2008T HD SYS. CDX BLUESTAR (191124); 2008T HD SYS. CDX W/bibag BLUESTAR (191126); 2008T HD SYS. W/O CDX BLUESTAR (191128); 2008T HD SYS. W/O CDX W/bibag BLUESTAR (191130)
K243237 · Fresenius Medical Care Renal Therapies Group, LLC · KDI · Jun 13, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K243237
Device Name
2008T HD SYS. CDX BLUESTAR (191124); 2008T HD SYS. CDX W/bibag BLUESTAR (191126); 2008T HD SYS. W/O CDX BLUESTAR (191128); 2008T HD SYS. W/O CDX W/bibag BLUESTAR (191130)
Applicant
Fresenius Medical Care Renal Therapies Group, LLC
Product Code
KDI · Gastroenterology, Urology
Decision Date
Jun 13, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The 2008T Machine is intended for use in acute and chronic hemodialysis therapy.
Device Story
Electromechanical hemodialysis system; controls dialysate and extracorporeal blood circuits. Pumps blood through dialyzer; semi-permeable membrane uses diffusion for toxin removal and ultrafiltration for water removal. Dialysate circuit mixes concentrates with purified water, heats, degasses, and delivers to dialyzer; balancing chambers regulate fluid flow. Monitors venous/arterial pressures, air, and blood presence. Used in hospitals/clinics by clinicians/nurses. Includes automation features: conductivity testing, auto-priming, auto-start, and assisted reinfusion. Optional CLiC monitor provides real-time hematocrit/blood volume data; clinician uses output to adjust fluid removal rates to prevent complications like nausea/cramping. Benefits patient by optimizing fluid balance and treatment efficacy.
Clinical Evidence
No clinical studies were performed. Bench testing included biocompatibility (ISO 10993-1), electrical safety, and electromagnetic compatibility (IEC 60601-1-2).
Technological Characteristics
Electromechanical hemodialysis delivery system. Hydraulics assembly materials: PPE+PS, EPDM, PP, PVDF, PVC, PPSU, PESU, PPS, PTFE, PEEK, Polyamide, silicone, titanium, stainless steel, tantalum, borosilicate glass, graphite, ceramic. Features volumetric fluid control, conductivity monitoring (±1.5%), temperature control (35-39°C, ±0.3°C), and heparin delivery. Connectivity via Patient Card system. Software-controlled fluid flow, mixing, heating, and alarms.
Indications for Use
Indicated for acute and chronic dialysis therapy in healthcare facilities for patients weighing ≥ 20kg. Includes Isolated Ultrafiltration, SLED (patients > 40kg), and low volume hemodialysis (20-40kg). Not for plasma replacement or patients < 20kg. Optional bibag system for bicarbonate hemodialysis. Optional Crit-Line Clip Monitor (CLiC) for real-time non-invasive measurement of hematocrit, oxygen saturation, and percent change in blood volume to assist clinicians in fluid removal management.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 13, 2025
Fresenius Medical Care Renal Therapies Group, LLC
Tara Kenney
Principal Regulatory Affairs Associate
920 Winter Street
Waltham, Maryland 02451
Re: K243237
Trade/Device Name: 2008T HD SYS. CDX BLUESTAR (191124); 2008T HD SYS. CDX W/bibag BLUESTAR (191126); 2008T HD SYS. W/O CDX BLUESTAR (191128); 2008T HD SYS. W/O CDX W/bibag BLUESTAR (191130)
Regulation Number: 21 CFR 876.5860
Regulation Name: High permeability hemodialysis system
Regulatory Class: Class II
Product Code: KDI
Dated: October 10, 2024
Received: May 14, 2025
Dear Tara Kenney:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243237 - Tara Kenney
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K243237 - Tara Kenney
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Maura Rooney -S
Maura Rooney
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243237
Device Name
2008T HD SYS. CDX BLUESTAR (191124);
2008T HD SYS. CDX W/bibag BLUESTAR (191126);
2008T HD SYS. W/O CDX BLUESTAR (191128);
2008T HD SYS. W/O CDX W/bibag BLUESTAR (191130)
Indications for Use (Describe)
2008T BlueStar Hemodialysis Machine: The 2008T BlueStar Hemodialysis Machine is indicated for acute and chronic dialysis therapy in a healthcare facility.
Additional therapy options for patients receiving hemodialysis include: Isolated Ultrafiltration, Sustained Low Efficiency Dialysis (SLED), and low volume hemodialysis (patients weighing $\geq 20\mathrm{kg}$ and $\leq 40\mathrm{kg}$). This machine accommodates the use of both low flux and high flux dialyzers. The SLED therapy option is not to be used for patients weighing $\leq 40\mathrm{kg}$. The 2008T BlueStar Hemodialysis Machine is not to be used for plasma replacement therapies, for patients weighing less than $20\mathrm{kg}$, or for renal therapies using substitution fluid.
bibag System (Optional):
The bibag system is used with three stream proportioning Hemodialysis Machines equipped with the bibag module such as the 2008T BlueStar Hemodialysis Machine and intended for use in bicarbonate hemodialysis for acute and chronic renal failure. The bibag is intended for extracorporeal bicarbonate hemodialysis according to a physician's prescription.
Crit-Line Clip Monitor (CLiC) (Optional):
The Crit-Line Clip Monitor is used with the 2008T BlueStar Hemodialysis Machine to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume. The CLiC device measures hematocrit, percent change in blood volume and oxygen saturation in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the clinician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e., increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common complications of dialysis which include nausea, cramping and vomiting.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}
FRESENIUS MEDICAL CARE
K243237
510(k) Summary
2008T BlueStar Hemodialysis Machine Traditional 510(k)
# 1. 510(K) SUMMARY
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR § 807.92.
## 1.1. Submitter’s Information
Name: Fresenius Medical Care Renal Therapies Group, LLC
Address: 920 Winter Street
Waltham, MA
02451-1457
Contact Person: Tara Kenney, Principal Regulatory Affairs Associate
Phone: (781) 250-7689
Fax: (781) 699-9635
Alternate Contact: Laura Reed, Regulatory Affairs Director
Phone: (781) 491-7581
Preparation Date: 10 October 2024
## 1.2. Device Name
Trade Name: 2008T BlueStar™ Hemodialysis Machine
Common Name: Hemodialysis Delivery Device
Regulation Name: High Permeability Hemodialysis System
Regulatory Class: Class II per 21 CFR §876.5860
Product Code: KDI
Product Code Name: Dialyzer, high permeability with or without sealed dialysate system
FDA Review Panel: Gastroenterology/Urology
## 1.3. Legally Marketed Predicate Device
The 2008T BlueStar Hemodialysis Machine (K231125) is the legally marketed predicate device.
## 1.4. Device Description
### 1.4.1. Device Identification
The device is identified as the 2008T BlueStar Hemodialysis Machine. The 2008T BlueStar Hemodialysis Machine is available in four (4) configurations (Table 1).
{5}
FRESENIUS MEDICAL CARE
K243237
510(k) Summary
2008T BlueStar Hemodialysis Machine Traditional 510(k)
Table 1: 2008T BlueStar Hemodialysis Machine Configurations
| Part Number | Part Number Description |
| --- | --- |
| 191124 | 2008T HD SYS. CDX BLUESTAR |
| 191126 | 2008T HD SYS. CDX W/bibag BLUESTAR |
| 191128 | 2008T HD SYS. W/O CDX BLUESTAR |
| 191130 | 2008T HD SYS. W/O CDX W/bibag BLUESTAR |
## 1.4.2. Device Characteristics
The 2008T BlueStar Hemodialysis Machine is an electromechanical device. Software controls the machine during hemodialysis treatment, including fluid flow, mixing, heating, and alarms.
## 1.4.3. Environment of Use
The 2008T BlueStar Hemodialysis Machine is intended for use in healthcare facilities, such as hospitals and dialysis clinics where intermittent dialysis treatment is performed.
## 1.4.4. Brief Written Description of the Device
The 2008T BlueStar Hemodialysis Machine provides hemodialysis treatment by controlling and monitoring both the dialysate circuit and the extracorporeal blood circuit. The machine pumps blood from the patient's body through an extracorporeal circuit, one component of which is the dialyzer. The dialyzer contains a semi-permeable membrane that uses diffusion to transfer toxins and ultrafiltration to transport excess water from the blood into the dialysate circuit. In this separate dialysate circuit, the dialysate concentrates are mixed with purified water, heated, degassed, and delivered to the dialyzer. Balancing chambers control the incoming flow and outgoing flow of the dialysate fluid during ultrafiltration. During hemodialysis, the extracorporeal blood circuit is monitored for venous and arterial blood pressures as well as for the presence of air and blood.
The 2008T BlueStar Hemodialysis Machine has automation features (independent internal conductivity testing, auto priming, auto start, CDX auto on, assisted reinfusion) to minimize user workload and improve user experience. The 2008T BlueStar Hemodialysis Machine includes SLED (Sustained Low Efficiency Dialysis), Low Volume, and Isolated Ultrafiltration therapy options. The 2008T BlueStar Hemodialysis Machine also accommodates the use of the Patient Card system, a dialysis treatment information system.
The 2008T BlueStar Hemodialysis Machine accommodates the following accessory devices and options:
### Accessories
- DIASAFE® plusUS Filter (K231534)
- Patient Card (K173972)
- Patient Card Reader (K173972)
Page 2
{6}
FRESENIUS MEDICAL CARE
K243237
510(k) Summary
2008T BlueStar Hemodialysis Machine Traditional 510(k)
- Bloodlines: 6.35 mm and 8 mm (K962081, K000451, K001107, K022536, K120823, K201207, and K213992)
- Dialyzers: Any commercially available dialyzer equipped with ISO 8637 standard dialysis connectors
- 2008T BlueStar Hemodialysis Machine Field Upgrade Kits (K173972, K231125)
## Options
- bibag® – K162716 (stand-alone disposable) and K121341 (bibag disposable cleared with 2008T HD Machine)
- CDX (Clinical Data Exchange) – K093902 (CDX cleared with 2008T HD Machine)
- CLiC (Crit-Line in a Clip Monitor) – K121599 (Stand-alone CLiC) and K131908 (CLiC with 2008T HD Machine)
- BTM (Blood Temperature Monitor) – K941460 (Stand-alone BTM) and K080964 (BTM with 2008T HD Machine)
- BVM (Blood Volume Monitor) – K982926 (Stand-alone BVM) and K994267 (BVM with 2008K HD Machine)
- Single Needle System – K080964 (Single Needle with 2008T HD Machine)
## 1.4.5. Materials of Use
The 2008T BlueStar Hemodialysis Machine hydraulics assembly is classified as an externally communicating, blood path, indirect, long-term contact (>30 days) duration, (Category C) device in accordance with FDA guidance Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process (08 September 2023). A list of the hydraulic materials for the machine is provided in Table 2.
Table 2: Machine Hydraulic Materials
| Component Material | Material Type |
| --- | --- |
| Plastic/Rubber | PPE+PS (Polyphenylene ether + polystyrene)
EPDM (Ethylene Propylene Diene Monomer Rubber)
PP (Polypropylene)
PVDF (Polyvinylidene fluoride)
PVC (Polyvinyl chloride)
PPSU (Polyphenylsulfone)
PESU (Polyethersulfone)
PPS (Polyphenylene Sulfide)
PTFE (Polytetrafluoroethylene)
PEEK (Polyetheretherketone)
Polyamide (glass fiber reinforced nylon) |
{7}
FRESENIUS MEDICAL CARE
K243237
510(k) Summary
2008T BlueStar Hemodialysis Machine Traditional 510(k)
Table 2: Machine Hydraulic Materials
| Component Material | Material Type |
| --- | --- |
| | Polyester Monofilament
Silicone |
| Metals | Titanium
Stainless Steel
Tantalum |
| Other | Borosilicate Glass
Graphite
Ceramic |
## 1.4.6. Key Performance Specifications/Characteristics
The key performance specifications and characteristics for the 2008T BlueStar Hemodialysis Machine are outlined in Table 3.
Table 3: Key Performance Characteristics
| Feature | Specification/Characteristic | | |
| --- | --- | --- | --- |
| Blood Flow Rates | Blood Line | Blood Flow Rate | |
| | 8 mm | 20–600 mL/min* | |
| | 6.35 (displayed as 6.4) mm | 20–465 mL/min | |
| | 4.8 mm | 10–274 mL/min | |
| | 2.6 mm | 6–86 mL/min | |
| | *Not available with the Low Volume feature enabled.
Accuracy: ± 10% tested at -200 mmHg | | |
{8}
FRESENIUS MEDICAL CARE
K243237
510(k) Summary
2008T BlueStar Hemodialysis Machine Traditional 510(k)
Table 3: Key Performance Characteristics
| Feature | Specification/Characteristic | | |
| --- | --- | --- | --- |
| Maximum Dialysate Flow Rate | Dialysate flow rates are selectable on the Home screen in the following mL/min increments: (0)/100 †‡/150†‡/200†‡/300†/400/500/600/700/800 †Sustained Low Efficiency Dialysis (SLED) ‡Flow rate requires that the Allow Slow Flow option is selected in Service mode. | | |
| | Dialysate flow rates (Qd) for both 1.5x or 2.0x dialysate flow (Auto Flow), based on the Blood Pump rate (Qb): | | |
| | Qb w/1.5x Qd (mL/min) | Qb w/2.0x Qd (mL/min) | Qd (mL/min) |
| | 0–165* | 0–150* | 300 |
| | 166–215* | 151–215* | 400 |
| | 216–315* | 216–265* | 500 |
| | 315 and below** | 265 and below** | 500 |
| | 316–415 | 266–315 | 600 |
| | 416–480 | 316–365 | 700 |
| | 481 and above | 366 and above | 800 |
| | Note: All flow rates are approximate. Dialysate flow will not adjust unless the blood pump is adjusted at least 15–20 mL/min. | | |
| | * If Auto Flow Minimum of 300 Qd is set in Service mode ** If Auto Flow Minimum of 500 Qd is set in Service mode | | |
| Net Fluid Removal | 0–4000 mL/hr | | |
| | Dialysate Flow Rate | Accuracy (on total vol. removed) | |
| | 100 mL/min | ± (1% UF rate + 18 mL/hr) | |
| | 500 mL/min | ± (1% UF rate + 30 mL/hr) | |
| | 800 mL/min | ± (1% UF rate + 48 mL/hr) | |
| Dialysis Time | Hemodialysis: 0 – 9:59 hours, time can be adjusted manually SLED: Fixed at 12 hours Accuracy: ± 1 second per hour | | |
Page 5
{9}
FRESENIUS MEDICAL CARE
K243237
510(k) Summary
2008T BlueStar Hemodialysis Machine Traditional 510(k)
Table 3: Key Performance Characteristics
| Feature | Specification/Characteristic |
| --- | --- |
| Dialysis Fluid Composition | Volumetric, selectable
Acid adjustment range: 130–155 mEq/L Na+
Bicarbonate adjustment range: 20–40 mEq/L Bicarbonate (post-reaction, after mixing with the acid and purified water)
Monitoring conductivity average accuracy: ± 1.5% |
| Dialysis Fluid Temperature | Range 35°C–39°C with alarm limit window automatically adjusted to 2°C above and below set point. Alarm window will not adjust to below 34°C (or 30°C during BTM recirculation measurement) or above 41°C.
Accuracy: ± 0.3°C |
| Heparin Delivery Rate | 0 – 9.9 mL/hr
Accuracy: ± 5% |
## 1.5. Intended Use
The intended use of the 2008T BlueStar Hemodialysis Machine is identical to the intended use of the predicate device (K231125):
The 2008T Machine is intended for use in acute and chronic hemodialysis therapy.
## 1.6. Indications for Use
2008T BlueStar Hemodialysis Machine: The 2008T BlueStar Hemodialysis Machine is indicated for acute and chronic dialysis therapy in a healthcare facility.
Additional therapy options for patients receiving hemodialysis include: Isolated Ultrafiltration, Sustained Low Efficiency Dialysis (SLED), and low volume hemodialysis (patients weighing ≥ 20kg and ≤ 40 kg). This machine accommodates the use of both low flux and high flux dialyzers. The SLED therapy option is not to be used for patients weighing ≤ 40 kg. The 2008T BlueStar Hemodialysis Machine is not to be used for plasma replacement therapies, for patients weighing less than 20 kg, or for renal therapies using substitution fluid.
**bibag System (Optional):**
The bibag system is used with three stream proportioning Hemodialysis Machines equipped with the bibag module such as the 2008T BlueStar Hemodialysis Machine and intended for use in bicarbonate hemodialysis for acute and chronic renal failure. The bibag is intended for extracorporeal bicarbonate hemodialysis according to a physician's prescription.
**Crit-Line Clip Monitor (CLiC) (Optional):**
The Crit-Line Clip Monitor is used with the 2008T BlueStar Hemodialysis Machine to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume. The CLiC device measures hematocrit, percent change in blood volume and oxygen saturation in real time for application in the treatment of dialysis patients with the intended purpose of providing a
{10}
FRESENIUS MEDICAL CARE
K243237
510(k) Summary
2008T BlueStar Hemodialysis Machine Traditional 510(k)
more effective treatment for both the dialysis patient and the clinician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e., increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common complications of dialysis which include nausea, cramping and vomiting.
## 1.7. Comparison of Technological Characteristics with the Predicate Device
The following technological characteristics of the 2008T BlueStar Hemodialysis Machine are substantially equivalent to those of the predicate 2008T BlueStar Hemodialysis Machine (K231125):
- Intended Use
- Indications for Use
- Design Specifications
- Technological Characteristics
- Principle of Operation
- Performance Requirements
## 1.8. Performance Data
Performance testing was conducted for the 2008T BlueStar Hemodialysis Machine. Results of performance testing support the substantial equivalence, safety, and efficacy of the 2008T BlueStar Hemodialysis Machine.
## 1.8.1. Biocompatibility Testing
Biocompatibility testing for the 2008T BlueStar Hemodialysis Machine’s patient contacting components was conducted in accordance with ISO 10993-1:2018 and FDA guidance document *Use of International Standard ISO 10993-1, “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process”* (08 September 2023). The following endpoints were evaluated to support the biological safety of the 2008T BlueStar Hemodialysis Machine:
- Cytotoxicity
- Sensitization
- Irritation
- Pyrogenicity (Material Mediated)
- Hemocompatibility
- Chemical Characterization
- Toxicological Risk Assessment
{11}
FRESENIUS MEDICAL CARE
K243237
510(k) Summary
2008T BlueStar Hemodialysis Machine Traditional 510(k)
## 1.8.2. Human Factors Validation Testing
The changes to the 2008T BlueStar Hemodialysis Machine do not impact the usability of the device. Human Factors Validation Testing is not required.
## 1.8.3. Electrical Safety and Electromagnetic Compatibility (EMC)
Electromagnetic compatibility (EMC) testing was conducted on the 2008T BlueStar Hemodialysis Machine in accordance with IEC 60601-1-2 (2020) Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic disturbances - Requirements and tests.
## 1.8.4. Software Verification and Validation Testing
There are no changes to the 2008T BlueStar Hemodialysis Machine’s software. Software verification and validation testing are not required.
## 1.8.5. Animal Studies
No animal studies were performed for the 2008T BlueStar Hemodialysis Machine.
## 1.8.6. Clinical Studies
No clinical studies were performed for the 2008T BlueStar Hemodialysis Machine.
## 1.9. Conclusion
The intended use, design, principle of operation, and technological characteristics of the 2008T BlueStar Hemodialysis Machine are substantially equivalent to those of the predicate device. Differences between the 2008T BlueStar Hemodialysis Machine and the predicate do not raise new concerns with regard to safety or efficacy. FMCRTG concludes that within the meaning of the Medical Device Amendments Act of 1976, the 2008T BlueStar Hemodialysis Machine is safe and effective for its intended use.
Page 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.