The LUSTRE 3XPRESS Light Beauty Therapy patches (Model: PR6001, PR5001) is an Over-the-Counter (OTC) device intended for treatment of mild to moderate inflammatory acne.
Device Story
LUSTRE 3XPRESS Light Beauty Therapy patches (models PR6001, PR5001) are OTC LED-based devices for treating mild to moderate inflammatory acne. The device consists of two patches containing a total of 24 LEDs (415 nm and 630 nm) per model, providing approximately 30 mw/cm² power intensity. The user applies the patches to the treatment area; an integrated controller via USB cable manages the operation. The device features an automatic 3-minute shut-off timer. Components include mini dots for fixation, protective goggles, and a travel pouch. The device is intended for home use by the patient. The light therapy aims to reduce acne inflammation. No clinical data was required; safety and effectiveness are supported by non-clinical bench testing, including electrical safety, electromagnetic compatibility, photobiological safety, and biocompatibility.
Clinical Evidence
No clinical data was provided. Safety and effectiveness were established through non-clinical bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), photobiological safety (IEC 62471), and biocompatibility (ISO 10993-5, -10, -11, -23).
Technological Characteristics
LED-based light therapy device emitting 415 nm and 630 nm wavelengths. Power intensity ~30 mw/cm². Powered via USB (100-240Vac input, 5V/2A output). Components include triangular (PR6001) or crescent (PR5001) patches. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57, IEC 62471, and ISO 10993 standards. Software is classified as moderate level of concern.
Indications for Use
Indicated for the treatment of mild to moderate inflammatory acne in the general population using an Over-the-Counter (OTC) device.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue rectangle with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 3, 2023
Light Tree Ventures Europe B.V. Kim Laurens Manager Laan van Ypenburg 108, 2497 GC, The Hague, The Netherlands Hague, Netherlands
Re: K230720
Trade/Device Name: LUSTRE 3XPRESS Light Beauty Therapy patches, model: PR6001, PR5001 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OLP Dated: June 6, 2023 Received: June 6, 2023
Dear Kim Laurens:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows the name "Jianting Wang -S" in a large, bold font. The text is black and is set against a white background. There is a faint, light blue watermark in the background.
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230720
Device Name
LUSTRE 3XPRESS Light Beauty Therapy patches (Model: PR6001, PR5001)
Indications for Use (Describe)
The LUSTRE 3XPRESS Light Beauty Therapy patches (Model: PR6001) is an Over-the-Counter (OTC) device intended for treatment of mild to moderate inflammatory acne.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary of K230720
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
#### 1. Date of the summary prepared: June 13, 2023
#### 2. Submitter's Information
Sponsor Name: Light Tree Ventures Europe B.V. Establishment Registration Number: 3017422691 Address: Laan van Ypenburg 108, 2497 GC, The Hague, The Netherlands Contact Person (including title): Alain Dijkstra (Manager) Tel: +86-135-10378748 Fax: +86-755-25024651 E-mail: regulation@kaiyanmedical.com
# Distributor
Distributor Name: Lustre Skin Ltd Address: Alba Innovation Centre, Alba Campus, Livinqston Scotland, UK, EH547GA
# Factory
Company Name: Shenzhen Kaiyan Medical Equipment Co., Ltd Address: Building#3 and Building#5, 40th of Fuxin Street, Huaide Community Fuyong Town, Baoan District, Shenzhen, Guangdong 518103, China
# Application Correspondent:
Contact Person: Alain Dijkstra Company: Light Tree Ventures Europe B.V. Address: Laan van Ypenburg 108, 2497 GC, The Hague, The Netherlands Tel: +86 755 82129361 Fax: +86 755 25024651 Email: regulation@kaiyanmedical.com
#### 3. Subject Device Information
Trade Name: LUSTRE 3XPRESS Light Beauty Therapy patches, model: PR6001, PR5001 Classification Name: Over-The-Counter Powered Light Based Laser For Acne Review Panel: General & Plastic Surgery Product Code: OLP Regulation Number: 21 CFR 878.4810 Regulation Class: II
# 4. Predicate Device Information
Sponsor: Shenzhen Kaiyan Medical Co Ltd Trade Name: DemarkQ WOW, DemarkQ POP Classification Name: Over-The-Counter Powered Light Based Laser For Acne 510(K) Number: K203214 Review Panel: General & Plastic Surgery Product Code: OLP
{4}------------------------------------------------
Regulation Number: 21 CFR 878.4810 Regulation Class: II
# 5. Device Description
The LUSTRE 3XPRESS Light Beauty Therapy patches, model: PR6001 and PR5001 is an over-thecounter light emitting diode (LED) device that emits energy to treat mild to moderate acne. The device uses two types of LEDs: 415 nm and 630nm.
The LUSTRE 3XPRESS Light Beauty Therapy patches components include the device containing two Light Beauty Therapy Patches, an USB Cable with an integrated controller, 120 Mini Dots, a Goggles, a Travel pouch and a User Manual. Among them, mini dots are used for fixed device, goggles are used to protect eyes, and travel pouch are used to store device.
There is a total of 24 LEDs for each model to provide a power intensity of about 30 mw/cm².
The user pastes the device on the treatment area, and the device will shut down automatically after a 3minute after finishing treatment.
The two models are only different in shape. PR6001 is triangular and PR5001 is crescent.
# 6. Intended Use / Indications for Use
The LUSTRE 3XPRESS Light Beauty Therapy patches (Model: PR6001, PR5001) is an Over-the-Counter (OTC) device intended for treatment of mild to moderate inflammatory acne.
# 7. Comparison to predicate device
Compare with the predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between the subject device and predicate devices do not raise new questions of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | | Predicate Device (Primary<br>predicated device) | | Remark |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|---------------------|
| Company | Light Tree Ventures Europe B.V. | | Shenzhen Kaiyan Medical Co Ltd | | -- |
| Trade Name | LUSTRE 3XPRESS Light Beauty<br>Therapy patches | | DemarkQ | | -- |
| Model | PR6001 | PR5001 | DemarkQ<br>WOW (PB<br>B) | DemarkQ<br>POP(SJ-72) | -- |
| Classification Name | Over-The-Counter Powered<br>Light Based Laser For Acne | | Over-The-Counter Powered<br>Light Based Laser For Acne | | Same |
| 510(k) Number | K230720 | | K203214 | | -- |
| Product Code | OLP | | OLP | | Same |
| Intended Use /<br>Indications for Use | The LUSTRE 3XPRESS Light<br>Beauty Therapy patches (Model:<br>PR6001, PR5001) is an Over-<br>the-Counter (OTC) device<br>intended for treatment of mild to<br>moderate inflammatory acne. | | The DemarkQ is an Over the-<br>Counter (OTC) device<br>intended for treatment of mild<br>to moderate inflammatory<br>acne. | | Same |
| Power Supply | Input: 100-240Va.c., 50Hz/60Hz, | output: DC 5V, 2A | Power by: 2 x<br>(3.7Vdc, 2700<br>mAh Lithium<br>battery, | Power by:<br>3.7Vdc,<br>1500 mAh<br>lithium | Different<br>Note 1 |
| | | 10.26Wh)<br>Adapter Input:<br>100- 240Vac,<br>50/60Hz<br>Adapter<br>Output: DC<br>5V, 2A<br>Main unit<br>input: DC 5V,<br>2A | battery,<br>10.26Wh<br>Adapter<br>Input: 100-<br>240Vac,<br>50/60Hz<br>Adapter<br>Output: DC<br>5V, 2A Main<br>unit input:<br>DC 5V, 2A | | |
| Wavelengths | 630 ± 10nm<br>415 ± 10nm | 630nm±10nm<br>415nm±10nm | | Same | |
| Irradiance source | LED | LED | | Same | |
| Number of LEDs | For each model:<br>Blue: 12<br>Red: 12<br>Total :24 | 112 each<br>panel | Total 72 | Different<br>Note 2 | |
| Irradiance (mw/cm²) | Red: 5<br>Blue: 25 | Red: 5<br>Blue: 25 | Red: 25<br>Blue: 25±5 | Same | |
| Treatment Time | 3 minutes per treatment | 3 minutes per treatment | | Same | |
| Location for USE | OTC | OTC | | Same | |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | | Same | |
| Safety | IEC 60601-1<br>IEC 60601-2-57<br>IEC 60601-1-11<br>IEC 62471 | IEC 60601-1<br>IEC 60601-2-57<br>IEC 60601-1-11<br>IEC 62471 | | Same | |
| Biocompatibility | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-23<br>ISO 10993-11 | ISO 10993-1,<br>ISO 10993-5,<br>ISO 10993-10 | | Similar<br>Note 3 | |
{5}------------------------------------------------
# Comparison in Detail(s):
Note 1: Although the "Power Supply" is a little different from the predicate devices, they all complied with the IEC 60601-1 and IEC 60601-1-2 safety standards' requirements. So, these slight differences will not raise any safety or effectiveness issues.
Note 2: The "Number of LEDs" is different from the predicate devices, it's because the area of them is different. However, both of them have same "Irradiance". So, these differences will not raise any safety or effectiveness issues.
Note 3: Although the "Biocompatibility" is a little different from the predicate devices, they all meet the
{6}------------------------------------------------
requirements of the ISO 10993-1 guidance requirement. So, these slight differences will not raise any safety or effectiveness issues.
# 8. Test Summary
# 8.1 Non-Clinical Tests Performed
Non-clinical tests were performed on the subject device in order to validate the design and to assure conformance with the following voluntary design standards in connection with medical device electrical safety, and electromagnetic compatibility:
- AAMI/ANSI ES60601-1 2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) � [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- � IEC 60601-1-11 Edition 2.1 2020-07 Medical Electrical Equipment --Part 1: General Requirements for Basic Safety and Essential Performance --Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment.
- � IEC 60601-2-57 Edition 1.0 2011-01 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use.
- � IEC 60601-1-2 Edition 4.1 2020-09 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests.
- � IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems.
- � IEC 62133-2 Edition 1.0 2017-02 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems.
- � IEC 60601-1-6 Edition 3.2 2020-07 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- � IEC 62366-1 Edition 1.1 2020-06 Medical devices - Part 1: Application of usability engineering to medical devices
# 1) Biocompatibility
The subject device is biocompatible for its intended use. They are complied with biocompatibility standards ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Sensitization), ISO 10993-23 (Skin Irritation) and ISO 10993-11 (Acute Systemic Toxicity and Material-mediated Pyrogens).
# 2) Software verification and validation testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA'S Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level concern, since a malfunction of, or a latent design flaw in, the Software Device leads to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.
# 3) Cybersecurity
The subject device does not have any external interfaces, according to FDA guidance "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices", there's no need for a cybersecurity evaluation.
#### Usability Testing 4)
{7}------------------------------------------------
Usability testing was conducted on the LUSTRE 3XPRESS Light Beauty Therapy patches (Models: PR6001, PR5001), which complies with IEC 62366-1 and IEC 60601-1-6.
# 8.2 Summary of Clinical Performance
Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively.
# 9. Final Conclusion:
The subject device is as safe, as effective, and performs as well as or better than the legally marketed predicated devices K203214.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.