K233864 · Kestra Medical Technologies, Inc. · MWJ · May 7, 2024 · Cardiovascular
Device Facts
Record ID
K233864
Device Name
ASSURE Wearable ECG
Applicant
Kestra Medical Technologies, Inc.
Product Code
MWJ · Cardiovascular
Decision Date
May 7, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2800
Device Class
Class 2
Attributes
AI/ML
Indications for Use
The ASSURE Wearable ECG continuously monitors heart rate information. Event data, which is automatically stored for low and high heart rates and patient-triggered events, is intended to aid medical professionals as they monitor various clinical conditions, events, and trends. The ASSURE Wearable ECG is intended for use by a patient during their normal daily activities primarily in the home or community setting, but also hospitals, medical clinics, healthcare facilities and transport.
Device Story
Reusable ambulatory ECG system; monitors heart rate and activity; utilizes 5-electrode SensorFit garment (previously used with ASSURE WCD). Inputs: ECG signals via electrodes; 3-axis accelerometer data for activity/wear time. Operation: continuous recording; embedded algorithm identifies high/low heart rate episodes; patient-triggered events via Alert Button. Output: stored ECG waveforms and activity reports transmitted to Kestra server. Used in home, community, or clinical settings by patients. Healthcare providers access reports via server to assist in clinical management and diagnosis. Benefits: provides cardiac monitoring for patients transitioning from WCD therapy; non-invasive; no therapy delivery or life support.
Clinical Evidence
Bench testing only. No clinical data presented. Testing verified continuous ECG monitoring, event data storage, and data transmission to Kestra server. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-47, ISO 10993-1, UL 2054, IEC 62133-2, and AIM 7351731 confirmed.
Technological Characteristics
Reusable ambulatory ECG system; 5-electrode SensorFit garment; 3-axis accelerometer; rechargeable Li-Ion battery. Connectivity: Bluetooth for data transmission to mobile client/server. Software: embedded arrhythmia detection algorithm (Tachy/Brady capture). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-47, ISO 10993-1, UL 2054, IEC 62133-2, AIM 7351731.
Indications for Use
Indicated for adult patients prescribed this device by a medical professional who were previously prescribed the ASSURE WCD system, and who may be asymptomatic or suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, presyncope, syncope, fatigue, or anxiety. Contraindicated for patients with an active implantable pacemaker or defibrillator.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size, also in blue.
May 7, 2024
Kestra Medical Technologies, Inc. Jay Wiese Sr. Regulatory Affairs Specialist 3933 Lake Washington Boulevard NE Suite 200 Kirkland, Washington 98033
Re: K233864
Trade/Device Name: ASSURE Wearable ECG Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: MWJ, DXH Dated: April 9, 2024 Received: April 9, 2024
Dear Jay Wiese:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jennifer W. Shih -S
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K233864
Device Name
ASSURE Wearable ECG (80553-001)
Indications for Use (Describe)
The ASSURE Wearable ECG is indicated for adult patients who have been prescribed this device by a medical professional, who were previously prescribed the ASSURE WCD system, and who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, presyncope, fatigue, or anxiety. The signal acquired by the ASSURE Wearable ECG is not intended and should not be used for automated or semiautomated analysis. The device does not deliver any therapy, administer any drugs, provide interpretive or diagnostic statements, or provide for any life support.
The ASSURE Wearable ECG is contraindicated for use in patients with an active implantable pacemaker or defibrillator.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
ver-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Kestra. The logo consists of two parts: a graphic element on the left and the word "kestra" on the right. The graphic element is made up of three curved lines that intersect each other, with the top line being gray, the middle line being yellow, and the bottom line being gray. The word "kestra" is written in a sans-serif font and is gray in color. There is a registered trademark symbol to the upper right of the word "kestra".
#### 510(k) Summary (K233864)
Date Prepared: April 9, 2024
#### I. General Information
#### Applicant:
Kestra Medical Technologies, Inc. 3933 Lake Washington Blvd NE Ste 200 Kirkland, WA 98033 USA Phone: 425-526-4927 Fax: 415-632-5701
#### Contact Person:
Bev Magrane VP, Quality and Requlatory Affairs Email: beverly.magrane@kestramedical.com
#### II. Device Information
#### Trade Name:
| ASSURE Wearable ECG | |
|----------------------------|------------------------------------------------------------------------------------------------------------------------|
| Generic/Common Name: | |
| Classification Regulation: | 21 CFR 870.2800 |
| Classification Name: | Medical Magnetic Tape Recorder |
| Regulatory Class: | Class II |
| Product Codes: | MWJ Electrocardiograph, Ambulatory (Without Analysis)<br>DXH Transmitters and Receivers, Electrocardiograph, Telephone |
#### III. Predicate Devices
The following predicate devices have been selected:
| Primary Predicate Device: | Biotricity Biotres system |
|-----------------------------|---------------------------------------------------------------|
| Predicate Clearance | K211709 (19 Jan 2022) |
| Classification Regulation: | 21 CFR 870.2800 |
| Classification Name: | Medical Magnetic Tape Recorder |
| Regulatory Class: | Class II |
| Product Code: | MWJ Electrocardiograph, Ambulatory (Without Analysis) |
| Secondary Predicate Device: | Biotricity Bioflux system |
| Predicate Clearance | K172311 (15 Dec 2017) |
| Classification Regulation: | 21 CFR 870.2920 |
| Classification Name: | Telephone Electrocardiograph Transmitter and Receiver |
| Regulatory Class: | Class II |
| Product Code: | DXH Transmitters and Receivers, Electrocardiograph, Telephone |
#### IV. Indications for Use
The ASSURE Wearable ECG is indicated for adult patients who have been prescribed this device by a medical professional, who were previously prescribed the ASSURE WCD system, and who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, lightheadedness, presyncope, syncope, fatigue, or anxiety. The signal acquired by the ASSURE Wearable ECG is
kestramedical.com
Kestra Medical Technologies, Inc. 3933 Lake Washington Blvd NE Suite 200 Kirkland, WA 98033
+1 (425) 279-8002 main
+1 (800) 957-0028 toll-free
+1 (425) 279-8040 fax
{4}------------------------------------------------
not intended and should not be used for automated or semi-automated analysis. The device does not deliver any therapy, administer any drugs, provide interpretive or diagnostic statements or provide for any life support.
The ASSURE Wearable ECG is contraindicated for use in patients with an active implantable pacemaker or defibrillator.
#### V. Intended Use
The ASSURE Wearable ECG continuously monitors heart rate information. Event data, which is automatically stored for low and high heart rates and patient-triggered events, is intended to aid medical professionals as they monitor various clinical conditions, events, and trends. The ASSURE Wearable ECG is intended for use by a patient during their normal daily activities primarily in the home or community setting, but also hospitals, medical clinics, healthcare facilities and transport.
#### VI. Device Description
The ASSURE Wearable ECG is a reusable, ambulatory electrocardiography-based, cardiac- and physiologicmonitoring, medical-electrical system whose intended purpose is to inform clinical management of options for diagnosing, monitoring and/or mitigating cardiac conditions after patient's improvement following ASSURE® Wearable Cardioverter Defibrillator (WCD) prescriptive use. The system utilizes the same five-electrode SensorFit™ Garment worn previously with the WCD prescription. The system continuously records ECG data and upon detection, it identifies and records episodes as high and low heart rate, as well as patient-triggered events. The system utilizes the same algorithm detection and episode reporting software marketed in the ASSURE WCD with high (Tachy) and low (Brady) capture for later transmission to the medical professional for interpretation. The system captures and stores ECG episodes, and non-ECG patient activity and wear information to be displayed and reported in counters and trends. Recorded events include ECG waveforms and reports identifying high and low heart rates, as well as patient-triggered events. The system uses a 3axis accelerometer to monitor non-ECG patient activity (steps and wear time).
The ASSURE Wearable ECG event reports do not contain diagnostic interpretation. The reported events are provided for review by the prescriber to assist in diagnosis of the recently transitioned WCD patient and to assess care options based on the healthcare professional's judgment and experience.
The ASSURE Wearable Cardiac ECG is a prescription use device. The ASSURE Wearable ECG is intended for use by a patient during their normal daily activities primarily in the home or community setting, but also hospitals, medical clinics, healthcare facilities and transport. The device is non-invasive, and intended to be used on one patient at a time.
The Wearable Cardiac ECG System is comprised of the following reusable patient-worn components:
- Monitor Cable Assembly Houses the primary electronics, embedded application software, . communication modules, and ECG signal acquisition hardware. It includes a Hub, Alert Button and Battery Connector.
- Hub Physical housing for the primary Wearable ECG electronics and 3-axis accelerometer. The Hub ● provides a means for connecting the Monitor Cable to the Garment.
- Alert Button Provides audio, tones and voice prompts as well as vibratory alerts and a button for the ● user to provide input.
- Battery Pack - Rechargeable Lithium-lon battery.
- SensorFit™ Garment This is the same PMA-approved (P200037) garment previously worn with the . WCD and fitted according to body size and body style. The Garment includes built-in ECG electrodes and provides a method of affixing the Hub, Alert Button, and Battery Pack for proper placement during operation.
- Charger Medical-grade AC-DC power adaptor and USB-C power cable. ●
{5}------------------------------------------------
#### VII. Comparison of Technological Characteristics with Predicate Device
The subject ASSURE Wearable ECG has the same intended use as the predicate devices. The differences in the technological characteristics between the subject and predicate devices do not raise any issues of safety or effectiveness as the fundamental scientific technology and intended use is unchanged.
The ASSURE Wearable ECG is considered substantially equivalent to the predicate devices. A comparison table outlining the similarities and differences between the subject device and the predicate devices is provided in the Summary Table below.
| Category | Wearable ECG (subject) | Biotres (predicate) K211709 | Bioflux (secondary) K172311 |
|--------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The ASSURE Wearable ECG is<br>indicated for adult patients who<br>have been prescribed this<br>device by a medical<br>professional, who were<br>previously prescribed the<br>ASSURE WCD system, and<br>who may be asymptomatic or<br>who may suffer from transient<br>symptoms such as palpitations,<br>shortness of breath, dizziness,<br>light-headedness, presyncope,<br>syncope, fatigue, or anxiety. The<br>signal acquired by the Wearable<br>ECG is not intended and should<br>not be used for automated or<br>semi-automated analysis. The<br>device does not deliver any<br>therapy, administer any drugs,<br>provide interpretive or diagnostic<br>statements or provide for any life<br>support. | The Biotres is indicated for<br>use on adult patients 18 years<br>or older who may be<br>asymptomatic or who suffer<br>from transient symptoms such<br>as palpitations, shortness of<br>breath, dizziness, light<br>headedness, pre-syncope,<br>syncope, fatigue, chest pain<br>and/or anxiety and may<br>require cardiac recording on a<br>continuous basis for up to 30<br>days. The signal acquired by<br>the Biotres is not intended<br>and should not be used for<br>automated or semi-automated<br>analysis. | The bioflux Device is intended for<br>use by patients who either have or<br>are at risk of having cardiac disease<br>and those that demonstrate<br>intermittent symptoms indicative of<br>cardiac disease and require cardiac<br>monitoring on a continuing basis for<br>up to 30 days. The device<br>continuously records ECG data and<br>upon detection by an ECG analysis<br>algorithm or manually initiated by the<br>patient, automatically delivers the<br>recorded cardiac activity to the<br>server where it is presented and can<br>be reviewed by a medical<br>professional.<br>The data received from the bioflux<br>device can be used by another<br>device for arrhythmia analysis,<br>reporting and signal measurements.<br>The Bioflux device is not intended to<br>sound any alarms. The device does<br>not deliver any therapy, administer<br>any drugs, provide interpretive or<br>diagnostic statements or provide for<br>any life support.<br>bioflux is for prescription use only. |
| Classification<br>Name | Recorder, Magnetic Tape,<br>Medical | Recorder, Magnetic Tape,<br>Medical | Telephone electrocardiograph<br>transmitter and receiver |
| Product Code | MWJ, DXH | MWJ | DXH |
| Category | Wearable ECG (subject) | Biotres (predicate) K211709 | Bioflux (secondary) K172311 |
| Defibrillator<br>Protection | The ECG Electrodes in the<br>WCD-approved Garment are<br>classified as defibrillation-proof<br>applied parts. | Not Defibrillator Proof, 8.5.5.2<br>of IEC 60601-1 passed | |
| Wear Time | Up to 30 Days | Up to 30 Days | Up to 30 Days |
| Recording Format | Continuously monitors ECG and<br>activity data. Data is auto-<br>triggered for events such as<br>Bradycardia and Tachycardia, as<br>identified by an embedded<br>arrhythmia detection algorithm;<br>and stored as low and high heart<br>rate episodes and patient-<br>triggered events | Continuous | Captures patient-activated and auto-<br>triggered events such as<br>Bradycardia, Tachycardia, pause and<br>Atrial Fibrillation as identified by an<br>embedded arrhythmia detection<br>algorithm |
| Delivered device<br>includes | ECG monitor with 5-electrode<br>WCD-supplied garment<br>Internal rechargeable battery<br>Wall Battery charger | -3 lead ECG monitor<br>-internal rechargeable battery<br>-Wall Battery charger | Device has at least 2 ECG channels<br>and 3-lead electrodes |
| Monitor functional<br>blocks | Analog ECG front end,<br>accelerometer, MCU, eMMC<br>data storage, BLE modem for<br>data transmission, LED<br>indicator, and event record<br>button | Analog ECG front end, MCU,<br>Flash data storage, BLE<br>modem for data transmission,<br>LED indicator, and Record<br>button | Analog ECG front end, MCU, Flash<br>data storage, RF modem for data<br>transmission, LCD screen, and<br>Record button |
| Data transfer | Bluetooth for transfer of patient<br>data by mobile data client to a<br>Wi-Fi-connected server | Mobile App, iOS, Android | Data can be delivered to the server<br>wirelessly via mobile network |
| Server | Subsequent access to ECG<br>display and episode report<br>viewer for evaluation by<br>prescribing medical professional | Facilitate data communication<br>with the Biotres device,<br>provide data storage, and<br>present the data for<br>evaluation by a medical<br>professional | Facilitate data communication with<br>the Bioflux device, provide data<br>storage, and present the data for<br>evaluation by a medical professional |
| Device form factor | Utilizes Kestra PMA-approved<br>WCD garment electrodes | Small, lightweight body worn<br>ambulatory cardiac monitors | Small, lightweight body worn<br>ambulatory cardiac monitors |
| Category | Wearable ECG (subject) | Biotres (predicate) K211709 | Bioflux (secondary) K172311 |
| Wireless<br>technology used<br>to transmit data to<br>server | Yes | Yes | Data is delivered to the server<br>wirelessly via mobile cellular<br>dedicated network connection |
| Power | Device is battery powered by a<br>rechargeable Li-Ion battery | Device is battery powered by<br>a rechargeable Li-Ion battery | Device is battery powered by a<br>rechargeable Li-Ion battery |
| Device Program<br>Parameters | ASSURE Wearable ECG has no<br>user-programmable functions or<br>features | Using an Android or IOS<br>based app, clinician can<br>adjust device programming<br>parameters such as pre-post<br>recording times and auto-<br>triggering configuration | A medical professional, using the<br>server, can adjust and program the<br>device configuration and auto-<br>triggering parameters |
| Monitor Multiple<br>Parameters | ECG signal, high and low heart<br>rate and accelerometer-based<br>patient data (steps and wear<br>time) | No (ECG only) | Captures patient activated and auto-<br>triggered events such as<br>Bradycardia, Tachycardia, pause and<br>Atrial Fibrillation as identified by an<br>embedded arrhythmia detection<br>algorithm |
| Manual Event<br>Recording | Alert button for patient-triggered<br>event recordings and visual,<br>audio and haptic modes of<br>alerting patient to device status | Devices have Record button<br>for manual event recordings<br>and a user LED to indicate<br>device status and mode of<br>operation. | Devices have Record button for<br>manual event recordings and a user<br>LED to indicate device status and<br>mode of operation. |
| Prescriber Review | The ECG recorded data from<br>the ASSURE Wearable ECG is<br>transferred to the clinician from<br>the Kestra Secure Server when<br>the device is connected to the<br>server at the end of a patient<br>study | The ECG recorded data from<br>the Biotres Recorder can be<br>transferred, by the clinician,<br>from the Biotres Configured<br>Secure Server when the<br>recorder is connected to the<br>Biotres Gateway App at the<br>end of a patient study. | Automatically delivers the recorded<br>cardiac activity to the server where it<br>is presented and can be reviewed by<br>a medical professional |
{6}------------------------------------------------
{7}------------------------------------------------
## VIII. Performance Data
Safety and performance of the subject ASSURE Wearable ECG system has been evaluated and verified in accordance with design specifications and to support a determination of substantial equivalence to the predicate device. Because the device is designed and intended to monitor low-risk patients using proprietary algorithm detection software developed for the ASSURE Wearable Cardioverter Defibrillator, it intentionally does not meet all performance clauses of IEC 60601-2-47:2015 that are associated with diagnostic ECG.
The following performance and safety tests have passed successfully:
{8}------------------------------------------------
| Recognition<br>Number | Number | Document Description (Title) |
|-----------------------|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 19-49 | IEC 60601-<br>1:2005+A1:2012+A2:2020 | Medical electrical equipment - Part 1: General requirements<br>for basic safety and essential performance (Edition 3.2) |
| 19-36 | IEC 60601-1-2:2014+A1:2020 | Medical electrical equipment - Part 1-2: General requirements<br>for basic safety and essential performance - Collateral<br>standard: Electromagnetic compatibility - Requirements and<br>tests |
| 19-38 | IEC 60601-1-11:2015+A1:2021 | Medical electrical equipment - Part 1-11: General<br>requirements for basic safety and essential performance -<br>Collateral Standard: Requirements for medical electrical<br>equipment and medical electrical systems used in the home<br>healthcare environment |
| 3-155 | IEC 60601-2-47:2015 | Medical electrical equipment - Part 2-47: Particular<br>requirements for the basic safety and essential performance<br>of ambulatory electrocardiographic systems |
| 2-258 | ISO 10993-1:2018 | Biological evaluation of medical devices Part 1: Evaluation<br>and testing within a risk management process |
| 19-11 | UL 2054:2004(R2011) | Standard for Household and Commercial Batteries, 2nd<br>Edition |
| 19-33 | IEC 62133-2:2017 | Secondary cells and batteries containing alkaline or other<br>non-acid electrolytes - Safety requirements for portable<br>sealed secondary cells, and for batteries made from them, for<br>use in portable applications, Part 2: Lithium systems (IEC<br>62133-2:2017) |
| 19-30 | AIM 7351731:2017 | Medical Electrical Equipment and System Electromagnetic<br>Immunity Test for Exposure to Radio Frequency Identification<br>Readers - An AIM Standard (Rev. 2.00) |
Bench test results verify that the ASSURE Wearable ECG system can continuously monitor ECG signal, store ECG event data in the device memory, and transmit recorded data to a Kestra display server for clinician review.
# IX. Conclusion
Based on a review of the test results and a comparison to the predicate devices characteristics and known specifications, the results show that the ASSURE Wearable ECG is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.