The Biotres is indicated for use on adult patients 18 years or older who may be asymptomatic or who suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety and may require cardiac recording on a continuous basis for up to 30 days. The signal acquired by the Biotres is not intended and should not be used for automated or semi-automated analysis.
Device Story
Biotres is a lightweight, 3-channel ambulatory ECG recorder for home use. Device captures continuous ECG data via three standard electrodes placed on the patient's chest. Patient operates power and event-marking buttons; LED provides status feedback. Data is stored locally on the device and transmitted via Bluetooth to a mobile Gateway App (iOS/Android), then to a secure SFTP server. Clinicians access recordings via the server to interpret data; device provides no automated or semi-automated analysis, diagnostic, or interpretive statements. System includes a dedicated AC wall charger. Intended for use by patients at home; clinicians manage study initiation and data review. Output is raw ECG data for physician interpretation to aid in cardiac diagnosis.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified against IEC 60601-2-47, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 62366-1. Biocompatibility testing performed per ISO 10993-1. Bench results confirm continuous ECG recording, data storage, and transmission capabilities.
Technological Characteristics
3-channel ambulatory ECG recorder; rechargeable Li-Ion battery; Bluetooth connectivity; LED status indicator; manual event-marking button. Complies with IEC 60601-1 (3rd Ed), IEC 60601-1-2 (4th Ed), IEC 60601-1-11, and IEC 60601-2-47. Biocompatible materials per ISO 10993-1. No automated analysis algorithms; rule-based status signaling only.
Indications for Use
Indicated for adult patients 18+ years, asymptomatic or experiencing transient symptoms (palpitations, dyspnea, dizziness, lightheadedness, pre-syncope, syncope, fatigue, chest pain, anxiety) requiring continuous cardiac recording up to 30 days. Contraindicated for patients with life-threatening arrhythmias requiring inpatient monitoring, patients requiring hospitalization, infants <10kg, patients with implanted pacemakers, or defibrillator use.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 19, 2022
Biotricity Spencer Ladow VP Engineering 275 Shoreline Drive, Suite 150 Redwood City, California 94065
Re: K211709
Trade/Device Name: Biotres Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: MWJ Dated: December 20, 2021 Received: December 22, 2021
Dear Spencer Ladow:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211709
Device Name Biotres
Indications for Use (Describe)
The Biotres is indicated for use on adult patients 18 years or older who may be
asymptomatic or who suffer from transient symptoms such as palpitations, shortness of
breath, dizziness, light headedness, pre-syncope, fatigue, chest pain and/or anxiety
and may require cardiac recording on a continuous basis for up to 30 days. The signal acquired by the Biotres is not intended and should not be used for automated or semi-automated analysis.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
(as required by 21CFR 807.92)
# Date Prepared: January 10, 2022
#### SUBMITTER l.
Biotricity 275 Shoreline Drive, Suite 150 Redwood City CA 94065 Contact Person: Spencer LaDow sladow@biotricity.com Phone: 585-414-7407
#### II. DEVICE
Name of Device: Classification Regulation: Common Name: Device Panel: Regulatory Class: Product Code:
Biotres 21 CFR§870.2800 Medical Magnetic Tape Recorder Cardiovascular Class II MWJ Electrocardiograph, Ambulatory (without Analysis)
#### III. PREDICATE DEVICE
The ePatch (Braemar Manufacturing, LLC) was cleared under K171410.
Name of Device: Classification Regulation: Common Name: Device Panel: Regulatory Class: Product Code:
ePatch 21 CFR§870.2800 Medical Magnetic Tape Recorder Cardiovascular Class II DSH
{4}------------------------------------------------
### IV. DEVICE DESCRIPTION
The Biotres system consists of 1) the Biotres Recorder, 2) a dedicated Biotres AC wall charger, 3) Biotres Gateway application for use on mobile devices using Android or iOS, and 4) Biotres Configured Secure Server.
## Biotres Recorder
The Biotres Recorder is an ECG recording only device and does not include ECG analysis, an ECG viewer/EGG display, alarms and is not a real time ECG monitor. The Biotres recorder is a small, lightweight, ambulatory, reusable 3 channel ECG recorder that is placed on the center of the chest (Reference Below).
Biotres Recorder – Front View
Image /page/4/Figure/5 description: This image is a diagram of a device with several labeled parts. The diagram labels the 'Status Indicator Light' and 'Marking Button' at the top of the device. The diagram also labels the 'Power Button' and 'Charging Port' on the side of the device, as well as the 'Snap for ECG electrode' at the bottom.
Biotres Recorder – Back View
BACK VIEW
Image /page/4/Figure/8 description: The image shows a top-down view of a Biometric device with three electrode connectors. The device has a circular body with the word "biotricity" and a logo printed on the top. There is also a QR code and some text that reads "Model: Blotree 1.0, Serial: 1230000089, Contains FCC ID: 2AKGEBIOTRES1, FDA CLEARED". The bottom of the device has the text "FC IP22 B".
## FRONT VIEW
{5}------------------------------------------------
The Biotres Recorder has two buttons: 1) power button and 2) patient event marking button. The pash the patient event marking button when the patient is feeling symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, fatigue, chest pain and/or anxiety. There is an LED "status indicator light" in the middle of the patient event marking button that provides recorder status indications when powering on, electrodes are disconnected, low battery, charging, fully charged, encountered internal error, acknowledgment of the patient event marking button, and study not engaged/start/running/paused/completed state.
The Biotres Recorder attaches to the patient's chest through three standard commercially available ECG electrodes (not include with the system) supplied by a clinic or a monitoring center. Biotres includes a Bluetooth modem to communicate with the Biotres Configured Secure Server through the Biotres Gateway Application. The Biotres Recorder integrated memory can store up to 30-days of ECG data. This data can be uploaded by the clinician to the Biotres Configured Secure Server.
## Biotres AC Wall Charger
The dedicated Biotres AC Wall charger is used to charge the Biotres Recorder daily when disconnected from the patient.
## Biotres Gateway Application
The ECG recorded data is transferred via the Biotres Gateway App (Downloaded from the app store for iOS or Android respectively) via a Bluetooth connection and subsequently to the Biotres Configured Secure Server. Clinicians can use the Biotres Gateway Application using their privileged login credentials to connect Biotres Recorders to the Biotres Configured Secure Server so that patient studies may be started and completed. The Biotres Gateway Application can be used by the patient once registered (nonprivileged login credentials) and the patient can use the application to monitor the status indicators of the Biotres Recorder and to enter symptoms if the Event Marking Button is pressed on the recorder. The ECG recorded data from the Biotres Recorder can be transferred, by the clinician, from the Biotres Configured Secure Server when the recorder is connected to the Biotres Gateway App at the end of a patient study.
## Biotres Secure Server
The Biotres Configured Secure Server is an SFTP server that is configured to work with the Biotres System. The server enables the Clinician, through secured login, to command the Biotres Recorder (e.g. initiate ECG data collection) and to access uploaded MIT16 ECG data recordings when the ECG data collection is complete. Biotres does not provide the patient or clinician with diagnostic or interpretive statements based on the ECG data. Diagnosis and interpretation of ECG data is dependent upon the medical advice and guidance provided by a physician or trained healthcare professional. The Biotres is for home use and not for use in the hospital. Clinical judgment and experience are used to check and interpret the data.
#### INDICATIONS FOR USE V.
The Biotres is indicated for use on adult patients 18 years or older who may be asymptomatic or who suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety and may require cardiac recording on a continuous basis for up to 30 days. The signal acquired by the Biotres is not intended and should not be used for automated or semi-automated analysis.
{6}------------------------------------------------
#### CONTRAINDICATIONS FOR USE VI.
The Biotres device is not intended for use under the following conditions:
- Patients with potentially life-threatening arrhythmias who require . inpatient monitoring.
- Patients who the attending physician thinks should be hospitalized. .
- Infants weighing less than 10 kg. (22 lbs.).
- . Patients with implanted pacemakers.
- Defibrillator Use. .
{7}------------------------------------------------
## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS WITH THE VII. PREDICATE DEVICE
| Category | Identical/<br>Different | Biotres | ePatch |
|--------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Identical | K211709 | K171410 |
| Classification Name | Identical | Medical<br>Magnetic Tape Recorder | Medical<br>Magnetic Tape Recorder |
| Product Code | Similar | MWJ | DSH |
| Intended Use | Similar | The Biotres is indicated for use on adult patients 18 years or older who may be asymptomatic or who suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety and may require cardiac recording on a continuous basis for up to 30 days. The signal acquired by the Biotres is not intended and should not be used for automated or semi-automated analysis. | ePatch® is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety. The ePatch® is intended for use by adolescents 18-21 and adults. |
| Defibrillator Protection | Similar | Not Defibrillator Proof, 8.5.5.2 of IEC 60601-1 passed | Not Defibrillator Proof |
| Wear Time | Different | Up to 30 Days | Up to 5 days |
| Recording Format | Identical | Continuous | Continuous |
| Delivered device includes | Similar | -3 lead ECG monitor<br>-internal rechargeable battery<br>-Wall Battery charger | ePatch monitor<br>Battery<br>Sensor Patch<br>Skin prep kit |
| Monitor functional blocks | Similar | Analog ECG front end,<br>MCU,<br>Flash data storage,<br>BLE modem for<br>data transmission,<br>LED indicator, and | Analog ECG front end,<br>Accelerometer,<br>MCU,<br>Flash data storage,<br>USB cable for<br>data transmission |
| Category | Identical/<br>Different | Biotres | ePatch |
| | | Record button | LED indicator, and<br>Record button |
| The App: | Similar | Mobile App, iOS, Android | PC based |
| The server: | Similar | Facilitate data communication<br>with the Biotres device,<br>provide data storage, and<br>present the data for evaluation by<br>a medical professional | Facilitate data communication<br>with the ePatch device, provide<br>data storage, and<br>present the data for evaluation by<br>a medical professional |
| Device form factor | Identical | Small, lightweight body worn<br>ambulatory cardiac monitors. | Small, lightweight body worn<br>ambulatory cardiac monitors. |
| Wireless technology<br>used to transmit data<br>to server | Different | Yes | No |
| Device is battery<br>powered by a<br>rechargeable Li-Ion<br>battery | Identical | Yes | Yes |
| Using an app, can<br>adjust device<br>programming<br>parameters such as<br>pre-post recording<br>times and auto-<br>triggering<br>configuration. | Similar | Yes, app is android or IOS based | Yes, app is PC based |
| Devices have Record<br>button for manual<br>event recordings and<br>a user led to indicate<br>device status and<br>mode of operation. | Similar | Yes | Yes |
| Device has at least 2<br>ECG channels and 3-<br>lead electrodes | Similar | Yes | Yes |
| Functional,<br>Environmental and<br>Electrical<br>characteristics | Similar | Yes | Yes |
| USB connection | Different | Yes, used for battery charging<br>cannot be connected during ECG<br>recording | Yes, used for data download and<br>communication, battery charging<br>cannot be connected during ECG<br>recording |
{8}------------------------------------------------
{9}------------------------------------------------
#### PERFORMANCE TESTING VIII.
The following performance and safety tests have been passed successfully:
- IEC 60601-2-47:2012 Medical electrical equipment Part 2-47: . Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems.
- IEC 60601-1:2012 3rd Edition with amendment 1. Medical electrical equipment ● - Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-2:2014 4th Edition, Medical Electrical Equipment Part ● 1-2: General requirements for basic safety and essential performance-Collateral standard: Electromagnetic compatibility - Requirements and tests.
- IEC 60601-1-11:2015 Edition 1.1. Medical electrical equipment Part ● 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- IEC 62366-1:2015 Edition 1.0. Medical devices Part 1: Application ● of usability engineering to medical devices
- Biocompatibility testing of patient contacting materials according to ISO 10993-1.
- Bench test results verify that Biotres system can continuously record ECG signal, store ECG data in the device memory, and transmit manual activated event recordings to the server via Bluetooth Biotres Gateway app connection for evaluation by a medical professional. Test results verify that all requirements were met and that the Biotres System performs as designed.
#### CONCLUSION IX.
The analysis of the differences between Biotres and the predicate device does not raise new questions of safety and effectiveness. Based on device performance test results. Biotricity determined that the Biotres system performed within its design specifications and is considered to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.