K091673 · Ge Medical Systems, Inc. · JAK · Jun 24, 2009 · Radiology
Device Facts
Record ID
K091673
Device Name
GE LIGHTSPEED ACT FP16
Applicant
Ge Medical Systems, Inc.
Product Code
JAK · Radiology
Decision Date
Jun 24, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
When configured as a CT System, the GE LightSpeed ACT FP16 is indicated for head and whole body X-ray computed tomography applications. When configured as a fluoroscopic system, the GE LightSpeed ACT FP16 is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, cardiology, diagnostic and interventional procedures, and optionally, rotational imaging procedures. It is intended to replace fluoroscopic images obtained through image intensifier technology and is not intended for mammography applications.
Device Story
GE LightSpeed ACT FP16 is a modified CT system designed for integration with GE Innova 4100 Single Plane Digital Fluoroscopic Imaging system. Device features a movable gantry allowing CT scanning while patient remains on a shared x-ray table. Systems operate independently, one at a time. Used in clinical settings by healthcare professionals for diagnostic and interventional procedures. Modification enables rapid transition between CT and fluoroscopic imaging modalities without patient repositioning, improving workflow efficiency and patient comfort during complex interventional procedures.
Clinical Evidence
No clinical data. Verification of modification performed via bench testing and imaging phantoms on a fully configured installation. Performance of equipment movements, user controls, and status displays evaluated by test personnel.
Technological Characteristics
Computed tomography x-ray system with movable gantry. Integrates with digital fluoroscopic imaging system. Conforms to thermal, electrical, radiation, and mechanical safety standards. Design and development conform to 21 CFR 820, ISO 9001, and ISO 13485.
Indications for Use
Indicated for head and whole body CT imaging and fluoroscopic imaging of human anatomy for vascular angiography, cardiology, diagnostic, and interventional procedures. Not for mammography.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
KC91673
Special 510(k) Premarket Notification GE LiahtSpeed ACT FP16 Modification Revised June 15, 2009
## Attachment B:
JUN 2 4 2009
Summary of Safety and Effectiveness Prepared in accordance with 21 CFR Part 807.92(c).
| | | <i>GE Healthcare, P.O. Box 414, Milwaukee, WI 53201</i> |
|-------------|------------------|-----------------------------------------------------------------------------------------------------------|
| Section a): | | |
| 1. | Submitter: | GE Medical Systems, LLC doing business as GE Healthcare<br>3000 N. Grandview Blvd.<br>Waukesha, WI 53188 |
| | Contact Person: | Allen Schuh, Regulatory Affairs Manager<br>9900 Innovation Drive, Mail Stop RP2138<br>Wauwatosa, WI 53226 |
| | Date Prepared: | June 16, 2009 |
| 2. | Device Name: | GE LightSpeed ACT FP16<br>Computed Tomography X-Ray System, 21 CFR 892.1750, 90J--AK |
| 3. | Marketed Device: | GE HiSpeed X/i Smart Gantry Option |
- 4. Device Description: The GE LightSpeed ACT FP16 is a modified CT system intended to be located in close proximity to a GE Innova 4100 Single Plane Digital Fluoroscopic Imaging system utilizing a single patient table which can be configured for interventional x-ray or CT imaging procedures. This modification allows the patient to remain on the x-ray table which can be quickly configured for CT scanning via a movable gantry. Both the CT and X-ray systems are used separately, one at a time, in accordance with their respective indications for use.
- 5. Indications for Use: When configured as a CT System, the GE LightSpeed ACT FP16 is indicated for head and whole body X-ray computed tomography applications. When configured as a fluoroscopic system, the GE LightSpeed ACT FP16 is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, cardiology, diagnostic and interventional procedures, and optionally, rotational imaging procedures. It is intended to replace fluoroscopic images obtained through image intensifier technology and is not intended for mammography applications.
- 6. Comparison with Predicate Device: GE LightSpeed ACT FP16 is substantially equivalent in tems of safety and effectiveness to the GE HiSpeed X/i Smart Gantry Option (K012385). Both systems utilize a movable CT gantry to scan the patient while placed on a stationary table. They have the same technological characteristics, are comparable in key safety and effectiveness features, utilize similar design, construction, and materials and share the same intended use and clinical purpose.
#### Section b):
1. Non-clinical Tests: The modifications have been evaluated for conformance to specifications and overall performance as well as for thermal, electrical, radiation and mechanical safety as part of the design verification by employees of the manufacturer and by an independent test agency. The product has been found to conform with applicable medical device safety standards.
2. Clinical Tests: None required. Verification of conformance to specifications for the modification was completed with bench testing and imaging phantoms on a fully configured installation. Performance of the equipment movements, user controls and status display was easily evaluated with test equipment and test personnel familiar with clinical practice and procedures.
3. Conclusion: Intended uses and other key features are consistent with traditional clinical practice, FDA guidelines, and established methods of patient examination. The design and development process of the manufacturer conforms to 21 CFR 820, ISO 9001 and 13485 quality systems. The device conforms to applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Therefore, it is the opinion of GE Healthcare that the GE LightSpeed ACT FP16 is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES ..
Public-Health-Service_
JUN 2 4 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
GE Medical Systems LLC % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLS 1394 25" Street NW BUFFALO MN 55313
Re: K091673
Trade/Device Name: GE LightSpeed™ ACT FP16 Modification Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK, JAA Dated: June 5, 2009 Received: June 9, 2009
#### Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{2}------------------------------------------------
Page 2
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sincerely yours,
Jadine M. Morris
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Special 510(k) Premarket Notification GE LightSpeed ACT FP16 Modification Revised June 15, 2009
# Attachment E
### Indications for Use
510(k) Number (if known): K091673
Device Name: GE LightSpeed™ ACT FP16 Modification
Indications For Use of GE LightSpeed™ ACT FP16 when configured as a CT System:
The GE LightSpeed™ ACT FP16 is indicated for head and whole body X-ray computed tomography applications.
Indications For Use of GE LightSpeed™ ACT FP16 when configured as a Fluoroscopic System: .
GE LightSpeed™ ACT FP16 is indicated for use in generating fluoroscopic images of SE Elghtopeda. nonedures, and optionally, rotational imaging procedures. It is intended to replace procedures, and optionally, rethnology image intensifier technology and is not intended for mammography applications.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X | AND/OR | Over-The-Counter Use<br>(21 CFR 801 Subpart C) |
|-------------------------------------------------|---|--------|------------------------------------------------|
|-------------------------------------------------|---|--------|------------------------------------------------|
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-----------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal and<br>Radiological Devices | |
| 510(k) Number | K091673 |
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.