K240158 · Modular Medical, Inc. · LZG · Sep 4, 2024 · General Hospital
Device Facts
Record ID
K240158
Device Name
Modular Medical MODD1 Insulin Delivery System
Applicant
Modular Medical, Inc.
Product Code
LZG · General Hospital
Decision Date
Sep 4, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Modular Medical MODD1 Insulin Delivery System is indicated for the subcutaneous delivery of insulin at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 18 years of age and greater.
Device Story
MODD1 Insulin Delivery System is a body-worn, software-controlled infusion pump for diabetes management. System components include a reusable pump, single-use 3.0mL insulin cartridge with integrated battery, adhesive pad, and infusion set. User programs basal delivery schedules via MMI App on smartphone (iOS); pump interface manages bolus delivery, status checks, and suspension. Pump uses gear motor rotation of a fixed-position camshaft to drive insulin from reservoir to infusion set. Software monitors dispense output via multiple sensors; provides audio-visual alarms for malfunctions, low volume, or occlusions. Device communicates via BLE and NFC. Healthcare providers use device history for clinical decision-making; patients benefit from automated, programmable insulin delivery. Used in home environment.
Clinical Evidence
Bench testing only. No clinical trials performed. Evidence includes risk management (ISO 14971), human factors validation (IEC 62366-1), insulin compatibility (Humalog U-100), biocompatibility (ISO 10993-1), sterility validation, electrical safety/EMC (IEC 60601-1/1-2, RTCA DO-160G), wireless coexistence (AAMI TIR69, IEEE/ANSI C63.27), and software/cybersecurity verification/validation.
Technological Characteristics
Motor-driven (multiple piston) infusion pump; internal non-rechargeable coin cell battery; dimensions 1.51" x 2.38" x 0.56"; weight 28g; IP24 water resistance. Connectivity via BLE and NFC. Software-controlled gear motor camshaft mechanism. Sterilization via gamma radiation for disposable components. Complies with IEC 60601-1, IEC 60601-1-2, ISO 14971, ISO 10993-1, IEC 62304.
Indications for Use
Indicated for subcutaneous insulin delivery at set/variable rates for diabetes mellitus management in adults (18+ years).
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K240158
B Applicant
Modular Medical, Inc.
C Proprietary and Established Names
Modular Medical MODD1 Insulin Delivery System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LZG | Class II | 21 CFR 880.5725 | CH – Clinical Chemistry |
E Purpose for Submission:
New device
## II Intended Use/Indications for Use:
A Intended Use(s):
See Indications For Use below
B Indication(s) for Use:
The Modular Medical MODD1 Insulin Delivery System is indicated for the subcutaneous delivery of insulin at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 18 years of age and greater.
C Special Conditions for Use Statement(s):
This device is for prescription use only.
The MODD1 System is contraindicated for:
1. Diagnosing Diabetes Mellitus.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
2. Use by patients who do NOT have adequate hearing and/or vision to allow recognition of all functions of the MODD1 System including Status and Alarms.
3. Use by patients who cannot manage their Diabetes therapy.
4. Use by patients unwilling to take a minimum of four (4) blood glucose readings per day.
5. Use by patients who are unable to use the MODD1 System in accordance with this User Guide.
6. Use by patients who are not capable of following the User Guide
7. Use by patient populations requiring basal rates greater than 4 U/hr or less than 0.5U/hr
Consistent drops of the Pump and/or exposure to personal care products such as lotions and sunscreens can result in significantly reduced Pump lifetime if also exposed to water. This may lead to severe hyperglycemia or Diabetic Ketoacidosis (DKA).
## III Device Description
The device is an insulin pump that is worn adhered to the user's body through an adhesive patch with a short infusion set. The subject device consists of a pump mechanism (reusable, up to 90 days), an insulin cartridge assembly (sterile single use, up to 72 hours), an adhesive pad (single use, up to 72 hours), an infusion set (sterile single use, up to 72 hours), and a smartphone application (reusable, indefinite length of time) contained on an iPhone operating iOS 16 or later. The infusion set is distributed by Modular Medical. There are previously cleared 3rd party components that are able to be used with the subject device including a syringe (BD; K110771), a needle (BD; K021475), and the Ypsomed Orbit Inserter (Ypsomed; K163400).
The pumping mechanism consists of a linear peristaltic pump that pushes a fixed amount of insulin (5 microliters or 0.5 Units) per cycle of the pump operating. The pump must be programmed through the MMI App installed on an Apple (iOS) smartphone by the pump user. Once a pump is programmed it can operate independently from the app and does not require a phone to be connected or paired with the pump unless new insulin delivery rates are being programmed.
The insulin cartridge exterior is a rigid cyclic olefin copolymer and supports the pistons that are manipulated by the pump and supports the coin cell battery. The interior of the insulin cartridge that has contact with insulin is flexible and has membranes that allows for air to enter the cartridge and push against the flexible membrane and for insulin to exit the cartridge via the infusion set.
## IV Substantial Equivalence Information:
### A Predicate Device Name(s):
The t:slim® Insulin Delivery System
### B Predicate 510(k) Number(s):
K160056
K240158 - Page 2 of 9
{2}
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K240158 | K160056 |
| --- | --- | --- |
| Device Trade Name | Modular Medical MODD1
Insulin Delivery System | The t:slim® Insulin
Delivery System |
| General Device
Characteristic
Similarities | | |
| Intended
Use/Indications For
Use | Subcutaneous delivery of
insulin at set and variable
rates, for the management of
diabetes mellitus in persons
requiring insulin. | Same |
| Operating
Environment | Home Use | Same |
| Insulin Delivery
Modes | Basal and Bolus | Same |
| Reservoir Volume | 3 mL | Same |
| General Device
Characteristic
Differences | | |
| Intended User Age
Range | 18 years and greater | 6 years and greater |
| Bolus Dose Range | 2-20 units | 0.05-25 units |
| Basal Delivery Flow
Rates | 0.5 – 4 units/hour | 0.1-15 units/hour |
| Insulin Bolus
Calculator | No | Yes |
| Connectivity | BLE and NFC | BLE |
V Standards/Guidance Documents Referenced:
Standards
- AAMI TIR101:2021 Fluid delivery performance testing for infusion pumps
K240158 - Page 3 of 9
{3}
- IEC 60601-1-11 Edition 2.1 2020-07 CONSOLIDATED VERSION, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-1-8 Edition 2.2 2020-07 CONSOLIDATED VERSION, Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION, Medical devices - Part 1: Application of usability engineering to medical devices
- IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- ISO 14971 Third Edition 2019-12, Medical devices - Application of risk management to medical devices
- AAMI HE75:2009/(R)2018, Human factors engineering - Design of medical devices
- RTCA DO-160 Environmental Conditions and Test Procedures for Airborne Equipment
FDA Guidance Documents:
- Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (September 2023)
- Applying Human Factors and Usability Engineering to Medical Devices (February 2016)
- Electromagnetic Compatibility (EMC) of Medical Devices (June 2022)
- Content of Premarket Submissions for Device Software Functions (June 2023)
- Use of International Standard ISO 10993-1 “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process (September 2023)
VI Performance Characteristics:
A. Analytical Performance
1. Basal Accuracy
The basal delivery accuracy of the pump was tested with water to simulate insulin for minimum, intermediate and maximum basal rate delivery performance for both new and aged pumps without any warm-up period. Both new pumps (n=15) and aged pumps (n=15) were tested for accuracy. Table 1 reports the typical basal performance (median) observed, along with the lowest and highest results observed for minimum, intermediate, and maximum basal rate settings for all pumps tested.
K240158 - Page 4 of 9
{4}
Table 1.: Amount of fluid delivered after 1, 6, and 12 hours at minimum, intermediate, and maximum basal rate settings
| Median Insulin Units Delivered, New Pumps n = 15 | | | |
| --- | --- | --- | --- |
| Basal Rate | 1 hour | 6 hours | 12 hours |
| 0.5 U/hour [min, max] | 0.62 U [0.57, 0.65] | 3.16 U [3.03, 3.26] | 6.17 U [5.93, 6.39] |
| 2.0 U/hour [min, max] | 2.10 U [2.02, 2.25] | 12.05 U [11.53, 12.78] | 24.05 U [22.88, 25.56] |
| 4.0 U/hour [min, max] | 4.20 U [4.05, 4.45] | 24.51 U [23.47, 25.82] | 48.81 U [46.53, 51.58] |
| Median Insulin Units Delivered, Aged Pumps n = 15 | | | |
| 0.5 U/hour [min, max] | 0.63 U [0.59, 0.65] | 3.19 U [3.03, 3.33] | 6.22 U [5.92, 6.58] |
| 2.0 U/hour [min, max] | 2.17 U [2.03, 2.25] | 12.16 U [11.32, 12.82] | 24.14 U [22.48, 25.47] |
| 4.0 U/hour [min, max] | 4.22 U [4.00, 4.51] | 24.33 U [22.94, 25.85] | 48.40 U [45.57, 51.37] |
# 2. Bolus Accuracy
The bolus delivery accuracy of the pump was tested with water to simulate insulin for the minimum, intermediate, and maximum bolus amounts. Both new pumps $(n = 15)$ and aged pumps $(n = 15)$ were tested for accuracy. The pumps each were tested by delivering at minimum, intermediate, and maximum bolus volumes (0.05 U, 6.0 U, and 30 U).
Table 2 below shows the number (and %) of boluses within the specified range of the minimum bolus, one intermediate bolus and the maximum bolus volume.
Table 2.: Amount of fluid delivered after bolus requests of 2, 10, and 20 units for new and aged pumps
| | 2U Insulin Bolus Request, n=450 boluses, n=15 new pumps | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | <25% | 25-75% | 75-90% | 90-95% | 95-105% | 105-110% | 110-125% | 125-175% | 175-250% | >250% |
| # in range | 0/450 | 0/450 | 13/450 | 63/450 | 350/450 | 24/450 | 0/450 | 0/450 | 0/450 | 0/450 |
| % in range | 0% | 0% | 2.9% | 14.0% | 77.8% | 5.3% | 0% | 0% | 0% | 0% |
| | 10U Insulin Bolus Request, n=423 boluses n= 15 new pumps | | | | | | | | | |
| # in range | 0/423 | 0/423 | 1/423 | 41/423 | 332/423 | 47/423 | 2/423 | 0/423 | 0/423 | 0/423 |
| % in range | 0% | 0% | 0.2% | 9.7% | 78.5% | 11.1% | 0.5% | 0% | 0% | 0% |
K240158 - Page 5 of 9
{5}
K240158 - Page 6 of 9
| | 20U Insulin Bolus Request, n=209 boluses n= 15 new pumps | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| # in range | 0/209 | 0/209 | 1/209 | 16/209 | 178/209 | 14/209 | 0/209 | 0/209 | 0/209 | 0/209 |
| % in range | 0% | 0% | 0.5% | 7.7% | 85.2% | 6.7% | 0% | 0% | 0% | 0% |
| | 2U Insulin Bolus Request, n=450 boluses, n=15 aged pumps | | | | | | | | | |
| # in range | 0/450 | 1/450 | 15/450 | 49/450 | 371/450 | 14/450 | 0/450 | 0/450 | 0/450 | 0/450 |
| % in range | 0% | 0.2% | 3.3% | 10.9% | 82.4% | 3.1% | 0% | 0% | 0% | 0% |
| | 10U Insulin Bolus Request, n=423 boluses n= 15 aged pumps | | | | | | | | | |
| # in range | 0/419 | 0/419 | 1/419 | 58/419 | 318/419 | 42/419 | 0/419 | 0/419 | 0/419 | 0/419 |
| % in range | 0% | 0% | 0.2% | 13.8% | 75.9% | 10.0% | 0% | 0% | 0% | 0% |
| | 20U Insulin Bolus Request, n=209 boluses n= 15 aged pumps | | | | | | | | | |
| # in range | 0/210 | 0/210 | 1/210 | 17/210 | 180/210 | 12/210 | 0/210 | 0/210 | 0/210 | 0/210 |
| % in range | 0% | 0% | 0.5% | 8.1% | 85.7% | 5.7% | 0% | 0% | 0% | 0% |
## 3. Occlusion Detection
The time to detection of complete occlusions was measured. Pumps were physically occluded by closing the patient end of the fluid path. After pumps alarmed, the occlusions were cleared, and the total amount of fluid delivered was measured. Unintended bolus after release of an occlusion is less than 3 U of insulin. Table 3 shows typical time to occlusion detection for new pumps under three different situations.
Table 3. Timing of occlusion detection alarms
| Operating Rate | Typical | Maximum |
| --- | --- | --- |
| Bolus (2U or greater) | 1 Minute 51 Seconds | 2 Minutes 19 Seconds |
| Basal (2U/hr) | 30 Minutes 29 Seconds | 31 Minutes 31 Seconds |
| Basal (0.5U/hr) | 2 Hours 0 Minutes | 2 Hours 0 Minutes |
{6}
B. Other Supportive Instrument Performance Characteristics Data
1. Hazard Analysis
A comprehensive hazard analysis was reviewed, in which design inputs and outputs, risks, and risk mitigations for hardware and software associated with proper functioning of the insulin pump were reviewed. The sponsor performed a hazard analysis to account for the unique intended use, design elements, and risks of their infusion pump. This analysis identified hazards which could reasonably be anticipated to impact the proper use of the device, traced all identified risks to adequate design controls, and demonstrated that design features were appropriately implemented and validated.
2. Human Factors
Human Factors validation testing was conducted in the United States and validated the safe and effective use of the device with 16 people who had Diabetes Mellitus aged 18 years or older and treated as an adult. There was a mix of insulin pump experience and naïve. Users had 5-42 years of diabetes management experience. The use of the system was validated only for users who can self-manage an insulin pump.
Users were provided 1 on 1 training by a trainer from Modular Medical that lasted up to 90 minutes and consisted of using the subject device with the trainer, using the quick start guide and the instructions for use, and reviewing the training slides. Training consisted of walking users through how to use the system and instructional materials followed by users demonstrating to the trainer how they would use the system and being able to complete all of the tasks associated with training. The trainer then made a final determination on a user being ready to use the subject device followed by a 1-hour decay period and subsequent Human Factors validation testing in a simulated home use environment.
During simulate use testing, users were provided with a smartphone that was capable of dialing a customer support number that was listed in the instructions for use and an employee of Modular Medical was available to answer the call if needed. Users were validated on their ability to use the subject device when provided clinical context without prompting from the moderator to complete specific tasks and included their ability to notice and resolve alarms.
Critical tasks were identified based on the clinical harm severities that could occur due to use error such as hypoglycemia, hyperglycemia, diabetic ketoacidosis, and infection.
3. Biocompatibility
Biocompatibility testing was performed per FDA Guidance Document: Use of International Standard ISO 10993-1 “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process,” and results of the testing were found adequate.
4. Sterility
The insulin cartridge was sterilized with radiation and validated ANSI/AAMI/ISO 11137-1:2006/(R)2015 Sterilization of health care products - Radiation - Part 1: Requirements for
K240158 - Page 7 of 9
{7}
development, validation, and routine control of a sterilization process for medical devices including Amendment 1 (2013) and Amendment 2 (2019).
## 5. Insulin Compatibility and Stability
Insulin compatibility studies were conducted on the subject device with Humalog U-100 insulin to demonstrate device material compatibility with the insulin, physicochemical stability of insulin when delivered through the device under simulated use conditions, and the continued ability of insulin to provide antimicrobial effectiveness after having been stored in the device.
## 6. Additional Bench Testing
In addition to the performance testing described above, mechanical testing and simulated use testing, the subject device was subjected to additional testing to verify its reliability to demonstrate that it could function for 90 days. This testing included cyclic reliability testing to simulate actual use of the pump including accelerated aging of insulin cartridges, simulation of packaging and shipping, vibration and shock testing, exposure to personal care products, cleaning, button presses, insulin cartridge changes, drops, and ingress testing. Functional testing included the ability of the pump to deliver insulin and detect occlusions. The reliability testing results supported that the device could function as intended for the intended duration, but that exposure to personal care products, drops, and subsequent exposure to water could impact pump survivability.
| Testing to Support System Reliability |
| --- |
| Pump lifetime testing |
| Transient Temperature |
| Transient Pressure |
| Ingress Protection |
| Cleaning |
| Against mechanical stressors (e.g., vibrations, shock) |
| Against personal care products such as sunscreen and lotions |
| Delivery Accuracy and Occlusion Detection Testing |
## 7. Electromagnetic Compatibility and Wireless Coexistence
Electromagnetic compatibility, electromagnetic immunity and wireless coexistence testing was performed for the pump. The subject device meets immunity requirements per RTCA DO-160 (Susceptibility Category T). The subject device was also tested per FDA recognized standard IEC 60601-1-2:2020. All tests demonstrated that the device would perform as expected in the home healthcare environment.
K240158 - Page 8 of 9
{8}
8. **Basic Safety and Essential Performance**
The sponsor provided verification evidence for compliance with the IEC 60601-1 and applicable collateral standards. Verification results support the finding of substantial equivalent for this device.
9. **Software and Cybersecurity**
Detailed information on software was reviewed and found to be acceptable. The cybersecurity documentation supports that the device is cybersecure.
**VII Proposed Labeling:**
The labeling supports the finding of substantial equivalence for this device.
**VIII Conclusion:**
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K240158 - Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.