Remunity Pump for Remodulin (treprostinil) Injection
K202690 · Deka Research and Development · QJY · Dec 30, 2020 · General Hospital
Device Facts
Record ID
K202690
Device Name
Remunity Pump for Remodulin (treprostinil) Injection
Applicant
Deka Research and Development
Product Code
QJY · General Hospital
Decision Date
Dec 30, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Remunity Subcutaneous Infusion System (the Remunity System) is intended for continuous subcutaneous delivery of Remodulin® (treprostinil) Injection for use in adults (greater than 22 years of age).
Device Story
Wearable infusion pump for continuous subcutaneous delivery of Remodulin (treprostinil) to treat pulmonary arterial hypertension; consists of reusable pump, disposable pharmacy-filled cassette, remote interface, and accessories. Uses micro-dosing pump mechanism with acoustic volume sensor (AVS) feedback for delivery accuracy. Used in professional healthcare facilities and home environments by patients. Pharmacy-filled cassettes are stored up to 14 days and used for up to 72 hours. Healthcare providers program delivery rates; device provides audible/vibratory alerts for errors, occlusions, or battery status. Benefits include precise, automated drug delivery for chronic condition management.
Clinical Evidence
No clinical data was obtained in support of this premarket submission. Bench testing only.
Technological Characteristics
Microprocessor-controlled micro-dosing pump; acoustic volume sensor (AVS) feedback. Materials: cassette fluid path (polycarbonate, bromobutyl, SEBS, polyurethane), pump (ABS, polycarbonate, aluminum), luer lock cap (ABS). Dimensions: 6cm x 6cm x 2cm; weight: 50g. Power: rechargeable Li-ion battery. Connectivity: remote interface. Ingress protection: IP58. Biocompatibility: ISO 10993-1 compliant. Sterilization: SAL 10^-6 for luer lock cap.
Indications for Use
Indicated for continuous subcutaneous delivery of Remodulin (treprostinil) injection in adults >22 years of age with pulmonary arterial hypertension.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Unity Subcutaneous Delivery System for Remodulin (K191313)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
December 30, 2020
DEKA Research and Development Paul Smolenski Regulatory Affairs 340 Commercial Street Manchester, New Hampshire 03101
Re: K202690
Trade/Device Name: Remunity Subcutaneous Delivery System for Remodulin Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: QJY Dated: December 4, 2020 Received: December 4, 2020
Dear Paul Smolenski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For CAPT. Alan Stevens Acting Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202690
Device Name
Remunity Subcutaneous Delivery System for Remodulin®
Indications for Use (Describe)
The Remunity Subcutaneous Infusion System (the Remunity System) is intended for continuous subcutaneous delivery of Remodulin® (treprostinil) Injection for use in adults (greater than 22 years of age).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K202690
#### Submitter Information
| 510(k) Sponsor: | DEKA Research & Development<br>340 Commercial Street<br>Manchester, NH 03101 |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Paul Smolenski<br>Regulatory Affairs<br>DEKA Research & Development Corporation<br>Phone: (603) 669-5139<br>Fax: (603) 624-0573<br>psmolenski@dekaresearch.com |
| Date Prepared: | December 18, 2020 |
#### Proposed Device
| Common/Usual Name: | Infusion Pump |
|-------------------------|----------------------------------------------------|
| Trade/Proprietary Name: | Remunity Subcutaneous Delivery System for |
| | Remodulin® |
| Classification Name: | Infusion Pump |
| Device Classification: | 880.5725 |
| Product Code: | QJY, Infusion pump, drug specific, pharmacy-filled |
| Class | II |
| Device Panel: | General Hospital |
#### Predicate Device
Unity Subcutaneous Delivery System for Remodulin K191313.
No reference devices are used in this submission.
#### Device Description
The Remunity Subcutaneous Delivery System for Remodulin® (Remunity System) is a wearable infusion pump designed to deliver Remodulin® for the treatment of pulmonary arterial hypertension (PAH). It is intended for continuous subcutaneous delivery of FDA-approved Remodulin® (treprostinil) (hereinafter referred to as 'Remodulin®' or 'Remodulin® (treprostinil)'), NDA 021272. The Remunity System consists of several components: a wearable pump assembly, a remote interface, a filling and priming aid, and accessories (e.g., rechargeable batteries, battery charger, charging cable, power adapter). A commercially available subcutaneous infusion set is connected to the pump assembly via a standard luer connector for the delivery of Remodulin® from the Remunity System to the patient.
{4}------------------------------------------------
The pump assembly is composed of a reusable pump and a disposable single-use cassette with a pharmacy-filled drug reservoir, which infuses Remodulin® subcutaneously into the patient based on an individualized programmed rate. Each disposable cassette may be used for up to 72 hours after attachment to the pump. The Remunity System utilizes a micro-dosing pump mechanism supplemented with acoustic volume sensor feedback to ensure delivery accuracy.
The pharmacy-filled cassette is intended to be stored (inclusive of shipping time) up to 14-days and is intended to be in use for up to 72-hours. The cassettes are identical to those cleared under K191313, with the exception that the luer lock cap material was changed from Polycarbonate to ABS.
The device is prescription use only.
# Indications for Use
The Remunity Subcutaneous Infusion System (the Remunity System) is intended for continuous subcutaneous delivery of Remodulin® (treprostinil) Injection for use in adults (greater than 22 years of age).
## Substantial Equivalence Discussion
## Intended Use Comparison
The table below includes a summation matrix of the intended use between the new device and those of the current device:
| Characteristic | Unity Subcutaneous Deliver<br>System for Remodulin®<br>K191313 | Proposed<br>Remunity Subcutaneous Deliver<br>System for Remodulin® |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Indications for Use | The Unity Subcutaneous<br>Delivery System for<br>Remodulin® (the Unity System)<br>is intended for continuous<br>subcutaneous delivery of<br>Remodulin® (treprostinil)<br>injection for use in adults<br>(greater than 22 years of age). | Same Indications for Use. Branding<br>has been updated from "Unity" to<br>"Remunity". |
| Prescription Only<br>or Over the Counter | Prescription Only | No Change |
| Intended Population | Adults (>22 years of age) | No Change |
| Environment of Use | In professional healthcare<br>facility and home healthcare<br>environments | No Change |
Discussions of differences in Indications for Use statement
{5}------------------------------------------------
The indications for use for the subject device is the same as the device cleared under K191313. The branding of the device has been updated from "Unity" to "Remunity".
#### Discussions of differences in intended population
The intended population for the subject device is identical to the predicate device cleared under K191313.
The Remunity system is indicated for adults (greater than 22 years of age)
#### Discussions of differences in environment of use
The environment of use for the subject device is identical to the predicate device cleared under K191313.
## Comparison of Technological Characteristics with the Predicate Device
The below table compares the characteristics of the subject device to the Unity System cleared under K191313 and includes an assessment of differences between them and why the differences between the subject device and predicate device do not introduce new or different questions of safety and effectiveness.
| Characteristic | Predicate Device<br>Unity Subcutaneous<br>Deliver System for<br>Remodulin®<br>K191313 | Subject Device<br>Remunity<br>Subcutaneous Deliver<br>System for<br>Remodulin® | Discussion of<br>Differences |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mechanism of<br>action | Microprocessor controlled<br>Micro- dosing pump<br>mechanism supplemented<br>with acoustic volume sensor<br>(AVS) feedback for<br>monitoring delivery<br>accuracy. | Same | N/A |
| Infusion<br>Accuracy | $\pm$ 6% | Same | N/A |
| Maximum<br>Infusion pressure | <16.4 psi (<113 kPa) | Same | N/A |
| Programmable<br>Flow rate ranges | 16 µl/hr to 225 µl/hr with<br>increments of 1 μl/hr | Same | N/A |
| Time to occlusion<br>alarm | Maximum time to occlusion<br>alarm:<br><12 min. at rates ≥ 100 µl/hr<br>within 8 hr, at rates < 100<br>µl/hr | Same | N/A |
| Characteristic | Predicate Device<br>Unity Subcutaneous<br>Deliver System for<br>Remodulin®<br>K191313 | Subject Device<br>Remunity<br>Subcutaneous Deliver<br>System for<br>Remodulin® | Discussion of<br>Differences |
| Post-occlusion<br>bolus | <40 µl at all rates. | Same | N/A |
| Alarms & Alerts | • Battery depleted<br>• Battery Low (pump)<br>• Battery Low (remote)<br>• Cassette Depleted<br>• Cassette Problem<br>• Cassette Removed<br>• Depletes Soon<br>• Pump Error<br>• Pump Failure<br>• Occlusion<br>• Delivery Stopped<br>• Basal Not Started<br>• Idle<br>• Software Version Error<br>• Tech<br>• Excessive Noise<br>• No Communication<br>• Message Timeout<br>• Pairing Failed<br>• Walkaway | Same | N/A |
| Device Service<br>Life | 3 years | Same | N/A |
| Characteristic | Predicate Device<br>Unity Subcutaneous<br>Deliver System for<br>Remodulin®<br>K191313 | Subject Device<br>Remunity<br>Subcutaneous Deliver<br>System for<br>Remodulin® | Discussion of<br>Differences |
| Dimensions &<br>Weight | 6 cm x 6 cm x 2 cm<br>(2.4 in x 2.4 in x 0.4 in)<br>50 g (1.76 oz) | Same | N/A |
| Materials | Cassette fluid path:<br>Polycarbonate, Bromobutyl,<br>SEBS, polyurethane<br><br>Pump: ABS, Polycarbonate,<br>Aluminum<br><br>Cartridge: Polycarbonate,<br>Acrylic, polyurethane<br><br>Filling Aid: PC-ABS<br><br>Luer Lock Cap:<br>Polycarbonate | Cassette fluid path:<br>Polycarbonate, Bromobutyl,<br>SEBS, polyurethane<br><br>Pump: ABS, Polycarbonate,<br>Aluminum<br><br>Cartridge: Polycarbonate,<br>Acrylic, polyurethane<br><br>Filling Aid: PC-ABS<br><br>Luer Lock Cap: ABS | The change in luer lock cap<br>material was evaluated to the<br>same standards for<br>biocompatibility under ISO-<br>10993-1 as the predicate.<br>The revised material<br>demonstrates equivalent<br>biocompatibility.<br><br>Functional (container<br>closure) testing was<br>performed. Subject device<br>was found to be equivalent<br>to the predicate. |
| Environment of<br>Use | In professional healthcare<br>facility and home healthcare<br>environments | Same | N/A |
| Ingress protection | IP58 when connected to the<br>reservoir | Same | N/A |
| Power source | Rechargeable Lithium-Ion<br>Battery | Same | N/A |
| Storage<br>Conditions | Temperature:<br>-13°F to 158°F (-25°C to<br>70°C)<br><br>Non-condensing humidity:<br>up to 90%.<br><br>Pressure: 500 hPa to 1060<br>hPa | Same | N/A |
| Characteristic | Predicate Device<br>Unity Subcutaneous<br>Deliver System for<br>Remodulin®<br>K191313 | Subject Device<br>Remunity<br>Subcutaneous Deliver<br>System for<br>Remodulin® | Discussion of<br>Differences |
| Operating<br>Conditions | Temperature:<br>41°F to 104°F (5°C to 40°C)<br>Non-condensing humidity:<br>up to 90%<br>Pressure: 700 hPa to 1060<br>hPa | Same | N/A |
| Remote user<br>feedback | Audible, vibratory | Same | N/A |
| Administration<br>Set | Medtronic Quick-set<br>Infusion Set Medtronic<br>Silhouette and Infusion Set<br>Smiths Medical Cleo 90<br>Infusion Set | Same | N/A |
| Cassette | Remodulin Unity cassettes,<br>3 ml, Specialty Pharmacy<br>filled | Same | N/A |
| Expiration -<br>Cassette | Pharmacy-Fill, 14 Days<br>(consistent with USP 797) | Same | N/A |
| End User<br>Packaging - | Aseptically filled cassette<br>with female luer lock fluid<br>path closure, placed in<br>plastic clamshell tray and<br>sealed in foil pouch | Same | N/A |
| Biocompatibility | Complies with ISO 10993-1<br>and the Agency's "Use of<br>International Standard ISO<br>10993-1, 'Biological<br>Evaluation of Medical<br>Devices - Part 1: Evaluation<br>and testing within a Risk<br>Management Process' -<br>Guidance for Industry and<br>Food and Drug<br>Administration Staff" dated<br>June 16, 2016 | Same | N/A |
| Luer Lock Cap<br>packaging | Tyvek/LDPE pouch | Same | N/A |
| Characteristic | Predicate Device<br>Unity Subcutaneous<br>Deliver System for<br>Remodulin®<br>K191313 | Subject Device<br>Remunity<br>Subcutaneous Deliver<br>System for<br>Remodulin® | Discussion of<br>Differences |
| Luer Lock Cap<br>sterility assurance<br>level | SAL 10-6 | Same | N/A |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
# Non-Clinical/ Performance Testing:
The safety assurance case provided for the Remunity Subcutaneous Delivery System for Remodulin® as recommended in the FDA guidance document, "Infusion Pumps Total Product Life Cycle" and cleared under premarket notification K191313 applies to the subject device in all aspects.
The stated goal of the safety assurance case ("safety case") is: "The Remunity Subcutaneous Infusion System for Remodulin® is adequately safe for its intended use."
The assurance case defined the device system, including the indications for use, system definition, operational description, patient populations, and use environments. The assurance case was updated to reflect the new luer lock cap material and continues to cover all system attributes that were stated for the predicate:
- All hazards associated with the system have been identified and adequately addressed
- . Device reliability is adequate
- The device design requirements are adequately verified and validated ●
The arguments specific to the luer lock cap material change were updated from that of the predicate K191313. Since the assurance case builds on the original case, only evidence used to provide assurance related to the changes are listed here:
| Container Closure<br>performance | Performance of the device to maintain adequate assurance of<br>protection from microbial ingress was evaluated through container<br>closure testing. |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | The materials used for the Remunity System comply with<br>biocompatibility requirements outlined in ISO 10993-1:2009 and the<br>"Guidance for Industry and Food and Drug Administration Staff, Use<br>of International Standard ISO 10993 Biological Evaluation of Medical<br>Devices Part 1: Evaluation and Testing within a Risk Management<br>Process" and are considered to be biocompatible. |
{10}------------------------------------------------
## Clinical Study
No clinical data was obtained in support of this premarket submission.
#### Design Control
The Remunity System was specified and developed by DEKA. DEKA is a registered medical device specification developer and complies with the FDA Quality System Regulation as specified in 21 CFR 820, as well as registered to ISO 13485.
#### Conclusion
The modifications in the luer lock cap material do not change the intended use or technological characteristics of the subject device compared to the Unity System cleared under premarket notification K191313. The changes summarized in this submission do not raise different questions of safety and effectiveness.
The performance of the device is supported by DEKA's design control process which included non-clinical testing and risk management activities. The Remunity Subcutaneous Delivery System for Remodulin® is Substantially Equivalent (SE) to the Remunity Subcutaneous Delivery System for Remodulin® cleared under premarket notification K191313.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.