K212275 · Aesthetic Technology, Ltd. · OHS · Nov 18, 2021 · General, Plastic Surgery
Device Facts
Record ID
K212275
Device Name
Dermalux Flex MD
Applicant
Aesthetic Technology, Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Nov 18, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Intended to deliver heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. The blue spectrum light is intended to reduce mild to moderate inflammatory acne vulgaris. Intended to emit energy in the red and infrared spectrum for the treatment of periorbital and full face wrinkles.
Device Story
Dermalux Flex MD is a light-based phototherapy system using blue (415nm), red (633nm), and near-infrared (830nm) LEDs. Device consists of a flexible LED array and a capacitive touch controller. Used in clinical or home settings; operated by patient or clinician. User selects treatment via controller; device emits specific wavelengths to target skin/tissue. For facial treatments, array is placed in a base; for body, array is applied directly to the area. Device uses pre-set timers to control treatment duration. Output provides topical heating, pain relief, acne reduction, and wrinkle treatment. Benefits include non-invasive management of dermatological and musculoskeletal conditions.
Clinical Evidence
No clinical or animal studies were performed. Substantial equivalence is based on bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), photobiological safety (IEC 62471), and software lifecycle processes (IEC 62304).
Indicated for patients seeking temporary relief of minor muscle/joint pain, arthritis, muscle spasm, stiffness, and promotion of muscle relaxation/local blood circulation. Also indicated for reduction of mild to moderate inflammatory acne vulgaris and treatment of periorbital and full-face wrinkles.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 18, 2021
Aesthetic Technology Ltd. % Richard Hamer US Agent/Consultant Richard Hamer Associates, LLC 705 Spring Lakes Blvd Bradenton, Florida 34210
Re: K212275
Trade/Device Name: Dermalux Flex MD Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS, OLP, ILY Dated: October 19, 2021 Received: October 25, 2021
Dear Richard Hamer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212275
Device Name
Dermalux Flex MD
Indications for Use (Describe)
Intended to deliver heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
The blue spectrum light is intended to reduce mild to moderate inflammatory acne vulgaris.
Intended to emit energy in the red and infrared spectrum for the treatment of periorbital and full face wrinkles.
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K212275
## 510(k) SUMMARY
### I. ADMINISTRATIVE
### Submitter:
Aesthetic Technology Ltd. 211 Europa Blvd. Warrington, Cheshire WA5 7TN United Kingdom +44 (0)845 689 1789
Contact Person: Dale Needham
Date of Preparation: November 18, 2021
### II. DEVICE NAME
Proprietary Name: Dermalux® Flex MD
Common Name: Laser Surgical Instrument for Use in General and Plastic Surgery and in Dermatology
Classification Name: Over-The-Counter Powered Light Based Laser For Acne (OLP), Light Based Over-The Counter Wrinkle Reduction (OHS) and Lamp Infrared, Therapeutic Heating (ILY),
Regulation Number: 21 CFR §878.4810
Regulatory Class: Class II
Product Code: OLP, OHS, ILY
### III. PREDICATE DEVICES
Dermalux® Flex MD; K202028, Aesthetic Technology Ltd.
Flexible LED Light Therapy Model HK209; K200983, Guangzhou GLOOMED Biological Technology Co. Ltd.
In addition, Omnilux CLEAR; K210948, GlobalMed Technologies and Biophotas Celluma3; K152280, Biophotas, Inc. are cited as reference devices for the purpose of establishing labeling equivalence.
{4}------------------------------------------------
# IV. DEVICE DECRIPTION
The Dermalux® Flex MD is a Class II medical device for use on a treatment bed. The device emits specific wavelengths of low level, narrow band light for the treatment of certain dermatological and musculoskeletal indications. The wavelengths used in the Dermalux® Flex MD system are Blue 415nm, Red 633nm and Near Infrared 830nm.
The Dermalux® Flex MD enables treatment of the face and the body via a flexible LED Array. The system is operated by a capacitive touch Controller, which allows one of 3 treatments to be selected through selecting various wavelengths. The light is generated by Light Emitting Diodes (LED's) that are contained within the flexible LED Array. The LED Array can be placed into the Base, holding in the appropriate position to administer a facial treatment. For use on the body, the LED Array is removed from the Base unit and placed over the appropriate body part.
## V. INDICATIONS FOR USE
Intended to deliver heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
The blue spectrum light is intended to reduce mild to moderate inflammatory acne vulgaris.
Intended to emit energy in the red and infrared spectrum for use in dermatology for the treatment of periorbital and full face wrinkles.
| Parameter | Subject Device | Predicate Device (K202028) | Secondary Predicate (K200983) |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product name | Dermalux® Flex MD<br>Phototherapy System | Dermalux® Flex MD<br>Phototherapy System | Flexible LED Light Therapy<br>Model HK209 |
| Product code | OHS, OLP. ILY | ILY, OHS, OLP | OHS, OLP, ILY |
| Regulation No. | 21 CFR 878.4810<br>21 CFR 890.5500 | 21 CFR 878.4810<br>21 CFR 890.5500 | 21 CFR 878.4810<br>21 CFR 890.5500 |
| Class | Class II | Class II | Class II |
| Intended use | Use of the red, blue and near-<br>infrared regions of the spectrum<br>to emit energy to treat dermato-<br>logical and musculoskeletal<br>conditions. | Use of the red, blue and near-<br>infrared regions of the spectrum<br>to emit energy to treat dermato-<br>logical and musculoskeletal<br>conditions | Use of the red, blue and near-<br>infrared regions of the spectrum<br>to emit energy to treat dermato-<br>logical and musculoskeletal con-<br>ditions |
| Indications for use | Intended to deliver heat in the IR<br>spectrum to provide topical<br>heating for the purpose of ele-<br>vating tissue temperature; for the<br>temporary relief of minor muscle<br>and joint pain, arthritis and<br>muscle spasm; relieving stiff- | Intended to deliver heat in the IR<br>spectrum to provide topical<br>heating for the purpose of ele-<br>vating tissue temperature; for the<br>temporary relief of minor muscle<br>and joint pain, arthritis and<br>muscle spasm; relieving stiff- | The device emitting energy in the<br>blue spectrum is intended to re-<br>duce mild to moderate inflamma-<br>tory acne vulgaris.<br>The device emitting energy in the<br>red and infrared spectrum is in-<br>tended for the treatment of full- |
# VI. COMPARISON TO PREDICATE DEVICES
{5}------------------------------------------------
| | ness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. The blue spectrum light is intended to reduce mild to moderate inflammatory acne vulgaris. Intended to emit energy in the red and infrared spectrum for the use in dermatology for the treatment of periorbital and full face wrinkles. | ness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. The blue spectrum light is intended to reduce mild to moderate inflammatory acne vulgaris. Intended to emit energy in the red and infrared spectrum for the use in dermatology for the treatment of periorbital and full face wrinkles. | face wrinkles. The device is Intended to deliver heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Power supply | 100-240Vac, 50/60Hz, 4.6 - 1.85A, 460-430W. | 100-240Vac, 50/60Hz, 4.6 - 1.85A, 460-430W. | 100-240 Vac<br>2.0 A, 50/60Hz |
| Wavelength | Red light: 633nm±5nm<br>Blue light: 415nm±5nm<br>NIR light: 830nm±5nm | Red light: 633nm±5nm<br>Blue light: 415nm±5nm<br>NIR light: 830nm±5nm | Red light: 640nm<br>Blue light: 465nm<br>NIR Light: 880nm |
| Panels Type | 1 Panel | 1 Panel | 1 Panel |
| Light frequency | N/A as DC Power | N/A as DC Power | N/A as DC Power |
| Output Power | Red - 633nm (11.5mW/cm²)<br>Blue - 415nm (5.52mW/cm²)<br>NIR - 830nm (5.5mW/cm²) | Red - 633nm (11.5mW/cm²)<br>Blue - 415nm (5.52mW/cm²)<br>NIR - 830nm (5.5mW/cm² | 6.5 mW/cm² |
| Standard dose in Joules | Red: 20J/cm²<br>Blue: 10J/cm²<br>NIR: 10J/cm² | Red: 20J/cm²<br>Blue: 10J/cm²<br>NIR: 10J/cm² | 11.7 J/cm² |
| Treatment area | Face and Body<br>792 cm² | Face and Body<br>792 cm² | Face and Body<br>890 cm² |
| Treatment time | Up to 30 minutes | Up to 30 minutes | 30 minutes |
| Treatment protocol<br>(Treatment time) | Acne and Wrinkles: 3-4x weekly; 4 wks. Pain relief: 3x weekly | Not stated | 3x weekly; 4wks |
| Operation interface | Device uses a pre-set timer and software to control treatment duration. | Device uses a pre-set timer and software to control treatment duration. | Not stated |
| Software | Yes | Yes | Not stated |
Substantial equivalence is readily evidenced by comparison of product attributes summarized in the table above. Apparent differences in output power and standard dose do not impact safety or effectiveness because the amount of energy delivered to the skin is similar due to differences in treatment duration and distance from the LED source.
# VII. PERFORMANCE TESTING
Bench: Performance testing of the Dermalux® Flex MD was conducted to verify that the device met all design specifications. The test results demonstrate that the Dermalux® Flex MD complies with all requirements, including international and FDA-recognized consensus standards:
EN/IEC 60601-1 (FDA #19-4) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance: 2005
EN/IEC 60601-1-2 (FDA #19-8) Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests: 2018
{6}------------------------------------------------
EN/IEC 60601-1-11 (FDA #19-14) Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment: 2015
EN/IEC 62304 (FDA #13-79) Medical device software - Software life cycle processes: 2006
EN/IEC 62471 (FDA #12-249) Photobiological safety of lamps and lamp systems: 2008
IEC 60601-2-57 (FDA #12-242) Medical electrical equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use: 2011
IEC 62366-1 (FDA #5-113) Medical devices - Part 1: Application of usability engineering to medical devices: 2015
EN/IEC 60601-1-6 (FDA #5-89) Medical electrical equipment -Part 1-6: General requirements for basic safety and essential performance - Collateral Standard: Usability: 2013
Photometric Testing of LED Light Panel
ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.
Animal: No studies were performed.
Clinical: No studies were performed.
### VIII. CONCLUSION
The Dermalux Flex MD is technologically identical to the predicate device cleared under K202028, intended for over-the-counter use, and similar to secondary predicate device K200983. Labeling is equivalent to reference devices K210948 and K152280. Based on design considerations and testing of product attributes, we conclude that the Dermalux® Flex MD performs at least as well as the predicate devices. The Dermalux® Flex MD is therefore considered to be substantially equivalent to the above-mentioned predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.