K230807 · Ge Healthcare Japan Corporation · JAK · Apr 20, 2023 · Radiology
Device Facts
Record ID
K230807
Device Name
Deep Learning Image Reconstruction
Applicant
Ge Healthcare Japan Corporation
Product Code
JAK · Radiology
Decision Date
Apr 20, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
AI/ML, Pediatric
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
CT Image Reconstruction
Deep Neural Network (DNN)
—
—
—
—
—
—
Indications for Use
The Deep Learning Image Reconstruction software is a deep learning based reconstruction method intended to produce cross-sectional images of the head and whole body by computer reconstruction of X-ray transmission data taken at different angles and planes, including Axial, (Volumetric), and Cardiac acquisitions, for all ages. Deep Learning Image Reconstruction software can be used for head, whole body, cardiac, and vascular CT applications.
Device Story
Deep Learning Image Reconstruction (DLIR) is a software-based image reconstruction method for CT scanners. It processes raw X-ray transmission data to generate cross-sectional images branded as 'TrueFidelity CT Images'. The device utilizes a dedicated Deep Neural Network (DNN) trained to model noise propagation through the imaging chain, allowing for noise removal while maintaining image quality metrics like low-contrast detectability and spatial resolution. The software is integrated into the CT scanner's existing reconstruction pipeline. Clinicians operate the device via the CT console, selecting from three reconstruction strengths (Low, Medium, High) based on clinical needs. The output provides enhanced image quality, which assists radiologists in diagnostic interpretation. The device is intended for use in clinical settings where CT scans are performed, benefiting patients by potentially enabling high-quality imaging at standard doses.
Clinical Evidence
Bench testing only. No clinical data. Performance was validated using phantom and raw clinical datasets to compare DLIR against ASiR-V. Metrics evaluated included Low Contrast Detectability (LCD), image noise (pixel standard deviation), high-contrast spatial resolution (MTF), streak artifact suppression, longitudinal spatial resolution (FWHM), Noise Power Spectrum (NPS), CT number uniformity/accuracy, contrast-to-noise ratio (CNR), and artifact analysis (metal, motion, truncation). Pediatric phantom and low-dose lung cancer screening protocol performance were also verified.
Technological Characteristics
Software-based deep learning reconstruction. Utilizes a dedicated Deep Neural Network (DNN) to model photon and electronic noise propagation. Integrated into CT scanner raw data reconstruction chain. Operates on CT console and GE Edison platform. DICOM compatible output. Software level of concern: Moderate.
Indications for Use
Indicated for head, whole body, cardiac, and vascular CT imaging in patients of all ages, including axial, volumetric, and cardiac acquisitions.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
April 20, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
GE Healthcare Japan Corporation % He Haibo Regulatory Affairs Leader 7-127, 4-chome, Asahigaoka Hino, Tokyo 191-8503 JAPAN
Re: K230807
Trade/Device Name: Deep Learning Image Reconstruction Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: March 23, 2023 Received: March 23, 2023
Dear He Haibo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K230807
Device Name Deep Learning Image Reconstruction
#### Indications for Use (Describe)
The Deep Learning Image Reconstruction software is a deep learning based reconstruction method intended to produce cross-sectional images of the head and whole body by computer reconstruction of X-ray transmission data taken at different angles and planes, including Axial, (Volumetric), and Cardiac acquisitions, for all ages.
Deep Learning Image Reconstruction software can be used for head, whole body, cardiac, and vascular CT applications.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the GE Healthcare logo. The logo consists of a purple circular emblem with a stylized "GE" monogram inside. To the right of the emblem is the text "GE HealthCare" in a sans-serif font, also in purple. The logo is clean and modern, representing the company's brand identity.
#### K230807 510(k) SUMMARY OF SAFETY AND EFFECTIVNESS
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.92:
| Date: | April 20, 2023 |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare Japan Corporation<br>7-127, Asahigaoka, 4-chome<br>Hino-shi, Tokyo, 191-8503, Japan |
| Primary Contact: | He Haibo<br>Regulatory Affairs Leader<br>Phone: 86-010-5708-3413<br>Email: haibo.he1@ge.com |
| Secondary Contacts: | Helen Peng<br>Senior Regulatory Affairs Director<br>Phone: 1-262-424-8222<br>Email: hong.peng@med.ge.com<br>Laura Turner<br>Regulatory Affairs Manager<br>Phone: 1-262-200-1044<br>Email: laura.turner@ge.com |
| Subject Device Name: | Deep Learning Image Reconstruction |
| Device Classification | Class II |
| Regulation Number/<br>Product Code: | 21 CFR 892.1750 Computed tomography x-ray system / JAK |
# Predicate Device Information
| Device Name: | Deep Learning Image Reconstruction |
|-------------------------------------|--------------------------------------------------------|
| Manufacturer: | GE Medical Systems, LLC |
| 510(k) Number: | K213999 cleared on February 18, 2022 |
| Regulation Number/<br>Product Code: | 21 CFR 892.1750 Computed tomography x-ray system / JAI |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the GE HealthCare logo. The logo consists of a purple circular emblem with the letters "GE" intertwined inside. To the right of the emblem, the words "GE HealthCare" are written in purple, with "GE" slightly larger than "HealthCare".
#### Reference Devices Information
| Device Name: | ASiR-V |
|-------------------------------------|--------------------------------------------------------|
| Manufacturer: | GE Medical Systems, LLC |
| 510(k) Number: | K133640 cleared on March 25, 2014 |
| Regulation Number/<br>Product Code: | 21 CFR 892.1750 Computed tomography x-ray system / JAK |
| Device Name: | Deep Learning Image Reconstruction |
| Manufacturer: | GE Healthcare Japan Corporation |
| 510(k) Number: | K212067 cleared on September 17, 2021 |
| Regulation Number/<br>Product Code: | 21 CFR 892.1750 Computed tomography x-ray system / JAK |
#### Device Description
Deep Learning Image Reconstruction is an image reconstruction method that uses a dedicated Deep Neural Network (DNN) that has been designed and trained specifically to generate CT Images to give an image appearance, as shown on axial NPS plots, similar to traditional FBP images while maintaining the performance of ASiR-V in the following areas: image noise (pixel standard deviation), low contrast detectability, high-contrast spatial resolution, and streak artifact suppression.
The images produced are branded as "TrueFidelity™ CT Images". Reconstruction times with Deep Learning Image Reconstruction support a normal throughput for routine CT.
The deep learning technology is integrated into the scanner's existing raw data-based image reconstruction chain to produce DICOM compatible "TrueFidelity™ CT Images".
The system allows user selection of three strengths of Deep Learning Image Recon: Low, Medium or High. The strength selection will vary with individual users' preference for the specific clinical need.
The DLR algorithm was modified on the Revolution CT/Apex platform for improved reconstruction speed and image quality and cleared in February 2022 with K213999. The same modified DLIR is now being ported to Revolution EVO (K131576) /Revolution Maxima (K192686), Revolution Ascend (K203169, K213938) and Discovery CT750 HD family CT systems including Discovery CT750 HD, Revolution Frontier and Revolution Discovery CT (K120833).
Compared to the predicate device, the intended use and indications for use of Deep Learning Image Reconstruction are identical.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the GE HealthCare logo. The logo consists of a purple circular emblem with a stylized "GE" monogram inside. To the right of the emblem, the text "GE HealthCare" is written in a simple, sans-serif font, also in purple. The logo is clean and modern, representing the company's brand identity.
### Intended Use
The Deep Learning Image Reconstruction software is intended for head, whole body, cardiac, and vascular CT scans.
#### Indications for Use
The Deep Learning Image Reconstruction software is a deep-learning based reconstruction method intended to produce cross-sectional images of the head and whole body by computer reconstruction of X-ray transmission data taken at different angles and planes, including Axial, Helical (Volumetric), and Cardiac acquisitions, for all ages.
Deep Learning Image Reconstruction software can be used for head, whole body, cardiac, and vascular CT applications.
#### Comparisons
The GEHC Deep Learning Image Reconstruction (DLIR) option for Revolution Maxima, Revolution Ascend and Discovery CT750 HD family CT systems is substantially equivalent to the predicate device K213999 for Revolution CT/Apex platform. The fundamental technology, i.e., the DLR algorithm, remains unchanged from the predicate. The table below summarizes the substantive feature/ technological similarities and differences between the predicate device and the proposed device.
| Specification/<br>Attribute | Predicate Device<br>Deep Learning Image Reconstruction<br>(K213999) | Proposed Device<br>Deep Learning Image<br>Reconstruction |
|----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| Technology | DLIR uses a dedicated Deep Neural<br>Network (DNN) which is trained on the<br>CT scanner and therefore models the<br>propagation of noise through the<br>system to identify and remove the<br>noise | Same |
| System statistics -<br>Noise modeling of<br>the data collection<br>imaging chain<br>(photon noise and<br>electronic noise) | Characterization of the photon<br>statistics as it propagates through the<br>preprocessing and calibration imaging<br>chain | Same |
| System statistics -<br>Noise characteristics<br>of the reconstructed<br>images | DLIR uses a trained DNN which models<br>the scanned object using information<br>obtained from extensive phantom and<br>clinical data to identify the noise<br>characteristics and remove it | Same |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the GE HealthCare logo. The logo consists of a purple circular emblem on the left and the text "GE HealthCare" in purple on the right. The emblem appears to be a stylized version of the letters "GE" intertwined.
| Specification/<br>Attribute | Predicate Device<br>Deep Learning Image Reconstruction<br>(K213999) | Proposed Device<br>Deep Learning Image<br>Reconstruction |
|-----------------------------|---------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clinical Workflow | Select recon type and strength (Low,<br>Medium, High). | Same |
| Reference<br>protocols/dose | Using the same reference protocols<br>provided on the Revolution CT/Apex<br>platform systems for ASIR-V | Using the same reference protocols<br>provided on the Revolution EVO,<br>Revolution Maxima, Revolution<br>Ascend and Discovery CT750 HD<br>family systems for ASIR-V |
| Deployment<br>Environment | On CT Console | On CT Console<br>On GE's Edison Platform. |
The subject device Deep Learning Image Reconstruction does not introduce any additional risks/hazards, warnings, or limitations.
## Determination of Substantial Equivalence
### Summary of Non-Clinical Testing
Deep Learning Image Reconstruction has successfully completed the design control testing per our quality system. No additional hazards were identified, and no unexpected test results were observed. Deep Learning Image Reconstruction was designed under the Quality System Regulations of 21CFR 820 and ISO 13485. GEHC believes that the extensive bench testing is sufficient for FDA's substantial equivalence determination.
The following quality assurance measures have been applied to the development of the system:
- . Requirement Definition
- Risk Analysis and Control
- Technical Design Reviews
- Formal Design Reviews
- Software Development Lifecycle
- O Code Review
- Software Unit Implementation O
- Software Integrations and Integration Testing O
- System Testing
- Safety Testing (Verification) o
- O Image Performance Testing (Verification)
- Simulating Use Testing (Validation) O
- Software Release
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the GE Healthcare logo. The logo consists of the GE monogram in a purple circle on the left, followed by the words "GE HealthCare" in purple text on the right. The text is in a sans-serif font and is slightly blurred.
The testing and results did not raise different questions of safety and effectiveness than associated with predicate device. We consider the proposed device is substantially equivalent to the predicate device, DLIR for Revolution CT/Apex platform systems.
The substantial equivalence is also based on the software documentation for a "Moderate" level of concern.
Engineering bench testing was also performed to support substantial equivalence and the product performance claims. The evaluation and analysis used the same test methodologies and acceptance criteria with the identical raw datasets obtained on GEHC's Revolution Ascend, Revolution Frontier and Discovery CT750 HD systems and then applying the Deep Learning Image Reconstruction or ASiR-V reconstruction (hence the dose (CTDIvol) is identical for both). The resultant images were then compared for:
- Low Contrast Detectability (LCD)
- Image Noise (pixel standard deviation)
- High-Contrast Spatial Resolution (MTF)
- Streak Artifact Suppression
- Spatial Resolution, longitudinal (FWHM slice sensitivity profile)
- Noise Power Spectrum (NPS) and Standard Deviation of noise
- CT Number Uniformity
- CT Number Accuracy
- Contrast to Noise (CNR) ratio
- Artifact analysis metal objects, unintended motion, truncation ●
- Pediatric Phantom IQ Performance Evaluation
- Low Dose Lung Cancer Screening Protocol IQ Performance Evaluation
#### Substantial Equivalence
Deep Learning Image Reconstruction for Revolution Maxima, Revolution Ascend and Discovery CT750 HD family was developed under GEHC Healthcare's quality system. The changes associated with Deep Learning Image Reconstruction (DLIR) do not change the Intended Use from the predicate, and represent equivalent technology characteristics, with no impact on control mechanism, operating principle, and energy type. Design verification, along with bench testing provided in this submission demonstrates that the Deep Learning Image Reconstruction (DLIR) met all of its design requirement and performance criteria. GEHC's quality system's design, verification, and risk management processes did not identify any additional hazards, unexpected results, or adverse effects stemming from the changes to the predicate.
GE HealthCare believes that Deep Learning Image Reconstruction software is substantially equivalent to the legally marketed predicate device and hence is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.