The Unomedical Comfort™ Subcutaneous Infusion Set and Unomedical Comfort Short™ Subcutaneous Infusion Set are indicated for subcutaneous infusion of insulin administered by an external pump. The Unomedical Neria Soft Subcutaneous Infusion Set is indicated for subcutaneous infusion of medication administered by an external pump. The Medtronic Silhouette® Subcutaneous (Luer Lock) Infusion Set is indicated for subcutaneous infusion of insulin from and infusion pump. The Medtronic Silhouette Paradigm® Subcutaneous Infusion Set is indicated for use with Medtronic Paradigm Insulin Subcutaneous Infusion Pumps for continuous subcutaneous insulin infusion by patients or caregivers in the home environment. The Roche Accu-Chek Tender™ Subcutaneous Infusion Set is indicated for subcutaneous infusion of insulin administered with microdosage insulin pumps. The Asante Comfort™ Subcutaneous Infusion Set is indicated for subcutaneous infusion of insulin administered by the Snap™ Insulin Pump System. The Abbott Comfort™ Subcutaneous Infusion Set is indicated for infusion of fluids into the body below the surface of the skin when attached to a fluid reservoir of a compatible Abbott pump.
Device Story
Sterile, single-use subcutaneous infusion sets; consist of indwelling soft cannula assembly (PFOA-free PTFE cannula, adhesive fixation, injection port, click-lock connector) and infusion tubing set. Insertion facilitated by stainless-steel needle; needle withdrawn post-insertion. Tubing connects to external insulin/medication pumps via luer lock or proprietary connectors. Used in clinical or home environments by patients/caregivers. Provides continuous delivery of medication/insulin to subcutaneous tissue. Benefits include reliable, consistent drug delivery for chronic conditions like diabetes.
Clinical Evidence
Bench testing only. Testing included flow rate verification (40 ml/min), leak testing at various pressures (0.56-3 bar), dynamic pull tests (3N-15N) for component bonds, and bend testing (3000 cycles) to ensure structural integrity. Biocompatibility of the new PFOA-free PTFE material was evaluated per ISO 10993.
Indicated for subcutaneous infusion of insulin or medication via external infusion pumps in patients requiring such therapy, including home use by patients or caregivers.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Maersk Medical Pureline Comfort Subcutaneous Infusion Set (K972135)
Abbott Comfort Subcutaneous Infusion Set (K051264)
Asante Comfort Subcutaneous Infusion Set (K120872)
Medtronic Silhouette Paradigm Infusion Set (K160648)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines extending from the profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 1, 2017
Unomedical A/S % Lee Leichter President P/1 Biomedical 10882 Stonington Avenue Fort Myers, Florida 33913
Re: K162812
Trade/Device Name: Unomedical Comfort Subcutaneous Infusion Set, Unomedical Comfort Short Subcutaneous Infusion Set. Unomedical Neria Soft Subcutaneous Infusion Set, Medtronic Silhouette Subcutaneous Infusion Set, Medtronic Silhouette Paradigm Subcutaneous Infusion Set, Roche Accu-chek Tender Subcutaneous Infusion Set, Asante Comfort Subcutaneous Infusion Set. Abbott Comfort Subcutaneous Infusion Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: August 31, 2017 Received: September 1, 2017
Dear Lee Leichter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image contains a signature that reads "Tina Kiang". The signature is written in a cursive style. There is also a scribble to the left of the name.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162812
#### Device Name
Unomedical Comfort™ Subcutaneous Infusion Set Unomedical Comfort Short™ Subcutaneous Infusion Set Unomedical Neria Soft Subcutaneous Infusion Set
Indications for Use (Describe)
The Unomedical Comfort™ Subcutaneous Infusion Set and Unomedical Comfort Short™ Subcutaneous Infusion Set are indicated for subcutaneous infusion of insulin administered by an external pump.
The Unomedical Neria Soft Subcutaneous Infusion Set is indicated for subcutaneous infusion of medication administered by an external pump.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CER 801 Subpart D) | <span style="font-family: Arial, sans-serif;">☐</span> Over-The-Counter Use (21 CER 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162812
#### Device Name
Medtronic Silhouette® Subcutaneous (Luer Lock) Infusion Set Medtronic Silhouette Paradigm® Subcutaneous Infusion Set
#### Indications for Use (Describe)
The Medtronic Silhouette® Subcutaneous (Luer Lock) Infusion Set is indicated for subcutaneous infusion of insulin from and infusion pump.
The Medtronic Silhouette Paradigm® Subcutaneous Infusion Set is indicated for use with Medtronic Paradigm Insulin Subcutaneous Infusion Pumps for continuous subcutaneous insulin infusion by patients or caregivers in the home environment.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Indications for Use
### 510(k) Number (if known) K162812
#### Device Name
Roche Accu-Chek Tender™ Subcutaneous Infusion Set Asante Comfort™ Subcutaneous Infusion Set Abbott Comfort™ Subcutaneous Infusion Set
#### Indications for Use (Describe)
The Roche Accu-Chek Tender™ Subcutaneous Infusion Set is indicated for subcutaneous infusion of insulin administered with microdosage insulin pumps.
The Asante Comfort™ Subcutaneous Infusion Set is indicated for subcutaneous infusion of insulin administered by the Snap™ Insulin Pump System.
The Abbott Comfort™ Subcutaneous Infusion Set is indicated for infusion of fluids into the body below the surface of the skin when attached to a fluid reservoir of a compatible Abbott pump.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## "510(K) SUMMARY"
Submitted By/Contact Person: John M. Lindskog President Unomedical A/S Infusion Devices Aaholmvej 1-3, Osted DK-4320 Leire, Denmark Telephone: + 45 46 42 78 10 (CET 0800-1700) Fax: + 45 46 42 78 65 E-mail: John.Lindskog@convatec.com
Date Prepared: 27 August 2017
- 7.1 510(k) Number: K162812
- 7.2 Trade/Proprietary Name: Unomedical Comfort™ Subcutaneous Infusion set
Unomedical Comfort Short™ Subcutaneous Infusion Set Unomedical Neria Soft Subcutaneous Infusion Set Medtronic Silhouette® Subcutaneous (Luer Lock) Infusion Set Medtronic Silhouette Paradigm® Subcutaneous Infusion Set Roche Accu-Chek Tender™ Subcutaneous Infusion Set Asante Comfort™ Subcutaneous Infusion Set Abbott Comfort™ Subcutaneous Infusion Set
- 7.3 Common/Usual Name Subcutaneous Infusion Set
- 7.4 Classification Name Intravascular Administration Set
- 7.5 Classification Class: II Panel: 80 Product Code: FPA Regulation: 21 CFR 880.5440
#### 7.6 Purpose of Submission
To accurately describe the currently marketed iterates of the Comfort Subcutaneous Infusion set line manufactured in multiple configurations, in multiple facilities and marketed under several brand names by several different companies.
The only changes to the devices included in this submission that were not previously cleared were the change in sterilization facility and the change to a PFOA free silicone cannula, which was done under the U.S. Environmental Protection Agency PFOA Stewardship Program. These changes are supported by this submission.
- 7.7 Substantial Equivalence
{6}------------------------------------------------
The Comfort™ subcutaneous infusion Sets are substantially equivalent to the Subcutaneous Infusion Sets cleared under the following 510(k)s: K972135 - Maersk Medical Pureline Comfort Subcutaneous Infusion Set K051264 - Abbott Comfort™ Subcutaneous Infusion Set K120872 – Asante Comfort™ Subcutaneous Infusion Set K160648 – Medtronic Silhouette® Paradigm Infusion Set
#### 7.8 Predicate Devices
The devices that serve as predicates for each of the subject devices is identified in Table 1.
#### 7.9 Device Description
The Comfort Subcutaneous Infusion Sets are sterile, non-pyrogenic, single use subcutaneous infusion sets. The current sets are designed to be used with commercially available infusion devices or are indicated for a specific pump where the set has a proprietary pump reservoir connection (e.g. Medtronic, Asante). Each has two basic components provided for each device. The first is a stand-alone subcutaneous indwelling Soft Cannula. This component of the set is provided as an integral assembly with a PFOA-Free PTFE soft cannula, adhesive backed fixation tape, an injection port and the female portion of a proprietary plastic "click-lock" connector. The assembly comes with a stainless-steel insertion needle. The insertion needle is mounted to a male portion of the proprietary plastic "click-lock" connector. The insertion needle comes to the user inserted through the injection port and the inner lumen of the soft cannula with the needle end protruding past the tip of the soft cannula. The male connector is locked to the female connector on the indwelling soft cannula. A needle protector is assembled over the soft cannula and the insertion needle. A separate male portion of the proprietary connector without the insertion needle is provided in the package. This component is used to attach to the female connector after the indwelling soft cannula has been inserted and the steel insertion needle has been withdrawn and protects the indwelling cannula when the infusion set is not attached. Each Soft cannula set comes individually packaged in its own blister pack sealed with paper lid stock.
The second component is the infusion tubing set. The infusion tubing set for all sets are comprised of a co-extruded tube with a stainless-steel needle incorporated into the male portion of the proprietary plastic "click-lock" connector at the patient end. The proximal end of the current Comfort set terminates in either a standard luer lock connector or a proprietary connector compatible with the specified pump. The sets come individually packaged in blister packs sealed with paper lid stock.
#### 7.10 -Indication for Use
The Indications for use for each set is provided in Table 1.
#### 7.11 Intended Use
The intended use of these sets has not changed. These sets remain intended to be used for subcutaneous infusion of insulin or medication when connected to a compatible pump. The specific drug (insulin or medication) and the specific compatible pump differ depending on the intended use of each infusion set.
## 7.12 Technological Characteristics
{7}------------------------------------------------
The technological characteristics of all the Comfort Subcutaneous infusion Sets (regardless of the branding) have not changed. The device's operating principle(s) or mechanism of action, configurations (models, lengths, etc.), performance specifications, and packaging have not changed. All changes in the materials of construction or design of any of the devices have already been submitted on at least one of the family of devices. All of these changes have been previously cleared except the change from of the PFTE Soft Cannula to a PFOA-Free PTFE Soft Cannula (due to mandatory material discontinuation by the suppliance with the U.S. Environmental Protection Agency PFOA Stewardship Program); which is cleared in this submission. This new material was extracted and analyzed per ISO 10993 and found to substantially equivalent to the PTFE material. This submission is to ensure that the information at the FDA on these devices is consistent with the latest configurations. Table 1 provides a comparison of the key characteristics of the devices.
#### 7.13 -Performance Data
The following verification testing was performed:
- 7.8.1 Flow test:
7.8.1.1 Tubing Set - 40 ml/min at 1 bar pressure
- 7.8.2 Leak test:
- 7.8.2.1 Luer-Lock - No Leaks at 3 bar pressure for 30 seconds
- 7.8.2.2 Pcap - No Leaks at 1.4 bar pressure for 30 seconds
- Asante Hub No Leaks at 0.56 bar pressure for 30 seconds 7.8.2.3
## 7.8.3 Pull tests:
- Soft Cannula to Cannula housing: 3 N, Dynamic Pull 7.8.3.1
- 7.8.3.2 Tubing to tubing Luer-lock: 15 N, Dynamic pull
- Tubing to Needle Connector: 15 N. Dynamic pull 7.8.3.3
- Cannula housing to Needle Connector: 15 N, Dynamic pull 7.8.3.4
- 7.8.3.5 Cannula housing to Adhesive: 15 N, Dynamic pull
## 7.8.4 Bend test:
- No leaks and no break at the tubing after 3000 bendings 7.8.4.1
#### 7.14 Conclusion
Unomedical A/S confirmed that the devices passed all testing and concluded that the Subject Devices are substantially equivalent to the predicate devices currently legally marketed in the USA.
{8}------------------------------------------------
# TABLE 1
| SUBJECT<br>DEVICE BRAND<br>NAMES | UNOMEDICAL<br>COMFORT | UNOMEDICAL<br>COMFORT -<br>SHORT | MEDTRONIC<br>SILHOUETTE®<br>LUER LOCK<br>INFUSION SET | ROCHE<br>ACCUCHECK<br>TENDER | UNOMEDICAL<br>NERIA SOFT | MEDTRONIC<br>SILHOUETTE®<br>PARADIGM® | ABBOTT COMFORT | ASANTE<br>COMFORT |
|---------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PREDICATE<br>510(K) FILE<br>NUMBER | K972135 | K972135 | K972135 | K972135 | K972135 | K160648 | K051264* | K120872 |
| PREDICATE<br>NAME | MAERSK<br>MEDICAL<br>PURELINE<br>COMFORT<br>SUBCUTANEOUS<br>INFUSION SET | MAERSK<br>MEDICAL<br>PURELINE<br>COMFORT<br>SUBCUTANEOUS<br>INFUSION SET | MAERSK<br>MEDICAL<br>PURELINE<br>COMFORT<br>SUBCUTANEOU<br>S INFUSION SET | MAERSK<br>MEDICAL<br>PURELINE<br>COMFORT<br>SUBCUTANEO<br>US INFUSION<br>SET | MAERSK<br>MEDICAL<br>PURELINE<br>COMFORT<br>SUBCUTANEOU<br>S INFUSION SET | MEDTRONIC<br>SILHOUETTE®<br>PARADIGM® | ABBOTT COMFORT<br>SUBCUTANEOUS<br>INFUSION SET | ASANTE<br>COMFORT<br>SUBCUTANEOUS<br>INFUSION SET |
| CLEARED<br>INDICATIONS<br>FOR USE | INFUSION<br>AND/OR<br>INJECTION OF<br>FLUIDS INTO THE<br>BODY BELOW<br>THE SURFACE<br>OF THE SKIN. | INFUSION<br>AND/OR<br>INJECTION OF<br>FLUIDS INTO THE<br>BODY BELOW<br>THE SURFACE<br>OF THE SKIN. | INFUSION<br>AND/OR<br>INJECTION OF<br>FLUIDS INTO<br>THE BODY<br>BELOW THE<br>SURFACE OF<br>THE SKIN. | INFUSION<br>AND/OR<br>INJECTION OF<br>FLUIDS INTO<br>THE BODY<br>BELOW THE<br>SURFACE OF<br>THE SKIN. | INFUSION<br>AND/OR<br>INJECTION OF<br>FLUIDS INTO<br>THE BODY<br>BELOW THE<br>SURFACE OF<br>THE SKIN. | FOR USE WITH<br>MEDTRONIC<br>PARADIGM<br>INSULIN<br>SUBCUTANEOUS<br>INFUSION PUMPS<br>FOR<br>CONTINUOUS<br>SUBCUTANEOUS<br>INSULIN<br>INFUSION BY<br>PATIENTS OR<br>CAREGIVERS IN<br>THE HOME<br>ENVIRONMENT. | INFUSION OF<br>FLUIDS INTO THE<br>BODY BELOW THE<br>SURFACE OF THE<br>SKIN WHEN<br>ATTACHED TO A<br>FLUID RESERVOIR<br>OF A COMPATIBLE<br>ABBOTT PUMP | SUBCUTANEOUS<br>INFUSION OF<br>MEDICATION,<br>INCLUDING<br>INSULIN,<br>ADMINISTERED<br>BY THE PEARL™<br>EXTERNAL<br>INSULIN PUMP<br>SYSTEM |
| PROPOSED<br>INDICATIONS<br>FOR USE<br>(K162812) | SUBCUTANEOUS<br>INFUSION OF<br>INSULIN<br>ADMINISTERED<br>BY AN<br>EXTERNAL PUMP | SUBCUTANEOUS<br>INFUSION OF<br>INSULIN<br>ADMINISTERED<br>BY AN<br>EXTERNAL PUMP | SUBCUTANEOU<br>S INFUSION OF<br>INSULIN FROM<br>AN INFUSION<br>PUMP | SUBCUTANEO<br>US INFUSION<br>OF INSULIN<br>ADMINISTERED<br>WITH<br>MICRODOSAG<br>E INSULIN<br>PUMPS. | SUBCUTANEOU<br>S INFUSION OF<br>MEDICATION<br>ADMINISTERED<br>BY AN<br>EXTERNAL<br>PUMP | NO CHANGE | NO CHANGE (IF<br>COMMERCIALIZED) | SUBCUTANEOUS<br>INFUSION OF<br>INSULIN<br>ADMINISTERED<br>BY THE SNAP™<br>INSULIN PUMP<br>SYSTEM |
| DETAIL OF<br>DIFFERENCES | SUBCUTANEOUS<br>INFUSION<br>AND/OR<br>INJECTION OF<br>FLUIDS-INSULIN<br>INTO THE BODY | SUBCUTANEOUS<br>INFUSION<br>AND/OR<br>INJECTION OF<br>FLUIDS-INSULIN<br>INTO THE BODY | SUBCUTANEOU<br>S INFUSION<br>AND/OR<br>INJECTION OF<br>FLUIDS-INSULIN<br>INTO THE BODY | SUBCUTANEO<br>US INFUSION<br>AND/OR<br>INJECTION OF<br>FLUIDS<br>INSULIN INTO | SUBCUTANEOU<br>S INFUSION<br>AND/OR<br>INJECTION OF<br>FLUIDS<br>MEDICATION | NO CHANGE | NO CHANGE (IF<br>COMMERCIALIZED) | SUBCUTANEOUS<br>INFUSION OF<br>MEDICATION,<br>INCLUDING<br>INSULIN,<br>ADMINISTERED |
| | BELOW THE<br>SURFACE OF<br>THE SKIN<br>ADMINISTERED<br>BY AN<br>EXTERNAL<br>PUMP. | BELOW THE<br>SURFACE OF<br>THE SKIN<br>ADMINISTERED<br>BY AN<br>EXTERNAL<br>PUMP. | BELOW THE<br>SURFACE OF<br>THE SKIN-FROM<br>AN INFUSION<br>PUMP. | THE BODY<br>BELOW THE<br>SURFACE OF<br>THE SKIN<br>ADMINISTERED<br>WITH<br>MICRODOSAG<br>E INSULIN<br>PUMPS. | INTO THE BODY<br>BELOW THE<br>SURFACE OF<br>THE SKIN<br>ADMINISTERED<br>BY AN<br>EXTERNAL<br>PUMP. | | | BY THE<br>PEARL™SNAP™<br>EXTERNAL<br>INSULIN PUMP<br>SYSTEM |
| INSTRUCTION<br>FOR USE<br>CHANGES | NO<br>SUBSTANTIVE<br>CHANGE -<br>INSTRUCTIONS<br>FOR USE<br>PROVIDED IN<br>ORIGINAL<br>SUBMISSION<br>HAVE BEEN<br>UPDATED AND<br>STANDARDIZED<br>(ALL ARE<br>PROVIDED IN<br>APPENDIX II). | NO<br>SUBSTANTIVE<br>CHANGE -<br>INSTRUCTIONS<br>FOR USE<br>PROVIDED IN<br>ORIGINAL<br>SUBMISSION<br>HAVE BEEN<br>UPDATED AND<br>STANDARDIZED<br>(ALL ARE<br>PROVIDED IN<br>APPENDIX II). | NO<br>SUBSTANTIVE<br>CHANGE -<br>INSTRUCTIONS<br>FOR USE<br>PROVIDED IN<br>ORIGINAL<br>SUBMISSION<br>HAVE BEEN<br>UPDATED AND<br>STANDARDIZED<br>(ALL ARE<br>PROVIDED IN<br>APPENDIX II). | NO<br>SUBSTANTIVE<br>CHANGE -<br>INSTRUCTIONS<br>FOR USE<br>PROVIDED IN<br>ORIGINAL<br>SUBMISSION<br>HAVE BEEN<br>UPDATED AND<br>STANDARDIZE<br>D (ALL ARE<br>PROVIDED IN<br>APPENDIX II). | NO<br>SUBSTANTIVE<br>CHANGE -<br>INSTRUCTIONS<br>FOR USE<br>PROVIDED IN<br>ORIGINAL<br>SUBMISSION<br>HAVE BEEN<br>UPDATED AND<br>STANDARDIZED<br>(ALL ARE<br>PROVIDED IN<br>APPENDIX II). | NO CHANGE | IF<br>COMMERCIALIZED,<br>DRAFT<br>INSTRUCTIONS FOR<br>USE PROVIDED IN<br>ORIGINAL<br>SUBMISSION WILL<br>BE FINALIZED AND<br>STANDARDIZED | NO<br>SUBSTANTIVE<br>CHANGE -<br>DRAFT<br>INSTRUCTIONS<br>FOR USE<br>PROVIDED IN<br>ORIGINAL<br>SUBMISSION<br>HAVE BEEN<br>FINALIZED AND<br>STANDARDIZED<br>(ALL ARE<br>PROVIDED IN<br>APPENDIX II). |
| STERILIZATION/<br>STERILIZATION<br>LOCATION/<br>SUBCONTRACT<br>ORS<br>CHANGES | CHANGE OF<br>STERILIZATION<br>SITE ONLY. NO<br>CHANGE TO<br>STERILIZATION<br>METHOD OR SAL. | CHANGE OF<br>STERILIZATION<br>SITE ONLY. NO<br>CHANGE TO<br>STERILIZATION<br>METHOD OR SAL. | CHANGE OF<br>STERILIZATION<br>SITE ONLY. NO<br>CHANGE TO<br>STERILIZATION<br>METHOD OR<br>SAL. | CHANGE OF<br>STERILIZATION<br>SITE ONLY. NO<br>CHANGE TO<br>STERILIZATION<br>METHOD OR<br>SAL. | CHANGE OF<br>STERILIZATION<br>SITE ONLY. NO<br>CHANGE TO<br>STERILIZATION<br>METHOD OR<br>SAL. | CHANGE OF<br>STERILIZATION<br>SITE ONLY. NO<br>CHANGE TO<br>STERILIZATION<br>METHOD OR SAL. | IF<br>COMMERCIALIZED,<br>STERILIZATION<br>SITE WILL BE<br>CHANGED<br>WITHOUT CHANGE<br>TO STERILIZATION<br>METHOD OR SAL. | CHANGE OF<br>STERILIZATION<br>SITE ONLY. NO<br>CHANGE TO<br>STERILIZATION<br>METHOD OR SAL. |
| MATERIALS/<br>COMPONENTS<br>CHANGES | THE SOFT<br>CANNULA<br>MATERIAL WAS<br>MODIFIED TO A<br>PFOA FREE<br>PTFE, (U.S.<br>ENVIRONMENTA<br>L PROTECTION<br>AGENCY PFOA<br>STEWARDSHIP<br>PROGRAM). | THE SOFT<br>CANNULA<br>MATERIAL WAS<br>MODIFIED TO A<br>PFOA FREE<br>PTFE, (U.S.<br>ENVIRONMENTAL<br>PROTECTION<br>AGENCY PFOA<br>STEWARDSHIP<br>PROGRAM). | THE SOFT<br>CANNULA<br>MATERIAL WAS<br>MODIFIED TO A<br>PFOA FREE<br>PTFE,<br>(U.S.<br>ENVIRONMENTA<br>L PROTECTION<br>AGENCY PFOA<br>STEWARDSHIP<br>PROGRAM). | THE SOFT<br>CANNULA<br>MATERIAL<br>WAS MODIFIED<br>TO A PFOA<br>FREE PTFE,<br>(U.S.<br>ENVIRONMENT<br>AL<br>PROTECTION<br>AGENCY PFOA<br>STEWARDSHIP<br>PROGRAM). | THE SOFT<br>CANNULA<br>MATERIAL WAS<br>MODIFIED TO A<br>PFOA FREE<br>PTFE,<br>(U.S.<br>ENVIRONMENTA<br>L PROTECTION<br>AGENCY PFOA<br>STEWARDSHIP<br>PROGRAM). | THE SOFT<br>CANNULA<br>MATERIAL WAS<br>MODIFIED TO A<br>PFOA FREE<br>PTFE,<br>(U.S.<br>ENVIRONMENTAL<br>PROTECTION<br>AGENCY PFOA<br>STEWARDSHIP<br>PROGRAM). | IF<br>COMMERCIALIZED,<br>THE SOFT<br>CANNULA<br>MATERIAL WILL BE<br>MODIFIED TO A<br>PFOA FREE PTFE,<br>(U.S.<br>ENVIRONMENTAL<br>PROTECTION<br>AGENCY PFOA<br>STEWARDSHIP<br>PROGRAM). | THE SOFT<br>CANNULA<br>MATERIAL WAS<br>MODIFIED TO A<br>PFOA FREE<br>PTFE,<br>(U.S.<br>ENVIRONMENTAL<br>PROTECTION<br>AGENCY PFOA<br>STEWARDSHIP<br>PROGRAM). |
| | THE SEPTUM<br>WAS CHANGED<br>FROM<br>THERMOPLASTIC<br>ELASTOMER<br>(TPE) TO<br>SILICONE<br>RUBBER. THIS<br>MATERIAL WAS<br>IDENTIFIED AND<br>CLEARED IN<br>SUBMISSION<br>K051264 | THE SEPTUM<br>WAS CHANGED<br>FROM<br>THERMOPLASTIC<br>ELASTOMER<br>(TPE) TO<br>SILICONE<br>RUBBER. THIS<br>MATERIAL WAS<br>IDENTIFIED AND<br>CLEARED IN<br>SUBMISSION<br>K051264 | THE SEPTUM<br>WAS CHANGED<br>FROM<br>THERMOPLASTI<br>C ELASTOMER<br>(TPE) TO<br>SILICONE<br>RUBBER. THIS<br>MATERIAL WAS<br>IDENTIFIED AND<br>CLEARED IN<br>SUBMISSION<br>K051264 | THE SEPTUM<br>WAS<br>CHANGED<br>FROM<br>THERMOPLAST<br>IC ELASTOMER<br>(TPE) TO<br>SILICONE<br>RUBBER. THIS<br>MATERIAL<br>WAS<br>IDENTIFIED<br>AND CLEARED<br>IN SUBMISSION<br>K051264 | THE SEPTUM<br>WAS CHANGED<br>FROM<br>THERMOPLASTI<br>C ELASTOMER<br>(TPE) TO<br>SILICONE<br>RUBBER. THIS<br>MATERIAL WAS<br>IDENTIFIED AND<br>CLEARED IN<br>SUBMISSION<br>K051264 | THE SEPTUM<br>WAS CHANGED<br>FROM<br>THERMOPLASTIC<br>ELASTOMER<br>(TPE) TO<br>SILICONE<br>RUBBER. THIS<br>MATERIAL WAS<br>IDENTIFIED AND<br>CLEARED IN<br>SUBMISSION<br>K051264 | NO CHANGE | NO CHANGE |
| | THE LUER LOCK<br>CONNECTOR<br>CHANGED FROM<br>PVC (POLYVINYL<br>CHLORIDE) TO<br>MABS<br>(METHYLMETHA<br>CRYLATE<br>ACRYLONITRILE<br>BUTADIENE<br>STYRENE).<br>THIS MATERIAL<br>WAS IDENTIFIED<br>AND CLEARED IN<br>SUBMISSION<br>K032854 | THE LUER LOCK<br>CONNECTOR<br>CHANGED FROM<br>PVC (POLYVINYL<br>CHLORIDE) TO<br>MABS<br>(METHYLMETHAC<br>RYLATE<br>ACRYLONITRILE<br>BUTADIENE<br>STYRENE).<br>THIS MATERIAL<br>WAS IDENTIFIED<br>AND CLEARED IN<br>SUBMISSION<br>K032854 | THE LUER LOCK<br>CONNECTOR<br>CHANGED<br>FROM PVC<br>(POLYVINYL<br>CHLORIDE) TO<br>MABS<br>(METHYLMETHA<br>CRYLATE<br>ACRYLONITRILE<br>BUTADIENE<br>STYRENE).<br>THIS MATERIAL<br>WAS IDENTIFIED<br>AND CLEARED<br>IN SUBMISSION<br>K032854 | THE LUER<br>LOCK<br>CONNECTOR<br>CHANGED<br>FROM PVC<br>(POLYVINYL<br>CHLORIDE) TO<br>MABS<br>(METHYLMETH<br>ACRYLATE<br>ACRYLONITRIL<br>E BUTADIENE<br>STYRENE).<br>THIS<br>MATERIAL<br>WAS<br>IDENTIFIED<br>AND CLEARED<br>IN SUBMISSION<br>K032854 | THE LUER LOCK<br>CONNECTOR<br>CHANGED<br>FROM PVC<br>(POLYVINYL<br>CHLORIDE) TO<br>MABS<br>(METHYLMETHA<br>CRYLATE<br>ACRYLONITRILE<br>BUTADIENE<br>STYRENE).<br>THIS MATERIAL<br>WAS IDENTIFIED<br>AND CLEARED<br>IN SUBMISSION<br>K032854 | NO CHANGE | NO CHANGE | NO CHANGE |
| SPECIFCATION/<br>TEST METHODS<br>CHANGES | A BENDING TEST<br>HAS BEEN<br>INTRODUCED TO<br>VERIFY<br>ROBUSTESS OF<br>TUBING BONDS<br>WHEN<br>SUBJECTED TO<br>STRESSES SEEN<br>DURING USE. | A BENDING TEST<br>HAS BEEN<br>INTRODUCED TO<br>VERIFY<br>ROBUSTESS OF<br>TUBING BONDS<br>WHEN<br>SUBJECTED TO<br>STRESSES SEEN<br>DURING USE. | A BENDING<br>TEST HAS BEEN<br>INTRODUCED<br>TO VERIFY<br>ROBUSTESS OF<br>TUBING BONDS<br>WHEN<br>SUBJECTED TO<br>STRESSES<br>SEEN DURING<br>USE. | A BENDING<br>TEST HAS<br>BEEN<br>INTRODUCED<br>TO VERIFY<br>ROBUSTESS<br>OF TUBING<br>BONDS WHEN<br>SUBJECTED<br>TO STRESSES<br>SEEN DURING<br>USE. | A BENDING<br>TEST HAS BEEN<br>INTRODUCED<br>TO VERIFY<br>ROBUSTESS OF<br>TUBING BONDS<br>WHEN<br>SUBJECTED TO<br>STRESSES<br>SEEN DURING<br>USE. | A BENDING TEST<br>HAS BEEN<br>INTRODUCED TO<br>VERIFY<br>ROBUSTESS OF<br>TUBING BONDS<br>WHEN<br>SUBJECTED TO<br>STRESSES SEEN<br>DURING USE. | IF<br>COMMERCIALIZED,<br>A BENDING TEST<br>WILL BE<br>INTRODUCED TO<br>VERIFY<br>ROBUSTESS OF<br>TUBING BONDS<br>WHEN SUBJECTED<br>TO STRESSES<br>SEEN DURING USE. | A BENDING TEST<br>HAS BEEN<br>INTRODUCED TO<br>VERIFY<br>ROBUSTESS OF<br>TUBING BONDS<br>WHEN<br>SUBJECTED TO<br>STRESSES SEEN<br>DURING USE. |
| A TENSION TEST<br>HAS BEEN<br>INTRODCUED<br>AFTER PRE-<br>CONDITIONING<br>OF TEST<br>SAMPLES<br>IMMERSED IN<br>37°C WATER FOR<br>18-24 HRS. | A TENSION TEST<br>HAS BEEN<br>INTRODCUED<br>AFTER PRE-<br>CONDITIONING<br>OF TEST<br>SAMPLES<br>IMMERSED IN<br>37°C WATER FOR<br>18-24 HRS. | A TENSION<br>TEST HAS BEEN<br>INTRODCUED<br>AFTER PRE-<br>CONDITIONING<br>OF TEST<br>SAMPLES<br>IMMERSED IN<br>37°C WATER<br>FOR 18-24 HRS. | A TENSION<br>TEST HAS<br>BEEN<br>INTRODCUED<br>AFTER PRE-<br>CONDITIONING<br>OF TEST<br>SAMPLES<br>IMMERSED IN<br>37°C WATER<br>FOR 18-24 HRS. | A TENSION<br>TEST HAS BEEN<br>INTRODCUED<br>AFTER PRE-<br>CONDITIONING<br>OF TEST<br>SAMPLES<br>IMMERSED IN<br>37°C WATER<br>FOR 18-24 HRS. | A TENSION TEST<br>HAS BEEN<br>INTRODCUED<br>AFTER PRE-<br>CONDITIONING<br>OF TEST<br>SAMPLES<br>IMMERSED IN<br>37°C WATER FOR<br>18-24 HRS. | IF<br>COMMERCIALIZED,<br>A TENSION TEST<br>WILL BE<br>INTRODCUED<br>AFTER PRE-<br>CONDITIONING OF<br>TEST SAMPLES<br>IMMERSED IN 37°C<br>WATER FOR 18-24<br>HRS. | A TENSION TEST<br>HAS BEEN<br>INTRODCUED<br>AFTER PRE-<br>CONDITIONING<br>OF TEST<br>SAMPLES<br>IMMERSED IN<br>37°C WATER FOR<br>18-24 HRS. | |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
*The Abbott Set was cleared but has not been commercialized
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.