K241516 · BTL Industries, Inc. · KPI · Feb 10, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K241516
Device Name
BTL-398
Applicant
BTL Industries, Inc.
Product Code
KPI · Gastroenterology, Urology
Decision Date
Feb 10, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is intended to provide entirely non-invasive electromagnetic stimulation of pelvic flor the purpose of rehabilitation of weak pelvic muscles and restoration of neuromuscular control for the treatment of male and female urinary incontinence.
Device Story
BTL-398 is a non-invasive therapeutic device for pelvic floor muscle stimulation; treats urinary incontinence in adults. System comprises main unit with electromagnetic field generators, microprocessor control, and touch-screen interface; includes Chair Applicator with integrated magnetic coil. Physician/clinician operates device in healthcare facility. Before therapy, user adjusts seat height and coil position via remote controller; Dynamic HFEM Technology (DHT) ensures optimal coil positioning. During session, Pressure Wave Technology (PWT) circulates air through seat cushion to assist patient positioning; intensity adjustable based on patient feedback. Device delivers magnetic pulses to initiate nerve action potentials, causing muscle contraction. Touch-screen guides user through procedure; displays therapy parameters and remaining time. Therapy administered through clothing. Benefits include non-invasive rehabilitation of pelvic muscles and restoration of neuromuscular control.
Clinical Evidence
Bench testing only. No clinical data presented. Testing included electrical safety (IEC 60601-1:2020), electromagnetic compatibility (IEC 60601-1-2:2020), software verification/validation, magnetic field strength, coil surface temperature, and PWT feature performance. All pre-determined acceptance criteria were met.
Technological Characteristics
Non-implanted electrical continence device; magnetic field stimulation. Energy source: 100-240V AC, 50-60Hz. Output: 1 channel, 0.7-2.5T intensity, 1-150Hz pulse rate, 280μs pulse width, dual-phase sine pulses. Features: Dynamic HFEM Technology (DHT) for coil positioning; Pressure Wave Technology (PWT) for patient positioning. Connectivity: Touch-screen control panel. Standards: IEC 60601-1, IEC 60601-1-2.
Indications for Use
Indicated for adult males and females with weak pelvic muscles and urinary incontinence requiring rehabilitation and neuromuscular control restoration.
Regulatory Classification
Identification
A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out as "U.S. Food & Drug Administration".
February 10, 2025
BTL Industries, Inc. David Chmel CEO North America 362 Elm Street Marlborough, Massachusetts 01752
Re: K241516 Trade/Device Name: Btl-398 Regulation Number: 21 CFR 876.5320 Regulation Name: Nonimplanted Electrical Continence Device Regulatory Class: II Product Code: KPI Dated: May 28, 2024 Received: January 8, 2025
Dear David Chmel:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica K. Nguyen -S
Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K241516
Device Name Btl-398
Indications for Use (Describe)
The device is intended to provide entirely non-invasive electromagnetic stimulation of pelvic flor the purpose of rehabilitation of weak pelvic muscles and restoration of neuromuscular control for the treatment of male and female urinary incontinence.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for BTL Industries. The logo consists of three interconnected squares with the letters "BTL" inside. To the right of the logo is the text "BTL Industries" in a sans-serif font.
# Traditional 510(k) Summary
# 1. Submitter Information
| Applicant | BTL Industries, Inc.<br>362 Elm Street<br>Marlborough, MA 01752<br>Tel: +1-866-285-1656<br>Fax: +1-888-499-2502 |
|------------------------|-----------------------------------------------------------------------------------------------------------------|
| Official Correspondent | David Chmel<br>BTL Industries, Inc.<br>chmel@btlnet.com |
| Date Prepared: | February 7, 2025 |
# 2. Device Name
| Trade/Proprietary Name: | BTL-398 |
|------------------------------|-------------------------------------------|
| Common Name: | BTL-398 |
| Primary Classification Name: | Nonimplanted electrical continence device |
| Classification Regulation: | 21 CFR 876.5320, Class II |
| Classification Product Code: | KPI |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of a geometric design on the left, followed by the text "BTL Industries" on the right. The geometric design is made up of three nested squares, with the letters "B", "T", and "L" inside the innermost square. The text "BTL Industries" is in a sans-serif font and is a dark gray color.
#### 3. Predicate Device Identification
K181497 HPM-6000UF BTL Industries, Inc.
The predicate device was never subjected to a design related recall.
#### 4. Device Description
The BTL-398 is a non-invasive therapeutic device that delivers magnetic pulses to stimulate the pelvic floor muscles for the treatment of urinary incontinence in adult population. The BTL-398 device is intended for professional use in healthcare facilities as a prescription use only.
The BTL-398 consists of a main unit and a Chair Applicator. The main unit consists of the electromagnetic field generators, the microprocessor-driven control unit, and the touch-screen control panel. The Chair Applicator is designed for non-invasive treatment of urinary incontinence. The magnetic coil is in the center of the seat, and its position is user-adjustable during therapy. Ongoing therapy is indicated by illuminated segments.
The Chair Applicator is designed for precise patient positioning, ensuring the most appropriate posture before and during therapy. Prior to High-Intensity Focused Electromagnetic (HIFEM) therapy, the seat height and coil position can be adjusted using an attached remote controller. The Dynamic HFEM Technology (DHT) feature ensures the coil is optimally positioned at the right location beneath the patient before starting therapy. At the start of therapy, air is circulated through the seat cushion with the Pressure Wave Technology (PWT) feature (intensity of the PWT function is set based on patient feedback). The PWT feature helps patients maintain the correct therapeutic position during therapy. Both the PWT and DHT features can be readjusted, and the PWT feature can be turned off, at any time during the session. The touch screen display guides the user step-by-step through the entire therapy procedure. During therapy, the screen displays information about the applied therapy, remaining therapy time, and main therapy parameters. Therapy is administered through clothing.
#### 5. Indications for Use
The device is intended to provide entirely non-invasive electromagnetic stimulation of pelvic floor musculature for the purpose of rehabilitation of weak pelvic muscles and restoration of neuromuscular control for the treatment of male and female urinary incontinence.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of a blue geometric design on the left and the text "BTL Industries" on the right. The geometric design is made up of three interlocking squares, with the letters "B", "T", and "L" inside the squares. The text "BTL Industries" is written in a gray, sans-serif font.
# 6. Comparison with the Predicate Device
| 510(k) number | K241516 | K181497 |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | BTL-398 | HPM-6000UF |
| Company<br>name | BTL Industries, Inc. | BTL Industries, Inc. |
| Type | Subject device | Primary predicate |
| Product Code and<br>Regulation | Gastroenterology-Urology<br>21 CFR 876.5320<br>Nonimplanted Electrical Continence<br>Device<br>KPI - Stimulator, Electrical, Non-<br>Implantable, For Incontinence | Gastroenterology-Urology<br>21 CFR 876.5320<br>Nonimplanted Electrical Continence<br>Device<br>KPI - Stimulator, Electrical, Non-<br>Implantable, For Incontinence |
| Indications for<br>Use | The device is intended to provide<br>entirely non-invasive electromagnetic<br>stimulation of pelvic floor musculature<br>for the purpose of rehabilitation of weak<br>pelvic muscles and restoration of<br>neuromuscular control for the treatment<br>of male and female urinary<br>incontinence. | HPM-6000UF is intended to provide<br>entirely non-invasive electromagnetic<br>stimulation of pelvic floor musculature<br>for the purpose of rehabilitation of<br>weak pelvic muscles and restoration of<br>neuromuscular control for the<br>treatment of male and female urinary<br>incontinence. |
| Clinical Use | Prescription Use | Prescription Use |
| Intended<br>population | Adult Male and Female | Adult Male and Female |
| Primary<br>Function | Pelvic floor muscle stimulation | Pelvic floor muscle stimulation |
| Principle of<br>Action | Initiating the action potential of nerves<br>results in muscle contraction | Initiating the action potential of<br>nerves results in muscle<br>contraction |
| Type of Energy | Magnetic field (Therapy) + air (PWT<br>feature for proper patient positioning) | Magnetic field (Therapy) |
| Energy Source | 100 - 240 V AC, 50 - 60 Hz | 100 - 240 V AC, 50 - 60 Hz |
| Number of<br>output<br>channels | 1 | 1 |
| | | |
| Magnetic Field<br>Intensity | 0.7 - 2.5 T | 0.7 - 2.5 T |
| Pulse Repetition<br>Rate | 1 – 150 Hz | 1 – 150 Hz |
| Pulse Width | 280 μς (± 20%) | 280 μς (± 20%) |
| Shape of<br>Stimulation Pulse | Dual-phase, sine pulses | Dual-phase, sine pulses |
| Therapy Time | Up to 30 min | Up to 30 min |
| PWT feature | Yes | No |
| PWT intensity | 0-100% | N/A |
| PWT repetition rate | Up to 40 Hz | N/A |
| Operating<br>Temperature | +10 to +30 °C<br>(50 - 86 °F) | +10 to +30 °C<br>(50 - 86 °F) |
| Applicator type | Chair applicator | Chair applicator |
| Main Unit<br>Dimensions<br>(W×H×D) | 580 × 1380 × 580 mm<br>(23 × 54 × 23 in) | 500 × 970 × 580 mm<br>(20 × 38 × 23 in) |
| Applicator<br>Dimensions<br>(W×H×D) | 730 x 850 x 730 mm<br>(29 x 33,5 x 29 in) | 730 x 730 x 730 mm<br>(29 x 29 x 29 in) |
| System Weight | 99 kg (216 lb) | 82 kg (181 lb) |
| Applicator height<br>adjustment | 17 to 22 in (42 to 55 cm) | 17 to 22 in (42 to 55 cm) |
| Coil adjustment<br>(DHT Feature) | Yes<br>100±5mm in horizontal direction | No |
| Therapy Modes | • Basic Mode:<br>Frequency: 1-150 Hz, Pulse: 0.01-60 s,<br>Pause: 0-60s, PWT: 1-40 Hz<br><br>• Sequence Mode:<br>○ Amplitude modulation with PWT<br>1-40 Hz<br>○ Frequency modulation with PWT<br>1-40 Hz<br><br>• Single Mode:<br>Frequency: 1-150 Hz, Number of pulses:<br>1-1000 | • Basic Mode:<br>Frequency: 1-150 Hz, Pulse: 0.01-60 s,<br>Pause: 0-60s<br><br>• Sequence Mode:<br>○ Amplitude modulation<br>○ Frequency modulation<br><br>• Single Mode:<br>Frequency: 1-150 Hz, Number of pulses:<br>1-1000 |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of a stylized graphic of three nested squares rotated to form a diamond shape, with the letters "BTL" inside the innermost square. To the right of the graphic is the text "BTL Industries" in a simple, sans-serif font.
K241516
Page 4 of 5
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three blue squares nested inside each other, with the letters "BTL" in white inside the innermost square. To the right of the square logo is the text "BTL Industries" in a gray sans-serif font.
Image /page/8/Picture/2 description: The image is a rectangular shape divided into three equal sections. The rectangle is bordered by a thick black line on the top and bottom. The three sections are separated by thin black vertical lines.
As evidenced by the above table, both the subject and the predicate devices have similar intended use, but the subject and predicate devices have different technological characteristics. The differences in technological characteristics between the subject and the predicate does not raise different questions of safety or effectiveness. Performance testing was conducted to establish substantial equivalence of the subject device to the predicate device.
# 7. Summary of Non-clinical Testing
Below is a list of the tests that were performed and successfully completed for the subject device per the specified guidance and standards:
- Electrical Safety testing, mechanical strength testing and thermal safety testing to ● IEC 60601-1: 2020 - Medical electrical equipment -Basic safety and essential performance.
- Electromagnetic Compatibility testing according to IEC 60601-1-2: 2020 General ● requirements for basic safety and essential performance -- Collateral Standard: Electromagnetic disturbances -- Requirements and tests.
- . Software Verification and Validation Testing according to FDA's Guidance (2023) . "Content of Premarket Submissions for Device Software Functions | FDA"
Additionally, performance bench data was submitted for device performance i.e., magnetic field strength testing, coil surface temperature testing/controls and PWT feature testing.
All pre-determined acceptance criteria were met.
## 8. Conclusion
Based on the information presented in this submission, it can be concluded that the subject device is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.