Retrospective clinical study data (patient medical records/clinical outcomes)
Two retrospective studies were used to support the clinical performance, survival rates, and success of the JDZygoma dental implant system (diameters 3.9mm and 4.3mm) using the extra-maxillary surgical method.
Retrospective study of JDZygoma 3.9mm and 4.3mm implants; Retrospective study; Follow-up/Duration: 1 year (3.9mm implants) and 3 years (4.3mm implants)
Male and female adult patients with partially or fully edentulous maxilla; Sample Size: 66 patients (143 implants total)
JDZygoma Dental implant System is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxilla, in order to restore patient esthetics and chewing function. The JDZygoma Implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Device Story
JDZygoma dental implants are endosseous, root-form implants designed for placement in the zygomatic bone; intended for patients with partially or fully edentulous maxillae. Implants are manufactured from commercially pure titanium (Grade 4) with an apical SLA surface (sandblasted/acid-etched) and a machined coronal surface. Abutments are titanium Grade 5 with 45°, 52.5°, and 60° angulations. The system uses an internal hexagon connection. Devices are gamma-sterilized (implants) or non-sterile (abutments). Used by qualified medical personnel in clinical settings to support fixed or removable prosthetics. Immediate loading is permitted given sufficient primary stability. Benefits include restoration of chewing function and esthetics in patients with limited maxillary bone volume.
Clinical Evidence
Clinical evidence consists of two retrospective studies involving 66 adult patients (male and female) with 143 JDZygoma implants (3.9mm and 4.3mm diameters). Follow-up was 1 year for 3.9mm implants and 3 years for 4.3mm implants. Primary endpoint was implant survival rate, which was 100% for both groups. Secondary outcomes included mPLI, mBI, MSEE, and ZICL. Results demonstrate successful osseointegration and no significant differences in clinical indices across abutment inclinations.
Indicated for adult patients with partially or fully edentulous maxilla requiring support for fixed or removable prosthetic devices in the upper jaw arch to restore esthetics and chewing function.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is stacked, with "U.S. FOOD & DRUG" on top and "ADMINISTRATION" on the bottom.
October 11, 2024
JDentalCare Srl % Pantaleoni Maurizio Consultant Maurizio Pantaleoni Via Borgo Santa Cristina 12 Imoa, 40026 ITALY
Re: K240143
Trade/Device Name: JDZygoma dental implants Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: September 16, 2024 Received: September 16, 2024
Dear Pantaleoni Maurizio:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sherrill Lathrop Blitzer
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K240143
Device Name
JDZygoma dental implants
### Indications for Use (Describe)
JDZygoma Dental implant is a JDentalCare implant system.
JDZvgoma Dental implant System is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxilla, in order to restore patient esthetics and chewing function.
The JDZygoma Implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary JDZYGOMA DENTAL IMPLANTS
This 510(k) Summary is being submitted per 21 CFR 807.92.
### 1. General Information
| Submitter: | J DENTAL CARE S.r.l. is located at:<br>Via Dino Campana, 2<br>MODENA<br>ITALY<br>Tel. +39 059 454255<br>Fax +39 059 450045 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------|
| Consultant/ Contact: | Maurizio Pantaleoni<br>Via Borgo Santa Cristina 12<br>40026 Imola (BO) |
Summary Prepared Date:
October 11, 2024
Mobile: +39 348-4435155
Email: maurizio.pantaleoni@gmail.com
- 2. Names
| Device Name: | JDZYGOMA DENTAL IMPLANTS |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name: | Implant, Endosseous, Root-form |
| Product Code: | DZE |
| Secondary Product Code: | NHA |
| Regulation number: | 872.3640 |
| CLASS | ============================================================================================================================================================================== |
{5}------------------------------------------------
### 3. Predicate and Reference Devices
The JDZYGOMA DENTAL IMPLANTS, is substantially equivalent to the following predicate device legally marketed in US:
| Applicant | Device name | 510(k) Number |
|---------------|--------------------------------------------------------------------|---------------|
| Noris Medical | Noris Medical Dental Implants System<br>(Primary Predicate device) | K151909 |
Reference devices that have been used for comparison of technological features are the following:
| Applicant | Device name | 510(k)<br>Number |
|--------------------------------|--------------------------------------|------------------|
| Noris Medical | Noris Medical Dental Implants System | K210356 |
| Nobel Biocare AB | NobelZygoma 45° | K152093 |
| Nobel Biocare AB | Zygoma Angled Abutments | K052885 |
| Southern Implants (Pty)<br>Ltd | Zygomatic Implant System<br>(ZYGAN) | K173343 |
| J DENTAL CARE S.r.l. | JDIcon | K182081 |
#### 4. Device Description
JDZygoma dental implants are implantable devices produced in commercially pure titanium, intended to be integrated in the Zygomatic bone to support prosthetic devices, such as artificial teeth, in order to restore chewing function.
The JDZygoma implant bodies are single use, gamma sterilized devices that shall be used by qualified medical personnel. The abutments are supplied non sterile and must be sterilized prior to use.
The JDZYGOMA DENTAL IMPLANTS includes 2 lines of devices with different body diameters of implants (3.9 mm and 4.3 mm) with different implant lengths:
- Diameter(mm) 3.9 Lengths (mm): 30 / 35 / 37.5 / 40 / 42.5 / 45 / 47.5 / 50 / 52.5 / 55 / 57.5 ●
- Diameter(mm) 4.3 Lengths (mm): 30 / 35 / 37.5 / 40 / 42.5 / 45 / 47.5 / 50 / 52.5 / 55 / 57.5 .
The surfaces JDZYGOMA DENTAL IMPLANTS, both 3.9 mm and 4.3 mm of diameter, is machined in the distal part of the fixture while the apical part, the threaded part, is characterized by a surface that is a SLA surface obtained through a sandblasting process followed by acid etching.
Machined surface has been analyzed form a morphological point of view using scanning electron microscope (SEM) with different level of magnification (100x, 500x, 1000x and 250x) and also performing a 3D analysis, that allow to define roughness profile with an average height of selected area (Sa) equal to 0.64 um, with a Root mean square gradient (Sdq) of 1.06 and a developed interfacial area ratio (Sdr) of 47.64%.
{6}------------------------------------------------
The abutments are made of titanium grade 5 and have the following angulations:
- . 45° 52.5° and 60°
The connection implant / abutment is done through an internal hexagon.
The conical abutments surface is completely treated with anodic oxidation resulting in perception of yellow color.
The anodization process is performed by immersing the anodization solution and by providing an electrical voltage for a time necessary to activate the process with the material (titanium grade 5) and then performing washing process and drying the devices.
The result of this process has been evaluated from a morphological point of view using scanning electron microscope (SEM), measuring the roughness parameters and performing qualitative analysis of the chemical composition of the abutments surface with X-Ray detector considering and comparing both a non anodized abutments and the same anodized abutments in their finished forms.
These evaluations show a relatively uniform surface with characteristic textures of typical machined titanium, clean and free from significant residues or contaminants, with a morphology of the surface before and after anodization that doesn't show appreciable differences also for roughness and X-ray analysis identify a presence of only of Ti, Al, V, O, as expected.
### 5. Indications for Use
JDZygoma Dental implant is a JDentalCare implant system.
JDZygoma Dental implant System is intended to be implanted in the upper jaw arch to provide support for fixed or removable prosthetic devices in patients with partially or fully edentulous maxilla, in order to restore patient esthetics and chewing function.
The JDZygoma Implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
| | Subject device | Primary Predicate device | Reference device |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | JDENTALCARE | Noris Medical | Nobel Biocare AB |
| | Implant System JDZygoma | Noris Medical Zygomatic Implants<br>System (K151909) | Nobel Zygoma 45°<br>(K152093) |
| Regulation<br>Number: | 872.3640 - Endosseous dental<br>implant | 872.3640 - Endosseous dental implant. | 872.3640 - Endosseous<br>dental implant. |
| Classification | II | II | II |
| Product Code | DZE (implant) | DZE (implant) | DZE (implant) |
| | NHA (Abutments) | NHA (Abutments) | NHA (Abutments) |
| | JDZygoma dental implant is a<br>JDentalCare implant system.<br>JDZygoma Dental implant System<br>is intended to be implanted in the<br>upper jaw arch to provide support | Noris Medical Dental Implants System is<br>intended to replace missing tooth/teeth in<br>either jaw for supporting prosthetic devices<br>that may aid in restoring the patient's<br>chewing function. The procedure can be<br>accomplished in a one-stage or two-stage<br>surgical operation. All implants are<br>appropriate for immediate loading when<br>good primary stability is achieved and with<br>appropriate occlusal loading.<br>Noris Medical Zygomatic Dental Implant<br>System is intended to be implanted in the | Nobel Biocare's Zygoma<br>implants are endosseous dental<br>implants intended to be<br>surgically placed in the bone of<br>the upper jaw arches to provide<br>support for prosthetic devices<br>such as artificial teeth, in order<br>to restore patient esthetics<br>and chewing function. |
| Intended<br>Use | for fixed or removable prosthetic<br>devices in patients with partially<br>or fully edentulous<br>maxillae, in order to restore<br>patient esthetics and chewing<br>function.<br>The JDZygoma Implants are | | The Zygoma Implants may be<br>put into immediate function |
### 6. Comparison of the technological characteristics with the predicate and reference devices
{7}------------------------------------------------
# JDENTALCARE®S.R.L. Traditional 510(k) Submission
| | appropriate for<br>immediate<br>loading when good primary<br>stability is achieved and with<br>appropriate occlusal loading. | upper jaw arch to provide support for fixed<br>or removable prosthetic devices in patients<br>with partially or fully edentulous<br>maxillae. | provided that stability<br>requirements detailed in the<br>directions for use are satisfied. |
|---------------------|------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Indication | Immediate or delayed load | Immediate or delayed load | Immediate or delayed load |
| Placement<br>method | Extra maxillary | Extra maxillary | Inside or outside the maxillary<br>sinus |
Table 1 - Regulatory Comparison Table
| | Subject device<br>JDENTALCARE | | Primary Predicate device<br>Noris Medical | Reference device<br>Nobel Biocare AB |
|--------------------------------------------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| | Implant System<br>JDZygoma<br>Diameter 3.9 | Implant System<br>JDZygoma<br>Diameter 4.3 | Noris Medical<br>Zygomatic Implants<br>System (K151909) | Nobel Zygoma 45°<br>(K152093) |
| Design & Mechanical features | | | | |
| Materials | Titanium grade 4<br>Two pieces | Titanium grade 4<br>Two pieces | Titanium alloy (Grade 5)<br>Two pieces | Titanium grade 4<br>Two pieces |
| Shape | Cylindrical screw<br>straight 0° | Cylindrical screw<br>straight 0° | Cylindrical screw<br>straight 0° | Cylindrical screw<br>preangled 45° |
| Connections | Internal Hexagon<br>Conical connection<br>Platform Diam: 3.4 mm<br>Int. Hex Diam: 2,425mm | Internal Hexagon<br>Conical connection<br>Platform Diam: 3.4 mm<br>Int. Hex Diam: 2,425mm | Internal Hexagon<br>Conical connection<br>Platform Diam: 3,75 mm<br>Int. Hex Diam: 2,42mm | External Hexagon |
| Diameters (mm) | 4.35 max diameter<br>(at platform level)<br>3.9 body<br>3.9 apex | 4.5 max diameter<br>(at platform level)<br>4.3 body<br>4.3 apex | 4.2 max diameter<br>3.75mm platform<br>4.2 body<br>3.5 apex | 4.5 platform-<br>4.0 body<br>3.9 apex |
| Lengths | 30 / 35 / 37.5 / 40 /<br>42.5 / 45 / 47.5 / 50 /<br>52.5/ 55 / 57.5 | 30 / 35 / 37.5 / 40 / 42.5 /<br>45 / 47.5 / 50 / 52.5/ 55/<br>57.5 | Diameter 4.2<br>30 / 35 / 37.5 / 40 / 42.5/<br>45 / 47.5 / 50 / 52.5/ 55/<br>57.5 | Diameter 4.0<br>30 / 35 / 37.5 / 40 / 42.5<br>/ 45 / 47.5 / 50 / 52.5 |
| Abutment | 45°; 52.5°; 60° | 45°; 52.5°; 60° | 17°, 30°, 45°<br>(Extended at 52° and 60°<br>with K210356) | 45°, 62°<br>(62° with the use of<br>K052885) |
| Design of the Thread | | | | |
| Thread Length<br>(starts at apex<br>and extends<br>coronally | 16 mm | 18 mm | 13 mm | 18 mm |
| Thread design | Double lead | Double lead | Double lead | Single lead |
| Thread Pitch | 1.295mm | 0.645mm | 1.295mm | 0.600mm |
| Thread Depth | 0.400mm (mean) | 0.330mm (mean) | 0.500mm (mean) | 0.320mm |
| Thread Width | 0.200mm | 0.100mm | 0.200mm | 0.100mm |
| Design of the surfaces | | | | |
{8}------------------------------------------------
| | Subject device<br>JDENTALCARE | | Primary Predicate<br>device<br>Noris Medical | Reference device<br>Nobel Biocare AB |
|--------------------------------|--------------------------------------------|--------------------------------------------|---------------------------------------------------------|--------------------------------------|
| | Implant System<br>JDZygoma<br>Diameter 3.9 | Implant System<br>JDZygoma<br>Diameter 4.3 | Noris Medical<br>Zygomatic Implants<br>System (K151909) | Nobel Zygoma 45°<br>(K152093) |
| Apex parts | Sandblasting followed<br>by acid etching | Sandblasting followed by<br>acid etching | RBM<br>(Resorbable Blast<br>Media) | TiUnite |
| Treated<br>surface height | 18 mm | 18 mm | 16 mm | TiUnite |
| Crestal part of<br>the implant | Machined | Machined | Machined | TiUnite |
| Design - Dimensions (mm) | | | | |
| Packaging | | | | |
| kind of<br>package | plastic vial + blister | plastic vial + blister | plastic vial + blister | plastic vial + blister |
| Sterile | Yes<br>Gamma Radiation | Yes<br>Gamma Radiation | Yes<br>Gamma Radiation | Yes<br>Gamma Radiation |
Table 2 - Comparison Table for Endosseous Implants fixtures
| | Subject device<br>JDENTALCARE | Primary Predicate device<br>Noris Medical | Reference device |
|-------------------------------------|--------------------------------------------|-------------------------------------------------------------------|-------------------------------------------------------------|
| Features | Implant System JDZygoma | Noris Medical Zygomatic<br>Implants System (K151909 +<br>K210356) | Zygoma Angled Abutments<br>(K052885) |
| Design | | | |
| Images of<br>angulated<br>abutments | Image: angulated abutments for JDENTALCARE | Image: angulated abutments for Noris Medical | Image: angulated abutments for Zygoma |
| Angles | 45°, 52.5°, 60° | 17°, 30°, 45°, 52° (*), 60° (*) | 0°, 17°<br>(work angle from 45° to 62° with<br>the implant) |
| Abutments<br>Height | 4.5 and 5.0mm | 2, 5 mm | 2, 3, 5 mm |
| Materials | | | |
| Materials | Titanium grade 5 | Titanium grade 5 | Titanium grade 5 |
| Packaging | | | |
| Sterility | Non sterile | Non Sterile | Non Sterile |
Table 3 - Comparison Table for Abutments
The fixtures included in the JDentalCare implant system JDZygoma are considered substantially equivalent to the predicate and reference devices due to the following:
- 1) the difference in material respect to the primary predicate device (titanium grade 4 vs 5) doesn't affect the safety and effectiveness of the device, and material is the same material of the reference device.
- 2) are two-pieces implants with a cylindrical shape, straight 0° platform exactly as the primary predicate device Noris Medical Zygomatic Implants System (K151909).
{9}------------------------------------------------
- 3) has a conical connection with internal hexagon, exactly as the primary predicate device with an internal Diameter of the hexagon very similar to the one of the primary predicate device.
- diameter, length and work angles are within the range of the predicate devices and/or of the 4) reference device.
- 5) the thread of JDentalCare implant system JDZygoma can be considered substantially equivalent to the thread of the predicate and reference devices.
- ର) the surface treatment used in JDentalCare implant system JDZygoma doesn't introduce any new issues related to safety and effectiveness of the devices
Finally, where differences are present these are supported also by the availability of the reference devices and by the results of:
- The comparative mechanical fatigue test executed on the worst cases configuration in comparison . with the predicate and reference device, that confirms that fatigue performances of the subject device are comparable to or better than both predicate and reference Devices.
- Real World Evidence
#### 7. Performance Data (non clinical)
A program of design verification and validation testing was performed that includes the following:
- Biocompatibility was conducted according to ISO 10993-1:2018 and ISO 10993-5
- Verification of Washing Process testing was conducted to demonstrate that no residuals substances were present on implant or abutment surface
- Packaging shelf life accelerated aging testing was conducted according to ISO 11607-1
- Comparative mechanical fatigue testing was performed according to a method modified from ISO 14801, on worst-case test articles from the subject, predicate, and reference devices, to best represent the extra-maxillary surgical method
- . Gamma Ray Sterilization Validation in accordance with ISO 11137-1:2006 and ISO 11137-2: 2013
- . Endotoxin tests is performed on all the batches of JDZygoma implants with the highest dental implant surface, to demonstrate that the pyrogen limit specifications through LAL test according to the United States Pharmacopeial Convention is always respected
Results of the evaluations demonstrate that the subject device demonstrated substantially equivalence performance as per its indication for use.
### 8. Clinical data
To support the use of the JDZygoma dental system, including different proposed angulations, implant diameter, and length for use with the extra-maxillary surgical method, real world evidence was provided in the form of two retrospective studies, respectively relevant for JDZygoma lines diameter 3.9mm and 4.3mm with follow-up after placement of 1 year and 3 years.
143 JDZygoma dental implants of the two diameters and different lengths were placed in male and female adult population for a total of 66 patients. Distribution of the site, position, length, and diameter varied. Implant/abutment success, prosthetic survival, general postoperative complications have been evaluated as Iso Modified Plaque Index (mPLI), modified Bleeding Index (mBI), Mucosal Seal Efficacy Evaluation (MSEE) and Zygomatic Implants Classification Level (ZICL), as secondary outcomes.
Outcome measures were evaluated, for the 3.9mm JDZygoma implants at 1 year post-surgery and for 4.3mm implants at 3 years post-surgery.
{10}------------------------------------------------
Based on results of these studies, the survival rates of the 3.9mm and 4.3mm implants which met the criteria for successful implant osseointegration, were 100% for both the studies and no differences were registered among clinical indices related to the different abutments inclinations.
### 9. Conclusions
The JDZYGOMA DENTAL IMPLANTS was evaluated for substantial equivalence using standards, comparative testing and real world evidences.
The totality of information combined between the real world evidence and bench testing was provided to support the performance of the implant-abutment configuration up to 60° with lengths of 55mm and 57.5mm for the diameter of 3.9mm for the demonstration of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.