Published clinical literature (retrospective study)
Retrospective clinical results from 40 implants in 20 subjects were used to support the performance of the Oncology implant design, demonstrating 100% success and no soft tissue complications.
Southern Implants Zygomatic System Standard implants, Zygan (narrow apex) implants, and Oncology implants are intended to be implanted in the upper jaw arch to provide support for fixed or removable dental prostheses in patients with partially or fully edentulous maxillae. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Device Story
Zygomatic Implant System consists of root-form dental implants and mating abutments for placement into zygomatic bone. Implants include Standard, Zygan (narrow apex), and Oncology designs; all feature 55° angled external hex abutment interface. Abutments include Compact Conical and Titanium Cylinder designs for screw-retained overdenture support. Used by dental surgeons in clinical settings to rehabilitate edentulous maxillae. Implants provide structural support for dental prostheses; immediate loading possible with adequate primary stability. Benefits include restoration of oral function and aesthetics for patients with severe maxillary bone loss.
Clinical Evidence
Bench testing only. Includes engineering analysis, dimensional analysis, and dynamic compression-bending fatigue testing per ISO 14801. Biocompatibility referenced from predicates. Bacterial endotoxin testing per USP <85>. Clinical data for oncology implant design referenced from retrospective study (Boyes-Varley et al., 2007) of 40 implants in 20 subjects, showing 100% success at 96 months.
Technological Characteristics
Materials: Unalloyed titanium (ASTM F67) for implants/cylinder abutments; titanium alloy (ASTM F136) for conical abutments. Surface: Aluminum oxide grit-blasted. Design: Root-form, threaded, 55° head angulation. Dimensions: 3.4mm-4.3mm diameters, 27.5mm-57.5mm lengths. Sterilization: Validated per ISO 17665-1/2. Connectivity: None (mechanical device).
Indications for Use
Indicated for patients with partially or fully edentulous maxillae requiring support for fixed or removable dental prostheses via implantation in the upper jaw arch (zygomatic bone). Appropriate for immediate loading given sufficient primary stability and occlusal loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
February 27, 2018
Southern Implants (Pty) Ltd % Kevin Thomas Vice President and Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real Suite 400 San Diego, California 92130
Re: K173343
Trade/Device Name: Zygomatic Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: January 25, 2018 Received: January 26, 2018
## Dear Kevin Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Andrew I. Steen -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
K173343
Device Name
Zygomatic Implant System
Indications for Use (Describe)
Southern Implants Zygomatic System Standard implants, Zygan (narrow apex) implants, and Oncology implants are intended to be implanted in the upper jaw arch to provide support for fixed or removable dental prostheses in patients with partially or fully edentulous maxillae. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
## 510(k) Summary - K173343 Zygomatic Implant System Southern Implants (Pty) Ltd February 27, 2018
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Southern Implants (Pty) Ltd<br>1 Albert Road<br>Irene, Gauteng, 0062 South Africa<br>Telephone +27 12 667 1046<br>Fax +27 12 667 1029 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | Lauranda G. Breytenbach<br>Head of Regulatory Affairs and Quality |
| Representative/Consultant | Kevin A. Thomas, PhD<br>Floyd G. Larson<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone: +1 858-792-1235<br>Fax: +1 858-792-1236<br>Email: kthomas@paxmed.com<br>flarson@paxmed.com |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | Zygomatic Implant System |
|---------------------------|---------------------------|
| Common Name | Dental implant |
| Classification Name | Endosseous dental implant |
| Classification Regulation | 21 CFR 872.3640, Class II |
| Primary Product Code | DZE |
| Secondary Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
#### PREDICATE DEVICE INFORMATION
The primary predicate device is K093562. The reference devices are K151909, K070841, K053478, and K141777.
#### INDICATIONS FOR USE STATEMENT
Southern Implants Zygomatic System Standard implants, Zygan (narrow apex) implants, and Oncology implants are intended to be implanted in the upper jaw arch to provide support for fixed or removable dental prostheses in patients with partially or fully edentulous maxillae. All implants are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
{4}------------------------------------------------
## SUBJECT DEVICE DESCRIPTION
This submission includes fully threaded and partially threaded root-form dental implants and mating abutments designed for placement into the zygomatic bone. All implants are provided with an external hexagon abutment interface angled 55° at the head of the implants are provided in three designs: Standard implant, fully threaded, diameter 4.3 mm (coronal) tapering to 3.8 mm (apical), in lengths of 30 mm and 57.5 mm; Zygan implant, partially threaded, diameter 4.3 mm (coronal) tapering to 3.4 mm (apical), in lengths from 30 mm to 57.5 mm; and Oncology implant, partially threaded diameter 4.3 mm (coronal) tapering to 3.8 mm (apical), in lengths from 30 mm to 47. 5 mm.
This submission includes additional designs of Compact Conical Abutments (gingival heights 2, 3, 4, and 5.5 mm) for use with any of the implants. This submission also includes a Titanium Cylinder Abutment, for use with the Compact Conical Abutments, with a collar (gingival) height of 5 mm, and a prosthetic platform diameter of 3.4 mm. All subject device abutments are for support of screw-retained overdenture prosthetic restorations.
All subject device implants are manufactured from unalloyed titanium conforming to ASTM F67. The threaded portions of the implants have the identical aluminum oxide grit-blasted surface as the implants cleared in K093562. The subject device Compact Conical Abutments are manufactured from titanium alloy conforming to ASTM F136. The subject device Titanium Cylinder Abutment is manufactured from unalloyed titanium conforming to ASTM F67. All of the subject device components are manufactured in the same facilities using the same manufacturing processes as used for the previously cleared predicate devices in K093562 and K070841.
## PERFORMANCE DATA
Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence include: biocompatibility (referenced from K093562 and K070841), engineering analysis, dimensional analysis, sterilization validation testing according to ISO 17665-1, ISO TS 17665-2, and dynamic compressionbending testing according to ISO 14801 Dentistry - Implants - Dynamic fatigue test for endosseous dental implants. The Limulus Amebocyte Lysate (LAL) test for detection and quantitation of bacterial endotoxin was conducted in accordance with USP 39-NF 34 <85> Bacterial Endotoxin Test, using the kinetic chromogenic test method. Clinical data referenced in this submission to support the oncology implant design included the retrospective results from 40 implants (in 20 subjects) demonstrating 100% implant success and no soft tissue complications at up to 96 months [Boyes-Varley JG, et al., Int J Prosthodont 2007; 20:521-531].
#### EQUIVALENCE TO MARKETED DEVICE
Southern Implants (Ptv) Ltd submits the information in this Premarket Notification to demonstrate that. for the purposes of FDA's regulation of medical devices, the subject device is substantially equivalent in indications and design principles to the following legally marketed devices:
K093562, Zygomatic Implant System, Southern Implants, Inc.;
K151909, Noris Medical Zygomatic Dental Implant System, Noris Medical Ltd .;
K070841, Endosseous Dental Implant System, Southern Implants, Inc .;
K053478, Endosseous Dental Implant System, Northern Implants, LLC; and
K141777. Neodent Implant System. JJGC Indústria e Comércio de Materiais Dentários SA.
{5}------------------------------------------------
A comparison of the technological characteristics of the subject device and the primary predicate device K093562 is provided in the following table.
| | Subject Device | Primary Predicate Device |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison | Zygomatic Implant System<br>Southern Implants (Pty) Ltd | K093562<br>Zygomatic Implant System<br>Southern Implants, Inc. |
| Indications for Use<br>Statement | Southern Implants Zygomatic System Standard<br>implants, Zygan (narrow apex) implants, and Oncology<br>implants are intended to be implanted in the upper jaw<br>arch to provide support for fixed or removable dental<br>prostheses in patients with partially or fully edentulous<br>maxillae. All implants are appropriate for immediate<br>loading when good primary stability is achieved and<br>with appropriate occlusal loading. | The Zygomatic implant is intended to be implanted in<br>the upper jaw arch to provide support for fixed or<br>removable dental prostheses in patients with partially<br>or fully edentulous maxillae. |
| Implants | | |
| Design | Fully and partially threaded root-form implants for<br>placement into the zygoma | Fully threaded root-form implants for placement into<br>the zygoma |
| Platform Ø | 4.05 mm | 4.05 mm |
| Implant Ø | Standard implant:<br>4.3 mm (coronal) taper to 3.8 mm (apical)<br>Zygan implant:<br>4.3 mm (coronal) taper to 3.4 mm (apical)<br>Oncology implant:<br>4.3 mm (coronal) taper to 3.8 mm (apical) | Standard implant:<br>4.3 mm (coronal) taper to 3.8 mm (apical) |
| Implant Lengths | Standard implant: 30 mm, 57.5 mm<br>Zygan implant: 30 mm - 57.5 mm<br>Oncology Implant: 27.5 mm - 47.5 mm | Standard implant: 35 mm - 55 mm |
| Threaded Lengths | Standard implant: Fully threaded<br>Zygan implant: 6 mm coronal + 15 mm apical<br>Oncology implant: 20 mm (apical only) | Standard implant: Fully threaded |
| Implant body-abutment<br>connection angle | 55° | 55° |
| Abutments | | |
| Design | One-piece, compact conical design | One-piece, compact conical design |
| Implant Interface | External hex; 55° angulation at head of implant | External hex; 55° angulation at head of implant |
| Gingival Height | 2.0 mm - 5.5 mm | 1 mm |
| Abutment Angle | 0° (straight) | 0° (straight) |
| Prosthesis Attachment | Screw-retained, multi-unit | Screw-retained, multi-unit |
| Materials | | |
| Implants | Unalloyed titanium, ASTM F67 | Unalloyed titanium, ASTM F67 |
| Abutments | Unalloyed titanium, ASTM F67;<br>Titanium alloy, ASTM E136 | Unalloyed titanium, ASTM F67:<br>Titanium alloy, ASTM E136 |
The Indications for Use Statements for the subject device and the primary predicate device K093562 are similar with the subject device statement including the additional Zygan and Oncology implants. The slight differences in wording between the Indications for Use Statements for the subject device and the primary predicate device do not affect the intended use as dental implants placed into the zygoma for rehabilitation of the edentulous maxilla.
The primary predicate device K093562 is for substantial equivalence of the subject device implant designs. The subject device Standard implants have the identical design as implants in K093562 except
{6}------------------------------------------------
for the additional lengths. The subject device Zygan implants have a design that is substantially equivalent design to implants in K093562, with the differences being the non-threaded region (starting 6 mm below the head of the implant), the diameter tapering to 3.4 mm at the apex (versus 3.8 mm for Standard implants), and the range of implant lengths. Similarly, the subject device Oncology implants have a design that is substantially equivalent to implants in K093562, with the differences being the nonthreaded region starting below the head of the implant, and the range of implant lengths.
The reference device K151909 is for substantial equivalence of the subject device Oncology implant design. The implants cleared in K151909 and the subject device Oncology implants both have a nonthreaded region starting at the coronal end (K151909) or just below the implant head (subject device), and a threaded region that tapers to the apex. The reference device K151909 also is for the range of implant lengths.
These differences in design between the subject devices and primary predicate devices were accounted for by including additional instructions in the labeling; therefore, the differences do not impact cafety and effectiveness.
The reference device K070841 is for substantial equivalence of the subject device Compact Conical Abutment design. The subject device Compact Conical Abutments have the identical design as components cleared in K070841, with the only difference being additional sizes of gingival height.
The reference device K053478 is for substantial equivalence of the subject device Titanium Abutment design. The subject device Titanium Abutment has the identical design as components cleared in K053478, with the only difference being a change in gingival height.
The reference device K141777 is for substantial equivalence of the range of implant lengths, and for substantial equivalence of mechanical performance discussed below.
The subject device and the predicate devices all incorporate the same materials and encompass similar ranges of dimensions. The surface treatment applied to the subject device implants is identical to that cleared in K093562. The subject device and all predicate devices are provided sterile for single-patient, single-use.
The subject device implants are packaged preassembled with a titanium fixture mount. The fixture mount is made from the same unalloyed titanium as the implant-fixture mount assembly is attached to a stainless steel clip within a polyethylene terephthalate (PET) thermoformed tray, and sealed with a Tyvek® lid. The actual implant does not have any contact with the packaging material. The thermoformed tray with Tyvek® lid forms the primary sterile barrier and is placed in a PET box. The product label is placed on the Tyvek® lid. The outer box is sealed with an address label and a product description label. For the predicate device K093562, the sterile barrier is provided by the same packaging. a PET tray (blister) sealed with a Tyvek® Lid. The difference between the subject device packaging and the predicate device K093562 packaging is in the additional packaging of the implant and mount within a rigid plastic cylinder which is placed in the PET tray and sealed with the Tyvek® lid. For both the subject device and the predicate device, the primary sterile barrier is placed in a clear rigid PET box, which does not provide an additional sterile barrier but functions to protect the tray/Tyvek® lid package. The change in packaging has been addressed by specific labeling precautions, with images, for packaging handling and removal of the device into the surgical field, in order to maintain sterility of the device
{7}------------------------------------------------
Substantial equivalence of the subject device components in terms of biocompatibility is supported by the materials being identical in formulation, processing, component interactions, and storage conditions to the predicate devices in K093562 and K070841.
In support of substantial equivalence in terms of mechanical performance, dynamic compression-bending testing according to ISO 14801 was performed. Dynamic testing was performed on the worst-case subject device constructs. The results from the testing demonstrated fatigue performance substantially equivalent to that of the reference device K141777.
#### CONCLUSION
The subject device and the predicate devices have the same intended use, have similar technological characteristics, and are made of similar materials. The subject device and the predicate devices encompass the same range of physical dimensions, including diameter and length of the implants, and the diameter and angulation of the abutments. The subject device and the predicate devices are packaged in similar materials and sterilized using similar methods.
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.