K982926 · Fresenius Medical Care North America · KDI · Sep 3, 1999 · Gastroenterology, Urology
Device Facts
Record ID
K982926
Device Name
FRESENIUS BLOOD VOLUME MONITOR
Applicant
Fresenius Medical Care North America
Product Code
KDI · Gastroenterology, Urology
Decision Date
Sep 3, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Indications for Use
To non-invasively provide real time information about relative blood volume, only in conjunction with the Fresenius Dialysis System 2008H, so that adjustments in the rate at which fluid is removed from the patient can be made to limit or eliminate the common symptoms of dialysis such as nausea, cramping and vomiting.
Device Story
Fresenius Blood Volume Monitor (BVM) integrates with Fresenius Dialysis System 2008H; utilizes ultrasound pulses transmitted through polycarbonate cuvette in arterial blood line; measures sound pulse transit time to calculate sound velocity; sound velocity correlates with blood density and water content; temperature-controlled block compensates for ambient temperature effects; provides real-time relative blood volume (RBV), hematocrit, hemoglobin, and blood water content estimates; operated by dialysis technicians; output used to adjust ultrafiltration rates during hemodialysis; helps prevent dialysis-related symptoms like nausea, cramping, and vomiting.
Clinical Evidence
Clinical evidence includes two studies. Study 1: 6 centers, 45 patients, 3 measurements each; compared device RBV/hematocrit to standard methods; correlation r=0.98. Study 2: 9 patients, 104 comparisons against centrifuge method; mean difference 0.1% ± 1.8%.
Technological Characteristics
Ultrasonic sensor head (20mW/cm2); polycarbonate cuvette; silicone rubber sound coupling; temperature-controlled block; ultrasound transit-time measurement principle; integrated with Fresenius Dialysis System 2008H; non-invasive.
Indications for Use
Indicated for patients undergoing hemodialysis to monitor relative blood volume, hematocrit, hemoglobin, and blood water content in real time to assist technicians in adjusting fluid removal rates to prevent symptoms like nausea, cramping, and vomiting.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
SEP -3 1999
Fresenius Medical Care
North America
Renal Product Technologies
510(k) Summary Fresenius Blood Monitor August 19, 1998
| Contact: | Arthur E. Eilinsfeld<br>Manager, Regulatory Affairs<br>Fresenius Medical Care North America<br>two Ledgemont Center<br>95 Hayden Drive<br>Lexington, MA 02173 |
|----------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
Telephone: 781-402-9000
- Prepared by: K. M. Associates 10 Oak Point Wrentham, MA 02093
Contact: Donald A. Lincoln
Telephone: 508-384-2546
Trade Name: FRESENIUS Blood Monitor
Common Name: Blood Monitor
Classification Name: System, Hemodialysis, Access Recirculation Monitoring
Substantial Equivalence is claimed to: CRIT-Line Monitor III (K972470)
Device Description: There are a variety of method for measuring relative blood volume (RBV), and all make use of the fact that certain constituents of blood (cells, haemoglobin, plasma protein, total protein) remain confined to the vascular compartment, whereas the plasma water can pass both the capillary membrane (into the interstitial compartment) and the dialyser membrane (into the dialysis fluid compartment). Therefore the blood volume change can be determined from the concentration change of these blood constituents.
The Fresenius Blood Volume Monitor is only to be installed in an available bay of the Fresenius
Page 1 of 2
{1}------------------------------------------------
Dialysis System 2008H machine. It uses ultrasound pulses from a transmitter through a polycarbonate cuvette installed into the arterial blood line and received on the opposite side. The primary signal measure is the transient time of the sound pulse through the blood sample, from which the sound velocity is calculated. Silicone rubber is used as sound coupling for the cuvette, with the blood passing through from bottom to top to avoid accumulation of air bubbles. Blood temperature is also measure, and a temperature controlled block prevents interference from ambient temperature changes.
The sound speed depends on the blood density, which in turn depends on the mass fraction of protein and the mass fraction of water or blood water content (BWC). Since the mass of blood protein is virtually unchanged during ultra filtration, the mass of blood water changes, and thereby the density changes. Changes in BWC can be detected by changes of sound velocity, and can be converted to relative blood volume changes. Since sound velocity not only depends on density, but also on temperature, a high precision temperature measurement is also required.
Intended use: To non-invasively provide real time information about relative blood volume, only in conjunction with the Fresenius Dialysis System 2008H, so that adjustments in the rate at which fluid is removed from the patient can be made to limit or eliminate the common symptoms of dialysis such as nausea, cramping and vomiting.
Comparison with predicate device: Both devices provide non-invasive information related to blood volume changes so that corrections in dialysis rates can be made to avoid unwanted side affects. The predicate device can be used with any dialysis system, this current device is for use only with the Fresenius Dialysis System 2008H. Both provide information with an accuracy of approximately ±2% and compare well with other methods such as centrifuging, which is a periodic measurement. The principal difference between the two BVM's is that the predicate device makes its measurements using infa-red technologies while the current device uses ultrasound.
Clinical testing: Several studies have been reported upon. One comprising 6 centers and 45 patients examined three times, each time six values of blood volume and/or hematrocite were compared to a standard method. The relative blood volume of the current device and the relative blood volume calculated from the hemoglobin concentration had a high correlation of r = . 98.
In another study, 9 patients underwent 104 comparison with the centrifuge method. The agreement to the centrifuge was excellent. The mean difference between ultrasonic and centrifuge measurement was 0.1%±1.8%.
Compliance to standards and regulations: The current device, developed in Europe and currently marketed outside the United States, has been tested and found to be in compliance with relevant standards within the EC for devices of this type.
{2}------------------------------------------------
Image /page/2/Picture/13 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its wings or body.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP -3 1999
Fresenius Medical Care North America c/o Mr. Donald A. Lincoln K. M. Associates 10 Oak Point Wrentham, Massachusetts 02093
Re: K982926 Fresenius Blood Volume Monitor Dated: February 4, 1999 Received: February 5, 1999 Regulatory Class: III 21 CFR 876.5860/Procode: 78 KDI
Dear Mr. Lincoln:
We have reviewed your section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act. You may, therefore, market the device, subject to the general controls provisions Act (Act). The general controls provisions of the Act include requirements for registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducer intended for use with the Fresenius Blood Volume Monitor, as described in your premarket notification:
| Device Model | Transducer Model Number |
|-------------------------|-----------------------------------|
| BVM with Control Module | Ultrasonic Sensor Head (20mW/cm2) |
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G,
{3}------------------------------------------------
Page 2 - Mr. Donald Lincoln
(enclosed) of the Center's February 17, 1993 "Revised 510(k) Diagnostic Ultrasound Guidance for 1993." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "dsmo@fdadr.cdrh.fda.gov".
If you have any questions regarding the content of this letter, please contact Richard J. Williams at (301) 594-1220.
Sincerely yours,
David A. Lezman
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Page 1 of 1
## 510(k) Number: K982926
## Device Name(s): Fresenius Blood Volume Monitor
## Indications For Use:
For non-invasive monitoring, in real time, of the estimated relative blood volume percent for patients undergoing hemodialysis so that technicians may make system adjustments in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.
The device can also provide estimated values of: (1) relative blood volume trend; (2) hematocrit; (3) hemoglobin and (4) blood water content or mass fraction of blood water in the total blood mass.
The device is ONL Y to be used in conjunction with the Fresenius Dialysis System 2008H.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONȚINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David R. Legnon
Division Sign Off
Division Sign-Of ivision of Reproductive, Abdomina and Radiological Devic 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.