NOVAE SUNFIT TH DUAL MOBILITY ACETABULAR CUP MODEL SUNFIT TH (METAL-BACK), NOVAE E TH DUAL MOBILITY ACETABULAR CUP MODEL
K111572 · Serf · LZO · Aug 29, 2011 · Orthopedic
Device Facts
Record ID
K111572
Device Name
NOVAE SUNFIT TH DUAL MOBILITY ACETABULAR CUP MODEL SUNFIT TH (METAL-BACK), NOVAE E TH DUAL MOBILITY ACETABULAR CUP MODEL
Applicant
Serf
Product Code
LZO · Orthopedic
Decision Date
Aug 29, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
NOVAE® Dual mobility Acetabular Cup is indicated for total hip replacement, which includes: - Osteoarthritis - Femoral neck fracture - Dislocation risk - Osteonecrosis of the femoral head - Revision procedures where other treatments or devices have failed and if bone reconstruction so permits SUNFIT TH, NOVAE E TH and COPTOS TH are intended for press-fit use and NOVAE STICK is indicated for cemented use
Device Story
NOVAE® Dual Mobility Acetabular Cup is a total hip prosthesis component; consists of metallic shell and mobile UHMWPE liner. Metallic shell fixed in acetabulum via cement (NOVAE STICK) or press-fit (SUNFIT TH, NOVAE E TH, COPTOS TH). Liner remains mobile within shell and is retained on prosthetic femoral head (Ø22.2 or 28 mm). Used by orthopedic surgeons in OR for primary or revision hip arthroplasty. Provides dual mobility to reduce dislocation risk. Fixation reinforced by pegs/cortical screws in specific models. Clinical benefit: restoration of hip function and stability.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by non-clinical bench testing including dimensional analysis, push-out force, head insertion/pull-out/lever-out force, bending cycles, flange fracture testing, coating characterization, range of motion, and wear analysis.
Technological Characteristics
Metallic shell: Stainless steel (ISO 5832-1:2007). Cementless coating: CP titanium and hydroxyapatite plasma spray. Liner: UHMWPE (ISO 5834-2:2006). Fixation: Cemented or press-fit with optional stainless steel pegs/cortical screws. Dual mobility design with mobile liner. Sterilization method not specified.
Indications for Use
Indicated for total hip replacement in patients with osteoarthritis, femoral neck fracture, dislocation risk, osteonecrosis of the femoral head, or requiring revision procedures where bone reconstruction permits. Available as cemented (NOVAE STICK) or press-fit (SUNFIT TH, NOVAE E TH, COPTOS TH) versions.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Tri-Polar system and RingLoc® + modular acetabular shell (K991990)
Submission Summary (Full Text)
{0}------------------------------------------------
AUG 2 9 2011
## 510(k) Summary
Date: August the 23th 2011
| Company name and address: | SERF |
|---------------------------|-------------------------------------|
| | 85 avenue des Bruyères |
| | 69153 Décines Cedex |
| | FRANCE |
| | Phone: +33 4 72 05 60 10 |
| | Fax: +33 4 72 02 19 18 |
| Contact person: | Caroline GELLY |
| | Regulatory Affairs Engineer |
| Date prepared: | August the 23rd, 2011 |
| Trade name: | NOVAE® Dual Mobility Acetabular Cup |
Classification name: Hip joint metal/ceramic/polymer semi-constrained cemented or
Total hip prosthesis – Acetabular component
nonporous uncemented prosthesis (21 CFR 888.3353, Product Code LZO/MEH)
## Device description
Common name:
The NOVAE® Dual Mobility Acetabular Cup is composed of a metallic shell fixed in the acetabulum and a polyethylene liner.
### Cemented metallic shell:
- NOVAE STICK: the cemented metallic shell is made of stainless steel according to ISO 5832-1:2007. The outer surface is sandblasted and presents macro-volumes to augment contact between the implant and the cement. The inner surface is highly polished.
### Cementless metallic shell:
The cementless metallic shell is made of stainless steel according to ISO 5832-1:2007 and is coated with a CP titanium and hydroxyapatite plasma spray (double layer coating) on the outer surface and is highly polished on the inner surface.
Three cementless metal-backs are available:
- SUNFIT TH: the SUNFIT TH version is available without the pegs and screw, and is a press fit onlv.
- NOVAE E TH: the NOVAE E TH has 3 fixation points (2 divergent pegs towards the pubis and ischium and 1 cortical screw through a flange towards the ilium).
- COPTOS TH: the COPTOS TH has 5 fixation points (2 divergent pegs towards the pubis and ischium, cortical screws through 2 flanges towards the ilium and 1 foramen hook).
{1}------------------------------------------------
## Liner:
The liner is made of Ultra-High-Molecular-Weight Polyethylene according to ISO 5834-2:2006. The liner is mobile (free) in the metallic shell and retained on the prosthetic femoral head. Liners can be used with Ø22,2 or 28 mm prosthetic femoral heads.
## Pegs and cortical screws:
When applicable, primary fixation can be reinforced by impacted pegs and self tapping cortical screws made of stainless steel according to ISO 5832-1:2007.
## Substantial equivalence claimed to predicate devices
NOVAE® Dual Mobility Acetabular Cup is substantially equivalent to the POLARCUP® dual mobility system (K070278, PLUS Orthopedics) in terms of intended use, design, range of sizes, materials used, mechanical safety and performances, the Restoration™ ADM system (K072020, Stryker) in terms of intended use, design and metal-back coating and the Tri-Polar system and RingLoc® + modular acetabular shell (K991990, BIOMET) in terms of intended use, design and range of sizes.
| Device | NOVAE® Dual<br>Mobility<br>Acetabular Cup | POLARCUP® dual<br>mobility system | Restoration™<br>ADM System | Tri-Polar system<br>and RingLoc®<br>+ modular<br>acetabular shell |
|----------------------------------|-------------------------------------------|-------------------------------------------------------|----------------------------------|-------------------------------------------------------------------|
| 510(k) number | / | K070278 | K072020 | K991990 |
| Intended use | | | | |
| Total hip<br>replacement | Yes | Yes | Yes | Yes |
| Cementless/<br>cemented | Yes/Yes | Yes/No | Yes/No | Yes/No |
| Primary/<br>Revision | Yes/Yes | Yes/Yes | Yes/No | Yes/Yes |
| Design | | | | |
| Dual mobility | Yes | Yes | Yes | Yes |
| Metal-back and<br>a mobile liner | Yes | Yes | Yes | No<br>(bipolar head +<br>acetabular cup) |
| Screws | Cortical | Cortical | No screws | Screws |
| Pegs | Grooved | Grooved | No pegs | No pegs |
| Liner is retained<br>on the head | Yes | Yes | Yes | Yes |
| Materials | | | | |
| Metal-back | Stainless steel (ISO<br>5832-1:2007) | High nitrogen<br>stainless steel (ISO<br>5832-9:1992) | Wrought Cobalt<br>Chromium alloy | Titanium alloy |
| Metal-back<br>coating | CP titanium and HA | CP titanium | CP titanium and HA | Titanium alloy |
| Liner | UHMWPE | UHMWPE | UHMWPE | Polyethylene |
| Pegs and Screws | Stainless steel | Stainless | Not applicable | Unknown |
{2}------------------------------------------------
## Intended use
NOVAE® Dual mobility Acetabular Cup is indicated for total hip replacement, which includes:
- . Osteoarthritis
- Femoral neck fracture .
- Dislocation risk .
- . Osteonecrosis of the femoral head
- Revision procedures where other treatments or devices have failed and if bone . reconstruction so permits
SUNFIT TH, NOVAE E TH and COPTOS TH are intended for press-fit use and NOVAE STICK is indicated for cemented use
## Non-clinical Test Summary
The following tests were conducted:
- . Dimensional analysis (acceptance criteria was met)
- . Uncemented metal-back push-out force (acceptance criteria was met)
- Head insertion force and Head Pull-Out testing (acceptance criteria was met) .
- . Head lever out force (acceptance criteria was met)
- Number of bending cycles to flange fracture (acceptance criteria was met) .
- Characterization of CP titanium and HA plasma spray (acceptance criteria was met) .
- Range Of Motion Analysis (acceptance criteria was met) .
- Wear Analysis (acceptance criteria was met)
## Clinical Test Summary
No clinical studies were performed
Conclusions Nonclinical and Clinical
The NOVAE® dual mobility acetabular cup is substantially equivalent to the predicate devices in terms of indications for use, design, material, function and performance.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three curved shapes resembling a bird or abstract design. The logo is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SERF % Mr. J.D. Webb Orthomedix Group, Inc. 1001 Oakwood Blvd. Round Rock, Texas 78681
AUG 2 9 2011
Re: K111572
Trade/Device Name: NOVAE® Dual Mobility Acetabular Cup Regulation Number: 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO, MEH Dated: May 5, 2011 Received: June 6, 2011
Dear Mr. Webb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, 7 ug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act . The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements 1/0 hast or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Mr. J.D. Webb
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
yours,
Melkerson
Mark N. Melkerso Director Division of Surgical, Orthopedic
and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): / K\\1572 (pg 1/1)
Device Name: NOVAE® Dual Mobility Acetabular Cup
Indications For Use:
NOVAE® Dual mobility Acetabular Cup is indicated for total hip replacement, which includes:
- . Osteoarthritis
- . Femoral neck fracture
- . Dislocation risk
- . Osteonecrosis of the femoral head
- Revision procedures where other treatments or devices have failed and if bone . reconstruction so permits
SUNFIT TH, NOVAE E TH and COPTOS TH are intended for press-fit use and NOVAE STICK is indicated for cemented use.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for m melleer
(Division Sign-Off) (Division Sign-Off)
Division of Surgical, Orthopedic, Orthopedic,
Division of Surgical, Orthopedic, Division of Surger
and Restorative Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.