PELVIPOWER Pelvic Functional Magnetic Stimulation (PELVIPOWER PelvicFMS) (033-0-1100-02)
Applicant
Pontemedag
Product Code
KPI · Gastroenterology, Urology
Decision Date
Apr 17, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5320
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K252334 · Apr 17, 2026
PELVIPOWER Pelvic Functional Magnetic Stimulation (PELVIPOWER PelvicFMS) (033-0-1100-02)
Pontemedag
Post-market field study
A post-market field study was utilized to support the service life verification of the device.
service life; post-market field study; durability
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Post-market field study; Post-market field study
—
Not applicable for this study
Service life verification
Indications for Use
The PELVIPOWER Pelvic Functional Magnetic Stimulation Device (PELVIPOWER PelvicFMS) is intended to provide entirely non-invasive electromagnetic stimulation of pelvic floor musculature for the purpose of strengthening and rehabilitating weak pelvic muscles and restoring neuromuscular control for the treatment of male and female urinary incontinence.
Device Story
PELVIPOWER PelvicFMS is a non-invasive pelvic floor muscle stimulator; used in healthcare facilities by clinicians. Device consists of a chair with an integrated electromagnetic coil; generates pulsed magnetic fields to induce pelvic floor muscle contractions. Features include adjustable armrests, seat inclination, leg rest, and horizontal coil positioning to optimize field delivery. Clinician selects therapy mode via 12-inch touchscreen; device displays therapy parameters and remaining duration. Operates in automatic or manual modes. Benefits include non-invasive strengthening of pelvic musculature and restoration of neuromuscular control for urinary incontinence treatment.
Clinical Evidence
Bench testing only. Included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2, IEC 60601-4-2), software verification/validation, and cybersecurity testing. Performance bench data included magnetic stimulation parameter verification, service life testing, stimulation coil surface temperature rise (per ASTM F2182-19E02), and magnetic flux density measurements. All acceptance criteria met.
Technological Characteristics
Magnetic stimulation via coil; 0.5–2 T intensity; 5–50 Hz pulse rate; 340μs pulse width. Adjustable coil position (150mm horizontal), seat inclination, and leg rest. 12" touchscreen interface. Power: 230 V~ (with US external transformer). Artificial leather applicator. Software-controlled. Sterilization: non-sterile.
Indications for Use
Indicated for male and female patients with urinary incontinence requiring strengthening and rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control.
Regulatory Classification
Identification
A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 17, 2026
PonteMed AG
% Adrienne Lenz
Principal Medical Device Regulatory Expert
Hyman, Phelps, & McNamara, P.C.
700 Thirteenth St, N.W.
Suite 1200
Washington, District of Columbia 20005
Re: K252334
Trade/Device Name: PELVIPOWER Pelvic Functional Magnetic Stimulation (PELVIPOWER PelvicFMS) (033-0-1100-02)
Regulation Number: 21 CFR 876.5320
Regulation Name: Nonimplanted electrical continence device
Regulatory Class: II
Product Code: KPI
Dated: March 18, 2026
Received: March 18, 2026
Dear Adrienne Lenz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252334 - Adrienne Lenz
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn
{2}
K252334 - Adrienne Lenz
Page 3
(https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JESSICA K. NGUYEN -S
Jessica K. Nguyen, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252334 | |
| Device Name PELVIPOWER Pelvic Functional Magnetic Stimulation (PELVIPOWER PelvicFMS) (033-0-1100-02) | |
| Indications for Use (Describe) | |
| The PELVIPOWER Pelvic Functional Magnetic Stimulation Device (PELVIPOWER PelvicFMS) is intended to provide entirely non-invasive electromagnetic stimulation of pelvic floor musculature for the purpose of strengthening and rehabilitating weak pelvic muscles and restoring neuromuscular control for the treatment of male and female urinary incontinence. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K252334
Page 1 of 4
# K252334
# 510(K) SUMMARY
## 1. SUBMITTER INFORMATION
510(k) Submitter: PonteMed AG
Address: Stickereistrasse 4
9320 Arbon
Switzerland
Applicant Contact Person: Dr. Remo Schneider
CEO
Phone: +41 71 333 60 7
E-Mail: schneider@pontemed.com
Correspondent Contact: Ms. Adrienne Lenz
Hyman, Phelps, & McNamara, P.C
1101 K St, N.W.; Suite 700
Washington, DC 200005
United States
Phone: (202)737-4292
E-Mail: alenz@hpm.com
Date Summary Prepared: April 17, 2026
## 2. DEVICE NAME
Trade Name of the Device: PELVIPOWER PelvicFMS (033-0-1100-02)
Common Name: Pelvic floor muscle stimulator
Classification Name: Stimulator, Electrical, Non-Implantable, For Incontinence
Classification Regulation: 21 CFR 876.5320
Device Class: Class II
Panel: Gastroenterology/Urology
Product Code: KPI
## 3. PREDICATE DEVICE
Predicate Devices: HPM-6000UF
510(k) number: K181497
Company Name: BTL Industries, Inc.
The predicate device has not been subject to a design related recall.
The BTL-398 (K241516) was used as a reference device to support testing of the coil movement feature of the subject device.
## 4. DEVICE DESCRIPTION
PELVIPOWER PelvicFMS is a non-invasive pelvic floor muscle stimulator device used for the purpose of strengthening, and rehabilitation of weak pelvic muscles and restoration of neuromuscular control. The device produces electromagnetic field impulses that stimulate and strengthen the pelvic floor and surrounding
{5}
K252334
Page 2 of 4
musculature. The coil under the seat of the PELVIPOWER PelvicFMS chair generates a pulsed magnetic field that elicits muscle contractions in the pelvic floor area. The device is a non-sterile, multi-patient, multi-use device intended for use in health care facilities.
The PELVIPOWER PelvicFMS chair includes an adjustable armrest that can be folded up as necessary to allow easier access for patients with limited mobility (e.g., wheelchair users). To adapt the magnetic field to the patient sitting above it and to achieve an optimal effect, the coil position, the angles of inclination of the seat, and the leg rest are electrically adjustable. The PELVIPOWER PelvicFMS includes a 12" touchscreen display. The on-screen information guides the user throughout the therapy session. The therapeutic parameters can be configured via the touch screen interface. Throughout the therapy session, the device displays therapy information, including the type of therapy, the remaining duration, and key therapy parameters. The chair applicator is made of artificial leather. The PELVIPOWER PelvicFMS chair therapy sessions can be delivered to patients following a pre-set (Automatic Setting) or healthcare professional (HCP)-determined Manual Setting.
# 5. INDICATION FOR USE
The PELVIPOWER Pelvic Functional Magnetic Stimulation Device (PELVIPOWER PelvicFMS) is intended to provide entirely non-invasive electromagnetic stimulation of pelvic floor musculature for the purpose of strengthening and rehabilitating weak pelvic muscles and restoring neuromuscular control for the treatment of male and female urinary incontinence.
# 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE
| Technical Characteristics | Subject Device
PonteMed AG's PELVIPOWER PelvicFMS
[K252334] | Predicate Device
BTL Industries, Inc.'s HPM-6000UF
[K181497] |
| --- | --- | --- |
| Indication for use | The PELVIPOWER Pelvic Functional Magnetic Stimulation Device (PELVIPOWER PelvicFMS) is intended to provide entirely non-invasive electromagnetic stimulation of pelvic floor musculature for the purpose of strengthening and rehabilitating weak pelvic muscles and restoring neuromuscular control for the treatment of male and female urinary incontinence. | HPM-6000UF is intended to provide entirely non-invasive electromagnetic stimulation of pelvic floor musculature for the purpose of rehabilitation of weak pelvic muscles and restoration of neuromuscular control for the treatment of male and female urinary incontinence. |
| Function | Pelvic floor muscle stimulation | Pelvic floor muscle stimulation |
| Type of Energy | Magnetic field | Magnetic field |
| Stimulation source | Magnetic coil | Magnetic coil |
| Prescription Use Only | Yes | Yes |
| Power Source | 230 V~, 50–60 Hz, max. 600 VA (2.5 A/230 V)
US external transformer: 115/120/125 V~, 50-60Hz, 600 VA
The external transformer provided with the system is required for connection to US mains to convert the voltage from 110/115 V to 230 V. | 100 – 240 V AC, 50 – 60 Hz, max 14 A |
| Therapy Process | Specific therapy mode is selected by the clinician based on the patient's symptoms. | Specific therapy mode is selected by the clinician based on the patient's symptoms. |
| Number of output channels | 1 | 1 |
{6}
K252334
Page 3 of 4
| Magnetic Field Intensity | 0.5 – 2 T | 0.7 – 2.5 T |
| --- | --- | --- |
| Pulse Repetition Rate | 5 to 50 Hz | 1 – 150 Hz |
| Pulse Width | 340μs (± 20%) | 280 μs (± 20%) |
| Pulse amplitude adjustment | 0-100% in 6 increments | 0-100% |
| Therapy Time | Up to 22 min | Up to 30 min |
| Interface | Touchscreen | Touchscreen |
| Firmware controlled | Yes | Yes |
| Environmental Specifications | Hospital/Clinics only | Hospital/Clinics only |
| Applicator | Chair | Chair |
| Applicator (seat) Dimensions | 51 x 54 cm (20 x 21 in) | 73 x 73 cm (29 x 29 in) |
| Coil position relative to pelvic floor muscles | Coil adjustment 150mm in horizontal direction | No coil adjustment provided, but patients can be positioned over the center of the coil. |
As evidenced by the above table, both the subject and the predicate devices have similar intended use, but the subject and predicate devices have different technological characteristics. The BTL-398 (K241516) was used as a reference device to support testing of the coil movement feature of the subject device which is absent in the predicate. Additionally, performance testing was conducted on the subject device, and it was established that the differences in technological characteristics between the subject and the predicate do not raise different questions of safety or effectiveness.
## 7. NON-CLINICAL TESTING
Below is a list of the tests that were performed and successfully completed for the subject device per the following guidance and standards.
- Electrical Safety testing according to IEC 60601-1: 2020 - Medical electrical equipment – Basic safety and essential performance
- Electromagnetic Compatibility testing according to IEC 60601-1-2: 2020 - General requirements for basic safety and essential performance -- Collateral Standard: Electromagnetic disturbances -- Requirements and tests and IEC 60601-4-2:2024-Medical electrical equipment – Part 4-2: Guidance and interpretation – Electromagnetic compatibility – Test plan and test report
- Software Verification and Validation Testing according to FDA’s Guidance “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices”
- Cybersecurity Verification and Validation Testing according to FDA’s Guidance “Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions”
Additionally, performance bench data was submitted for device performance and durability of the subject device. This data included:
- Verification testing of magnetic stimulation parameters in comparison to the predicate device
- Service life verification test supported by a post-market field study
- Stimulation coil surface temperature rise test (acceptance criterion per ASTM F2182-19E02)
- Magnetic flux density Test to evaluate maximum flux density
{7}
K252334
Page 4 of 4
All pre-determined acceptance criteria were met.
## 8. CONCLUSION
Based on the information presented in this submission, it can be concluded that the subject device is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.