t:slim Insulin Delivery System

K160056 · Tandem Diabetes Care, Inc. · LZG · Jul 25, 2016 · General Hospital

Device Facts

Record IDK160056
Device Namet:slim Insulin Delivery System
ApplicantTandem Diabetes Care, Inc.
Product CodeLZG · General Hospital
Decision DateJul 25, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic, Real-World Evidence, Pediatric

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K160056 · Jul 25, 2016t:slim Insulin Delivery SystemTandem Diabetes Care, Inc.Peer-reviewed clinical literatureThe sponsor utilized a summary of peer-reviewed literature to support the safety and clinical use of CSII therapy in pediatric patients aged 6 through 11, supporting the expansion of the device's indications for use.Pediatric use; CSII; Literature review

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Peer-reviewed literature review regarding CSII in ages 6-11; Literature reviewPediatric patients (ages 6-11) with diabetes mellitusNot applicable for this studySafety and efficacy of CSII therapy

Indications for Use

The t:slim® Insulin Delivery System is indicated for the subcutaneous delivery of insulin at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 6 years of age and greater. The device is indicated for use with NovoLog or Humalog U-100 insulin.

Device Story

The t:slim Insulin Delivery System is a software-controlled, programmable insulin infusion pump for subcutaneous insulin delivery. It consists of a pump, a 3.0 mL disposable insulin cartridge, an infusion set, and power accessories. The user programs basal and bolus delivery via a touch-screen graphical user interface. A motor-driven micro-syringe mechanism within the cartridge delivers insulin. The pump includes a pressure measurement sensor to estimate reservoir volume, detect occlusions, and monitor cartridge status. Used by patients for diabetes management; the device provides continuous insulin infusion. This submission expands the indicated age range to include pediatric patients 6 years and older. No hardware or software changes were made to the device; safety is supported by human factors testing and an assurance case.

Clinical Evidence

No new clinical trials were conducted. Evidence consists of a safety assurance case, human factors testing to validate use in the pediatric population (6+ years), and a review of peer-reviewed literature supporting continuous subcutaneous insulin infusion (CSII) in children aged 6 through 11.

Technological Characteristics

Programmable, battery-operated infusion pump; motor-driven micro-syringe delivery mechanism; 3.0 mL disposable cartridge; touch-screen GUI; pressure sensor for volume estimation and occlusion detection; USB-rechargeable internal battery; includes AC/DC power adapters. No changes to materials or hardware from predicate.

Indications for Use

Indicated for individuals 6 years of age and older with diabetes mellitus requiring subcutaneous insulin delivery using NovoLog or Humalog U-100.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are connected and appear to be part of a single, flowing design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 25, 2016 Tandem Diabetes Care, Inc. Mr. Michael Sarrasin Senior Director of Regulatory Affairs 11045 Roselle Street San Diego, California 92121 Re: K160056 Trade/Device Name: t:slim Insulin Delivery System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZG Dated: Julv 5. 2016 Received: July 6, 2016 Dear Mr. Sarrasin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Tina Kiang -S for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K160056 Device Name t:slim® Insulin Delivery System #### Indications for Use (Describe) The t:slim® Insulin Delivery System is indicated for the subcutaneous delivery of insulin at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 6 years of age and greater. The device is indicated for use with NovoLog or Humalog U-100 insulin. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------| | <span style="text-decoration: underline;"><b>Prescription Use</b></span> (Part 21 CFR 801 Subpart D) | <div style="display:inline-block; width:15px; height:15px; border:1px solid black; text-align:center;"><b>X</b></div> | | Over-The-Counter Use (21 CFR 801 Subpart C) | <div style="display:inline-block; width:15px; height:15px; border:1px solid black;"></div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) SUMMARY K160056 # Tandem Diabetes Care, Inc. #### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Tandem Diabetes Care, Inc. 11045 Roselle Street San Diego, CA 92121 Phone: 858-366-6900 Facsimile: 858-362-7070 Contact Person: Michael Sarrasin Contact Email: MSarrasin@tandemdiabetes.com Date Prepared: July 21, 2016 #### Name of Device t:slim® Insulin Delivery System #### Common or Usual Name Insulin infusion pump # Classification Name Infusion Pump per 21 CFR 880.5725 # Product Code LZG, Insulin infusion Pump #### Predicate Devices Tandem Diabetes Care, Inc. t:slim® Insulin Delivery System (K141758) # Purpose of the 510(k) notification The purpose of this 510(k) is to describe modified indications for use for the t:slim® Insulin Delivery System #### Intended Use The t:slim® Insulin Delivery System is indicated for the subcutaneous delivery of insulin at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, for individuals 6 years of age and greater. The device is indicated for use with NovoLog or Humalog U-100 insulin. {4}------------------------------------------------ # Device Description The t:slim® Insulin Delivery System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The scope of this Traditional 510(k) is limited to expansion of the indications for use to include individuals 6 years of age and greater. The device is indicated for use with NovoLog or Humalog U-100 insulin. No changes have been made to the device components to accommodate the modified indications for use. The t:slim System, like the predicate device, consists of the following components and accessories: - 1. a software-controlled, programmable insulin infusion pump ("t:slim Pump" or "Pump"); - 2. a dedicated, disposable 3.0 mL (300 unit) insulin cartridge ("cartridge"); - 3. an UnoMedical Comfort™ Infusion Set (K051264), or an equivalently FDA cleared infusion set ("infusion set"); and - 4. accessories, including a 3 mL sterile syringe and 26 gauge sterile needle, as well as an AC power supply and DC car adapter power supply with USB for charging the Pump's internal battery, cartridge Instructions for Use, and User's Guide, The t:slim Pump is a battery operated infusion pump capable of both basal and bolus delivery of insulin. The Pump utilizes a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient's subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient). The graphical user interface ("GUI") of the t:slim Pump is a touch screen that displays information used to control the t:slim System. The delivery of insulin is accomplished through a micro-syringe within the head of the cartridge. ldentical to the predicate device, the t:slim cartridge is a single-use device, individually packaged and sealed, and is provided sterile. The cartridge attaches to the t:slim Pump and is designed to hold up to 3.0 mL (300 units) of insulin. Both in the predicate and in the device that is the subject of this 510(k) notice, the delivery of insulin is accomplished through a micro-syringe within the head of the cartridge. The Pump has a pressure measurement sensor to estimate the volume of insulin in the insulin reservoir by measuring air pressure in the cartridge. This information is used to provide feedback to the software to determine the insulin volume, detect occlusions and cartridge removal. # Technological Characteristics The t:slim Insulin Delivery System that is the subject of the modified indication in this 510(k) submission has the same technological characteristics as the predicate device. The 510(k) device and the predicate Systems both consist of: (1) a software-controlled, programmable insulin infusion pump capable of both basal and bolus delivery of insulin ("t:slim Pump"); (2) a dedicated disposable 3.0 mL (300 units) insulin cartridge; (3) UnoMedical's Comfort™ Infusion Set (K051264), or an equivalently cleared set; and (4) additional device accessories including a sterile syringe and needle (for cartridge filling) an AC power supply and DC car adapter power supply with USB. # Performance Data A summary of peer reviewed literature supports the use of CSII in ages 6 through 11. {5}------------------------------------------------ A safety assurance case was provided for the t:slim Insulin Delivery Systems as recommended in the FDA's guidance document, Guidance for Industry and FDA Staff - Total Product Life Cycle: Infusion Pump - Premarket Notification [510(k)] Submissions Guidance, issued on December 2, 2014. The stated goal of the t:slim assurance case is: - The safety risk resulting from use of t:slim pump is reduced to acceptable levels and is As . Low As Reasonably Practicable (ALARP) through use of mitigating technologies and the risks are substantially equivalent to the predicate device. The assurance case defined the device system, including the indications for use, patient types, users, use conditions, environments of use, and list of specific devices covered by the assurance case with associated design specification and labeling documentation. The supporting assurance arguments covered the following attributes: - . Demonstrate acceptability of risk mitigations - Demonstrate adequate device reliability - Demonstrate adequate design verification and validation of device specifications . The assurance case included mitigations of risks related to the following hazards: - . Operational Hazards - . Environmental Hazards - Electrical Hazards - Hardware Hazards - Software Hazards - Mechanical Hazards - . Biological and Chemical Hazards - Use Hazards The following evidence was included in the assurance case to support the modifications of the subject device's indications for use in comparison to the predicate: Human Factors testing supports substantially equivalent use of the device in individual's from 6 years of age and greater. Updated labeling, warnings, and precautions provide further information regarding the risks of the device for pediatric users. # Substantial Equivalence The current t:slim System has the same intended use and very similar indications for use compared to the predicate device. The only change is the modification of the indications to include users as young as 6 years of age. The system also has identical technological characteristics. Human factors testing confirms that the current t:slim System, when indicated for younger users, continues to perform in a substantially equivalent manner.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...