K172311 · Biotricity, Inc. · DXH · Dec 15, 2017 · Cardiovascular
Device Facts
Record ID
K172311
Device Name
BioFlux Device
Applicant
Biotricity, Inc.
Product Code
DXH · Cardiovascular
Decision Date
Dec 15, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2920
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The bioflux Device is intended for use by patients who either have or are at risk of having cardiac disease and those that demonstrate intermittent symptoms indicative of cardiac disease and require cardiac monitoring on a continuing basis for up to 30 days. The device continuously records ECG data and upon detection by an ECG analysis algorithm or manually initiated by the patient, automatically delivers the recorded cardiac activity to the server where it is presented and can be reviewed by a medical professional. The data received from the bioflux device can be used by another device for arrhythmia analysis, reporting and signal measurements. The bioflux device is not intended to sound any alarms. The device does not deliver any therapy, administer any drugs, provide interpretive or diagnostic statements or provide for any life support. bioflux is for prescription use only.
Device Story
BioFlux is a portable, battery-powered, wireless cardiac monitor for up to 30-day continuous ECG recording. Input: ECG signals via 3-lead cable. Operation: Embedded arrhythmia detection algorithm identifies events (Bradycardia, pause, Atrial Fibrillation); device automatically transmits data via cellular network to a server. Medical professionals access the server to review data and configure device parameters (e.g., auto-triggering). Patients can manually initiate recordings. Output: ECG data presented on server for professional review. Benefit: Enables remote cardiac monitoring for patients with intermittent symptoms. Does not provide diagnostic statements, therapy, or alarms.
Portable ambulatory monitor; 3-lead ECG; internal rechargeable Li-Ion battery; cellular RF modem; LCD screen; record button. Materials biocompatibility per ISO 10993-1. Software includes embedded arrhythmia detection algorithm. Connectivity: wireless cellular to server. Standards: IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-47, ANSI/AAMI-EC 57.
Indications for Use
Indicated for patients with or at risk of cardiac disease, or those with intermittent symptoms requiring cardiac monitoring for up to 30 days. Prescription use only.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health and Human Services logo. On the right is the FDA logo, which is a blue square with the letters FDA in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 15, 2017
Biotricity Inc. % Mark Job Official Correspondent Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, Minnesota 55313
Re: K172311
Trade/Device Name: BioFlux Device Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter and Receiver Regulatory Class: Class II Product Code: DXH Dated: November 14, 2017 Received: November 21, 2017
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MS. Hillebrand
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Food and Drug Administration
## Indications for Use
Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K172311
Device Name BioFlux Device
Indications for Use (Describe)
The bioflux Device is intended for use by patients who either have or are at risk of having cardiac disease and those that demonstrate intermittent symptoms indicative of cardiac disease and require cardiac monitoring on a continuing basis for up to 30 days. The device continuously records ECG data and upon detection by an ECG analysis algorithm or manually initiated by the patient, automatically delivers the recorded cardiac activity to the server where it is presented and can be reviewed by a medical professional.
The data received from the bioflux device can be used by another device for arrhythmia analysis, reporting and signal measurements. The bioflux device is not intended to sound any alarms. The device does not deliver any therapy, administer any drugs, provide interpretive or diagnostic statements or provide for any life support.
bioflux is for prescription use only.
| Type of Use (Select one or both, as applicable) | <table><tr><td><div> <span> <svg height="16" width="16"> <rect height="16" style="fill:none;stroke:#000000;stroke-width:1" width="16"></rect> <path d="M2,2 L14,14 M14,2 L2,14" style="stroke:#000000;stroke-width:1"></path> </svg> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div></td><td><div> <span> <svg height="16" width="16"> <rect height="16" style="fill:none;stroke:#000000;stroke-width:1" width="16"></rect> </svg> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div></td></tr></table> | <div> <span> <svg height="16" width="16"> <rect height="16" style="fill:none;stroke:#000000;stroke-width:1" width="16"></rect> <path d="M2,2 L14,14 M14,2 L2,14" style="stroke:#000000;stroke-width:1"></path> </svg> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> <svg height="16" width="16"> <rect height="16" style="fill:none;stroke:#000000;stroke-width:1" width="16"></rect> </svg> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <rect height="16" style="fill:none;stroke:#000000;stroke-width:1" width="16"></rect> <path d="M2,2 L14,14 M14,2 L2,14" style="stroke:#000000;stroke-width:1"></path> </svg> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> <svg height="16" width="16"> <rect height="16" style="fill:none;stroke:#000000;stroke-width:1" width="16"></rect> </svg> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
(as required by 21CFR 807.92)
#### SUBMITTER l.
Biotricity - Measuring Vitals 275 Shoreline Drive, Suite 150 Redwood City CA 94065 Contact Person: Spencer LaDow sladow@biotricity.com Phone: 585-414-7407 Date Prepared: November 14, 2017
#### DEVICE II.
Name of Device: BioFlux Device Classification Name: 870.2920 Telephone electrocardiograph transmitter and receiver. Common or Usual Name: Mobile Cardiac Monitor Device Panel: Cardiovascular Regulatory Class: Class 2 Product Code: DXH
#### III. PREDICATE DEVICE
The bioflux System is substantially equivalent in intended use and similar technological characteristics the following device, Rhythmedix, LLC. RhythmStar System which was cleared under K141813.
{4}------------------------------------------------
#### DEVICE DESCRIPTION IV.
The bioflux system consists of the bioflux device and the server. The bioflux device is a portable, battery-powered, wireless cardiac monitor which may be worn by a patient to record ECG and activity level data for up to 30 consecutive days. The device can capture patient activated and auto-triggered events such as Bradycardia, pause and Atrial Fibrillation as identified by an embedded arrhythmia detection algorithm. The device is designed to automatically deliver the data to the server. The data is delivered to the server wirelessly via mobile cellular dedicated network connection. A medical professional, using the server, can adjust and program the device configuration and autotriggering parameters.
bioflux device is not a life-supporting or life-sustaining system. Clinical judgment and experience are used to check and interpret the data.
#### V. INDICATIONS FOR USE
The bioflux Device is intended for use by patients who either have or are at risk of having cardiac disease and those that demonstrate intermittent symptoms indicative of cardiac disease and require cardiac monitoring on a continuing basis for up to 30 days. The device continuously records ECG data and upon detection by an ECG analysis algorithm or manually initiated by the patient, automatically delivers the recorded cardiac activity to the server where it is presented and can be reviewed by a medical professional.
The data received from the bioflux device can be used by another device for arrhythmia analysis, reporting and signal measurements. The Bioflux device is not intended to sound any alarms. The device does not deliver any therapy, administer
any drugs, provide interpretive or diagnostic statements or provide for any life support. bioflux is for prescription use only.
{5}------------------------------------------------
## SUMMARY OF TECHNOLIGICAL CHARACTERISTICS WITH THE PREDICATE VI. DEVICE
The bioflux system is substantially equivalent in intended use and similar technological characteristics of RhythmStar System cleared under K141813.
| Category | Identical/<br>Different | bioflux | RhythmStar |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | | K172311 | K141813 |
| Classification | Identical | Medical | Medical |
| Name | Identical | Mobile Cardiac Monitor | Mobile Cardiac Monitor |
| Product Code | Similar | DXH | DXH |
| Intended Use | Similar | The <b>bioflux</b> Device is intended for<br>use by patients who either have or<br>are at risk of having cardiac disease<br>and those that demonstrate<br>intermittent symptoms indicative of<br>cardiac disease and require cardiac<br>monitoring on a continuing basis for<br>up to 30 days. The device<br>continuously records ECG data and<br>upon detection by an ECG analysis<br>algorithm or manually initiated by the<br>patient, automatically delivers the<br>recorded cardiac activity to the server<br>where it is presented and can be<br>reviewed by a medical professional.<br><br>The data received from the <b>bioflux</b><br>device can be used by another device<br>for arrhythmia analysis, reporting and<br>signal measurements. The <b>bioflux</b><br>device is not intended to sound any<br>alarms. The device does not deliver<br>any therapy, administer any drugs,<br>provide interpretive or diagnostic<br>statements or provide for any life<br>support. Bioflux is for prescription use<br>only. | The RhythmStar system is intended<br>for use by patients who either have<br>or are at risk of having cardiac<br>disease and those that demonstrate<br>intermittent symptoms indicative of<br>cardiac disease and require cardiac<br>monitoring on a continuing basis. The<br>device continuously records ECG data<br>and upon detection by an ECG<br>analysis algorithm or manually<br>initiated by the patient, automatically<br>delivers the recorded cardiac activity<br>to the server where it is presented<br>and can be reviewed by a medical<br>professional.<br><br>The data received from the<br>RhythmStar device can be used by<br>another device for arrhythmia<br>analysis, reporting and signal<br>measurements. The RhythmStar<br>system is not intended to sound any<br>alarms. The device does not deliver<br>any therapy, administer any drugs,<br>provide interpretive or diagnostic<br>statements or provide for any life<br>support. RhythmStar is for<br>prescription use only. |
| Delivered device<br>includes | Similar | <b>bioflux</b> device<br>-patient 3 lead ECG cable | RhythmStar monitor<br>- Patient 3 lead ECG lead cable |
| Category | Identical/Different | bioflux | RhythmStar |
| | | -internal rechargeable battery<br>-Wall Battery charger<br>-Carry pouch | - 2 externally re-chargeable batteries<br>- Wall battery charger |
| Monitor functional blocks | Similar | analog ECG front end,<br>MCU,<br>flash data storage,<br>RF modem for<br>data transmission,<br>LCD screen, and<br>Record button | analog ECG front end,<br>accelerometer,<br>MCU,<br>flash data storage,<br>RF modem for<br>data transmission,<br>LCD screen, and<br>Record button |
| The server: | identical | facilitate data communication with the bioflux device, provide data storage, and present the data for evaluation by a medical professional | facilitate data communication with the RhythmStar device, provide data storage, and present the data for evaluation by a medical professional |
| Device form factor | Similar | small, lightweight<br>ambulatory cardiac monitors. | small, lightweight<br>ambulatory cardiac monitors. |
| Wireless technology used to transmit data to server | Identical | Yes | Yes |
| Device is battery powered by a rechargeable Li-Ion battery | Identical | Yes | Yes |
| using a server, can adjust device programming parameters such as pre-post recording times and auto-triggering configuration. | Identical | Yes | Yes |
| Category | Identical/<br>Different | bioflux | RhythmStar |
| Devices incorporate an<br>accelerometer to<br>capture activity level<br>data related to patient<br>motion and device<br>orientation. | Different | No | Yes |
| devices have Record<br>button for manual<br>event recordings and a<br>user interface to<br>indicate device status<br>and mode of operation. | similar | Yes | Yes |
| Device incorporate<br>embedded ECG analysis<br>algorithm to auto-<br>capture Bradycardia,<br>Tachycardia and Atrial<br>Fibrillation and<br>arrhythmia events<br>between the signal<br>acquisition point and<br>the server. | Similar | Yes | Yes |
| device has at least 2<br>ECG channels and 3-<br>lead electrodes | Similar | Yes | Yes |
| Functional,<br>Environmental and<br>Electrical<br>characteristics | Similar | Yes | Yes |
| USB connection | different | Yes, used to charge the battery,<br>marks ECG data during recording in<br>case noise is introduced, not used for<br>data download. | Yes, used only for data download and<br>secondary communication, cannot be<br>connected during ECG recording |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
#### PERFORMANCE TESTING VII.
The following performance and safety tests have been passed successfully:
- . IEC 60601-2-47:2012 Medical electrical equipment - Part 2-47: Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems.
- IEC 60601-1:2012 3rd Edition with amendment 1, Medical electrical equipment -Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2:2014 4th Edition, Medical Electrical Equipment - Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility - Requirements and tests.
- IEC 60601-1-11:2015 Edition 1.1, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- ANSI/AAMI-EC 57:2012, Testing and Reporting Performance Results of Cardiac Rhythm and ST Segment Measurement Algorithms.
- IEC 62366-1:2015 Edition 1.0, Medical devices Part 1: Application of usability ● engineering to medical devices
- Biocompatibility testing of patient contacting materials according to ISO 10993-1.
- . Bench test results verify that Bioflux Monitor system can continuously record ECG signal, store ECG data in the device memory, and transmit manual or auto activated event recordings to the server via mobile network connection for evaluation by a medical professional. Test results verify that all requirements were met and that the Bioflux Monitor performs as designed.
#### SUBSTANTIAL EQUIVALENCE RATIONALE VIII.
The intended use, performance and technological characteristics of the bioflux Monitor system compared to the named predicate device demonstrates that the bioflux Monitor is substantially equivalent to the predicate.
#### CONCLUSIONS IX.
The analysis of the differences between bioflux Monitor and the predicate deice does not raise new questions of safety and effectiveness. Based on device performance test results, Biotricity determines that the bioflux Monitor system performs within its design specifications, and is substantially equivalent to the predicate devices.
The information in this 510(k) submission demonstrates that the bioflux Monitor system is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.