Clinical evaluation of 13 spinopelvic parameters: 146 AP images, 253 LAT images.
3 (US board certified radiologists)
Pedicle screw placement
Locked AI algorithm
True Positive Rate (TPR) > 80% with 95% confidence
TPR exceeded 80% with 95% confidence
160 studies (101 3D CT scans, 59 3D O-arm scans), 942 vertebrae, 5,795 unique screw placements.
34 (Surgical support technicians)
Clinical validation: 326 screw plans from 25 spine images.
3 (experts)
Indications for Use
The Mazor X is indicated for precise positioning of surgical instruments or spinal implants during general spinal surgery. It may be used in open or minimally invasive or percutaneous procedures. Mazor X 3D imaging capabilities provide a processing and conversion of 2D fluoroscopic projections from standard C-arms into a volumetric 3D image. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of high contrast objects. The Mazor X navigation tracks the position of instruments, during spinal surgery, in relation to the surgical anatomy and identifies this position on diagnostic or intraoperative images of a patient.
Device Story
Mazor X System provides robotic trajectory guidance and optical navigation for spinal surgery. Inputs include pre-operative CT scans and intra-operative 2D fluoroscopic projections. System transforms inputs via image registration (CT-to-fluoro) and 3D volumetric reconstruction; utilizes locked AI algorithms for automated landmark detection and screw placement planning. Output is visual guidance on workstation display and robotic arm positioning for instrument/implant placement. Used in hospital operating rooms by surgeons. System includes workstation, surgical arm, navigation camera, and disposable kits. Benefits include precise instrument positioning and simplified surgical planning. Modifications include software enhancements for registration/planning and compatibility with additional Medtronic tools.
Clinical Evidence
Bench testing and clinical validation performed. 2D Automatic Measurements validated using 146 AP and 253 LAT images; ICC > 0.75 (p<0.05). Plan Assist validated using 326 screw plans from 25 spine images; True Positive Rate > 80%. Datasets were independent from development data with site-level separation. No clinical outcomes data (e.g., patient recovery) provided; evidence focuses on technical accuracy of AI-assisted planning features.
Technological Characteristics
Stereotaxic instrument; robotic arm with 6-axis positioning; optical navigation (NDI Vega Polaris camera). Materials: Ultem 1000HU, Carbon Fiber Zacton 350, Ketron LSG PEEK, Stainless Steel, Aluminum 6061, Polyethylene, Medikote C12. Connectivity: Ethernet. Software: Locked ML-based algorithms for landmark detection and screw planning. Sterilization: Steam (reusable), Gamma (disposable), ETO (covers). Compliant with IEC 60601-1, IEC 60601-1-2, IEC 62304, and IEC 60825-1.
Indications for Use
Indicated for precise positioning of surgical instruments or spinal implants during general spinal surgery in orthopedic patients. May be used in open, minimally invasive, or percutaneous procedures. Contraindications include unique spinal conditions such as revisions, fractures, tumors, or hemivertebrae for specific AI-enabled features.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
Mazor X System (Mazor X Stealth Edition) (K230064)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 11, 2025
Mazor Robotics Ltd.
Anat Marom
Senior Regulatory Affairs specialist
1 HaEshel Street (Building C), Caesarea Buisness Park
Caeasarea, 3079830
Israel
Re: K251316
Trade/Device Name: Mazor X System / Mazor X Stealth Edition
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: OLO, LLZ
Dated: August 11, 2025
Received: August 11, 2025
Dear Anat Marom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251316 - Anat Marom
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251316 - Anat Marom
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tejen D. Soni -S
For
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Mazor X System/ Mazor X Stealth Edition
Page 1 of 1
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251316 | ? |
| Please provide the device trade name(s). | | ? |
| Mazor X System/ Mazor X Stealth Edition | | |
| Please provide your Indications for Use below. | | ? |
| The Mazor X is indicated for precise positioning of surgical instruments or spinal implants during general spinal surgery. It may be used in open or minimally invasive or percutaneous procedures. | | |
| Mazor X 3D imaging capabilities provide a processing and conversion of 2D fluoroscopic projections from standard C-arms into a volumetric 3D image. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of high contrast objects. | | |
| The Mazor X navigation tracks the position of instruments, during spinal surgery, in relation to the surgical anatomy and identifies this position on diagnostic or intraoperative images of a patient. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Medtronic Engineering the extraordinary
510(k) Summary – K251316
Applicant name: Mazor Robotics Ltd.
1 HaEshel Street (Building C)
Caesarea business Park 3079830, Israel
Contact person: Name: Anat Marom
Role: Sr. Regulatory Affairs Specialist
Email: anat.marom@medtronic.com
Phone: +972-506874994
Date Prepared: September 11, 2025
Name of Device: Mazor X Stealth Edition
Classification Name: Stereotaxic instrument
Classification Code: OLO (primary) and LLZ
Device class: II
Regulation number: 21 CFR 882.4560
Panel: Orthopedic
Predicate Devices: Mazor X System (Mazor X Stealth Edition) (K230064)
## Intended Use / Indications for Use
The Mazor X is indicated for precise positioning of surgical instruments or spinal implants during general spinal surgery. It may be used in open or minimally invasive or percutaneous procedures. Mazor X 3D imaging capabilities provide a processing and conversion of 2D fluoroscopic projections from standard C Arms into volumetric 3D image. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of high contrast objects.
The Mazor X navigation tracks the position of instruments, during spinal surgery, in relation to the surgical anatomy and identifies this position on diagnostic or intraoperative images of a patient.
K251316, Page 1 of 16
{5}
Medtronic Engineering the extraordinary
# Device Description
The Mazor X system combines robotic trajectory guidance with navigated surgical instruments (either guided or free hand navigation) to enable the surgeon to precisely position surgical instruments and/or implants according to predefined planning. With the imaging capabilities of the system, the user can also visualize the implants on the patients CT. Same as the predicate device the modified Mazor X consists of a workstation with dedicated software, the surgical system, navigation camera, accessories, instruments and disposable kits. The modified Mazor X, the subject of this 510(k) application, introduces software and hardware modifications to the Mazor X System cleared in 510(k) K230064.
# Technological Characteristics
The modified Mazor X device is similar in its technological features to its predicate device, the cleared Mazor X. Both the subject and predicate systems combine robotic and navigation technologies to enable the precise positioning and tracking of surgical instruments or spinal implants during general spinal surgery. Both systems allow registration of image data with patient anatomy by registering the pre-operative CT scan to the intra-operative fluoroscopic scans, and by localizing the position of the 3D patient volume to the robotic system. In both systems, the positioning of surgical instruments and their trajectories are guided by the system in accordance with the planning conducted by the surgeon on the pre-operative CT image. In both systems, the navigation feature provides the option of tracking the surgical instruments during the procedure. Both systems include very similar hardware and software components, including the workstation, the surgical system, navigation camera, accessories, instruments, and disposable kits. The modifications to the cleared Mazor X System, that are the subject of this premarket application, include software enhancements to enable extended functionality (in registration and planning stages), compatibility with additional, previously cleared, surgical tools and some minor hardware changes.
The minor differences in the technological characteristics between the modified Mazor X and its predicate device do not raise different questions of safety or efficacy. The performance bench testing and Human Factors validation, provided with this premarket application, demonstrate that the modified Mazor X is as safe and effective as its predicate devices.
K251316, Page 2 of 16
{6}
Medtronic
Engineering the extraordinary
Comparison of the Proposed Mazor X to the Cleared Mazor X System (K230064) is provided below:
Table 1: Comparison of the Proposed Mazor X to the Cleared Mazor X System (K230064)
| Technological Characteristic | Mazor X System
Mazor Robotics Ltd. | Mazor X System
Mazor Robotics Ltd.
K230064 (predicate device) | Comparison |
| --- | --- | --- | --- |
| Product Code, Class | OLO and LLZ,
Class II | OLO and LLZ,
Class II | Identical |
| Indications for Use | The Mazor X is indicated for precise positioning of surgical instruments or spinal implants during general spinal surgery. It may be used in open or minimally invasive or percutaneous procedures.
Mazor X 3D imaging capabilities provide a processing and conversion of 2D fluoroscopic projections from standard C-arms into a volumetric 3D image. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of high contrast objects.
The Mazor X navigation tracks the position of instruments, during spinal surgery, in relation to the surgical anatomy and identifies this position on diagnostic or intraoperative images of a patient. | The Mazor X is indicated for precise positioning of surgical instruments or spinal implants during general spinal surgery. It may be used in open or minimally invasive or percutaneous procedures.
Mazor X 3D imaging capabilities provide a processing and conversion of 2D fluoroscopic projections from standard C-arms into a volumetric 3D image. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of high contrast objects.
The Mazor X navigation tracks the position of instruments, during spinal surgery, in relation to the surgical anatomy and identifies this position on diagnostic or intraoperative images of a patient. | Identical |
K251316, Page 3 of 16
{7}
Medtronic
Engineering the extraordinary
| Technological Characteristic | Mazor X System
Mazor Robotics Ltd. | Mazor X System
Mazor Robotics Ltd.
K230064 (predicate device) | Comparison |
| --- | --- | --- | --- |
| Target Population | Orthopedic patients | Orthopedic patients | Identical |
| Anatomical Sites | Spine | Spine | Identical |
| Environment Used | Hospital setting (operating room) | Hospital setting (operating room) | Identical |
| Main system components | The Mazor X System consists of the following components:
• Workstation
• Surgical System 3Define Camera: SR300 model
• Bed Connecting Unit (e.g., Bed Frame)
• Device accessories for spine application
• Mazor X Navigation Camera and accessories | The Mazor X System consists of the following components:
• Workstation
• Surgical System 3Define Camera: SR300 model
• Bed Connecting Unit (e.g., Bed Frame)
• Device accessories for spine application
• Mazor X Navigation Camera and accessories | Similar.
X-Eye camera was removed, and its function to intra-operatively position the Surgical Arm for fluoro acquisition purposes was replaced by “Drag AP DOT” feature. |
| Mechanism of Action | Computer assisted Stereotaxy: Instrument position and trajectory calculation based on image data & instrument tracking based on optical navigation. Motorized positioning of the Surgical Arm (spine) with tool guide through 6 axes. | Computer assisted Stereotaxy: Instrument position and trajectory calculation based on image data & instrument tracking based on optical navigation. Motorized positioning of the Surgical Arm (spine) with tool guide through 6 axes. | Identical |
K251316, Page 4 of 16
{8}
Medtronic
Engineering the extraordinary
| Technological Characteristic | Mazor X System
Mazor Robotics Ltd. | Mazor X System
Mazor Robotics Ltd.
K230064 (predicate device) | Comparison |
| --- | --- | --- | --- |
| Compatibility with Medtronic authorized instruments and tools | Mazor X is compatible with the following Medtronic tools and instruments:
• Medtronic Spine instruments (K182121)
• Medtronic Navigation NavLock Trackers (K182104)
• Medtronic Stealth-Midas and Stealth Midas MR8 navigated drill systems (K160713, K183644)
• Medtronic navigated surgical instruments - Trials trackers, Inserters and Disc Prep Instruments (K131425, K163581, K150231) and Medtronic Pedicle Probe (K050438)
• Medicrea UNiD Analyzer (K212005)
• Stealth-Midas Rex Drill System (K202552)
• Medtronic Navigated Anterolateral Disc Prep Instruments (K211441)
• Medtronic Adaptix Interbody System with Titan nanoLOC Surface Technology (K201267)
• Medtronic CD HORIZON™ Spinal System (K211596)
• Medtronic Catalyft PL (K214011) | Mazor X is compatible with the following Medtronic tools and instruments:
• Medtronic Spine instruments (K182121)
• Medtronic Navigation NavLock Trackers (K182104)
• Medtronic Stealth-Midas and Stealth Midas MR8 navigated drill systems (K160713, K183644)
• Medtronic navigated surgical instruments - Trials trackers, Inserters and Disc Prep Instruments (K131425, K163581, K150231) and Medtronic Pedicle Probe (K050438)
• Medicrea UNiD Analyzer (K212005)
• Stealth-Midas Rex Drill System (K202552)
• Medtronic Navigated Anterolateral Disc Prep Instruments (K211441)
• Medtronic Adaptix Interbody System with Titan nanoLOC Surface Technology (K201267)
• Medtronic CD HORIZO NTM Spinal System (K211596) | Similar.
Compatibility with additional, cleared to market, Medtronic authorized instruments and tools using the same navigation technology, enabling tracking the position of the instruments in relation to the surgical anatomy and identify this position on the patient images.
The compatibility with additional, cleared to market, tools, does not change the system indications for use, design or its principles of operation and do not raise new questions of safety and efficacy. |
K251316, Page 5 of 16
{9}
Medtronic
Engineering the extraordinary
| Technological Characteristic | Mazor X System
Mazor Robotics Ltd. | Mazor X System
Mazor Robotics Ltd.
K230064 (predicate device) | Comparison |
| --- | --- | --- | --- |
| | • Medtronic Anteralign TL and Voyager FNS (K214011)
• Anteralign LS (K222383), Capstone PTC implants and Enhanced (size 36, length of 36 mm) Capstone implants (K172199).
• Anteralign™ Spinal System with Titan nanoLOCK™ Surface Technology (K212524)
• ModuLeX FAS drivers, FAS screws and Osteogrip Shanks (K231184)
• ModuLeX FNS Screw set (K232141)
• ModuLeX Degen Scoli (K233951)
• Adaptix™ PEEK Interbody System (K241605) | • Medtronic Catalyft PL (K214011)
• Medtronic Anteralign TL and Voyager FNS (K214011)
• Anteralign LS (K222383), Capstone PTC implants and Enhanced (size 36, length of 36 mm) Capstone implants (K172199). | |
| Features | The Mazor X System consists of the following features:
• Preoperative planning and operation
• Advanced 3D Visualization (volume rendering)
• Robotic guidance and optical navigation of surgical tools | The Mazor X System consists of the following features:
• Preoperative planning and operation
• Advanced 3D Visualization (volume rendering)
• Robotic guidance and optical navigation of surgical tools | Identical |
| Imaging Modalities | • CT and Fluoro based imaging
• X-ray based imaging (Planning) | • CT and Fluoro based imaging
• X-ray based imaging (planning) | Identical |
K251316, Page 6 of 16
{10}
Medtronic
Engineering the extraordinary
| Technological Characteristic | Mazor X System
Mazor Robotics Ltd. | Mazor X System
Mazor Robotics Ltd.
K230064 (predicate device) | Comparison |
| --- | --- | --- | --- |
| Registration Features | • CT-Fluoro Merge Registration
• Automatic 3D Image Registration (Scan and Plan, Automatic Registration) | • CT-Fluoro Merge Registration
• Automatic 3D Image Registration (Scan and Plan) | Similar.
The Automatic Registration with O-arm is an enhancement to the Automatic 3D Image Registration process. |
| Planning Features | • Plan Entry and Target Selection
• 3D Model Building | • Plan Entry and Target Selection
• 3D Model Building | Similar.
Three enhancements were introduced to the Plan Entry and Target Selection capabilities:
1. 2D Automatic Measurements: incorporates a locked machine learning based algorithm that automatically detects anatomical landmarks, used for calculating anatomical measurements, for the automatic pre-op planning capability.
2. Plan Assist feature, a locked machine learning based algorithm, enhances the initial screw placement by considering user preferences and clinical constrains, thus providing clinically acceptable initial screw placement.
3. Support for additional pre-bent rods type - Medtronic CD |
K251316, Page 7 of 16
{11}
Medtronic
Engineering the extraordinary
| Technological Characteristic | Mazor X System
Mazor Robotics Ltd. | | Mazor X System
Mazor Robotics Ltd.
K230064 (predicate device) | | Comparison |
| --- | --- | --- | --- | --- | --- |
| | | | | | Horizon (ModuleX) short pre-bent rods for planning and repositioning screws also in the Scan & Plan workflow, in addition to CT-FL workflow. All three (3) planning enhancements described above do not change the system’s indications for use, have no effect on system’s accuracy and do not raise new safety and effectiveness issues. |
| Medical Device Interfaces | • O- arm Imaging System
• 2D C-Arm
• 3D C-Arm | | • O- arm Imaging System
• 2D C-Arm
• 3D C-Arm | | Identical |
| Dimensions | Workstation | 1.85 x 1.20 x 0.64 m (6.07 x 3.94 x 2 ft) | Workstation | 1.85 x 1.20 x 0.64 m (6.07 x 3.94 x 2 ft) | Similar
The removal of the X-eye camera (a component of the surgical system) has a negligible impact on the system dimensions. |
| | Surgical System | 0.92 x 0.63 x 0.58 m (3.02 x 2.07 x 1.9 ft) | Surgical System | 0.92 x 0.63 x 0.58 m (3.02 x 2.07 x 1.9 ft) | |
| | Mazor X Navigation Camera (NDI Vega Polaris) | 1.9 x 0.65 x 0.61 m (6.2x2.1 x 2 ft) | Mazor X Navigation Camera (NDI Vega Polaris) | 1.9 x 0.65 x 0.61 m (6.2x2.1 x 2 ft) | |
| Weight | • Workstation - 185 kg (408 lbs) | | • Workstation - 185 kg (408 lbs) | | Similar |
K251316, Page 8 of 16
{12}
Medtronic
Engineering the extraordinary
| Technological Characteristic | Mazor X System
Mazor Robotics Ltd. | Mazor X System
Mazor Robotics Ltd.
K230064 (predicate device) | Comparison |
| --- | --- | --- | --- |
| | • Surgical System – 25 kg (55.1 lbs)
• Mazor X Navigation Camera (NDI Vega Polaris) - 75Kg (165.3lbs) | • Surgical System – 25 kg (55.1 lbs)
• Mazor X Navigation Camera (NDI Vega Polaris) - 75Kg (165.3lbs) | The removal of the X-eye camera (a component of the surgical system) has a negligible impact on the system weight. |
| Performance | • Mazor X System mean accuracy <1.5mm
• Navigation Accuracy mean positional error <2mm and mean trajectory error of 2°.
• Mazor X will also provide Facet Decortication depth accuracy within 1.5mm. | • Mazor X System mean accuracy <1.5mm
• Navigation Accuracy mean positional error <2mm and mean trajectory error of 2°.
• Mazor X will also provide Facet Decortication depth accuracy within 1.5mm. | Identical |
| Human Factors | The Mazor X System simplifies the planning and surgical procedure (pre-intra-operative planning) and provides the user with additional imaging modalities (3D Define Scan, Auto Segmentation Process, etc.). Mazor X Planning facilitates performing measurements and calculating angles within a specific spinal region of interest, in accordance with standard classifications. | The Mazor X System simplifies the planning and surgical procedure (pre-intra-operative planning) and provides the user with additional imaging modalities (3D Define Scan, Auto Segmentation Process, etc.). Mazor X Align facilitates performing measurements and calculating angles within a specific spinal region of interest, in accordance with standard classifications. | Identical |
K251316, Page 9 of 16
{13}
Medtronic
Engineering the extraordinary
| Technological Characteristic | Mazor X System
Mazor Robotics Ltd. | Mazor X System
Mazor Robotics Ltd.
K230064 (predicate device) | Comparison |
| --- | --- | --- | --- |
| Cybersecurity | Mazor X system contains external wired communication interface (ethernet). Mazor X system is developed in accordance with the Cybersecurity design controls to ensure Mazor X Cybersecurity and maintain safety and effectiveness, similarly to the currently cleared device. | Mazor X system contains external wired communication interface (ethernet). Mazor X system is developed in accordance with the Cybersecurity design controls to ensure Mazor X Cybersecurity and maintain safety and effectiveness, similarly to the currently cleared device. | Identical |
| Standards Met | IEC 62304 | IEC 62304 | Identical |
| Materials | • Ultem 1000HU
• Carbon Fiber Zacton 350
• Ketron LSG PEEK Natural Black/TECAPEEK
• Stainless Steel
• Aluminum 6061
• Polyethylene
• MedikoteTM C12 (Me-DLC) | • Ultem 1000HU
• Carbon Fiber Zacton 350
• Ketron LSG PEEK Natural Black/TECAPEEK
• Stainless Steel
• Aluminum 6061
• Polyethylene
• MedikoteTM C12 (Me-DLC) | Identical |
| Biocompatibility | Materials are biocompatible | Materials are biocompatible | Identical |
| Compatibility With the Environment and Other Devices | The Mazor X is compliant with the IEC 60601-1-2 (EMC Compatibility) standard. | The Mazor X is compliant with the IEC 60601-1-2 (EMC Compatibility) standard. | Identical |
K251316, Page 10 of 16
{14}
Medtronic
Engineering the extraordinary
| Technological Characteristic | Mazor X System
Mazor Robotics Ltd. | Mazor X System
Mazor Robotics Ltd.
K230064 (predicate device) | Comparison |
| --- | --- | --- | --- |
| Sterility | No new sterilization methods or products have been introduced in the Mazor X. The reusable accessories are sterilized by steam sterilization, single-use disposable kits by gamma radiation, and sterile covers (e.g., Sterile Sleeve for Surgical Arm) by ETO sterilization. Mazor X navigation tools are sterilized with the Medtronic Sofamor Danek spine instruments(K182121) and Medtronic Navigation NavLock tools (K182104). | No new sterilization methods or products have been introduced in the Mazor X. The reusable accessories are sterilized by steam sterilization, single-use disposable kits by gamma radiation, and sterile covers (e.g., Sterile Sleeve for Surgical Arm) by ETO sterilization. Mazor X navigation tools are sterilized with the Medtronic Sofamor Danek spine instruments(K182121) and Medtronic Navigation NavLock tools (K182104). | Identical |
| Electrical Safety | • Power Requirements:
• 110-120 VAC, 60 Hz
• 220-240 VAC, 50 Hz
• 100 VAC, 50/60 Hz
• Maximum Power = 1000 VA
The Mazor X System is compliant with the IEC 60601-1 (Electrical Safety) std. | • Power Requirements:
• 110-120 VAC, 60 Hz
• 220-240 VAC, 50 Hz
• 100 VAC, 50/60 Hz
• Maximum Power = 1000 VA
The Mazor X System is compliant with the IEC 60601-1 (Electrical Safety) std. | Identical |
| Mechanical Safety | The Mazor X System is compliant with the IEC 60601-1 (Electrical Safety) standard. | The Mazor X System is compliant with the IEC 60601-1 (Electrical Safety) standard. | Identical |
| Chemical Safety | Not Applicable | Not Applicable | Identical |
| Thermal Safety | The Mazor X System is compliant with the IEC 60601-1 standard. | The Mazor X System is compliant with the IEC 60601-1 standard. | Identical |
| Radiation Safety | The Mazor X System is compliant with the IEC 60601-1-2 (EMC) standard. | The Mazor X System is compliant with the IEC 60601-1-2 (EMC) standard. | Identical |
K251316, Page 11 of 16
{15}
Medtronic
Engineering the extraordinary
| Technological Characteristic | Mazor X System
Mazor Robotics Ltd. | Mazor X System
Mazor Robotics Ltd.
K230064 (predicate device) | Comparison |
| --- | --- | --- | --- |
| Laser Safety | The Mazor X System is compliant with IEC 60825-1 (2014) (Safety of laser) standard. | The Mazor X System is compliant with IEC 60825-1 (2014) (Safety of laser) standard. | Identical |
K251316, Page 12 of 16
{16}
Medtronic
SECTION 5.0
510(k) SUMMARY
# Performance Testing
The following testing was conducted to evaluate the device:
a) Software major level of concern verification and validation testing was conducted as required by IEC 62304 and FDA Guidance on General Principles of Software Validation, January 11, 2002.
b) Bench testing was conducted to verify that the design outputs meet the design inputs. The functionality of the new or modified features was tested to ensure that system requirements, software requirements and user needs were met as defined. Specifically, the Automatic registration feature and the new machine-learning based features (2D Automatic Measurements and Plan Assist) and the X-Eye camera removal were verified and validated through testing.
# AI-enabled device software functions
## 1. 2D Automatic Measurements
The software incorporates the 2D Automatic Measurements feature which automatically detects key spinopelvic landmarks and produces spinopelvic measurements on AP and LAT X-rays images. The feature utilizes two locked AI algorithms: the Endplate detection algorithm and the Femoral head detection algorithm.
Table 2: 2D Automatic Measurements algorithms development: demographic, technical and reference standard characteristics
| | Endplate detection development | Femoral heads detection development |
| --- | --- | --- |
| Sample size per view | 2327 AP images, 2651 LAT images. | 2233 LAT images |
| Gender | 42% female, 32% male, 26% unknown | 45% female, 29% male, 26% unknown |
| Age groups | 10% Children & adolescents, 31% Adults, 27% Geriatric, 32% unknown | 16% Children & adolescents, 25% Adults, 26% Geriatric, 33% unknown |
| Image quality distribution | 19% good, 32% medium, 49% poor | 13% good, 24% medium, 63% poor |
| Image quality definition | Image quality was categorized as Good, Medium, or Poor based on contrast, bone delineation, and artifact/noise levels. Images rated Poor were excluded from the validation dataset. | |
| Reference standard type & process | Endplate line and Circle femoral heads annotations were performed by trained labelers within a quality-controlled environment. In case of multiple annotations, they were aggregated. More than 65% of the data was reviewed by U.S. certified spine surgeons and radiologists. | |
K251316, Page 13 of 16
{17}
Medtronic
SECTION 5.0
510(k) SUMMARY
A non-clinical evaluation of the endplate and femoral heads detection accuracy was conducted as part of the verification effort.
A clinical evaluation of 13 spinopelvic parameters' measurements accuracy was conducted as part of the validation effort.
Table 3: 2D Automatic Measurements algorithms clinical testing process: demographic, technical, reference standard and statistical analysis characteristics
| | Spinopelvic parameters' measurements accuracy |
| --- | --- |
| Sample size per view | 146 AP images, 253 LAT images (for all spinopelvic parameters). Each spinopelvic parameter had its own sample size ranging from 24 images to 126 images. |
| Gender | 67% female, 30% male, 3% unknown |
| Age groups | 43% Children & adolescents, 33% Adults, 13% Geriatric, 11% unknown |
| Image quality distribution | ~30% good, ~70% medium (average across all spinopelvic parameters). |
| Image quality definition | Image quality was categorized as Good, Medium, or Poor based on contrast, bone delineation, and artifact/noise levels. Images rated Poor were excluded from the validation dataset. |
| Reference standard type & process | Each spinopelvic parameter annotations were comprised of endplate line annotations, endplate endpoint annotations, and/or femoral head circle annotations performed by Three U.S board certified radiologists using the Mazor X device and compared to the AI measurements value. |
| Test statistics | 1. Single measurement absolute agreement intraclass correlation coefficient (ICC) was used to assess agreement the AI measurement results and the reference standard. The acceptance criterion of the ICC of least 0.75 with a 95% confidence level was achieved for all spinopelvic measurements. 2. An F test was conducted to statistically demonstrate that the ICC was statistically larger than 0.75. For all spinopelvic measurements, the resulting p-value was below 0.05. 3. Bland Altman analysis – for all spinopelvic measurements, the 95% limits of agreement were in range of a defined clinical range. |
Non-clinical verification and clinical validation datasets were independent from the development dataset, with data source site-level separation to ensure data independence.
Out of scope for the 2D Automatic Measurements Feature: implants or foreign objects, images with a quality rank of "poor" or "very poor", and unique spinal conditions (revisions, fractures, tumors, hemivertebrae).
K251316, Page 14 of 16
{18}
Medtronic
SECTION 5.0
510(k) SUMMARY
# 2. Plan Assist
The software incorporates the Plan Assist feature, which automatically suggests initial pedicle screw positions during spinal surgery planning using locked AI algorithm
Table 4: Plan Assist algorithm development process: demographic, technical and reference standard Characteristics
| | Plan Assist Algorithm Development (Training + Tuning) |
| --- | --- |
| Sample size | The development dataset contains 160 studies (101 3D CT scans, 59 3D O-arm scans), 942 vertebrae, 5,795 unique screw placements and over 23,000 screw plannings. |
| Gender | 25% female, 11% male, 64% unknown |
| Age groups | 14% children and adolescents, 13% Adults, 9% Geriatric, 64% unknown |
| Image quality distribution | 47% good, 52% medium, 1% poor. |
| Image quality definition | Image quality was categorized as Good, Acceptable, Poor, or Very Poor based on contrast, bone delineation, and presence of artifacts and noise. Images rated Very Poor were excluded from the training dataset. |
| Reference standard type & process | Annotations for Pedicle screw planning were performed by 34 qualified Surgical support technicians. The reference Screw planning (“annotation”) was manually performed using a released Mazor X software. |
A clinical evaluation of screw placement was conducted as part of the validation effort.
Table 5: Plan Assist algorithm clinical testing - demographic, technical, reference standard, and statistical analysis characteristics
| | Plan Assist Algorithm Testing (Clinical Validation) |
| --- | --- |
| Sample size per view | 326 screw plans from 25 spine images. |
| Gender | 28% female, 12% male, 60% unknown |
| Age groups | 16% children and adolescents, 12% Adults, 12% Geriatric, 60% unknown |
| Image quality distribution | 56% good, 44% medium |
| Image quality definition | Image quality was classified as Good, Acceptable, Poor, or Very Poor based on contrast, bone delineation, and artifact/noise levels. Images rated Poor or Very Poor were excluded from the validation dataset. |
| Test statistics | Each AI-generated screw was evaluated by three experts for clinical acceptability. The True Positive Rate (TPR) exceeded 80% with 95% confidence, meeting the acceptance criteria. |
Development and validation datasets were separated at the site level to ensure data independence. Out of scope for the Plan Assist Feature: O-Arm scans in the range of T1-T8, implants or foreign objects of any kind, unique spinal conditions (revisions, fractures, tumors, hemivertebrae), and images with a quality rank of "poor" or "very poor".
K251316, Page 15 of 16
{19}
Medtronic
SECTION 5.0
510(k) SUMMARY
# Conclusion
The modified Mazor X is as safe and effective as the cleared Mazor X. The modified Mazor X has the same intended use and indications for use and similar technological characteristics, and principles of operation as its predicate device. The minor technological differences between the modified Mazor X and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the modified Mazor X is as safe and effective as its predicate device. Thus, the Mazor X is substantially equivalent.
K251316, Page 16 of 16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.