K233749 · Ge Hangwei Medical Systems Co., Ltd. · JAK · Feb 22, 2024 · Radiology
Device Facts
Record ID
K233749
Device Name
Revolution Ascend Sliding
Applicant
Ge Hangwei Medical Systems Co., Ltd.
Product Code
JAK · Radiology
Decision Date
Feb 22, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
AI/ML, Pediatric
Indications for Use
The system is intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories. This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further the images can be post processed to produce additional imaging planes or analysis results. The system is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
Device Story
Revolution Ascend Sliding is a head and whole-body CT system; modification of predicate Revolution Ascend (K213938). Input: X-ray transmission data acquired at various angles/planes (Axial, Cine, Helical, Cardiac, Gated). Transformation: Computer reconstruction of raw data into cross-sectional images; includes post-processing for reformatted planes/analysis. Output: Diagnostic CT images. Context: Gantry mounted on a transporter moving longitudinally on rails; patient remains stationary on a support device. Used in clinical settings (e.g., interventional radiology, radiation oncology) to allow imaging and procedures in the same room without patient transport. Clinicians use output for diagnosis, treatment planning, and monitoring. Benefit: Improved clinical workflow by enabling imaging and procedures in a single suite.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation testing performed under GE HealthCare quality system, including risk management, software V&V, dose specifications, and image quality testing per IEC 61223-3-5 ed.2.
Technological Characteristics
CT system with 75cm bore gantry on rail-based transporter. Components: Performix 40 Plus X-Ray Tube (liquid metal rotor), JEDI60DC High Voltage Generator (72 kW), NGX Collimator, Merc40H Detector (backlit diode, ASIC signal conversion). Reconstruction: FBP, ASIR-V, DLIR. Standards: IEC 60601-1 Ed. 3.2, IEC 60601-1-2 Ed 4.0, IEC 60601-2-44 Ed. 3.2, NEMA XR-25, XR-26, XR-28. Connectivity: Networked system.
Indications for Use
Indicated for head, whole body, cardiac, and vascular CT imaging in patients of all ages for diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the agency's name on the right. The symbol on the left is a stylized depiction of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The words "FOOD & DRUG" are larger than the other words.
February 22, 2024
GE Hangwei Medical Systems Co., Ltd. % Wenging Tian Lead Specialist, Regulatory Affairs West Area of Building No 3, No.1 Yongchang North Road Beijing Economic & Technological Development Area, Beijing 100176 CHINA
Re: K233749
Trade/Device Name: Revolution Ascend Sliding Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: January 23, 2024 Received: January 23, 2024
Dear Wenqing Tian:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233749
Device Name Revolution Ascend Sliding
#### Indications for Use (Describe)
The system is intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions. These images may be obtained either without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further the image can be post processed to produce additional imaging planes or analysis results.
The system is indicated for head, whole body, cardiac X-ray Computed Tomography applications in patients of all ages.
The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
Type of Use (Select one or both, as applicable)
| <span> <b> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> | <span> <b> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </b> </span> |
|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the GE Healthcare logo. The logo consists of the GE monogram enclosed in a circle on the left, followed by the text "GE HealthCare" on the right. The color of the logo and text is a light purple.
# 510(k) Summary
K233749
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | November 22, 2023 |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Hangwei Medical System Co., Ltd.<br>West Area of Building No.3, No.1 Yongchang North Road,<br>Beijing Economic and Technological Development Area, 100176 Beijing,<br>P.R. China. |
| Primary Contact: | Wenqing Tian<br>Lead Specialist, Regulatory Affairs<br>Phone: +86-155-9370-8328<br>Email: Wenqing.TIAN@ge.com |
| Secondary Contacts: | Helen Peng<br>Sr Regulatory Affairs Director<br>Phone: 262-4248222<br>Email: hong.peng@med.ge.com |
| | Haibo He<br>Regulatory Affairs Leader<br>Phone: +86-010-5708-3413<br>Email: haibo.he1@ge.com |
| Proposed Device/<br>Device Trade Name: | Revolution Ascend Sliding |
| Device Classification: | Class II |
| Regulation Number/<br>Product Code: | 21 CFR 892.1750 Computed Tomography X-ray System / JAK |
| Predicate Device Information | |
| Device Name: | Revolution Ascend |
| Manufacturer: | GE Healthcare Japan Corporation |
510(k) Number: K213938, Cleared on February 4, 2022
Device Classification: Class II
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the GE Healthcare logo. The logo consists of a purple circular emblem with a stylized "GE" monogram inside. To the right of the emblem, the words "GE HealthCare" are written in a simple, sans-serif font, also in purple.
Regulation Number/ 21 CFR 892.1750 Computed Tomography X-ray System / JAK Product Code:
### Reference Device Information
| Device Name: | GE LightSpeed ACT FP16 |
|-------------------------------------|--------------------------------------------------------|
| Manufacturer: | GE Medical System, LLC |
| 510(k) Number: | K091673 |
| Device Classification: | Class II |
| Regulation Number/<br>Product Code: | 21 CFR 892.1750 Computed Tomography X-ray System / JAK |
### Device Description
The proposed device Revolution Ascend Sliding is a head and whole-body CT system composed of a gantry, transporter, operator console with a host computer, power distribution unit, and interconnection cables. The system also includes image acquisition and reconstruction hardware/software, general system software, accompanying documents, and associated accessories/interconnections.
Revolution Ascend Sliding generates cross-sectional images of the body by computer reconstruction of xray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions modes.
CT is routinely used in interventional radiology (IR) and radiation oncology because it can provide outstanding spatial resolution and valuable geometric structure information for determining the treatment plan before a treatment procedure takes place. In some clinical scenarios in radiation oncology and/or IR, it is not optimal to transport a patient to a separate room and/or a separate table to get a diagnostic CT image. To simplify the clinical workflow, some hospitals desire to have the treatment procedure to be done in the CT suite where the clinicians can do CT imaging and the treatment procedure in the same room without moving the patient between rooms. To support such improved clinical workflow, we have modified predicate device Revolution Ascend (K213938) by mounting the gantry to a transporter which allows the gantry linear movement on rails in longitude direction to scan the desired region of interest instead of having the patient on a traditional CT tabletop (a.k.a. Cradle) that moves in and out of the gantry bore in the predicate device.
Revolution Ascend Sliding includes all the available features on predicate device (K213938) and maintains the identical/equivalent performance and image quality specifications. Its Intended Use and indications for use remain identical to those of the unmodified predicate device. Revolution Ascend Sliding remains compliant with IEC 60601-1 Ed. 3.2 and associated collateral and particular standards, NEMA XR25, XR26, XR28, and 21CFR Subchapter J performance standards.
### Intended Use
The system is intended to be used for head, whole body Computed Tomography applications.
{5}------------------------------------------------
# GE HealthCare 510(k) Premarket Notification Submission
Revolution Ascend Sliding
Image /page/5/Picture/2 description: The image shows the logo for GE Healthcare. The logo consists of the GE monogram in purple on the left, followed by the text "GE HealthCare" in purple. The monogram is a stylized version of the letters G and E intertwined.
# Indications for Use
The system is intended to produce cross-sectional images of the body by computer reconstruction of Xray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment supports, components and accessories.
This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further the images can be post processed to produce additional imaging planes or analysis results.
The system is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.
The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
## Technology
Revolution Ascend Sliding employs the same basic operating principles and fundamental technologies as the predicate device and reference device.
The table below summarizes the substantive feature/technological differences between the predicate device and the proposed device:
| Subsystem | Revolution Ascend<br>(Predicate Device, K213938) | Revolution Ascend Sliding<br>(Proposed Device) |
|-----------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Gantry | Revolution Ascend Gantry<br>- Bore size: 75cm<br>- Physical Tilt | Revolution Ascend Sliding Gantry<br>- Bore size: 75cm<br>- Physical Tilt: Disabled |
| | Performix 40 Plus X-Ray Tube<br>- Supports 40 mm beamwidth<br>- Liquid Metal rotor bearing<br>JEDI60DC High Voltage Generator<br>- Peak Power: 72 kW<br>-55 kW option (software limit<br>on mA) | Performix 40 Plus X-Ray Tube<br>- Supports 40 mm beam width<br>- Liquid Metal rotor bearing<br>JEDI60DC High Voltage Generator<br>- Peak Power: 72 kW |
| | NGX Collimator (75cm bore)<br>- 40 mm max z-coverage<br>Merc40H Detector<br>- Backlit Diode technology<br>- Chiclet Module design<br>- GE low noise ASIC technology<br>used for signal conversion | NGX Collimator (75cm bore)<br>- 40 mm max z-coverage<br>Merc40H Detector<br>- Backlit Diode technology<br>- Chiclet Module design<br>- GE low noise ASIC technology used<br>for signal conversion |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the GE Healthcare logo. The logo consists of a purple circular emblem with a stylized "GE" monogram inside. To the right of the emblem, the text "GE HealthCare" is written in a simple, sans-serif font, also in purple. The overall design is clean and corporate, representing the brand identity of GE Healthcare.
| Patient Support | - VT1700v Table<br>- Lite Table<br>- Table accessories | Patient Support equipment is stationary in<br>the horizontal axis, Gantry moves<br>horizontally with rails system. |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Horizontal<br>movement | The patient table moves horizontally<br>to the gantry bore during a CT exam<br>while the gantry is stationary | The gantry mounted on a transporter<br>moves horizontally on rails towards the<br>patient who is positioned on a patient<br>support device that is stationary. The<br>transporter uses substantial design and<br>manufacturing process as the reference<br>device cleared in K091673. |
| Reconstruction | FBP<br>ASIR-V (K133640)<br>DLIR cleared with Revolution Ascend<br>(K212067, K230807). | Same |
| Standards | IEC 60601-1 Ed. 3.1<br>IEC 60601-1-2 Ed 4.0<br>IEC 60601-1-3 Ed 2.1<br>IEC 60601-2-28 Ed 3.0<br>IEC 60601-2-44 Ed. 3.2<br>IEC 61223-3-5 Ed. 2.0<br>NEMA XR-25<br>NEMA XR-26<br>NEMA XR-28 | IEC 60601-1 Ed. 3.2<br>IEC 60601-1-2 Ed 4.0<br>IEC 60601-1-3 Ed 2.2<br>IEC 60601-2-28 Ed 3.0<br>IEC 60601-2-44 Ed. 3.2<br>IEC 61223-3-5 Ed. 2.0<br>NEMA XR-25<br>NEMA XR-26<br>NEMA XR-28 |
The changes described above do not change the fundamental control mechanism, operating principle, energy type, and do not change the intended use from the predicate device Revolution Ascend.
# Determination of Substantial Equivalence
The Revolution Ascend Sliding has completed testing and is in compliance with AAMI/ANSI ES 60601-1 and IEC60601-1 Ed. 3.2 and its associated collateral and particular standards, 21 CFR Subchapter J, and NEMA standards XR 25, XR 26, and XR 28. The device has successfully completed engineering design V &V and bench testing in support of substantial equivalence to the predicate device. The following quality assurance measures were applied to the development of the system:
- -Risk Analysis and Control
- -Required Reviews
- -Design Reviews
- -Testing on unit level (Module verification)
- -Integration testing (System verification)
- -Performance testing (Verification)
- -Safety testing (Verification)
GEHC believes the Revolution Ascend Sliding CT system is of comparable type and substantially equivalent to our currently marketed system Revolution Ascend (K213938).
{7}------------------------------------------------
# GE HealthCare 510(k) Premarket Notification Submission Revolution Ascend Sliding
Image /page/7/Picture/1 description: The image shows the GE Healthcare logo. The logo consists of a purple circular emblem with a stylized "GE" monogram inside. To the right of the emblem, the words "GE HealthCare" are written in a simple, sans-serif font, also in purple.
# Non-Clinical Testing
The verification and validation testing have been successfully completed as required by design control procedures under GE HealthCare's quality system. This includes risk management, software verification and validation testing as well as dose specifications and image quality testing in accordance with IEC 61223-3-5 ed.2 to demonstrate the overall system performance in a standardized and referenceable manner.
Non-clinical bench test results demonstrated the subject device performs equivalently to the predicate device.
## Clinical Testing:
The Revolution Ascend Sliding CT system can be fully tested on the engineering bench thus no additional clinical testing was required.
## Substantial Equivalence Conclusion
Based on the conformance to standards, development under our quality system, and the engineering testing provided, GE HealthCare believes that the Revolution Ascend Sliding is as safe and effective, and performs in a substantially equivalent manner to the unmodified predicate device Revolution Ascend (K213938).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.