K240089 · Shenzhen Kaiyan Medical Equipment Co., Ltd. · OHS · Apr 10, 2024 · General, Plastic Surgery
Device Facts
Record ID
K240089
Device Name
Face Patches (MT-12MA, MT-12MC)
Applicant
Shenzhen Kaiyan Medical Equipment Co., Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Apr 10, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Face Patches (Model: MT-12MA) is an Over-the-Counter (OTC) device intended for treatment of mild to moderate inflammatory acne. Face Patches (Model: MT-12MC) is an Over-the-Counter (OTC) device intended for treatment of mild to moderate inflammatory acne and full-face wrinkles (including periorbital wrinkles).
Device Story
Face Patches (MT-12MA, MT-12MC) are OTC LED-based light therapy devices for acne and wrinkle treatment. MT-12MA uses 415nm and 630nm LEDs; MT-12MC uses 415nm, 630nm, 830nm, and 590nm LEDs. Device components include patches, controller, charging cable, mini dots, and goggles. User applies patches to treatment area; therapy is initiated via physical button on controller. Device features automatic 3-minute shut-off; users adjust treatment time based on mode. Output irradiance ranges 30-35 mW/cm². Device provides non-invasive light therapy to skin surface. Benefits include reduction of inflammatory acne and facial wrinkles. Intended for home use by consumers.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), photobiological safety (IEC 62471), biocompatibility (ISO 10993 series), software validation, and usability testing (IEC 62366-1).
Indicated for treatment of mild to moderate inflammatory acne (MT-12MA, MT-12MC) and full-face wrinkles including periorbital wrinkles (MT-12MC). OTC use.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
April 10, 2024
Shenzhen Kaiyan Medical Equipment Co., Ltd Alain Diikstra Official Correspondent Building#3 and Building#5, 40th of Fuxin Street, Huaide Community Fuyong Town, Baoan District Shenzhen, Guangdong 518103 China
Re: K240089
Trade/Device Name: Face Patches (MT-12MA, MT-12MC) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS. OLP. GEX Dated: January 9, 2024 Received: January 12, 2024
Dear Alain Dijkstra:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tanisha L. Digitally signed by Hithe -S Date: 2024.04.10 14:46:04 -04'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240089
Device Name Face Patches (MT-12MA, MT-12MC)
Indications for Use (Describe)
Face Patches (Model: MT-12MA) is an Over-the-Counter (OTC) device intended for treatment of mild to moderate inflammatory acne.
Face Patches (Model: MT-12MC) is an Over-the-Counter (OTC) device intended for treatment of mild to moderate inflammatory acne and full-face wrinkles (including periorbital wrinkles).
Type of Use (Select one or both, as applicable)
| | <div> <span></span> </div> |
|--|----------------------------|
| | <div> <span></span> </div> |
| | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary of K240089
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
#### Submitter's Information 1.
Sponsor Name: Shenzhen Kaiyan Medical Equipment Co., Ltd Establishment Registration Number: 3011644607 Address: Building#3 and Building#5, 40th of Fuxin Street, Huaide Community Fuyong Town, Baoan District, Shenzhen, Guangdong 518103, China Contact Person (including title): Alain Dijkstra (Manager) Tel: +86-135-10378748 Fax: +86-755-25024651 E-mail: requlation@kaiyanmedical.com
### Application Correspondent:
Contact Person: Alain Diikstra Company: Shenzhen Kaiyan Medical Equipment Co., Ltd Address: Building#3 and Building#5, 40th of Fuxin Street, Huaide Community Fuyong Town, Baoan District, Shenzhen, Guangdong 518103, China Tel: +86 755 82129361 Fax: +86 755 25024651 Email: requlation@kaiyanmedical.com
#### 2. Subject Device Information:
Trade Name: Face Patches, model: MT-12MA, MT-12MC Classification Name: Over-The-Counter Powered Light Based Laser For Acne. Light based over the counter wrinkle reduction, Laser surgical instrument for use in general and plastic surgery and in dermatology Review Panel: General & Plastic Surgery Product Code: OHS, OLP, GEX Regulation Number:21 CFR 878.4810 Regulation Class: II
#### 3. Predicate Device Information Predicate Device 1 (K230720)
Sponsor: Light Tree Ventures Europe B.V.
Trade Name: LUSTRE 3XPRESS Light Beauty Therapy patches, model: PR6001, PR5001 Classification Name: Over-The-Counter Powered Light Based Laser For Acne, Light based over the counter wrinkle reduction, Laser surgical instrument for use in general and plastic surgery and in dermatology
Review Panel: General & Plastic Surgery Product Code: OLP Regulation Number: 21 CFR 878.4810 Regulation Class: II
# Predicate Device 2 (K221946)
Sponsor: Light Tree Ventures Europe B.V. Trade Name: LED LIGHT THERAPY MASK, model: MK66R-B Classification Name: Light based over the counter wrinkle reduction Review Panel: General & Plastic Surgery Product Code: OHS Requlation Number: 21 CFR 878.4810 Regulation Class: II
{5}------------------------------------------------
# Predicate Device 3 (K200104)
Sponsor: RAJA Trading Company, Inc. Trade Name: OxyLight Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Review Panel: General & Plastic Surgery Product Code: GEX, GFE Requlation Number: 21 CFR 878.4810, 21 CFR 878.4820 Regulation Class: II
#### Device Description 4.
The Face Patches, Model: MT-12MA and MT-12MC is an over-the-counter light emitting diode (LED) device.
Model: MT-12MA can emits 415nm and 630nm energy to treat mild to moderate acne. There is a total of 6 LEDs, providing a power intensity of approximately 30 mW/cm².
Model: MT-12MC has 3 modes (mode 1: 630+415nm, mode 2: 630+830nm, mode 3: 590nm), it can emit 630nm and 415 nm energy to treat mild to moderate acne and provide approximately 30 mW/cm²; 630nm and 830nm energy to treat full-face wrinkles and provide approximately 35 mW/cm²; and 590nm energy to treat wrinkles and provide approximately 35 mW/cm². There is a total of 18 LEDs.
The device components include the device containing two face patches, a controller, a charging cable, 120 Mini Dots, goggles and user manual. The user pastes the device on the treatment area, and the device will shut down automatically after a 3-minute after finishing treatment. Users should adjust the treatment time based on different modes and actual treatment situations. The device has one button, and delivery of therapy requires the operator to depress a physical switch on the device.
#### Intended Use / Indications for Use 5.
Face Patches (Model: MT-12MA) is an Over-the-Counter (OTC) device intended for treatment of mild to moderate inflammatory acne.
Face Patches (Model: MT-12MC) is an Over-the-Counter (OTC) device intended for treatment of mild to moderate inflammatory acne and full-face wrinkles (including periorbital wrinkles).
#### Comparison to predicate devices 6.
Compare with the predicate devices, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between the subject device and predicate devices do not raise new questions of safety or effectiveness.
| Elements of<br>Comparison | Subject device | Predicate device<br>(K230720) | Predicate<br>device<br>(K221946) | Predicate device<br>(K200104) | Remark |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Company | Shenzhen Kaiyan<br>Medical Equipment<br>Co., Ltd | Light Tree<br>Ventures Europe<br>B.V. | Light Tree<br>Ventures Europe<br>B.V. | RAJA Trading<br>Company, Inc. | -- |
| Trade Name | Face Patches | LUSTRE<br>3XPRESS Light<br>Beauty Therapy<br>patches | LED LIGHT<br>THERAPY<br>MASK | OxyLight | -- |
| Model | MT-12MA, MT-<br>12MC | PR6001, PR5001 | MK66R-B | -- | -- |
| 510 (K) Number | K240089 | K230720 | K221946 | K200104 | -- |
| Regulation<br>Class | Class II | Class II | Class II | Class II | Same |
| Indications for<br>Use / Intended | Face Patches<br>(Model: MT-12MA) | The LUSTRE<br>3XPRESS Light | The LED LIGHT<br>THERAPY | The Oxylight is<br>intended for | Different,<br>note 1 |
| Elements of<br>Comparison | Subject device | Predicate device<br>(K230720) | Predicate<br>device<br>(K221946) | Predicate device<br>(K200104) | Remark |
| use | is an Over-the-<br>Counter (OTC)<br>device intended for<br>treatment of mild<br>to moderate<br>inflammatory acne<br>Face Patches<br>(Model: MT-12MC) is<br>an Over-the-<br>Counter (OTC)<br>device intended for<br>treatment of mild<br>to moderate<br>inflammatory acne<br>and full-face<br>wrinkles (including<br>periorbital<br>wrinkles). | Beauty Therapy<br>patches (Model:<br>PR6001, PR5001) is<br>an Over-the-<br>Counter (OTC)<br>device intended for<br>the treatment of<br>mild to moderate<br>acne. | MASK<br>(MK66R-B) is an<br>over-the-counter<br>device that is<br>intended for the<br>use in<br>the treatment of<br>full-face wrinkles. | dermatological use<br>by physicians and<br>healthcare<br>professionals for the<br>following: LED<br>Technology is<br>intended for: -Blue<br>LED – 465nm – to<br>treat dermatological<br>conditions and<br>specifically<br>indicated to treat<br>moderate<br>inflammatory acne<br>vulgaris. -Red LED<br>625nm- for<br>treatment of<br>superficial, benign<br>vascular and<br>pigmented lesions. -<br>Yellow LED 590nm<br>- treatment of<br>periorbital wrinkles<br>and rhytides.<br>Microdermabrasion<br>is intended for<br>exfoliation of the<br>skin. Oxygen spray<br>is intended to<br>refresh the skin. | |
| Power Source | Controller: 3.6V,<br>65mAh lithium<br>battery, 0.234Wh | Input: 100-<br>240Va.c.,<br>50Hz/60Hz;<br>output: 5Vd.c., 2A | Input: 100-<br>240Va.c.,<br>50Hz/60Hz,<br>Output: 5Vd.c.,<br>2A. | not available | Different,<br>note 2 |
| Wavelengths | MT-12MA:<br>$630 \pm 10nm$<br>$415 \pm 10nm$<br>MT-12MC: $415nm \pm$<br>$10nm, 630nm \pm$<br>$10nm, 830nm \pm$<br>$10nm, 590nm \pm$<br>10nm | $630 \pm 10nm$<br>$415 \pm 10nm$ | Red: 630±5nm<br>NIR: 830nm | 590 nm +/- 5nm | Similar,<br>note 3 |
| Irradiance<br>source | LED | LED | LED | LED | Same |
| LED number | For MT-12MA:<br>415nm: 6<br>630nm: 6 | Blue: 12 Red: 12<br>Total 24 | 630+/-5nm: 66<br>830nm: 66 | not available | Different<br>note 4 |
| Elements of<br>Comparison | Subject device | Predicate device<br>(K230720) | Predicate<br>device<br>(K221946) | Predicate device<br>(K200104) | Remark |
| | For MT-12MC:<br>415nm: 6<br>630nm: 12 (two<br>type)<br>830nm: 6<br>590nm: 6<br>Total 18 LEDs | | | | |
| | Intended<br>location Use | Face | Face | Face | Face |
| Irradiance<br>(mW/cm2) | For MT-12MA:<br>630nm: 5<br>415nm: 25<br>630+415nm: 30<br>For MT-12MC:<br>630nm (type 1): 5<br>415nm: 25<br>Mode 1 - total: 30<br>830nm: 15<br>630nm (type 2): 20<br>Mode 2 - total: 35<br>Mode 3 - 590nm:<br>35 | Red: 5<br>Blue: 25<br>630+415nm: 30 | 630+/-5: 15-20<br>830nm: 10-15<br>630+830nm: 30 | Yellow (590nm): 35 | Similar,<br>note 5 |
| Dose / per<br>time(J/cm2) | For MT-12MA:<br>630+415nm: 5.4<br>For MT-12MC:<br>630+415nm: 5.4<br>630+830nm: 18.9<br>590nm: 44.1 | 630+415nm: 5.4 | 630+830nm: 18 | 590nm: 42 | Similar,<br>note 5 |
| LED distribution | Uniform<br>distribution | Uniform<br>distribution | Uniform<br>distribution | Uniform distribution | Same |
| Treatment Time | For MT-12MA<br>(630+415nm):<br>3 minutes per<br>treatment<br>For MT-12MC<br>(mode 1 -<br>630+415nm): 3<br>minutes per<br>treatment<br>For MT-12MC<br>(mode 2 -<br>630+830nm): 9<br>minutes per | 3 minutes per<br>treatment | 10 minutes per<br>treatment | 20 minutes | Similar,<br>note 5 |
| Elements of<br>Comparison | Subject device | Predicate device<br>(K230720) | Predicate<br>device<br>(K221946) | Predicate device<br>(K200104) | Remark |
| | treatment<br><br>For MT-12MC<br>(mode 3 - 590nm):<br>21 minutes per<br>treatment | | | | |
| EMC | IEC60601-1-2 | IEC60601-1-2 | IEC60601-1-2 | IEC 60601-1-2 | Same |
| Safety | IEC60601-1<br>IEC60601-2-57<br>IEC60601-1-11<br>IEC62471<br>IEC 60601-1-6 | IEC 60601-1<br>IEC 60601-2-57<br>IEC 60601-1-11<br>IEC 62471 | IEC 60601-1<br>IEC 60601-2-57<br>IEC 60601-1-11<br>IEC 62471<br>IEC 62366-1 &<br>IEC 60601-<br>1-6 | IEC 60601-1, EN<br>60601-1-2 | Same |
| Biocompatibility | ISO 10993-1<br>ISO10993-5<br>ISO10993-10<br>ISO 10993-23<br>ISO 10993-11 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-23<br>ISO 10993-11 | ISO 10993-1,<br>ISO 10993-5,<br>ISO 10993-10 | not available | Same |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
### Comparison in Detail(s):
Note 1: Although the "Indications for Use / Intended use" is a little different from the predicate devices, the acne removal function of the model MT-12MA and MT-12MC (mode1) of subject device is consistent with predicate device (K230720), the wrinkle removal function of the model MT-12MC (mode 2) of subject device is consistent with predicate device (K221946), the wrinkle removal function of the model MT-12MC (mode 3) of subject device is consistent with predicate device (K200104). So, this difference will not affect the safety and effectiveness of the subject device.
Note 2: Although the "Power Supply" is a little different from the predicate devices, they all complied with the IEC 60601-1, IEC 62133-2, and IEC 60601-1-2 safety standards' requirements. So, these slight differences will not raise any safety or effectiveness issues.
Note 3: The wavelengths of the subject device can be covered by the predicate device. The wavelength of MT-12MA and MT-12MC Mode 1 is 630nm and 415nm, which is consistent with the predicate device 1 (K230720); The wavelength of MT-12MC Mode 2 is 830nm which is consistent with the predicate device 2 (K221946); The wavelength of MT-12MC Mode 3 is 590nm which is consistent with the predicate device 3 (K200104).
So, the treatment wavelength of the subject device is acceptable.
Note 4: Although the "Number of LEDs" is different from the predicate devices, it will not change the irradiance which is similar. So, these differences will not raise any safety or effectiveness issues.
Note 5: Although the "Irradiance", "Dose / per time" and "Treatment Time" are a little different from the predicate devices, the values of each treatment for these parameters are very close. Dose is equal to Irradiance (W/cm²) * Time (s). Through calculation, it can be concluded that the dose of the subject device at 630+415nm is 0.03 * 3 * 60=5.4J/cm², which is consistent with the dose of the predicate device (K230720) at 0.03 * 3 * 60=5.4J/cm²; The dose of subject device 830+630nm is 0.035 * 9 * 60=18.9J/cm², which is similar to the dose of predicate device (K221946) of 0.03 * 10 * 60=18J/cm², and the dose of
{9}------------------------------------------------
590nm is 0.035 * 21 * 60=44.1J/cm², which is similar to the dose of predicate device (K200104) of 0.035 * 20 * 60=42J/cm². So, these differences will not raise any safety or effectiveness issues.
# 7. Test Summarv
### 7.1 Non-Clinical Tests Performed
## 1) Electrical safety, and electromagnetic compatibility Test
Non-clinical tests were performed on the subject device in order to validate the design and to assure conformance with the following voluntary design standards in connection with medical device electrical safety, and electromagnetic compatibility:
- IEC 60601-1 2020-08 Edition 3.2 Medical electrical equipment Part 1: General requirements for basic � safety and essential performance - Note: This standard is recognized with relevant US national differences applied, see references #1 and #2 in the Relevant FDA Guidance and/or Supportive Publication section below.
- � IEC 60601-1-11 Edition 2.1 2020-07 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance -- Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment.
- � IEC 60601-2-57 Edition 1.0 2011-01 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use.
- � IEC 60601-1-2 Edition 4.1 2020-09 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests.
- � IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems.
- � IEC 62133-2 Edition 1.0 2017-02 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems.
### 2) Biocompatibility Test
The subject device is biocompatible for its intended use. They are complied with biocompatibility standards ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Sensitization), ISO 10993-23 (Skin Irritation) and ISO 10993-11 (Acute Systemic Toxicity and Material-mediated Pyrogens) and USP 151.
### 3) Software verification and validation
Software verification and validation testing were conducted and documentation was provided as recommended by FDA'S Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level concern, since a malfunction of, or a latent design flaw in, the Software Device leads to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.
### 4) Usability validation
Usability testing was conducted on the Face Patches (Models: MT-12MC), which complies with IEC 62366-1 and IEC 60601-1-6.
# 5) Light output
The Light output test validated that the output of light from the devices are the same as stated in the specifications.
# 7.2 Summary of Clinical Performance
Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively.
### 8. Date of the summary prepared: April 9, 2024
# 9. Final Conclusion
{10}------------------------------------------------
The subject device is as safe, as effective, and performs as well as or better than the legally marketed predicated devices K230720, K221946 and K200104.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.