K110771 · Becton, Dickinson and Company · FMF · May 13, 2011 · General Hospital
Device Facts
Record ID
K110771
Device Name
BD SINGLE USE, HYPODERMIC SYRINGE
Applicant
Becton, Dickinson and Company
Product Code
FMF · General Hospital
Decision Date
May 13, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection.
Device Story
Three-piece manual hypodermic syringe; includes plastic barrel with graduated scale, synthetic rubber stopper, and plastic plunger rod. Features 6% (Luer) connector in 1ml Luer Slip, 3ml and 5ml Luer Lok and Luer Slip sizes. Used by healthcare professionals in clinical settings for fluid aspiration or injection. Modified from predicate via new synthetic stopper material and silicone-based lubricant formulation. Operates manually or with power-driven pumps. Provided sterile via EtO or irradiation. Performance verified against ISO 7886-1 and ISO 7886-2 standards.
Clinical Evidence
Bench testing only. No clinical data. Design verification testing included functional testing (sustaining force, break-out force, pump sticktion, stopper seal, autoclavability) and chemical extractables testing (zinc, lead, tin, iron, cadmium, pH shift) per ISO 7886-1/2. Biocompatibility testing included cytotoxicity, hemolysis, acute systemic toxicity, intracutaneous reactivity, murine local lymph node assay, and pyrogenicity per ISO 10993 standards.
Technological Characteristics
Three-piece manual piston syringe. Materials: plastic barrel, synthetic rubber stopper, plastic plunger rod. Lubricant: silicone-based. Connectivity: none. Sterilization: EtO or Irradiation (SAL 10^-6). Standards: ISO 7886-1 (manual use), ISO 7886-2 (power-driven pumps), ISO 10993 (biocompatibility).
Indications for Use
Indicated for use by healthcare professionals for general purpose fluid aspiration and injection.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Becton Dickinson Single Use Hypodermic Syringe (K980987)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(K) Summary of Safety and Effectiveness
MAY 13 2011
KU0771
Date Prepared: 16 March 2011
#### Submitted By: 1.
John Roberts Regulatory Affairs Specialist BD Medical - Medical Surgical Systems 1 Becton Drive Franklin Lakes, NJ 07417 201 847 5473 Tel: 201 847 5307 Fax:
#### 2. Device Name:
BD Single Use, Hypodermic Syringe Trade Name: Common Name: Piston Syringe Classification Name: Syringe, Piston Class II, 21 CFR 880.5860 Classificiation:
#### Predicate Device: 3.
Becton Dickinson Single Use Hypodermic Syringe Trade Name: Becton, Dickinson and Company Manufacturer: 510(k) Number: K980987
#### 4. Device Description:
The modified BD Single Use, Hypodermic Syringe is a three-piece single use, hypodermic syringe with a 6% (Luer) connector in 1ml Luer Slip, 3ml and 5ml Luer Lok and Luer Slip syringe sizes. The syringe assembly consists of a plastic barrel with a graduated scale, a synthetic rubber stopper, and a plastic plunger rod. The changes to the modified device from the predicate include a new synthetic stopper material and a new silicone based stopper lubricant formulation. The syringe performance characteristics are equivalent to the predicate device. The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection.
The modified BD Single Use, Hypodermic Syringe are provided sterile, by either EO or an Irradiation sterilization method, in a syringe only configuration or with a pre-attached or side-by-side hypodermic needle.
#### న్. Intended Use:
The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection.
{1}------------------------------------------------
#### Technological Characteristics: 6.
The principal device of this 510(k) premarket notification is the result of a design change to the predicate device (K980987) conducted in accordance with Quality System Regulations, 21 CFR 820. The BD Single Use, Hypodermic Syringe is Substantially Equivalent to the predicate device, given that:
- a) The BD Single Use, Hypodermic Syringe has the same intended use as the predicate device
- b) The BD Single Use, Hypodermic Syringe operates under the same operating principle as the predicate device
- c) The BD Single Use, Hypodermic Syringe barrel and plunger rod use an identical design and identical materials as the predicate device
- d) The BD Single Use, Hypodermic Syringe and the predicate device meet the requirements for manual use and use with power-driven pumps as defined by ISO 7886-1 and ISO 7886-2 respectively.
- e) The BD Single Use, Hypodermic Syringe and the predicate device component materials comply with ISO 10993 as applicable to the intended use of the device
- The BD Single Use, Hypodermic Syringe and the predicate device are sterilized to an f) SAL of 10-6 via an EtO or Irradiation sterilization process
- g) The BD Single Use, Hypodermic Syringe are assembled and packaged at the same manufacturing location utilizing the same equipment as the predicate device
- h) The BD Single Use, Hypodermic Syringe demonstrated equivalent performance to the predicate device during design verification testing.
#### 6. Performance:
Design Verification tests were performed based on the risk analysis performed, and the results of these tests demonstrate that the BD Single Use, Hypodermic Syringe performed in an equivalent manner to the predicate device and is safe and effective when used as intended.
Design Verification testing included the following:
000011
{2}------------------------------------------------
| Performance Characteristic | Test Performed | Acceptance Criteria |
|---------------------------------|----------------------------------------------------------------------------------------------|---------------------|
| Functional Testing | | |
| Sustaining Force | Determination of forces required<br>to operate plunger<br>ISO 7886-1 - Annex G | Per ISO 7886-1 |
| Break-Out Force | Determination of forces required<br>to operate plunger<br>ISO 7886-1 - Annex G | Per ISO 7886-1 |
| Pump Sticktion / Force | Determination of forces required<br>to move the piston<br>ISO 7886-2 - Annex C | Per ISO 7886-2 |
| Stopper Seal | Water Leakage Test<br>ISO 7886-1 - Annex D | Per ISO 7886-1 |
| Autoclavability | Water Leakage Test<br>ISO 7886-1 - Annex D<br>After Syringe autoclave for<br>15mins at 270°F | Per ISO 7886-1 |
| Chemical Testing (Extractables) | | |
| Zinc | ISO 7886-1 | Per ISO 7886-1 |
| Lead, Tin, Iron | ISO 7886-1 | Per ISO 7886-1 |
| Cadmium | ISO 7886-1 | Per ISO 7886-1 |
| pH shift | ISO 7886-1 | Per ISO 7886-1 |
# Biocompatibility Testing
.
! |
. "".
.. /
| Cytotoxicity | ISO10993-5:1999 | Non-Toxic |
|----------------------------------|--------------------------|----------------|
| Hemolysis | ISO10993-4:2002/A:2006 | Non-Toxic |
| Acute Systemic Toxicity | ISO10993-11:2006 | Non-Toxic |
| Intracutaneous Reactivity | ISO10993-10:2002/A1:2006 | Non-Irritant |
| Murine Local Lymph Node<br>Assay | ISO10993-10:2002/A1:2006 | Non-Sensitizer |
| Pyrogenicity | ISO10993-11:2006 | Non-Pyrogenic |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. John Roberts Regulatory Affairs Specialist Becton, Dickinson and Company BD Medicalsurgical 1 Becton Drive MC237 Franklin Lakes, New Jersey 07417
MAY 1 3 2011
Re: K110771
Trade/Device Name: BD Single Use, Hypodermic Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: April 19, 2011 Received: April 21, 2011
Dear Mr. Roberts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2- Mr. Roberts
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known): 长儿1077 |
BD Single Use, Hypodermic Syringe Device Name:
Indications for Use:
The BD Single Use, Hypodermic Syringe is intended for use by health care professionls for general purpose fluid aspiration/injection.
Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page | of |
Rilis C. Chapman 5/12/11
(Division Sign-Off)
יח vision of Anesthesiology, General Hospital intection Control, Dental Devices
:.0(k) Number: K110771
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.