K240256 · Deka Research and Development · QJY · Jun 12, 2024 · General Hospital
Device Facts
Record ID
K240256
Device Name
Remunity System
Applicant
Deka Research and Development
Product Code
QJY · General Hospital
Decision Date
Jun 12, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Remunity® Pump for Remodulin® (treprostinil) Injection (the Remunity System) is intended for continuous subcutaneous delivery of Remodulin (treprostinil) Injection for use in adults (greater than 22 years old).
Device Story
Wearable infusion pump for continuous subcutaneous delivery of Remodulin (treprostinil) to treat pulmonary arterial hypertension (PAH). System comprises a reusable pump, disposable single-use cassette (pharmacy or user-filled), and a touchscreen remote controller. Pump utilizes micro-dosing mechanism with acoustic volume sensor (AVS) feedback for delivery accuracy. Used in professional healthcare facilities and home environments; operated by patients or caregivers. Healthcare providers program delivery rates; pump provides visual, audible, and vibratory feedback. Touchscreen remote facilitates user interaction and monitoring. Benefits include precise, ambulatory drug delivery for chronic PAH management.
Clinical Evidence
No clinical or animal study data provided. Substantial equivalence is supported by bench testing, including functional performance, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), alarm testing (IEC 60601-1-8), and human factors validation.
Technological Characteristics
Wearable micro-dosing pump; AVS feedback for accuracy. Materials: PC, Bromobutyl, SEBS, polyurethane (cassette); ABS, PC, Aluminum (pump). Power: Rechargeable Li-ion battery. Connectivity: Wireless communication between pump and touchscreen remote. Sterilization: Gamma radiation (ISO 11137-1). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11, IEC 60601-2-24, ISO 10993-1, ISO 14971.
Indications for Use
Indicated for continuous subcutaneous delivery of Remodulin (treprostinil) in adults >22 years old with pulmonary arterial hypertension.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Remunity Subcutaneous Delivery System for Remodulin (K202690)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 12, 2024
Deka Research and Development Paul Smolenski Regulatory Affairs 340 Commercial Street Manchester, New Hampshire 03101
Re: K240256
Trade/Device Name: Remunity System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: Class II Product Code: QJY Dated: Mav 9, 2024 Received: May 10, 2024
Dear Paul Smolenski:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Jake K. Lindstrom -S
Jake Lindstrom, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K240256
Device Name Remunity System
Indications for Use (Describe)
The Remunity® Pump for Remodulin® (treprosimil) Injection (the Remunity System) is intended for continuous subcutaneous delivery of Remodulin (treprostinil) Injection for use in adults (greater than 22 years old).
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "DEKA" in a bold, sans-serif font. The letters are all capitalized and black. The "K" is stylized with an arrow shape pointing to the left.
## 510(k) Summary
This 510(k) Summary of Safety and Effectiveness information is prepared in accordance with the requirements of 21 CFR Part 807.92.
### Submitter Information
| 510(k) Sponsor | DEKA Research & Development<br>340 Commercial Street<br>Manchester, NH 03101 |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person(s) | Paul Smolenski (primary), Lauren Blake (secondary)<br>Regulatory Affairs<br>Phone: (603) 669-5139<br>Fax: (603) 624-0573<br>psmolenski@dekaresearch.com, lblake@dekaresearch.com |
Date Prepared January 30, 2024
### Proposed Device(s)
| Common/Usual Name: | Infusion Pump |
|-------------------------|------------------------------------------------------|
| Trade/Proprietary Name: | Remunity Subcutaneous Delivery System for Remodulin® |
| Classification Name: | Infusion pump, drug specific, pharmacy-filled |
| Device Classification: | 880.5725 |
| Product Code: | QJY |
| Class: | II |
| Device Panel: | General Hospital |
### Predicate Device(s)
The predicate device is the Remunity Subcutaneous Delivery System for Remodulin®, which was previously cleared under premarket application K202690 on December 30, 2020.
### Device Description
The Remunity Subcutaneous Delivery System for Remodulin® (Remunity System) is a wearable infusion pump designed to deliver Remodulin® for the treatment of pulmonary arterial hypertension (PAH). It is intended for continuous subcutaneous delivery of FDA-approved
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "DEKA" in a bold, sans-serif font. The letters are all capitalized and black. The "K" in the word is stylized with an arrow shape in the middle.
Remodulin® (treprostinil) (hereinafter referred to as 'Remodulin®' or 'Remodulin® (treprostinil)'), NDA 021272. The Remunity System consists of several components: a wearable pump assembly, a remote interface, a filling and priming aid, and accessories (e.g., rechargeable batteries, battery charger, charging cable, power adapter). A commercially available subcutaneous infusion set is connected to the pump assembly via a standard luer connector for the delivery of Remodulin® from the Remunity System to the patient.
The pump assembly is composed of a reusable pump and a disposable single-use cassette with a pharmacy-filled or user-filled drug reservoir, which infuses Remodulin® subcutaneously into the patient based on an individualized programmed rate. Each disposable cassette may be used for up to 72 hours after attachment to the pump. The Remunity System utilizes a micro-dosing pump mechanism supplemented with acoustic volume sensor feedback to ensure delivery accuracy.
The device is prescription use only.
The modified device utilizes a touchscreen remote controller.
### Indications for Use
The Remunity® Pump for Remodulin® (treprostinil) Injection (the Remunity System) is intended for continuous subcutaneous delivery of Remodulin (treprostinil) Injection for use in adults (greater than 22 years old).
### Substantial Equivalence Discussion
#### Intended Use Comparison
| Characteristic | Predicate Device - Remunity<br>System (K202690) | Subject Device – Remunity<br>System |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Remunity Subcutaneous<br>Infusion System (the Remunity<br>System) is intended for<br>continuous subcutaneous<br>delivery of Remodulin®<br>(treprostinil) Injection for use in<br>adults (greater than 22 years of<br>age). | The Remunity® Pump for<br>Remodulin® (treprostinil)<br>Injection (the Remunity System)<br>is intended for continuous<br>subcutaneous delivery of<br>Remodulin (treprostinil)<br>Injection for use in adults<br>(greater than 22 years old). |
The tables below includes a matrix of the intended use between the subject device and the predicate device.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "DEKA" in a bold, sans-serif font. The letters are all capitalized and black. The letters are closely spaced together, and the overall appearance is modern and industrial.
| Characteristic | Predicate Device – Remunity<br>System (K202690) | Subject Device – Remunity<br>System |
|--------------------------------------------------|-------------------------------------------------------------------|-------------------------------------|
| Prescription Use or<br>Over the Counter<br>(OTC) | Prescription | No change |
| Intended Patient<br>Population | Greater than 22 years of age | No change |
| Patient Environment | On-body wearable ambulatory<br>pump | No change |
| Environment of Use | In professional healthcare<br>facilities and home<br>environments | No change |
The intended use remains unchanged in the subject device with respect to the predicate device (K202690). The Indications for Use have been updated to reflect the current branding for the device. This update does not affect the intended patient population or environment of use.
#### Technological Characteristic Comparison
The table below compares the characteristics of the predicate and subject devices, and includes an assessment of the differences and why these differences do not introduce new or different questions of safety and effectiveness.
| Characteristic | Predicate Device –<br>Remunity System<br>(K202690) | Subject Device –<br>Remunity System | Equivalence |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Delivery<br>Method | Microprocessor<br>controlled micro-<br>dosing pump<br>mechanism<br>supplemented with<br>acoustic volume sensor<br>(AVS) feedback for<br>monitoring delivery<br>accuracy | Same | N/A |
| Delivery type | Subcutaneous infusion | Same | N/A |
| Dimensions | 6 cm x 6 cm x 2 cm<br>(2.4 in x 2.4 in x 0.4 in) | Same | N/A |
| Weight | 50 g (1.76 oz.) | Same | N/A |
| Basal Delivery<br>Rate Range | 16 μL/hr – 225 μL/hr,<br>with increments of 1<br>μL/hr | Same | N/A |
| Basal Accuracy | +6% | Same | N/A |
| Characteristic | Predicate Device –<br>Remunity System<br>(K202690) | Subject Device -<br>Remunity System | Equivalence |
| Bolus Volume<br>after Occlusion<br>Release | < 40 µL at all rates | Same | N/A |
| Time to<br>Occlusion<br>Alarm | Maximum time to<br>occlusion alarm:<br>• ≥ 100 µL/hr: 12 minutes<br>• ≤ 100 µL/hr: 8 hours | Maximum time to<br>occlusion alarm:<br>• ≥ 100 µL/hr: 12 minutes<br>• ≤ 100 µL/hr: 8 hours | N/A |
| Pump Service<br>Life | 3 years | Same | N/A |
| Ingress<br>Protection | IP58 when connected to<br>the Cassette | Same | N/A |
| Applicable<br>Safety Standards | • IEC 60601-1<br>• IEC 60601-1-2<br>• IEC 60601-1-8<br>• IEC 60601-1-11<br>• IEC 60601-2-24<br>• ISO 11137-1<br>(Sterilized via<br>Gamma Radiation)<br>• ISO 10993-1<br>• ISO 14971 | Same | N/A |
| Power Source | Rechargeable Lithium<br>Ion Battery | Same | N/A |
| Pump Storage<br>Conditions | Temperature: -25 °C to<br>70 °C (-13 °F to 158 °F)<br>Humidity: up to 90%<br>Pressure: 50 kPa to 106 kPa | Same | N/A |
| Operating<br>Conditions | Temperature: 5 °C to 40 °C (41 °F to 104 °F)<br>Humidity: up to 90%<br>Pressure: 70 kPa to 106 kPa | Same | N/A |
| Characteristic | Predicate Device –<br>Remunity System<br>(K202690) | Subject Device –<br>Remunity System | Equivalence |
| Remote<br>Interface | • Push-button device | • Touchscreen device | Equivalent<br>The subject device provides a touchscreen device for use with the pump. The text and format of information displayed to the user, as well as workflows and required user input remains unchanged. No additional risks have been introduced. |
| System User<br>Feedback | Visual, audible, vibratory | Same | N/A |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "DEKA" in a bold, sans-serif font. The letters are all capitalized and black. The letters are closely spaced together, and the overall impression is one of strength and stability.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "DEKA" in a bold, stylized font. The letters are black and have a unique design, with the "E" having three horizontal lines and the "K" resembling an arrow pointing to the left. The "A" has a flat top and a horizontal line through the middle, giving the word a distinctive and modern appearance. The overall design is simple yet eye-catching.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "DEKA" in a bold, sans-serif font. The letters are all capitalized and black. The letters are close together, and the overall impression is one of strength and stability. The background is white.
| Characteristic | Predicate Device –<br>Remunity System<br>(K202690) | Subject Device –<br>Remunity System | Equivalence |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|-------------|
| Common<br>Alarms | Battery Depleted<br>Cassette Depleted<br>Cassette Removed<br>Cassette Problem<br>Occlusion<br>Pump Error<br>Pump Failure<br>Basal Not Started<br>Adjust Pump<br>Delivery Stopped<br>Depletes Soon<br>Excessive Noise<br>Message Timeout<br>No Communication<br>Pairing Failed<br>Pairing Lost<br>Pump Idle<br>Pump Battery Low<br>Remote Battery Low<br>Request Refused<br>Software Version Error<br>Tech<br>Walkaway | Same | N/A |
| Battery<br>Operating Time | 72 hours | Same | N/A |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "DEKA" in a bold, stylized font. The letters are black and have a unique design, with rounded edges and geometric shapes. The overall appearance is modern and industrial. The letters are all capitalized and evenly spaced.
| Characteristic | Predicate Device –<br>Remunity System<br>(K202690) | Subject Device –<br>Remunity System | Equivalence |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|-------------|
| Infusion Set | Medtronic Quick-set<br>Infusion Set [K991759]<br>42" (MMT-390, MMT-391)23" (MMT-392, MMT-393)<br>Medtronic Silhouette<br>Infusion Set [K162812]<br>23" (MMT-373)<br>Smiths Medical Cleo 90<br>Infusion Set [K042172]<br>24" (21-7220-24, 21-7230-24)31" (21-7221-24, 21-7231-24)42" (21-7222-24, 21-7232-24) | Same | N/A |
| Compatible<br>Cassette | Remodulin Unity<br>Cassettes, 3 mL | Same | N/A |
| Cassette Filling<br>Method | Specialty pharmacyUser-filled | Same | N/A |
| Cassette Shelf<br>Life (empty) | 2 years | Same | N/A |
| Pharmacy-filled<br>Shelf-Life | 14 days (per <USP<br>797>) | Same | N/A |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "DEKA" in a bold, sans-serif font. The letters are all capitalized and black. The "K" is stylized with a sharp, arrow-like shape pointing to the left. The background is white.
| Characteristic | Predicate Device –<br>Remunity System<br>(K202690) | Subject Device –<br>Remunity System | Equivalence |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|-------------|
| Material<br>Biocompatibility | Cassette fluid path:<br>Polycarbonate (PC),<br>Bromobutyl, SEBS,<br>polyurethane | Same | N/A |
| | Pump: ABS, PC,<br>Aluminum | | |
| | Cartridge: PC, Acrylic,<br>polyurethane | | |
| | Filling Aid: PC-ABS | | |
| | Cap (Pharmacy-fill):<br>ABS | | |
| | | | |
| Pharmacy-fill<br>End-user<br>Packaging | Aseptically filled<br>cassette with female<br>luer lock fluid path<br>closure, placed in<br>plastic clamshell tray<br>and sealed in foil pouch | Same | N/A |
| Priming Method | Manually primed by<br>user before Cassette<br>attachment to the Pump | Same | N/A |
#### Performance Data
Performance testing was conducted in order to verify and validate the subject device remote interface for its intended use, and to establish substantial equivalence to the respective predicate device remote interface in terms of safety and effectiveness. The FDA Guidance Document, Infusion Pumps Total Product Life Cycle, issued December 2, 2014 was followed.
| Device<br>Performance | The following functional areas, specific to the remote interface, were evaluated:<br>Touchscreen Remote Reliability Touchscreen Remote Functionality Wireless Communication |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Software and<br>Cybersecurity | The subject device is considered to require an “Enhanced Documentation Level” per the 2023 FDA guidance document. |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the word "DEKA" in a bold, sans-serif font. The letters are all capitalized and black. The letters are stylized with rounded corners and a slightly futuristic look. The "K" is designed with an arrow-like shape pointing to the left.
| | Software and Cybersecurity testing was conducted per ANSI AAMI IEC<br>62304:2006/A1:2016 and the following FDA Guidance Documents: |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | • Content of Premarket Submissions for Device Software Functions,<br>issued June 14, 2023 |
| | • Off-The-Shelf Software Use in Medical Devices, issued September 27,<br>2019 |
| | • Content of Premarket Submissions for Management of Cybersecurity<br>in Medical Devices, issued September 27, 2023 |
| | • Cybersecurity in Medical Devices: Quality System Considerations and<br>Content of Premarket Submissions, issued April 8, 2020 |
| | • Cybersecurity in Medical Devices: Refuse to Accept Policy for Cyber<br>Devices and Related Systems Under Section 524B of the FD&C Act,<br>issued March 30, 2023 |
| Human Factors | The human factors evaluation followed the FDA Guidance Document,<br>Applying Human Factors and Usability Engineering to Medical Devices,<br>issued February 3, 2016, and the following standards: |
| | • IEC 60601-1-6:2020 |
| | • ANSI AAMI IEC 62366-1:2015+A1:2020 |
| Electrical Safety | Per IEC 60601-1:2020 |
| EMC | Per IEC 60601-1-2:2020 |
| Alarms | Per IEC 60601-1-8:2020 |
#### Animal and Clinical Studies
No animal study or clinical trial data was obtained in support of this premarket submission.
#### Design Control
The Remunity System was specified and developed by DEKA. DEKA complies with the FDA Quality System Regulation as specified in 21 CFR 820, as well as to ISO 13485:2016.
### Conclusion
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The Remunity Subcutaneous Delivery System for Remodulin® is substantially equivalent to the Remunity Subcutaneous Delivery System for Remodulin® cleared under K202690 with respect to the indications for use, target populations, treatment method, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.