K121599 · Fresenius Medical Care-North America · KOC · Nov 9, 2012 · Gastroenterology, Urology
Device Facts
Record ID
K121599
Device Name
CRIT-LINE CLIP (CLIC)
Applicant
Fresenius Medical Care-North America
Product Code
KOC · Gastroenterology, Urology
Decision Date
Nov 9, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Indications for Use
The Crit-Line Clip Monitor (CLiC) is used to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume. CLiC measures hematocrit, percent change in blood volume and oxygen saturation in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e. increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.
Device Story
CLiC Monitor is a miniaturized, non-invasive device for real-time monitoring of hematocrit, oxygen saturation, and blood volume changes during dialysis. It utilizes an optical sensor clip (LED array and photodetector) that attaches to a sterile, single-use cuvette installed in the arterial line of a dialysis tubing set. The device processes light absorption and scattering data to calculate blood parameters. Unlike the predicate, the CLiC miniaturizes the microprocessor and electronics directly into the sensor clip housing and replaces the multi-conductor cable with a USB interface. It connects to a host computer (Windows OS) which provides power, processes data, and displays results. Clinicians or dialysis technicians use the output to adjust fluid removal rates during treatment, aiming to maximize fluid removal while minimizing adverse symptoms like cramping, nausea, and vomiting.
Clinical Evidence
No clinical data. Evidence consists of in vitro bench testing, system verification, and validation testing to confirm performance against specifications. Safety and effectiveness were supported by adherence to IEC 60601-1 and IEC 60601-1-2 standards and a risk analysis per ISO 14971. Performance testing demonstrated that the CLiC Monitor's accuracy for hematocrit and oxygen saturation is equivalent to the predicate device specifications.
Technological Characteristics
Compact measurement device; sensor clip housing with LED illuminators and photodetector arrays; USB-powered via host computer (Windows OS). Uses light absorption/scattering principle. Connectivity: USB peripheral. Sterilization: Single-use sterile cuvette (K935958). Safety standards: UL 60601-1/IEC 60601-1, EMC 60601-1-2.
Indications for Use
Indicated for dialysis patients to non-invasively monitor hematocrit, oxygen saturation, and percent change in blood volume in real-time to assist clinicians in managing fluid removal rates and preventing dialysis-related symptoms like nausea, cramping, and vomiting.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
{0}------------------------------------------------
K121599
PAGE 1 OF 5
**NOV. 9 2012**
# 510k Summary
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR Part 807.92.
#### A. Submitter's Information
| Name: | Fresenius Medical Care North America |
|----------------------|--------------------------------------|
| Address: | 920 Winter Street |
| | Waltham, MA 02451-1457 |
| Phone: | (781) 699-4479 |
| Fax: | (781) 699-9635 |
| Contact Person: | Denise Oppermann, Senior Director |
| | Regulatory Affairs - Devices |
| | Renal Therapies Group |
| Date of Preparation: | 31 May 2012 |
#### B. Device Name
Trade Name: Classification Name
Classification Number Product Code/Classification Panel: CLiC Monitor (Crit-Line Clip Monitor) Non-invasive Hematocrit, Oxygen Saturation, Percent change in blood volume Monitor Class II per § 876.5820 78 MQS/Gastroenterology/Urology Panel
# Legally Marketed Predicate Device (unmodified device) C.
Crit-Line MI Monitor (K972470).
{1}------------------------------------------------
K121599
PAGE 2 OF 5
Image /page/1/Picture/1 description: The image shows the Fresenius Medical Care logo. The logo consists of a stylized graphic on the left and the text "FRESENIUS MEDICAL CARE" on the right. The graphic is made up of three chevron-like shapes stacked on top of each other, pointing downwards. The text is in bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.
#### D. Device Description
The intended use of the Crit-Line III (CLM III) Monitor is as a continuous real-time monitor for non-invasive hematocrit, oxygen saturation, and percent change in blood volume measurement. The CLM III is used primarily as a monitor for dialysis treatments.
The CLM III uses the principle of light absorption and scattering through the blood under test to measure oxygen saturation (O2 SAT) and hematocrit (HCT). The HCT values are then used to calculate the related Blood Volume (BV) percentage change relative to the starting BV based on the beginning HCT.
The CLM III consists of a microprocessor-controlled main circuit board located within the display housing. The display housing contains the sensor clip driver measurementcalculation circuitry and display screen circuitry. The sensor clip with optical system (LED array and a photodetector array) is tethered to the main housing and circuit board by a multi-conductor cable. The cable connects to a single-use, sterile cuvette (K935958) installed in-line with the arterial line of the dialysis tubing set.
Calculations performed on the main circuit board within the display housing yield the hematocrit, oxygen saturation and the related value of blood volume measurement results and are displayed on the LCD display screen.
Modifications to the previously cleared Crit-Line III Monitor include:
#### 3. Miniaturization/Re-packaging
The CLiC Monitor incorporates all hardware/electronic components - miniaturized and contained within the sensor clip housing, for measuring indicated parameters (HCT O2 SAT). The primary modification of the device involves integrating a miniaturized microprocessor with the required electronic circuitry and software to drive the existing optical system (LED and photo-detector array) within the plastic sensor clip housing. The CLiC utilizes the same, previously cleared Crit-Line technology (K972470), sensor clip and optical system. The operating principles and fundamental measurement methods of the CLiC are the same as the predicate device.
{2}------------------------------------------------
K121599
PAGE 3 OF 5
Image /page/2/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a stylized symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three chevron-like shapes stacked on top of each other, with the top two pointing upwards and the bottom one pointing downwards. The text is in bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.
#### 4. Distributed Process Technology
A standard USB cable replaces the tethering multi-conductor cable used in the CLM III Monitor. USB-compatible hardware and software are incorporated in the CLiC housing. The CLiC can be hosted on a medical grade computer system (Windows OS) that can interface with, power a USB peripheral and operate supporting drivers. CLiC interlocks with a singleuse, sterile cuvette (K935958) which is located in-line with the arterial line of the dialysis tubing set for indicated measurements. The lightwave measurements are performed autonomously by the CLiC. Blood parameters are then calculated and transmitted to the host computer display microprocessor in text format by means of the USB interface. The accompanying display software installed on the host computer system facilitates the display of measured parameters: hematocrit, oxygen saturation and blood volume changes on the display screen.
#### E. Indications for Use
The Crit-Line Clip Monitor (CLiC) is used to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume. The CLiC measures hematocrit, percent change in blood volume and oxygen saturation in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e. increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.
### Technological Characteristics F.
The CLiC Monitor is a compact measurement device which can be used with any medical grade computer system (Windows OS) that can interface with and power a USB peripheral. With the supporting display software, the Windows-based computer system provides the user interface.
The CLiC is a modification to the Crit-line III with equivalent indications for use, to noninvasively measure hematocrit, oxygen saturation and percent change in blood volume. Both devices non-invasively measure blood parameters using multiple wavelengths of light
{3}------------------------------------------------
**FRESENIUS**
**MEDICAL CARE**
which are scattered and absorbed, respectively, by different constituents in the blood. The CLiC Monitor has the same operating principle and fundamental scientific technology as the predicate device. It is also comparable to the predicate device in terms of key safety, effectiveness, and quality assurance features.
The CLiC Monitor is designed and developed using product specifications, construction and manufacturing methods functionally equivalent to the predicate device. The plastic sensor clip housing and optical system (LED illuminators and photodetectors) are same the as the predicate device (K972470).
The modifications described in this submission have not altered nor do they have the potential to alter the fundamental scientific technology of the device.
FMCNA has concluded that the CLiC Monitor is of comparable type and substantially equivalent to the currently marketed Crit-line III, complies with the same or equivalent standards and has the same intended use. In addition, the CLiC Monitor is certified to comply with safety requirements of UL 60601-1/IEC 60601-1, and EMC standard 60601-1-2.
The following technical specifications of the modified device remain the same as the unmodified device:
- . LED illuminators and Photodetector arrays
- . Safety system
- System performance ●
- Environmental Requirements .
- Transportation and Storage condition ●
- User Interface (except USB port and display software on host computer system) .
- Hardware (except miniaturization) .
- Environmental Design .
- . Alarms
- Accuracy and Controls ●
- . Protection against excessive temperature or other hazards
- . Manufacturing Location and manufacturing processes (testing, shipping, installation and service).
Safety and effectiveness of the CLiC Monitor is confirmed by:
- . System verification and validation testing - to verify performance to specifications, applicable referenced FDA regulations, and user requirements.
{4}------------------------------------------------
- . Adherence to FDA-recognized industry and international standards (IEC 60601-1 and IEC 60601-1-2) - Electrical Safety and Electromagnetic Compatibility testing was conducted to ensure the safety and effectiveness of the device after the proposed modifications.
A risk analysis, per ISO 14971, has been completed and potential hazards associated with the modifications are identified and mitigations are verified wherever applicable. All potential risks were deemed acceptable after mitigation.
## G. Performance Data
The performance of the modified device was evaluated according to existing procedures, protocols, declared performance standards and guidelines of the quality system regulation (21 CFR 820). Design verification and validation tests were conducted to ensure that the modifications described in this submission did not affect the essential performance of the device and the device functions as intended.
The CLiC Monitor accuracy is equivalent to the predicate device specifications for both hematocrit and oxygen saturation measurements. Because blood volume calculations are based on hematocrit and the CLiC Monitor matches the predicate specifications, the percentage blood volume comparisons are also equivalent.
### H. Conclusion
The CLiC Monitor does not pose any new potential safety risks and performs the same measurement functions as the predicate device.
Test results demonstrated that the modified CLiC Monitor functions as intended and met pre-determined acceptance criteria. Results of performance testing (in vitro bench testing), safety testing and usability testing indicate that the modified CLiC device is substantially equivalent to the named predicate device and remains safe and effective for its intended use.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. In the center of the seal is an abstract image of an eagle with its wings spread.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
November 9, 2012
Fresenius Medical Care - North America % Ms. Denise Oppermann Senior Director, Regulatory Affairs Devices 920 Winter Street WALTHAM MA 02451
Re: K121599
Trade/Device Name: Crit-Line Clip (CLIC) Monitor Regulation Number: 21 CFR\$ 876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: KOC Dated: October 10, 2012 Received: October 11, 2012
Dear Ms. Oppermann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{6}------------------------------------------------
Page 2 - Ms. Denise Oppermann
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
**Herbert R. Lerner**
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Fresenius Medical Care logo. The logo consists of a stylized symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three stacked chevrons, each pointing downwards. The text is in bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.
510(k) Number (if known):
K121599
Device Name:
Crit-Line Clip (CLiC) Monitor
# Indications for Use:
The Crit-Line Clip Monitor (CLiC) is used to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume. CLiC measures hematocrit, percent change in blood volume and oxygen saturation in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e. increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.
XPrescription Use (Per 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Herbert P. Lerner
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K121599
Page 75 of 209
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.