IEC 60601-1-2 | IEC 60601-1;<br>IEC 60601-1-2 | Same |
| Biocompatibility<br>feature | All body-contacting<br>materials are<br>complied with<br>ISO10993-5 and<br>ISO 10993-10 | All<br>body-contacting<br>materials are<br>complied with<br>ISO10993-5 and<br>ISO 10993-10 | All<br>body-contacting<br>materials are<br>complied with<br>ISO10993-5 and<br>ISO 10993-10 | Same |
{9}------------------------------------------------
| Item | Subject Device | Predicate<br>device3(K162489) | Predicate<br>Device4(K212155) | Comparison<br>Result |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| Trade name | LED light therapy<br>mask | RED Light Device | TheraFace LED | / |
| 510 (k) number | K223544 | K162489 | K212155 | / |
| Manufacturer | Guangdong<br>Newdermo Biotech<br>Co.,Ltd | Zhongshan Bisen<br>Plastic Electronic<br>Products Co.,Ltd. | Theragun, Inc. | / |
| Regulation<br>number | 21 CFR 878.4810,<br>21 CFR 890.5500 | 21 CFR 878.4810 | 21 CFR 878.4810 | Similar<br>Note1 |
| Regulation<br>Name | Light Based Over<br>The Counter<br>Wrinkle<br>Reduction(OHS),<br>Over-The-Counter<br>Powered Light<br>Based Laser For<br>Acne(OLP),<br>Infrared,<br>Therapeutic<br>Heating(ILY) | Light Based Over<br>The Counter<br>Wrinkle<br>Reduction(OHS), | Light Based Over<br>The Counter<br>Wrinkle<br>Reduction(OHS), | Similar<br>Note1 |
| Product code | OHS, OLP, ILY | OHS | OHS, OLP | Similar<br>Note1 |
| Classification | II | II | II | Same |
| Indications for<br>use/ Intended<br>use | Red light: | The RED Light<br>Device is an OTC<br>device indicated to<br>emit energy in the<br>red and IR region<br>of the<br>spectrum for use in<br>dermatology for<br>the treatment of<br>periorbital<br>wrinkles. | 1) The red light is<br>intended to treat<br>periorbital<br>wrinkles.<br>2) The blue light is<br>intended to treat<br>mild to moderate<br>inflammatory acne.<br>3) The Red + IR<br>is intended to treat<br>periorbital<br>wrinkles. | Similar<br>Note2 |
| Red light:<br>Treatment of<br>full-face wrinkles<br>Blue light:<br>Treatment of mild<br>to moderate<br>inflammatory acne.<br>Infrared light:<br>Provide topical<br>heating for the<br>purpose of<br>elevating tissue<br>temperature;<br>arthritis and muscle<br>spasm; relieving<br>stiffness; promoting<br>the relaxation of<br>muscle tissue; and<br>to temporarily<br>increase local blood<br>circulation.<br>Mixed light:<br>Treatment of mild<br>to moderate<br>inflammatory acne. | | | | |
| Location for use | Face and body | / | Face | Similar<br>Note3 |
| OTC or<br>prescription | OTC | OTC | OTC | Same |
| Power supply | Input:<br>100-240 V~, 50/60<br>Hz, 0,25 A<br>Output: DC 5 V,<br>500 mA | Adaptor:100~240V<br>AC 50/60Hz<br>Lithium battery:<br>2x3.7V | Lithium<br>battery :2x3.7V | Similar<br>Note4 |
| Light source | LED | LED | LED | Same |
| Wavelength | Red: 620nm<br>Blue: 460nm<br>Infrared: 850nm<br>Mixed: 620nm and<br>850nm and 460nm | Red: 633 ±5nm<br>Infrared: 830 ±5nm | Red light:<br>633nm±10nm<br>Blue light:<br>415nm±10nm<br>Red+IR:<br>633±10nm /<br>830nm±10nm | Compare<br>with<br>predicate<br>device 4<br>Similar<br>Note 5 |
| LED Intensity | Red light: 2.0~3.0 | 125 mW/cm² | Red light 73±5 | Compare<br>with |
| | Blue light:2.0~4.0<br>mW/cm²<br>Infrared light:<br>2.0~4.0 mW/cm²<br>Mixed light:<br>9.0~12.0 mW/cm² | nm); 55 mW/cm²<br>(830nm) | Blue light 64±5<br>mW/cm²<br>Red+IR:<br>73±5/55±5<br>mW/cm² | predicate<br>device 3<br>Similar<br>Note6 |
| Treatment time | Manual Mode:<br>15minutes each<br>time,<br>Automatic Mode:<br>10minutes each<br>time.<br>3-4 treatment a<br>week, reduce to 1-2<br>treatment a week<br>once the results<br>shown. | For the first month<br>(4 weeks),<br>treatment<br>should be<br>performed 3 times<br>a week for<br>15-20 minutes each<br>time.(5-7 minutes<br>on<br>each treatment<br>zone). | Red light: 5 - 7<br>minutes per<br>treatment zone<br>Blue: 5-7 minutes<br>per treatment zone<br>Red+IR: 5-7<br>minutes per<br>treatment zone | Compare<br>with<br>predicate<br>device 3<br>Similar<br>Note7 |
| Dimensions<br>(mm) | FM-01:<br>207X277X43mm,<br>FM-02:<br>198X383X33.5mm,<br>FM-03:<br>237.5X108X8.1mm | / | / | / |
| Electrical safety | IEC 60601-1;<br>IEC 60601-1-2 | IEC 60601-1;<br>IEC 60601-1-2 | IEC 60601-1;<br>IEC 60601-1-2 | Same |
| Biocompatibility<br>feature | All body-contacting<br>materials are<br>complied with<br>ISO10993-5 and<br>ISO 10993-10 | All<br>body-contacting<br>materials are<br>complied with<br>ISO10993-5 and<br>ISO 10993-10 | All<br>body-contacting<br>materials are<br>complied with<br>ISO10993-5 and<br>ISO 10993-10 | Same |
Guangdong Newdermo Biotech Co.,Ltd
{10}------------------------------------------------
{11}------------------------------------------------
Comparison in Detail(s):
Note1
The Regulation number, Regulation Name, Product code of predicate1 are not identical to the subject device, all these device are process similar intended use and use very similar light wavelengths to get the intended purpose. This slight deference will not raise safety and effective issue.
Note2
The Indications for use/ Intended use of Red light, Blue light , Infrared light of the subject device is the same as that of primary device K200983. The Indications for use/ Intended use of mixed light of the subject device is the same as that of secondary device K221151.
{12}------------------------------------------------
#### Note3
The treatment location of the subject device is the same as the predicate device K200983 but identical the other predicate device. The difference in applicable location will not raise safety and effective issue.
#### Note4
The power supply of devices are very similar but not identical, the IEC60601-1test demonstrated the safety of the power adapter. The slight difference will not raise safety and effective issue. Note5
The light wavelength is very similar but not identical. The subject device emits blue light, red light and IR light. In the mixed light mode, the subject device emits blue light, IR light simultaneously, while the reference device K222751 can emit red, blue, IR light simultaneously and it has been demonstrated that emit three lights simultaneously is safe and effect. Besides, the subject device has passed IEC60601-2-57 test, the slight difference will not raise safety and effective issue.
#### Note6
The LED Intensity of these devices are different, and the LED intensity of the subject device is low and subject device has passed IEC60601-2-57 test, the slight difference will not raise safety and effective issue.
#### Note7
The treatment time of these devices are very similar but not identical, treatment is one of the factors to get the intended purpose. What's more, the subject device has IEC60601-1, IEC60601-2-57 test, the slight difference will not raise safety and effective issue.
All the differences don't affect the safety and effectiveness which is concluded after all the required testing, so no safety and effectiveness issues relating to the product come into conclusion.
#### 10. Performance Data:
#### Non-clinical data:
Non-clinical tests have been conducted to verify that the Beauty LED Mask meets all design specifications which supports the conclusion that it's Substantially Equivalent (SE) to the predicate device. The testing results demonstrate that the subject device complies with the following standards:
- ANSI AAMI ES 60601-1, Medical electrical equipment Part 1: General A requirements for basic safety and essential performance
- A IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests
- > IEC 60601-1-11, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
{13}------------------------------------------------
- IEC 60601-2-57, Medical Electrical Equipment Part 2-57: Particular requirements A for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
#### Biocompatibility Test
The device has been tested for biocompatibility; it complies with the following standards.
- > ISO 10993-5, Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity
- A ISO 10993-10, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization.
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
#### Clinical data: Not applicable.
#### Summary
Based on the non-clinical and clinical performance as documented in the system development, the subject devices were found to have a safety and effectiveness profile that is similar to the predicate device.
#### 11. Conclusion
The non-clinical data support the safety of the device and the hardware and software verification and validation demonstrate that LED light therapy mask should perform as intended in the specified use conditions, and all the data demonstrate that the subject devices perform comparably to the predicate device that is currently marketed for the same intended use. In other words, the subject device (LED light therapy mask) is Substantial Equivalent to the predicate devices K200983, K221151, K162489 and K212155.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.