K152093 · Nobel Biocare AB · DZE · Mar 16, 2016 · Dental
Device Facts
Record ID
K152093
Device Name
NobelZygoma 45°
Applicant
Nobel Biocare AB
Product Code
DZE · Dental
Decision Date
Mar 16, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Nobel Biocare's Zygoma implants are endosseous dental implants intended to be surgically placed in the bone of the upper jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore patient esthetics and chewing function. The Zygoma Implants may be put into immediate function provided that stability requirements detailed in the directions for use are satisfied.
Device Story
NobelZygoma 45° is a threaded, root-form titanium dental implant; designed for surgical placement through the maxillary sinus into the Zygomaticus bone. Device supports prosthetic devices (e.g., artificial teeth) to restore chewing function and esthetics. Used by dental surgeons in clinical settings. Implant features 45° angled external hex connection; TiUnite surface treatment; lengths 30–52.5 mm. Provides stable anchorage in cases of advanced jaw bone resorption. Output is mechanical support for dental prosthetics; clinical decision-making relies on surgeon assessment of bone stability and prosthetic requirements. Benefits include restoration of oral function for patients with insufficient maxillary bone for standard implants.
Clinical Evidence
Bench testing only. Fatigue limit determined using modified ISO 14801 testing comparing subject and predicate devices under identical worst-case conditions. Predicate data leveraged for biocompatibility, sterilization (ANSI/AAMI/ISO 11137), packaging, and shelf-life (5-year expiration).
Technological Characteristics
Material: CP Titanium grade 4. Surface: TiUnite. Connection: 45° angled 4.0 mm external hex. Dimensions: 30–52.5 mm length; 4.5 mm platform diameter. Principle: Endosseous osseointegration. Sterilization: Gamma radiation (ANSI/AAMI/ISO 11137).
Indications for Use
Indicated for patients requiring dental implants in the upper jaw arches to support prosthetic devices for restoration of esthetics and chewing function.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are connected by a flowing, ribbon-like shape at the bottom. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 16, 2016
Nobel Biocare AB c/o Mr. Charlemagne Chua Senior Regulatory Affairs Manager Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda, California 92887
Re: K152093
> Trade/Device Name: NobelZygoma 45° Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: February 11, 2016 Received: February 12, 2016
#### Dear Mr. Chua:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang
-s
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Form Approved: OMB No. | 0910-0120 |
|--------------------------|------------------|
| Expiration Date: | January 31, 2017 |
| See PRA Statement below. | |
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use**
| 510(k) Number (if known) | K152093 |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | NobelZygoma 45° |
| Indications for Use (Describe) | Nobel Biocare's Zygoma implants are endosseous dental implants intended to be surgically placed in the bone of the upper jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore patient esthetics and chewing function. The Zygoma Implants may be put into immediate function provided that stability requirements detailed in the directions for use are satisfied. |
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| | <span style="display:inline-block; margin-right: 5px;"> ☑ </span> Prescription Use (Part 21 CFR 801 Subpart D) |
| | <span style="display:inline-block; margin-right: 5px;"> ☐ </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
| FORM FDA 3881 (1/14) | Page 1 of 1 |
|----------------------|-------------------------------------------|
| FEB 2016 | K152093-S001 |
| | Page 16 of 68 |
| | PSC Publishing Services (301) 443-6740 EF |
{3}------------------------------------------------
I. SUBMITTER
Nobel Biocare AB Vastra Hamngatan 1 Goteborg, SE-411 17 Sweden
Submitted by: Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda, CA 92887
Contact Person: Charlemagne Chua, Senior Regulatory Affairs Manager Phone: (714) 282-4800 x 7830 Fax: (714) 998-9348
Date Prepared: July 24, 2015 (revised March 10, 2016)
II. DEVICE
Name of Device: NobelZygoma 45° Common or Usual Name: Endosseous Dental Implant Classification Name: Endosseous Dental Implant (21 CFR 872.3640) Regulatory Class: II Product Code: DZE
III.PREDICATE DEVICE
Nobel Biocare – Zygoma Implant (K070182)
IV. DEVICE DESCRIPTION
Nobel Biocare's NobelZygoma 45° implants are threaded, root-form titanium dental implants intended to extend through the maxillary sinus into the Zygomaticus bone to support prosthetic devices, such as artificial teeth, in order to restore chewing function.
The NobelZygoma 45° implants are available in lengths between 30 and 52.5 mm and have a 4.5 mm external hex connection. They are made of commercially pure titanium and have the Nobel Biocare TiUnite surface treatment.
A.4.
{4}------------------------------------------------
# V. INDICATIONS FOR USE
Nobel Biocare's Zygoma implants are endosseous dental implants intended to be surgically placed in the bone of the upper jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore patient esthetics and chewing function. The Zygoma Implants may be put into immediate function provided that stability requirements detailed in the directions for use are satisfied.
{5}------------------------------------------------
#### VI. Comparison of Technological Characteristics
| characteristic | CANDIDATE | PREDICATE | |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | NobelZygoma 45° | Zygoma Implant (K070182) | |
| Features | Thread Design | Single lead thread | Single lead thread |
| | Implant Body Design | Parallel wall with 2 diameters transition at 2 mm from platform<br>- 4.5 mm at the platform<br>- 3.9 mm at the apex<br>Threads starting 18 mm from apex | Parallel wall with 2 diameters transition at 13.7 mm from platform<br>- 4.4 mm at the platform<br>- 3.9 mm at the apex<br>Threads starting 2.8 mm from apex |
| | Implant Tip Design | Tapered with cut out flutes. | Rounded with through hole |
| | Implant Length | 30, 32.5, 35, 37.5, 40, 42.5, 45, 47.5, 50, 52.5 mm | 30, 35, 40, 42.5, 45, 47.5, 50, 52.5 mm |
| | Connection Type | 45° angled 4.0 mm external hex | 45° angled 4.0 mm external hex |
| | Compatible Abutments | Zygoma system Multi-unit abutments | Zygoma system Multi-unit abutments |
| | Restoration compatibility | Fixed bar | Fixed Bar |
| | Device Material | CP Titanium grade 4 | CP Titanium grade 4 |
| | Surface | TiUnite | TiUnite |
| | | | |
| | Intended Use/ Principles of Operation | Nobel Biocare's Zygoma implants are endosseous implants and are integrated in the zygomatic bone (osseointegration). They are intended to be used for anchoring or supporting tooth replacements to restore chewing function. | Nobel Biocare's Zygoma implants are endosseous implants and are integrated in the zygomatic bone (osseointegration). They are intended to be used for anchoring or supporting tooth replacements to restore chewing function. |
| | Indications for Use | Nobel Biocare's Zygoma implants are endosseous dental implants intended to be surgically placed in the bone of the upper jaw arches to provide support for prosthetic devices such as artificial teeth in order to | Nobel Biocare's Zygoma implants are endosseous dental implants intended to be surgically placed in the bone of the upper jaw arches to provide support for prosthetic devices such as artificial teeth in order to |
| characteristic | CANDIDATE | PREDICATE | |
| | NobelZygoma 45° | Zygoma Implant (K070182) | |
| | restore patient esthetics and chewing function. The<br>Zygoma Implants may be put into immediate function<br>provided that stability requirements detailed in the<br>directions for use are satisfied. | restore patient esthetics and chewing function. The<br>Zygoma Implants may be put into immediate function<br>provided that stability requirements detailed in the<br>manual are satisfied. | |
{6}------------------------------------------------
{7}------------------------------------------------
Analysis of Differences Between Subject Device and Predicate
The subject NobelZygoma 45° implants have a fundamentally similar design as the predicate Zyqoma Implants (K070182). The two implant designs share the same materials, manufacturing methods, and use the TiUnite surface treatment. The two designs have the same 45° anqulated platform and use the same abutments and surgical tooling. The two designs are available in lengths from 30 to 52.5 mm. The illustrations below show the subject NobelZygoma 45° implant next to the predicate Zygoma Implant.
| of Personal of all be life of the labor of the | |
|------------------------------------------------|--|
|------------------------------------------------|--|
NobelZvgoma 45°
Image /page/7/Picture/4 description: The image shows a long, silver-colored, cylindrical object that appears to be a screw or bolt. The object has a series of ridges or threads running along its length, and one end has a hole or slot, while the other end is pointed. The screw is likely made of metal and is designed to be inserted into a material to hold it in place.
Zyqoma Implant (K070182)
Although the subject and predicate devices share many similarities, a number of design characteristics have been changed. These changes are listed below.
- -The tip of the NobelZygoma 45° has been modified to a tapered design with flutes to facilitate bone cutting. The predicate Zygoma Implant has a rounded tip with through hole.
- -The threads of the NobelZygoma 45° start at the implant apex and extend 18 mm. The threads of the predicate start 2.8 mm from the apex and extend to the platform.
- -The profile of the NobelZygoma 45° is different from the predicate. Both implants are parallel walled with 2 diameters. The NobelZygoma 45° has a diameter of 4.5 mm at the platform and a diameter of 3.9 mm starting from 2 mm below the platform. The predicate Zygoma Implant is 4.4 mm at the platform and 3.9 mm starting 13.7 mm from the platform
Summary:
These changes were made to modify the implant design to better adapt to situations of advanced jaw bone resorption in which Zygomatic implants are being used. The changes do not raise different questions of safety or effectiveness compared to the predicate. Differences in technology were evaluated through comparative performance testing.
{8}------------------------------------------------
## VII. PERFORMANCE DATA
#### Summary of Non-Clinical Testing:
Since the subject device does not represent a new worst case, data from the predicate device was leveraged in the following aspects of the 510(k).
- -Sterile Device Information
- The sterilization method for the subject device is the same as the o predicate. The sterilization method is Gamma radiation and has been validated in accordance with ANSI/AAMI/ISO 11137. Therefore, no additional testing was required.
- -Device Packaging
- The packaging for the subject device is the same as the predicate. o This is a titanium cylinder placed in a plastic vial with PVC shrinkwrap and tamper resistant strip. Therefore, no additional testing was required.
- Shelf Life
- o The packaging for the subject device is the same as the predicate and is labeled with a 5 year expiration date. Real time aging was used to determine the expiration dating. Therefore, no additional testing was required.
- Biocompatibility -
- The subject device is manufactured from the same material using o the same manufacturing method as the predicate, has the same intended use, and the same patient contact type and duration. Therefore, no additional testing was required.
The fatigue limit of the NobelZygoma 45° was determined using a modified version of ISO 14801. The modifications to ISO 14801 were done to reflect the likely worst-case clinical use of the device. Both the subject and predicate device were tested under identical conditions. The results of the testing were used to address questions related to substantial equivalence based on difference in design between the subject and predicate devices.
### VIII. CONCLUSIONS
The NobelZygoma 45° was evaluated for substantial equivalence using standard and/or comparative testing. In cases where the NobelZygoma 45° could be shown to not represent a worst-case with respect to the Zygoma Implant, data from the predicate device was leveraged to support the subject device. Based on technological characteristics and non-clinical test data included in this submission, the NobelZygoma 45° has been shown to be substantially equivalent to the Zygoma Implant.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.