Versacem Acetabular Shell and Double Mobility HC Liners
K241767 · Medacta International S.A. · LZO · Mar 6, 2025 · Orthopedic
Device Facts
Record ID
K241767
Device Name
Versacem Acetabular Shell and Double Mobility HC Liners
Applicant
Medacta International S.A.
Product Code
LZO · Orthopedic
Decision Date
Mar 6, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Versacem is designed for use in total hip arthroplasty, for selected primary or revision surgery. Total hip arthroplasty is indicated in the following cases: - Severely painful and/or disabled joint as a result of arthrosis, traumatic arthritis, rheumatoid polvarthritis or congenital hip dysplasia. - Avascular necrosis of the femoral head. - Acute traumatic fracture of the femoral head or neck. - Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, partial hip arthroplasty, hip resurfacing replacement or total hip arthroplasty. Versacem is also recommended when dislocation prevention is the main driver for the prosthesis choice and the acetabular bone quality is damaged, as in the following cases: - Elderly patients or poor bone quality. - Revision for recurrent dislocation in elderly patients with poor bone quality. - Patients treated with radiotherapy or chemotherapy. Versacem is intended for cemented use only.
Device Story
Versacem Acetabular Shell and Double Mobility HC Liners are sterile, implantable orthopedic devices for total hip arthroplasty; intended for cemented fixation. Device system consists of acetabular shell (sizes Ø40-Ø56) and compatible Double Mobility HC liners. Used by orthopedic surgeons in clinical settings to replace damaged hip joints; specifically indicated for patients with high dislocation risk or compromised bone quality. System functions as a semi-constrained hip prosthesis; provides stability and range of motion. Output is a reconstructed hip joint; clinical benefit includes pain relief, restoration of joint function, and reduced risk of dislocation. Device is implanted during surgery; performance verified via bench testing including wear, fatigue, range of motion, and impingement risk evaluations.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including wear testing, range of motion evaluation, fatigue testing, pull-off/lever-out tests, jumping distance evaluation, and impingement risk assessment. Pyrogenicity validated via LAL testing (ISO 11737-3, USP <85>/<161>) and in-vivo rabbit evaluation (USP <151>).
Technological Characteristics
Cemented semi-constrained hip prosthesis. Materials: Metal/Ceramic/Polymer. Components: Acetabular shell (Ø40-Ø56) and Double Mobility HC liners. Sterilization: Fully validated method. Biocompatibility: Assessed per standard requirements. Bench testing performed for mechanical integrity (fatigue, wear, pull-off, lever-out).
Indications for Use
Indicated for total hip arthroplasty in patients with painful/disabled joints due to arthrosis, traumatic arthritis, rheumatoid polyarthritis, congenital hip dysplasia, avascular necrosis, acute traumatic fracture, or failed previous hip surgery. Recommended for patients requiring dislocation prevention, including elderly patients, those with poor bone quality, or patients treated with radiotherapy/chemotherapy. Intended for cemented use only.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Predicate Devices
NOVAE® Stick Dual Mobility Acetabular Cup (K111572)
{0}------------------------------------------------
March 6, 2025
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of it is the FDA logo, with the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
Medacta International S.A. % Christopher Lussier Senior Director, Quality, Regulatory and Clinical Research Medacta USA 6386 Global Drive, Suite 101 Memphis. Tennessee 38141
Re: K241767
Trade/Device Name: Versacem Acetabular Shell and Double Mobility HC Liners Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO Dated: February 7, 2025 Received: February 7, 2025
Dear Christopher Lussier:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Picture/3 description: The image shows the text "Limin Sun -S" in a large, bold, sans-serif font. The text is black against a white background. The words are arranged horizontally, with "Limin" on the left, followed by "Sun" and then "-S". The text appears to be a title or heading. The background has a light blue watermark.
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K241767
Device Name
Versacem Acetabular Shell and Double Mobility HC Liners
#### Indications for Use (Describe)
Versacem is designed for use in total hip arthroplasty, for selected primary or revision surgery. Total hip arthroplasty is indicated in the following cases:
- Severely painful and/or disabled joint as a result of arthrosis, traumatic arthritis, rheumatoid polvarthritis or congenital hip dysplasia.
- · Avascular necrosis of the femoral head.
- · Acute traumatic fracture of the femoral head or neck.
· Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, partial hip arthroplasty, hip resurfacing replacement or total hip arthroplasty.
Versacem is also recommended when dislocation prevention is the main driver for the prosthesis choice and the acetabular bone quality is damaged, as in the following cases:
- · Elderly patients or poor bone quality.
- · Revision for recurrent dislocation in elderly patients with poor bone quality.
- · Patients treated with radiotherapy or chemotherapy.
### Versacem is intended for cemented use only.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Medacta International SA
# 510(k) Summary
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Senior Director, Quality, Regulatory, and Clinical Research, Medacta USA Date Prepared: June 18, 2024 Date Revised: February 07, 2025
#### II. Device
| Device Proprietary Name: | Versacem Acetabular Shell and Double Mobility HC Liners |
|--------------------------|--------------------------------------------------------------------------------------------------|
| Common or Usual Name: | Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer,<br>Cemented Or Non-Porous, Uncemented |
| Classification Name: | Hip joint metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis. |
| Primary Product Code | LZO |
| Regulation Number: | 21 CFR 888.3353 |
| Device Classification | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following predicate devices.
Primary Predicate device:
- . NOVAE® Stick Dual Mobility Acetabular Cup, K111572, Serf
Reference predicate devices:
- Double Mobility HighCross® highly crosslinked UHMWPE, K092265, Medacta International SA
- Double Mobility HighCross® highly crosslinked UHMWPE, K131458, Medacta International SA
- Double Mobility HighCross highly crosslinked UHMWPE Liners, K143453, Medacta International SA
- Medacta Bipolar Head, K091967, Medacta International SA
{5}------------------------------------------------
Medacta International SA
#### IV. Device Description
The purpose of this submission is to gain clearance for the Versacem Acetabular Shell and to add two new sizes of the compatible Double Mobility HC Liners previously cleared within K092265, K131458 and K143453.
The Versacem Acetabular Shell implants and the Double Mobility HC liners are sterile, individually packaged, implantable medical devices intended to be used in Total Hip Arthroplasty.
The devices subject of this submission are:
- . Versacem Acetabular Shell from size Ø40 to size Ø56;
- . Double Mobility HC liner Ø22.2/DMAZ;
- Double Mobility HC liner Ø28/DMC. ●
#### V. Indications for Use
Versacem is designed for use in total hip arthroplasty, for selected primary or revision surgery. Total hip arthroplasty is indicated in the following cases:
- . Severely painful and/or disabled joint as a result of arthrosis, traumatic arthritis, theumatoid polyarthritis or congenital hip dysplasia.
- Avascular necrosis of the femoral head. ●
- Acute traumatic fracture of the femoral head or neck.
- Failure of previous hip surgery: joint reconstruction, arthrodesis, partial hip ● arthroplasty, hip resurfacing replacement or total hip arthroplasty.
Versacem is also recommended when dislocation prevention is the main driver for the prosthesis choice and the acetabular bone quality is damaged, as in the following cases:
- . Elderly patients or poor bone quality.
- Revision for recurrent dislocation in elderly patients with poor bone quality.
- Patients treated with radiotherapy or chemotherapy. .
Versacem is intended for cemented use only.
#### VI. Comparison of Technological Characteristics
The subject Versacem Acetabular Shell and the predicate device NOVAE® Stick Dual Mobility Acetabular Cup (K111572) are substantially equivalent with respect to the following characteristics:
- General Design; ●
- Articular surface; ●
- Cementation; ●
- Material: ●
- Biocompatibility;
- Device usage ●
- Packaging ●
{6}------------------------------------------------
The subject Versacem Acetabular Shell and the predicate device NOVAE® Stick Dual Mobility Acetabular Cup (K111572) differ with respect to the following characteristics:
- . Range of sizes;
- Shelf-life; .
- Sterilization method. ●
### Discussion
The slight difference between subject and predicate devices range of sizes do not pose any new risk in terms of safety and performance as evaluated in the bench testing.
The subject device shelf-life does not affect in any way its safety and performance, as it is shared with the predicate device Medacta Bipolar Head (K091967) and in general with the majority of Medacta products cleared in US. The Medacta Bipolar Head (K091967) is substantially equivalent to the subject device in terms of manufacturing material and main steps of the production process.
Finally, the subject device sterilization method do not arise any new risk in terms of safety and performance as it is fully validated.
The subject Double Mobility HC liners are substantially equivalent to the predicate devices Double Mobility HC Liners (K092265; K131458; K143453) with respect to the following characteristics:
- Design:
- Articular Surface; ●
- Material; ●
- Biocompatibility: ●
- Device Usage;
- Packaging; ●
- Shelf-life: ●
- Sterilization method. ●
The subject Double Mobility HC liners differ from the predicate devices Double Mobility HC Liners (K092265; K131458; K143453) with respect to the following characteristics:
- Range of sizes. ●
## Discussion
The difference between subject and predicate devices range of sizes do not pose any new risk in terms of safety and performance as the inner diameter of subject and predicate devices is the same. The difference in the outer diameter between subject and predicate device do not pose any new risk in terms of safety and performance as evaluated in the bench testing.
#### VII. Performance Data
Based on the risk analysis, testing activities were conducted to written protocols. The following validations and tests are provided in support of the substantial equivalence determination.
## Nonclinical Studies
{7}------------------------------------------------
### DESIGN VALIDATION
- Versacem Acetabular Shell and double Mobility HC Liner Design Validation ●
# PERFORMANCE TESTING
- Versacem Acetabular Shell and Double Mobility HC Liners Wear Test ●
- . Versacem Acetabular Shell - Range of Motion Evaluation
- Versacem Acetabular Shell Fatigue Test ●
- Double Mobility HC Liner – Pull-off and Lever-out Tests
- Versacem Acetabular Shell Evaluation of Jumping Distance ●
- Versacem Acetabular Shell Evaluation of impingement risk ●
# PYROGENICITY
Medacta International does not intend to label the subject devices as non-pyrogenic or pyrogen free. Control of pyrogenicity is validated through the following primary test methods on all identified representative devices as part of their initial validation prior to submission:
- 1. Quantification of bacterial endotoxin units (EU) on the device surface after final cleaning and sterilization using the bacterial endotoxin test (LAL test) on the basis of a measured colourproducing reaction proportional to the interaction of Limulus amebocyte lysate (LAL) and endotoxin according to ISO11737-3 and European Pharmacopoeia \$2.6.14 (which is harmonized with USP chapter <85> and USP <161>)
- 2. In-vivo evaluation of pyrogenicity in rabbit, per USP <151>, in conjunction with biocompatibility studies.
## BIOCOMPATIBILITY assessment
### SHELF-LIFE evaluation
## VIII. Conclusion
The information provided above supports that the subject devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.