K181497 · BTL Industries, Inc. · KPI · Nov 14, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K181497
Device Name
HPM-6000UF
Applicant
BTL Industries, Inc.
Product Code
KPI · Gastroenterology, Urology
Decision Date
Nov 14, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
HPM-6000UF is intended to provide entirely non-invasive electromagnetic stimulation of pelvic floor musculature for the purpose of rehabilitation of weak pelvic muscles and restoration of neuromuscular control for the treatment of male and female urinary incontinence.
Device Story
HPM-6000UF is a non-invasive therapeutic device for pelvic floor muscle rehabilitation. It uses an electromagnetic field generator to induce neuromuscular tissue stimulation, triggering muscle contraction. The system consists of a console with a color touch-screen and an applicator chair containing an air-cooled coil. The device is operated by a clinician in a professional setting; the user follows on-screen, step-by-step guidance to set therapy parameters. The patient sits on the applicator chair, which can be adjusted for positioning. The electromagnetic field interacts with pelvic tissues to provide stimulation and tonization. This process aims to strengthen pelvic muscles and restore neuromuscular control, potentially benefiting patients suffering from urinary incontinence. Clinical safety regarding thermal effects on the testes in males was documented.
Clinical Evidence
Clinical study conducted to assess thermal safety in males. Results documented no meaningful rise in scrotal surface temperature during treatment, supporting safety for the male population.
Technological Characteristics
System console and applicator chair; electromagnetic field generator; air-cooled coil; color touch-screen interface; 100-240 V AC power source; magnetic field intensity 0.7–2.5 T; pulse repetition rate 1–150 Hz; pulse width 280 μs (± 20%); dual-phase rectangular pulses; firmware-controlled; moderate level of concern software.
Indications for Use
Indicated for male and female patients requiring rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control for the treatment of urinary incontinence.
Regulatory Classification
Identification
A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, with the letters "FDA" in a blue box. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 14, 2018
BTL Industries, Inc. David Chmel VP of Operations 362 Elm Street Marlborough, MA 01752
Re: K181497
> Trade/Device Name: HPM-6000UF Regulation Number: 21 CFR§ 876.5320 Regulation Name: Nonimplanted Electrical Continence Device Regulatory Class: II Product Code: KPI Dated: September 21, 2018 Received: September 26, 2018
Dear David Chmel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Glenn B. Bell -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K181497
Device Name HPM-6000UF
Indications for Use (Describe)
HPM-6000UF is intended to provide entirely non-invasive electromagnetic stimulation of pelvic floor musculature for the purpose of rehabilitation of weak pelvic muscles and restoration of neuromuscular control for the treatment of male and female urinary incontinence.
| <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|---------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Type of Use (Select one or both, as applicable)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three interconnected squares on the left side of the image. The text "BTL Industries" is on the right side of the image.
# 510(k) Summary
## General Information
| Sponsor: | BTL Industries, Inc.<br>362 Elm Street<br>Marlborough, MA 01752<br>Tel: +1-866-285-1656<br>Fax: +1-888-499-2502 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Applicant: | BTL Industries, Inc.<br>362 Elm Street<br>Marlborough, MA 01752<br>Tel: +1-866-285-1656<br>Fax: +1-888-499-2502 |
| Contact Person: | David Chmel<br>BTL Industries, Inc.<br>chmel@btlnet.com |
| Summary Preparation<br>Date: | November 13, 2018 |
#### Device Name
| Trade/Proprietary Name: | HPM-6000UF |
|------------------------------|-------------------------------------------|
| Primary Classification Name: | Nonimplanted electrical continence device |
| Common Name: | Pelvic floor muscle stimulator |
| Classification Regulation: | 21 CFR 876.5320, Class II |
| Classification Product Code: | KPI |
## Legally Marketed Predicate Devices
The HPM-6000UF is a non-invasive therapeutic device, and is substantially equivalent to the current products that are already cleared for distribution in the USA under the following 510(k) Premarket Notification numbers:
Primary predicate - HPM-6000U (K162010) Secondary predicate - InToneMV (K134020)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three blue squares, each containing the letters "BTL". The squares are arranged in a triangular pattern. To the right of the squares is the text "BTL Industries" in a gray, sans-serif font.
#### Product Description
The HPM-6000UF is a non-invasive therapeutic device. The device produces electromagnetic field that interacts with the tissues of the human body. The electromagnetic field delivered in the muscular or neuronal tissue area is triggering the stimulation and tonisation.
The HPM-6000UF is equipped with a color touch screen with wide view angle that significantly facilitates the use of the device. The on-screen information guides the user stepby-step through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen, buttons and knob on the device. During the therapy the device keeps information about the applied therapy type, remaining therapy time and main therapy parameters on the screen.
#### Intended Use
HPM-6000UF is intended to provide entirely non-invasive electromagnetic stimulation of pelvic floor musculature for the purpose of rehabilitation of weak pelvic muscles and restoration of neuromuscular control for the treatment of male and female urinary incontinence.
#### Side Effects
The intended use may be accompanied by various side effects such as muscular pain, temporary muscle spasm, temporary joint or tendon pain, temporary increase of local circulation, increased sensitivity during intercourse, and local erythema or skin redness.
#### Non-clinical Testing
The HPM-6000UF is identical in design to the HPM-6000U predicate. Therefore, no nonclinical testing was needed.
#### Clinical testing
In order to demonstrate safety of the HPM-6000UF in the treatment of males (specifically, thermal safety to the testes), a clinical study has been conducted. The clinical study documented no meaningful rise in scrotal surface temperature during treatment, and that the HPM-6000UF device is safe in the male population.
#### Technological Characteristics
The HPM-6000UF is identical in design to the HPM-6000U predicate device. The only difference relative to this predicate device is to expand the indications for use statement.
The HPM-6000UF device and its predicates are comprised of a system console and applicator. The system consists of the electromagnetic field generators, computer, and the touch-screen control panel. Applicator Chair is designed for non-invasive therapy of urinal incontinence. The coil is located in the middle of the seat and air cooled. Running
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for BTL Industries. The logo consists of three overlapping squares with the letters "BTL" inside. To the right of the squares is the text "BTL Industries" in a sans-serif font. The logo is simple and modern, and the colors are muted.
therapy is indicated by illuminated segments in armrests. For precise patient positioning prior to the therapy, the seat height can be electronically adjusted by a remote control.
The HPM-6000UF device is using a maqnetic field in order to induce neuromuscular tissue stimulation. The InToneMV predicate device is stimulating neuromuscular tissue via the direct current flow. In both cases there is an induction of electrical stimulus in the neuromuscular tissue resulting in the muscle contraction.
The HPM-6000UF device is intended for professional use as a prescription use only, while the InToneMV predicate device is intended also for a home use. The energy source of the HPM-6000UF device is electric power grid instead of battery source by the InToneMV predicate device. Further, the HPM-6000UF is equipped with larger colorful touch-screen control panel to increase user comfort.
The technological differences between the HPM-6000UF device and the predicate devices do not raise any new types of safety or effectiveness questions.
| 510(k) number | Not Assigned | K162010 | K134020 |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | HPM-6000UF | HPM-6000U | InToneMV |
| Company name | BTL Industries, Inc. | BTL Industries, Inc. | InControl Medical,<br>LLC |
| Product Code<br>and Regulation | Gastroenterology-<br>Urology Devices<br>21 CFR 876.5320<br>KPI - Stimulator,<br>Electrical, Non-<br>Implantable, For<br>Incontinence | Gastroenterology-<br>Urology Devices<br>21 CFR 876.5320<br>KPI - Stimulator,<br>Electrical, Non-<br>Implantable, For<br>Incontinence | Gastroenterology-<br>Urology Devices<br>21 CFR 876.5320<br>KPI - Stimulator,<br>Electrical, Non-<br>Implantable, For<br>Incontinence |
| Intended Use | HPM-6000UF is<br>intended to provide<br>entirely non-invasive<br>electromagnetic<br>stimulation of pelvic floor<br>musculature for the<br>purpose of rehabilitation<br>of weak pelvic muscles<br>and restoration of<br>neuromuscular control<br>for the treatment of male<br>and female urinary<br>incontinence. | HPM-6000U is intended<br>to provide entirely non-<br>invasive<br>electromagnetic<br>stimulation of pelvic<br>floor musculature for the<br>purpose of rehabilitation<br>of weak pelvic muscles<br>and restoration of<br>neuromuscular control<br>for the treatment of<br>urinary incontinence in<br>women. | InToneMV is intended<br>to provide electrical<br>stimulation and/or visual<br>biofeedback (via<br>manometry) for<br>treatment of male and<br>female urinary and fecal<br>incontinence. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three blue squares that are nested inside of each other. To the right of the squares is the text "BTL Industries" in a gray sans-serif font.
| | | K181497<br>Page 4 of 5 |
|-----------|---------------------------|------------------------|
| | 510(K) SUMMARY | |
| | | |
| | K134020 | |
| | InToneMV | |
| J | InControl Medical,<br>LLC | |
| ies, Inc. | | |
| 510(k) number | Not Assigned | K162010 | K134020 |
|---------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|
| Device name | HPM-6000UF | HPM-6000U | InToneMV |
| Company name | BTL Industries, Inc. | BTL Industries, Inc. | InControl Medical,<br>LLC |
| Clinical Use | Prescription Use | Prescription Use | Clinic or home use,<br>under direction of<br>physician |
| Applicable<br>Patients | Male and Female | Woman | Male and Female |
| Primary Function | Muscle stimulation | Muscle stimulation | Muscle stimulation |
| Principle of Action | Initiating action potential<br>of nerves results in<br>muscle contraction | Initiating action potential<br>of nerves results in<br>muscle contraction | Initiating action potential<br>of nerves results in<br>muscle contraction |
| Type of Energy | Magnetic field | Magnetic field | Electrical |
| Energy Source | 100 - 240 V AC, 50 - 60<br>Hz, max 14 A | 100 - 240 V AC, 50 -<br>60 Hz, max 14 A | 4/5 AA nickel metal<br>hydride battery |
| Number of output<br>channels | 1 | 1 | 1 |
| Magnetic Field<br>Intensity | 0.7 – 2.5 T | 0.7 – 2.5 T | N/A |
| Pulse Repetition<br>Rate | 1 – 150 Hz | 1 — 150 Hz | 0 - 50 Hz |
| Pulse Width | 280 $ μ $ s (± 20%) | 280 $ μ $ s (± 20%) | 200 $ μ $ s |
| Shape of<br>Stimulation Pulse | Dual phase, rectangular<br>pulses | Dual phase, rectangular<br>pulses | Dual phase, rectangular<br>pulses |
| Therapy Time | Up to 30 min | Up to 30 min | Up to 30 min |
| Interface | Touch-screen | Touch-screen | Buttons accompanied<br>by simple screen |
| Firmware<br>controlled | Yes | Yes | Yes |
| 510(k) number | Not Assigned | K162010 | K134020 |
| | | | |
| Device name | HPM-6000UF | HPM-6000U | InToneMV |
| | | | |
| Company name | BTL Industries, Inc. | BTL Industries, Inc. | InControl Medical,<br>LLC |
| Software Level of<br>Concern | Moderate | Moderate | Moderate |
| Environmental<br>Specifications | For indoor use only | For indoor use only | For indoor use only |
| Feedback | By Patient | By Patient | Manometric air pressure |
| Operating<br>Temperature | +10 to +30 °C<br>(50 - 86 °F) | +10 to +30 °C<br>(50 – 86 °F) | Unknown |
| Main Unit | 500 × 970 × 580 mm | 500 × 970 × 580 mm | 122 x 61 x 30 mm |
| Dimensions | | | |
| (W×H×D) | (20 × 38 × 23 in) | (20 × 38 × 23 in) | (4.8 x 2.4 x 1.1 in) |
| Applicator | 730 x 730 x 730 mm | 730 x 730 x 730 mm | 196 x 58 x 99 mm |
| Dimensions<br>(W×H×D) | (29 x 29 x 29 in) | (29 x 29 x 29 in) | (7.7 x 2.3 x 3.9 in) |
| System Weight | 46 kg (101 lb) | 46 kg (101 lb) | Unknown |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for BTL Industries. The logo consists of three overlapping blue squares with the letters "BTL" inside. To the right of the logo, the words "BTL Industries" are written in a dark gray sans-serif font. The logo is simple and modern, and the text is clear and easy to read.
510(K) SUMMARY
## Substantial Equivalence
Based upon the intended use and known technical information provided in this pre-market notification, the HPM-6000UF device has been shown to be substantially equivalent to currently marketed predicate devices.
## Conclusion
Based on the aforementioned information, the HPM-6000UF is safe and effective and substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.