The Navigated Disc Preparation Instruments are intended to facilitate a discectomy during spinal surgery. The Navigated CAPSTONE™ Trials are intended to be used to facilitate implant size selection of Medtronic intervertebral body fusion devices during spinal surgery. When the Navigated Rotating Shavers are used as trials for ELEVATE™ Spinal System, they are intended to be used to facilitate implant size selection. The Navigated Probe is intended to be used during pedicle and disc preparation during spinal fusion procedures. The Navigated Inserters are intended to be used for the placement of an implant. Navigated instruments are specifically designed for use with the StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
Device Story
Navigated surgical instruments (inserters, shavers, probes, trials) used during spinal surgery; designed for use with StealthStation® System. Instruments feature optical tracking markers; system tracks position relative to patient anatomy (CT/MR/fluoroscopy). Surgeon uses instruments to perform discectomy, prepare pedicles/discs, and place implants. Navigation provides real-time visual feedback on instrument position within surgical site; assists surgeon in accurate implant sizing and placement. Benefits include improved surgical precision and guidance during spinal procedures.
Clinical Evidence
Bench testing only. Testing included navigation accuracy analysis (2D/3D), anatomical simulated use, navigation simulated use, CAD model evaluation, implant/instrument mating verification, and software interface testing (Spine Tools package).
Technological Characteristics
Materials: stainless steel, silicone, aluminum (case/trays), nylon (brackets). Principle: optical navigation via StealthStation® System. Form factor: reusable surgical instruments. Connectivity: compatible with S7/Synergy Spine 2.1 application. Sterilization: non-sterile (reusable).
Indications for Use
Indicated for patients undergoing spinal surgery requiring discectomy, pedicle/disc preparation, or spinal fusion procedures where stereotactic navigation is appropriate.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 14, 2017
Medtronic Sofamor Danek USA, Inc Laveeda Leflore Regulatory Affairs Specialist 1800 Pyramid Place Memphis. Tennessee 38132
Re: K163581
Trade/Device Name: Navigated Elevate Inserter and Navigated Disc Prep Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: December 19, 2016 Received: December 20, 2016
Dear Laveeda Leflore:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163581
### Device Name
Navigated ELEVATE™ Inserter and Navigated Disc Prep Instruments.
### Indications for Use (Describe)
The Navigated Disc Preparation Instruments are intended to facilitate a discectomy during spinal surgery. The Navigated CAPSTONE™ Trials are intended to be used to facilitate implant size selection of Medtronic intervertebral body fusion devices during spinal surgery. When the Navigated Rotating Shavers are used as trials for ELEVATE™ Spinal System, they are intended to be used to facilitate implant size selection.
The Navigated Probe is intended to be used during pedicle and disc preparation during spinal fusion procedures. The Navigated Inserters are intended to be used for the placement of an implant.
Navigated instruments are specifically designed for use with the StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
| FORM FDA 3881 (8/14) | | |
|----------------------|--|--|
| | | |
Page of Page 1 of 1.
{3}------------------------------------------------
# 510(k) Summarv December 19, 2016
- l. Company: Medtronic Sofamor Danek, USA Inc. 1800 Pyramid Place Memphis. TN 38132 Telephone: (901) 396-3133
- Laveeda Leflore Contact: Regulatory Affairs Specialist Telephone: (901) 396-3133 Email: Laveeda.r.leflore@medtronic.com
- II. Proprietary Trade Name: Navigated ELEVATE™ Inserter and Navigated Disc Prep Instruments.
Common Name: Stereotaxic Instrument, Navigated Instruments
Classification Name: Stereotaxic Instrument (21 CFR 882.4560)
Classification: Class II
Product Code: OLO
#### III. Predicate Device:
- Navigated Disc Prep Instruments (K150231, S.E. 06/16/2015) primary predicate ●
- . Navigated Interbody Trials and Inserter (K131425, S.E. 08/14/2013) - secondary predicate
- ELEVATE™ Spinal System (K142559, S.E. 06/09/2015) - secondary predicate
The predicate devices have not been subject to a design-related recall.
#### IV. Product Description:
The Navigated ELEVATE™ Inserter and navigated rotating shavers are non-sterile, reusable surgical instruments that are intended to be used to facilitate a discectomy and placement of implants during spinal surgery procedure. The subject devices are designed specifically to be used with ELEVATE™ Spinal System and are also compatible with the StealthStation. These devices are offered in non-sterile form and are reusable. This subject devices are manufactured from stainless steel and silicone. The case and trays are manufactured from aluminum with either silicone or nylon coated brackets.
#### V. Indications for Use:
{4}------------------------------------------------
The Navigated Disc Preparation Instruments are intended to be used to facilitate a discectomy during spinal surgery. The Navigated CAPSTONE™ Trials are intended to be used to facilitate implant size selection of Medtronic intervertebral body fusion devices during spinal surgery. When the Navigated Rotating Shavers are used as trials for ELEVATE™ Spinal System, they are intended to facilitate implant size selection.
The Navigated Probe is intended to be used during pedicle and disc preparation during spinal fusion procedures. The Navigated Inserters are intended to be used for the placement of an implant.
Navigated instruments are specifically designed for use with the StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
#### VI. Comparison of the Technological Characteristics with the Predicate Device:
The subject Navigated ELEVATE™ Inserter is intended to be used for the placement of implants. The subject navigated rotating shavers will have a dual indication. They can be used as shavers to facilitate a discectomy or as trials to facilitate implant size (height) selection of the ELEVATE™ Spinal System implants during a spinal surgery procedure. These instruments are specifically designed for use with the StealthStation® System , which allows for optical navigation of the surgical instruments. These devices have similar designs and uses to the predicate devices and incorporate the same design features to enable navigation capabilities. Like the predicate devices, the subject Navigated ELEVATE™ Inserter and navigated rotating shavers are also made from stainless steel.
The instrument modifications detailed in this submission have no impact on the technological characteristics of either the existing instruments or the StealthStation® System.
{5}------------------------------------------------
#### VII. Performance Data
Testing was completed to ensure the functionality and compatibility with the identified Medtronic products. The following table summarizes the performance testing completed:
| Test | Description |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Navigation<br>Accuracy Analysis | Confirmed navigated instrument accuracy in both 2D and 3D space. |
| Anatomical<br>Simulated Use | Confirmed instrument functionality under expected use conditions. |
| Navigation<br>Simulated Use | Confirmed navigation system functionality under expected use conditions. |
| CAD Model<br>Evaluation | Verified that the CAD models are accurately reflected in the application software. |
| Implant/Instrument<br>Mating Conditions | Verified that the instruments can be assembled with the appropriate devices according to their intended use. |
| Spine Tools<br>Package<br>Functional Testing | Verified that the Spine Tools package has met the required interface needs of the spine application software. |
Table 5-1: Description of Performance Testing
The Navigated ELEVATE™ Inserter and the rotating shavers will use Tools Package Version 25, which will be compatible with the Synergy Spine 2.1 application on the S7.
#### VIII. Conclusions
That Navigated ELEVATE™ Inserter and the navigated rotating shavers have been shown through comparison and testing to be substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.