K221946 · Light Tree Ventures Europe B.V. · OHS · Nov 22, 2022 · General, Plastic Surgery
Device Facts
Record ID
K221946
Device Name
LED Light Therapy Mask (model: MK66R-B)
Applicant
Light Tree Ventures Europe B.V.
Product Code
OHS · General, Plastic Surgery
Decision Date
Nov 22, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LED LIGHT THERAPY MASK (Model:MK66R-B) is an over-the-counter device that is intended for the use in the treatment of full-face wrinkles.
Device Story
Home-use wearable LED phototherapy mask; treats full-face wrinkles. Flexible silicone mask contains 66 red (630nm) and 66 near-infrared (830nm) LEDs. Powered by rechargeable lithium battery via external controller; non-detachable cable. User wears mask on face, secured by velcro straps. Device operates via timer-controlled 10-minute sessions; 5x weekly for 6 weeks. Produces uniform light output (30 mW/cm² total intensity). No clinical data; safety established via bench testing and biocompatibility comparison to predicate. Intended for patient self-use.
Clinical Evidence
No clinical data. Safety and effectiveness supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), home healthcare environment safety (IEC 60601-1-11), photobiological safety (IEC 62471), and usability testing (IEC 62366-1).
Technological Characteristics
Flexible silicone mask; 66 red (630±5nm) and 66 NIR (830nm) LEDs. Rechargeable lithium battery; 10-minute timer-controlled operation. Total intensity 30 mW/cm². Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57, IEC 62471, IEC 62366-1. Software level of concern: moderate.
Indications for Use
Indicated for the treatment of full-face wrinkles in the over-the-counter (OTC) population.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services-USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
November 22, 2022
Light Tree Ventures Europe B.V. % Alain Dijkstra Official Correspondent Shenzhen Kaiyan Medical Equipment Co., Ltd Building 3, No.40, Fuxin street, Huaide Community, Fuyong Town, Baoan District Shenzhen, Guangdong 518103 China
Re: K221946
Trade/Device Name: LED Light Therapy Mask (model: MK66R-B) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS Dated: October 20, 2022 Received: October 20, 2022
Dear Alain Dijkstra:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Yan Yan Fu -S Date: 2022.11.22 11:42:54 -05'00'
For Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221946
Device Name LED Light Therapy Mask (Model: MK66R-B)
Indications for Use (Describe)
The LED Light Therapy Mask (Model:MK66R-B) is an over-the-counter device that is intended for the use in the treatment of full-face wrinkles.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary of K221946
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
### 1. Submitter's Information:
Sponsor Name: Light Tree Ventures Europe B.V. Establishment Registration Number: 3017422691 Address: Laan van Ypenburg 108, 2497 GC, The Hague, The Netherlands Contact Person (including title): Alain Dijkstra (Manager) Tel: +86-135-10378748 Fax: +86-755-25024651 E-mail: regulation@kaiyanmedical.com
#### Manufacturer:
Manufacturer Name: Shenzhen Kaiyan Medical Equipment Co., Ltd Address: Building 3, No.40, Fuxin street, Huaide Community, Fuyong Town, Baoan District, Shenzhen, Guangdong, 518103, China Contact Person (including title): Alain Dijkstra (CEO) Tel: 0755-82129361 Fax: 0755-25024651 E-mail: regulation@kaiyanmedical.com
#### Distributor:
CompanyCurrentBody.com Ltd Add: Unit D6,Stanley Green Business Park,Commercial Avenue,Cheadle Hulme,Cheshire, SK8 6QH
#### Application Correspondent:
Contact Person: Alain Dijkstra Company: Shenzhen Kaiyan Medical Equipment Co., Ltd Address: Building 3, No.40, Fuxin street, Huaide Community, Fuyong Town, Baoan District, Shenzhen, Guangdong, 518103, China Tel: +86 755 82129361 Fax: +86 755 25024651 Email: regulation@kaiyanmedical.com
#### Date of the summary prepared: November 20, 2022
#### 2. Subject Device Information
Classification Name: Light based over the counter wrinkle reduction Trade Name: LED LIGHT THERAPY MASK Model Name: MK66R-B Trademark: CurrentBody Skin™ Review Panel: General & Plastic Surgery Product Code: OHS Regulation Number: 21 CRF 878.4810 Regulatory Class: II
{4}------------------------------------------------
## 3. Predicate Device Information
Sponsor: ISMART Marketing Svcs Ltd Trade Name: FaceLITE Classification Name: Light based over the counter wrinkle reduction 510(K) Number: K191629 Review Panel: General & Plastic Surgery Product Code: OHS Requlation Number: 21 CRF 878.4810 Regulation Class: II
## 4. Device Description
The LED LIGHT THERAPY MASK is a home wearable light-emitting diode phototherapy device whose purpose is to produce an even, cool, narrow band of light for the treatment of full-face wrinkles.
The system consists of a flexible silicone mask that contains LEDs and a controller. The mask is worn on the face and is held in place by velcro straps. The mask compromises of 2 surfaces. An inner surface that contacts the skin and an outer surface. Both surfaces are constructed of sillcone.
The controller contains a rechargeable Lithium battery, the power supply (adaptor) is used to charge the Lithium battery and is connected to a suitable mains outlet via a 2 or 3 pin input socket and wall plug. The LED LIGHT THERAPY MASK cannot be operated while charging. The controller switches the LEDs ON/OFF and controls power to the mask. The device does not have an adapter, please use adapters of the following specifications: input: 100-240Va.c., 50Hz/60Hz, output: DC 5V, 2A.
The device produces red and near infra-red (NIR) light in the visible spectrum (Red: 630+/-5nm, NIR: 830nm) in intended to improve the appearance of wrinkles. The controller switches the LEDs ON/OFF and controls power to the mask. And the cable for connecting with the controller is non-detachable.
## 5. Intended Use / Indications for Use
The LED LIGHT THERAPY MASK (Model:MK66R-B) is an over-the-counter device that is intended for the use in the treatment of full-face wrinkles.
## 6. Comparison to predicate device
Compare with predicate devices, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device and predicate device do not raise and new questions of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device | Remark |
|------------------------------|---------------------------------------------------|---------------------------------------------------|--------|
| Company | Light Tree Ventures Europe B.V. | ISMART Marketing Svcs Ltd | -- |
| Trade Name | LED LIGHT THERAPY MASK | FaceLITE | -- |
| Classification<br>Name | Light Based Over The Counter<br>Wrinkle Reduction | Light Based Over The Counter<br>Wrinkle Reduction | -- |
| 510(k) Number | K221946 | K191629 | -- |
| Product Code | OHS | OHS | Same |
| FDA Device<br>Classification | Class II | Class II | Same |
| Use | Over the Counter | Over the counter | Same |
{5}------------------------------------------------
| Elements of<br>Comparison | Subject Device | Predicate Device | Remark |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|--------------------|
| Intended Use /<br>Indications for<br>Use | The LED LIGHT THERAPY MASK<br>(MK66R-B) is an over-the-counter<br>device that is intended for the use in<br>the treatment of full-face wrinkles. | The faceLITE LED mask is an<br>over-the-counter device that is<br>intended for the use in the<br>treatment of full-face wrinkles. | Same |
| Intended location<br>of use | Face | Face | Same |
| Energy Type | Light emitting diodes | Light emitting diodes | Same |
| Wavelengths | Red: 630±5 nm<br>NIR: 830nm | Red: 630nm±10nm<br>NIR: 830nm±10nm | Same |
| Total Intensity<br>(mW/cm²) | 30mw/cm² | 30mw/cm² total | Same |
| LED number | 630±5nm: 66<br>830nm: 66 | 630±5nm: 49<br>830nm: 49 | Different,<br>Note |
| LED distribution | Uniform distribution | Uniform distribution | Same |
| Power flux<br>(mW/cm²) | 630±5nm:15-20 mW/cm²<br>830nm:10-15 mW/cm² | 630±5nm: 15-20 mW/cm²<br>830nm: 6-12 mW/cm² | Similar<br>Note |
| Treatment Time | 10 minutes | 600 seconds | Same |
| Dose | 540 J/cm2 (cumulative does) | 540 J/cm2 (cumulative does) | Same |
| Treatment<br>protocol | 5 x weekly, 6 weeks | 5 x weekly, 6 weeks | Same |
| Software<br>controller | Device uses a timer and software to<br>control treatment duration | Yes | Same |
| Power supply | Rechargeable Lithium battery | Rechargeable Lithium battery | Same |
# Comparison in Detail(s):
Note: Although the "LED number" and "Power flux" is a little different from the predicate devices, both of them have the same out wavelengths, total Intensity, treatment time and dose, and they all complied with the IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57 and IEC 62471 safety standards' requirements. So, these slight differences will not raise any safety or effectiveness issues.
# 7. Test Summary
# 7.1 Summary of Non-Clinical Performance Testing
LED LIGHT THERAPY MASK (Model: MK66R-B) has been evaluated the safety and performance by lab bench testing as following:
| Title of the test | Device<br>Description/Sample Size | Test<br>Standards | Method/Applicable<br>Acceptance criteria | Unexpected<br>Results/Significant<br>Deviations | Test<br>results |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|-----------------|
| General<br>requirements<br>for basic<br>safety and | The test<br>sample is the<br>final, finished<br>product of | IEC 60601-<br>1:2005+A1:2012+A2:2020 Edition: 3.2 2020-08 | The test is carried out<br>under the test method<br>specified in the<br>standard, and the test | NA | Pass |
| essential<br>performance | representativ<br>e model:<br>MK66R-B | | result is within the test<br>acceptance range of<br>the standard. | | |
| Location:<br>VOL_010_A8.<br>Electrical<br>Safety and<br>EMC_<br>001_A8-1. IEC<br>60601-1, IEC<br>60601-1-11<br>Test Report<br>Electromagneti<br>c disturbances | The test<br>sample is the<br>final, finished<br>product of<br>representativ<br>e model:<br>MK66R-B | IEC 60601-1-<br>2:2014+A1:2020 Edition:<br>4.0 2020-09 | No degradation of<br>performance was<br>found during test or<br>Lower than limits of<br>measurement. | NA | Pass |
| Location:<br>VOL_010_A8.<br>Electrical<br>Safety and<br>EMC_<br>004_A8-4. IEC<br>60601-1-2 Test<br>Report | The test<br>sample is the<br>final, finished<br>product of<br>representativ<br>e model:<br>MK66R-B | IEC 60601-1-<br>2:2014+A1:2020 Edition:<br>4.0 2020-09 | No degradation of<br>performance was<br>found during test or<br>Lower than limits of<br>measurement. | NA | Pass |
| Basic safety<br>and<br>performance<br>requirements<br>for home<br>healthcare<br>environment<br>Location:<br>VOL_010_A8.<br>Electrical<br>Safety and<br>EMC_<br>001_A8-1. IEC<br>60601-1, IEC<br>60601-1-11<br>Test Report | The test<br>sample is the<br>final, finished<br>product of<br>representativ<br>e model:<br>MK66R-B | IEC 60601-1-<br>11:2015+A1:2020 Edition:<br>2.0 2020-07 | The device operates<br>normally, and can<br>provide basic safety<br>and essential<br>performance. | NA | Pass |
| Particular<br>requirements<br>for the basic<br>safety and<br>essential<br>performance<br>of non-laser | The test<br>sample is the<br>final, finished<br>product of<br>representativ<br>e model:<br>MK66R-B | IEC 60601-2-57 Edition<br>1.0 2011-01 | The test is carried out<br>under the test method<br>specified in the<br>standard, and the test<br>result is within the test<br>acceptance range of<br>the standard. | NA | Pass |
| light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use<br>Location:<br>VOL_010_A8.<br>Electrical Safety and EMC_<br>002_A8-2. IEC 60601-2-57<br>Test Report | | | | | |
| Photobiological safety of lamps and lamp systems<br>Location:<br>VOL_010_A8.<br>Electrical Safety and EMC_<br>003_A8-3. IEC 62471 Test Report | The test sample is the final, finished product of representativ<br>e model:<br>MK66R-B | IEC 62471 First edition 2006-07 | The test is carried out under the test method specified in the standard, and the test result is within the test acceptance range of the standard. | NA | Pass |
| Usability Study<br>Location:<br>VOL_009_A7.<br>Usability | The test sample is the final, finished product of representativ<br>e model:<br>MK66R-B | IEC 62366-1 Edition 1.0 2015-02 and IEC 60601-1-6 Edition 3.1 2013-10 | The subject device can meet the usability goal of IEC 62366-1 and IEC 60601-1-6 standards. | NA | Pass |
| Shelf Life Test | The test sample is the final, finished product of representativ<br>e model:<br>MK66R-B | The Shelf Life Test Report performs the following tests on the product before and after accelerated aging, and after use:<br>Performance test;<br>Key life test;<br>Mask bending test;<br>Plug and pull test; | Performance Test:<br>● Appearance:<br>The device should be no deformation, the surface is smooth and free of burrs, no color sports.<br>● The silicone | NA | Pass |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
| | Plug and pull test;<br>Power Density Test;<br>Leakage current test; | thickness is 9.42 $\pm$ 0.1 mm on<br>point A, 5 $\pm$ 0.1<br>mm on point B. |
|--|---------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Key life test: |
| | | There should be no key stuck. The key is sensitive. The device can be turned on and off normally. The LED indicator can work normally when pressing the key. |
| | | Mask bending test: |
| | | The device should be no cracks, no damage, can be charged and work normally. |
| | | Plug and pull test: |
| | | The device can be charged and work normally: The indicator can light up normally when plug the adapter into the power outlet. |
| | | Power Density Test: |
| | | Power Density: 30 mw/cm2 $\pm$ 10% (27 - 33 mw/cm2); |
| | | Leakage current test: |
| | | Should meet the clause 8.7.4 of IEC 60601-1 required. |
# 1) Biocompatibility statement
The component materials for Velcro Straps of the LED LIGHT THERAPY MASK (Models: MK66R-B) is identical to the corresponding component for Velcro Straps of the Aduro Light Beauty Mask (MK-660, MK-66USBO, MK-66USBA, MK-66USBB, MK-02O, MK-02B) in formulation, processing,
{9}------------------------------------------------
sterilization, and geometry, and no other chemicals have been added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents), that manufactured by SHENZHEN KAIYAN MEDICAL EQUIPMENT CO., LTD and has been clarified in K202390 on July 14, 2021.
The patient contacting material silicone of the outer surface and inner surface of model MK66R-B is identical to the materials for Aduro Light Beauty Mask (Model: MK-660, MK-66USBA, MK-66USBB, MK-020, MK-02B) in formulation, processing, sterilization, and geometry, and no other chemicals have been added (e.g., plasticizers, fillers, cleaning agents, mold release agents), that manufactured by SHENZHEN KAIYAN MEDICAL EQUIPMENT CO., LTD and has been clarified in K202390 on July 14, 2021.
# 2) Usability Testing
Usability testing was conducted on the LED LIGHT THERAPY MASK (Model: MK66R-B), which complies with IEC 62366-1 and IEC 60601-1-6.
# 3) Software verification and validation testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA'S Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level concern, since a malfunction of, or a latent design flaw in, the Software Device leads to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.
# 7.2 Summary of Clinical Performance
Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively.
# 8. Final Conclusion:
The subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate devices K191629.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.