K241735 · Epic Medical Pte. , Ltd. · ONB · Oct 10, 2024 · General Hospital
Device Facts
Record ID
K241735
Device Name
ProSeal™ In Line Pump Set (423850)
Applicant
Epic Medical Pte. , Ltd.
Product Code
ONB · General Hospital
Decision Date
Oct 10, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The ProSeal™ Closed System drug Transfer Device (CSTD) mechanically prohibits environmental contaminants from entering the system and the escape of drug or vapor concentrations from the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. The ProSeal™ system also prevents the introduction of microbial contaminations into the drug or fluid path for up to 7 days, when used as intended.
Device Story
ProSeal™ In Line Pump Set is a single-use, sterile, non-pyrogenic closed system transfer device (CSTD) component. It functions as an in-line IV infusate/drug transfer adapter. The device features an IV bag spike port at the proximal end and connects to a mating ProSeal™ Injection Site on other CSTD components; the distal end connects to a standard IV set. Operation involves manual manipulation to join mating devices, which are sealed with resealing membranes; pressing these membranes together creates a secured, closed fluid path. This physical barrier prevents environmental contaminants from entering and drug/vapor from escaping, minimizing exposure to hazardous drugs and preventing microbial ingress for up to 7 days. Used by healthcare professionals in clinical settings to facilitate safe drug administration and prevent needlestick injuries.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench performance testing, including leak integrity (ISO 8536-4), flow rate, biocompatibility (ISO 10993), and sterility/packaging validation (ASTM F1980, ASTM F88, ASTM F1929).
Technological Characteristics
Materials: PP (without DEHP), TPE. Sensing/Actuation: Mechanical resealing membranes. Form factor: In-line IV adapter with IV bag spike port. Connectivity: None. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Shelf-life: 3 years. Standards: ISO 8536-4, ISO 80369-7, ISO 15747, ISO 23908, ANSI AAMI CN27.
Indications for Use
Indicated for use by adequately trained healthcare professionals, including pharmacists, in clinical settings for the closed transfer of parenteral drugs. Contraindications not specified.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
October 10, 2024
Epic Medical Pte. Ltd. Freddie Lee CEO/md 105 Cecil Street #20-04, The Octagon Singapore, SG 069534 Singapore
Re: K241735
Trade/Device Name: ProSeal™ In Line Pump Set (423850) Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: ONB Dated: September 12, 2024 Received: September 12, 2024
Dear Freddie Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part
820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Porsche Bennett
Porsche Bennett For David Wolloscheck, Ph.D Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241735
Device Name ProSeal™ In Line Pump Set (423850)
#### Indications for Use (Describe)
The ProSeal™ Closed System drug Transfer Device (CSTD) mechanically prohibits environmental contaminants from entering the system and the escape of drug or vapor concentrations from thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. The ProSeal™ system also prevents the introduction of microbial contaminations into the drug or fluid path for up to 7 days, when used as intended.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------------|
| <span style="unicode-bidi:embed;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| □ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# cp10
#### K241735 – 510(k) Summary
#### I. Submitter
Epic Medical Pte. Ltd. 105 Cecil Street #20-04, The Octagon, Singapore 069534 Phone: +65 9635 2618 / +66 81 761 5292
Contact Person: Mr. Freddie LEE, Chief Executive Officer/ Managing Director Date Prepared: October 10, 2024 Content and Format: Prepared in accordance with 21 CFR 807.92 Type of Submission: Special
#### II. Subject Device
| 510(k) Number: | K241735 |
|------------------------------|--------------------------------------------------------------------------------|
| Trade/ Device Name: | ProSeal™ In Line Pump Set (423850) |
| Common/Usual Name: | Closed Antineoplastic and Hazardous Drug Reconstitution and<br>Transfer System |
| Regulation Number: | 21 CFR 880.5440 |
| Regulation Name: | Intravascular administration set |
| Regulatory Class: | Class: II |
| Classification Product Code: | ONB |
### III. Predicate
| 510(k) Number: | K240171 |
|---------------------|--------------------------------------------------------------------------------|
| Trade/ Device Name: | ProSeal™ Injector Plus (Model no. 421050) |
| Common/ Usual Name: | Closed Antineoplastic and Hazardous Drug Reconstitution and<br>Transfer System |
| Regulation Number: | 21 CFR 880.5440 |
| Regulation Name: | Intravascular administration set |
| Regulatory Class: | Class: II |
| Product Code: | ONB |
{5}------------------------------------------------
# IV. Device Description
The purpose of this Special 510(k) Submission is to seek FDA clearance prior to the introduction in the US, of an additional device in the ProSeal™ CSTD sub-grouping - ProSeal™ Adaptor/Connector - the Subject ProSeal™ In Line Pump Set. This device is an addition to the ProSeal™ CSTD system of devices (cleared K241823), enhancing the completeness of the portfolio of ProSeal™ CSTD system.
The ProSeal™ CSTD devices are single-use, sterile, non-pyrogenic CSTD component devices that are fitted to each other. They are used as sterile interfaces for the closed injections and withdrawals of liquids into and from the ProSeal™ CSTD component devices and external transfer devices.
The ProSeal™ In Line Pump Set is a closed system in-line IV infusate/drug transfer adapter/ connector for providing closed system protection during hazardous drug administration when connected to a standard IV set at its distal end, with the proximal end attached to a mating ProSeal™ Injection Site (K240433) of a ProSeal™ CSTD component device such as the eZSURE™ Empty Fluid Container with ProSeal™ Injection Site (K241442).
The change proposed in this 510(k) Submission to the existing ProSeal™ CSTD devices is as follows:
- Addition of I.V. bag spike port
## V. Indications for Use Statement (unchanged)
The ProSeal™ Closed System drug Transfer (CSTD) mechanically prohibits environmental contaminants from entering the system and the escape of drug or vapor concentrations from the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. The ProSeal™ system also prevents the introduction of microbial contaminations into the drug or fluid path for up to 7 days, when used as intended.
#### VI. Comparison of Technological Characteristics
The Predicate device (and a component/subassembly of the Subject device) is the ProSeal™ Injector Plus (cleared K240171). The Subject device and the Predicate device share the following key characteristics:
- . Principles of operation
- Closed system technology
- Preventing needlestick injuries ●
- Spike port subcomponent & material .
- Primary package top web ●
- Sterilization method
- Validated shelf-life
- Single-use only 0
- Labelling specifications .
An overview table summarizing the comparison of the key characteristics between the Subject and the Predicate device is provided hereunder.
{6}------------------------------------------------
| Characteristic<br>compared | Predicate Device (K240171)<br>ProSeal™ Injector Plus | Subject Device (K241735)<br>ProSeal™ In Line Pump Set | Comment/<br>Discussion |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Intended Use and<br>Indications for Use<br>statement | "The ProSeal" Closed System drug Transfer<br>Device (CSTD) mechanically prohibits<br>environmental contaminants from entering the<br>system and the escape of drug or vapor<br>concentrations from the system, thereby<br>minimizing individual and environmental<br>exposure to drug vapor, aerosols, and spills.<br>The ProSeal™ system also prevents the<br>introduction of microbial contaminations into<br>the drug or fluid path for up to 7 days, when<br>used as intended." | "The ProSeal™ Closed System drug Transfer<br>Device (CSTD) mechanically prohibits<br>environmental contaminants from entering the<br>system and the escape of drug or vapor<br>concentrations from the system, thereby<br>minimizing individual and environmental<br>exposure to drug vapor, aerosols, and spills.<br>The ProSeal™ system also prevents the<br>introduction of microbial contaminations into<br>the drug or fluid path for up to 7 days, when<br>used as intended." | Same |
| Common name | Closed Antineoplastic and Hazardous Drug<br>Reconstitution and Transfer System | Closed Antineoplastic and Hazardous Drug<br>Reconstitution and Transfer System | Same |
| Primary<br>product code | ONB | ONB | Same |
| Primary regulation<br>number | 21 CFR 880.5440 | 21 CFR 880.5440 | Same |
| Intended user<br>population/ intended<br>use environment | Adequately trained<br>health care professionals<br>including pharmacists/ clinical setting | Adequately trained<br>health care professionals<br>including pharmacists/ clinical setting | Same |
| Intended drug type | Parenteral drugs | Parenteral drugs | Same |
| R/ Prescription use | R only | R only | Same |
| Principles of<br>operation | A component device of a multi-component<br>system; devices are intended to be used in<br>combination with other component devices of<br>the ProSeal™ CSTD, manually manipulated | A component device of a multi-component<br>system; devices are intended to be used in<br>combination with other component devices of<br>the ProSeal™ CSTD, manually manipulated | Same |
| Characteristic<br>compared | Predicate Device (K240171)<br>ProSeal™ Injector Plus | Subject Device (K241735)<br>ProSeal™ In Line Pump Set | Comment/<br>Discussion |
| Closed system<br>technology | Mating ProSeal™ devices are sealed with<br>resealing membranes. When devices are joined<br>together, the two membranes are pressed, thus<br>creating a secured fluid path | Mating ProSeal™ devices are sealed with<br>resealing membranes. When devices are joined<br>together, the two membranes are pressed, thus<br>creating a secured fluid path | Same |
| | Physical barrier to prevent all drug masses from<br>crossing the system boundary | Physical barrier to prevent all drug masses from<br>crossing the system boundary | Same |
| Preventing<br>needlestick injuries | The ProSeal™ Injector Plus (K240171) Predicate<br>device, is encapsulated with a cannula and<br>inaccessible to users | The ProSeal™ Injector Plus (K240171), a<br>component/subassembly of the Subject<br>device, is encapsulated with a cannula and<br>inaccessible to users | Same |
| Components | ProSeal™ Injector Plus | ProSeal™ Injector Plus and I.V. bag spike port | Different,<br>See Comment #1 |
| Access<br>port/connector to<br>external transfer<br>device | Access to ProSeal™ Injector Plus is a female<br>Luer lock port by which an external MLL<br>syringe without needle could be connected | Access to ProSeal™ In Line Pump Set is a TPE<br>spiking port stopper, through which a bag<br>spike of an external administration set could be<br>connected | Different.<br>See Comment #2 |
| Spike port<br>subcomponent &<br>material | Epic Medical Pte Ltd's cleared eZSURE™ Empty<br>Fluid Container (K223674), 2-port sub-<br>assembly for IV Bag (consisting of an<br>integrated spike port) | Subject device's spike port is of the same<br>configuration as the eZSURE™ Empty Fluid<br>Container (K223674), 2-port sub-assembly for<br>IV Bag's integrated spike port | Same |
| | PP without DEHP*in K240433, MLL connector<br>of ProSeal™ Injection Site | PP without DEHP*, same as K240433, MLL<br>connector of ProSeal™ Injection Site | Same |
| | eZSURE™ EFC's spiking port stopper is made of<br>TPE | Subject device's spiking port stopper is made<br>of the same TPE | Same |
| Characteristic<br>compared | Predicate Device (K240171)<br>ProSeal™ Injector Plus | Subject Device (K241735)<br>ProSeal™ In Line Pump Set | Comment/<br>Discussion |
| Primary<br>package top web | Medical grade paper and<br>medical plastic film, heat sealed | Medical grade paper and<br>medical plastic film, heat sealed | Same |
| Sterilization method | Ethylene Oxide, EO, SAL 10-6 | Ethylene Oxide, EO, SAL 10-6 | Same |
| Validated shelf life | 3 years (36 months) | 3 years (36 months) | Same |
| Reuse or single-use | Single-use only | Single-use only | Same |
| Labeling<br>specifications | Meets the requirements specified in 21 CFR 801 | Meets the requirements specified in 21 CFR 801 | Same |
| * DEHP - Di (2-ethylhexyl) phthalate (DEHP), a plasticizer to make PVC soft and flexible. It is a substance known to cause cancer or reproductive toxicity, | | | |
{7}------------------------------------------------
{8}------------------------------------------------
#### Discussion of difference in technological characteristic Comment #1
Submission K241735 is for an FDA clearance of the Pump Set, a single component device to the ProSeal" CSTD system of component devices. Compared to Predication was for the addition of another single component device to the ProSeal" Injector Plus, to the same ProSeal" CSTD system of component devices. This dessions of safety and effectiveness as the Subject device is designed to and validated for use with other appropriate component devices of the same intended use. Bench top, biocompatibility, and sterility data defference does not raise different questions in safety and effectiveness. Further details are described in section VII.
#### Comment #2
Subject device's spiking port and spiking port stopper are materials and of the same configurations (design and dimensions) as the eZSURE™ Empty Fluid Container (K223674), 2-port sub-assembly for IV bag's integrated spiking port stopper. Subject device's ProSeal™ Injector Plus is the ProSeal™ Injector Plus (K240171) with no differences in material, design, and dimensions.
{9}------------------------------------------------
#### Performance Data Supporting Substantial Equivalence VII.
#### A. Functional Performance
The sterile, single-use, non-pyrogenic ProSeal™ In Line Pump Set described in this Summary was determined to be in conformance with the following ISO and FDA recognized standards:
- ANSI AAMI CN27:2021, General requirements for Luer activated valves (LAVs) . incorporated into medical devices for intravascular applications
- . ISO 7864: 2016, Sterile hypodermic needles for single use – Requirements and test methods.
- ISO 8536-4:2019, Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed
- . ISO 15747: 2018, Plastic containers for intravenous injections
- ISO 22413:2010, Transfer sets for pharmaceutical preparations — Requirements and test methods
- ISO 23908: 2011, Sharps injury protection Requirements and test methods Sharps ● protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
- . ISO 80369-7: 2016, Small-bore connectors for liquids and gases in healthcare application -Part 7, Connectors for intravascular or hypodermic applications
- . Intravascular-Administration-Sets-Premarket-Notifications-[510(k)]---Guidancefor-Industry-and-FDA-Staff
Bench performance verifications and validations referred to and performed on the Subject device are as follows:
- . Leak integrity testing per ISO 8536-4:2019, paragraph 7.2 and Annex A.3 performed on the Subject device
- . Liquid leakage testing, per ISO 8536-4:2019, paragraph 7.2, performed on the Subject device using NaCl 0.9%
- . Testing of flow rate, infusion set without air-inlet device per ISO 8536-4:2019, paragraph 7.1 and Annex A.5.1, performed on the Subject device
- The following performance testing were performed under K240171, per Luer lock connection ● tests specified in ISO 80369-7: 2021:
- Positive pressure fluid leakage test .
- . Sub-atmospheric pressure air leakage test
- Stress cracking test ●
- Resistance to separation from axial load test ●
- Resistance to separation from unscrewing test
- Resistance to overriding
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "EPIC" in a stylized, bold, sans-serif font. The "E" has a circular dot in the upper left corner. The letters are connected by a thick horizontal line at the bottom, giving the word a solid, unified appearance. The overall design is simple and modern.
- . The following sharps injury protection tests were performed under K240171 per ISO 23908:2011,
- Drop test ●
- Flexural force test ●
- Tensile force test
- Compression force test
- Testing access to the sharp in safe mode ●
- Device leakage integrity tests were performed under K240171 per ISO 8536-4:2019 ●
- Needle bonding strength test was performed under K240171 per ISO 7864:2016
- Injector Plus cap removal force test was performed under K240171 per ISO 8536-4: 2019
- The following performance testing were performed under K223674, per standards specified . hereunder:
- Resistance to temperature, pressure and leakage, per ISO 15747: 2018, Annex A.3 .
- . Resistance to dropping, per ISO 15747: 2018, Annex A.4
- Water vapor impermeability, per ISO 15747: 2018, Annex A.6
- Spiking port penetration ability, per ISO 15747: 2018, Annex A.8 ●
- Spiking port adhesion strength, per ISO 15747: 2018, Annex A.9 ●
- Particulate non-contamination, per ISO 15747: 2018, Annex A.6 ●
- Impermeability to microorganism, per ISO 15747: 2018, Annex C.2 ●
- Testing to (draft) NIOSH CSTD Test Protocol under K240171 ●
- Microbial ingress/ microbiological integrity testing, per the FDA guidance listed in paragraph ● 1 in this sub-section, and ANSI AAMI CN27:2021 (functional) from testing data on the existing devices, cleared under K240171 and K223674
#### B. Biocompatibility
In accordance with ISO 10993-1: 2018, the Subject device is classified as: Externally Communicating Device, Blood Path Indirect, Prolonged Contact (>24hr to 30d). The following testing are referred to from testing on the devices, under K240171 and K223674:
- . Cytotoxicity to ISO 10993-5
- Sensitization to ISO 10993-10 ●
- Intracutaneous Reactivity to ISO 10993-10 .
- Acute Systemic Toxicity to ISO 10993-11 .
- 14-day Subacute/ Subchronic Acute Systemic Toxicity to ISO 10993-11 .
- In-vitro Hemolysis Assessment to ISO 10993-4 ●
- Material Mediated Pyrogenicity to ISO 10993-11 ●
- Chemical Characterization and Toxicological Risk Assessment to ISO 10993-18 and ISO 10993-● 17
- . Particulate matter testing to USP <788> Particulate Matter in Injections
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows a logo for EPIC. The logo is black and white and features the word "EPIC" in a stylized font. The "E" is capitalized and has a circular dot in the upper left corner. A horizontal line runs beneath the letters "E," "P," and "I," connecting them. The "C" is also capitalized and is connected to the horizontal line.
### C. Sterility, Shipping, and Shelf-Life
The Subject device, like the Predicate device and the I.V. bag spike (a component/subassembly of the Subject device), complies with sterilization requirements of ISO 11135:2014, Sterilization of Health Care Products - Ethylene Oxide - Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices and the following testing/ evaluations:
- Package Integrity Tests per ASTM F1980-21, Standard guide for accelerated aging of sterile ● barrier systems for medical devices and Sterile Barrier Packaging Testing performed on the proposed device: Seal strength - ASTM F88/F88M-21, Standard test method for seal strength of flexible barrier materials; Dye Penetration - ASTM F1929-23, Standard test method for detecting seal leaks in porous medical device packaging by dye penetration; EN 868-5:2009, Packaging materials and systems for medical devices which are to be sterilized – Part 5: Heat and self-sealable pouches and reels of paper and plastic film construction – Requirements and test methods
- Pyrogen Tests per ANSI/AAMI ST72/2019, Bacterial endotoxins – Test methods, routing monitoring, and alternatives to batch testing, USP 40 <151>, Pyrogen test (USP rabbit test), USP-NF <161>, Medical Devices-Bacterial Endotoxin and Pyrogen Tests, USP-NF <85>, Bacterial Endotoxins Test
- Shelf-life of 3 years has been validated using the FDA recognized standard, ASTM 1980-21, ● Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
#### VIII. Clinical Tests
Not applicable.
#### IX. Conclusion
The Subject device was determined to be substantially equivalent to the Predicate, and there were no new questions of its safety or effectiveness. The Subject ProSeal™ In Line Pump Set is substantially equivalent to the Predicate, the ProSeal™ Injector Plus, with respect to the indications for use, principles of operation and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.