Dental Sensors, models Tuxdeluxe, 6100B-Size 1, and 6101B-Size 2 are intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed, and manipulated for diagnostic use by dentists. This device must only be used in hospital environments, clinics or dental offices by trained and qualified dental personnel, and not used in the oxygen rich environment. This device is suitable for providing dental radiography imaging for both adult and pediatric patients.
Device Story
Intraoral digital imaging sensor; collects dental x-ray photons; converts to digital signal via CMOS, scintillator, and optical fiber; transfers data via USB to computer workstation. Used in dental offices/clinics by trained personnel. Requires separate, cleared third-party software (e.g., XrayVision, Sota) for image storage, viewing, and manipulation. Provides high-quality dental images for diagnostic use; replaces traditional x-ray film. Benefits include improved resolution and diagnostic clarity for dental structures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), usability (IEC 60601-1-6), and IP68 ingress protection. Image quality verified by a USA Board Certified Dentist as clinically acceptable.
Technological Characteristics
CMOS sensor with scintillator and protective optical fiber. Sizes: 24.1x36.2x5.9mm (Size 1) and 30.5x42.8x5.7mm (Size 2). Connectivity: USB 2/3. Resolution: 14μm pixel size, 35 lp/mm. Operating temperature: +5°C to +35°C. Power: 5Vdc via USB. IP68 rated. Software: Compatible with third-party cleared dental imaging software.
Indications for Use
Indicated for dental radiography imaging in adult and pediatric patients. Must be used by trained dental personnel in hospitals, clinics, or dental offices. Contraindicated for use in oxygen-rich environments.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is the Department of Health & Human Services logo, and on the right, there is the FDA logo. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
October 20, 2023
Tuxedo Imaging LLC % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114
Re: K232552
Trade/Device Name: Tuxdeluxe (Size 1 6100B, Size 2 6101B) Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: MUH Dated: August 20, 2023 Received: August 23, 2023
Dear Mr. Daniel Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232552
Device Name Tuxdeluxe (Size 1 6100B, Size 2 6101B)
#### Indications for Use (Describe)
Dental Sensors, models Tuxdeluxe, 6100B-Size 1, and 6101B-Size 2 are intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed, and manipulated for diagnostic use by dentists. This device must only be used in hospital environments, clinics or dental offices by trained and qualified dental personnel, and not used in the oxygen rich environment. This device is suitable for providing dental radiography imaging for both adult and pediatric patients.
Type of Use (Select one or both, as applicable)
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------|
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary K232552 Tux Deluxe Intraoral Dental Digital Imaging Sensors
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 872.1800. Date prepared: September 18, 2023
#### 1. Company and Correspondent making the submission:
Tuxedo Imaging, LLC 313 S. High Street, Suite 200 Akron, OH 44308 Contact: Patrick Williams, President, Tel 330 382-5733
## 2. Trade /Proprietary Names: Tuxdeluxe (Size 1 6100B, Size 2 6101B)
Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: MUH
## 3. Legally Marketed Predicate Device Information: K220422
| Manufacturer: | iRay Technology Taicang Ltd. |
|--------------------|-----------------------------------|
| Trade/Device Name: | Dental Sensors NanoPix1, NanoPix2 |
| Regulation Number: | 21 CFR 872.1800 |
| Regulation Name: | Extraoral source x-ray system |
| Regulatory Class: | Class II |
| Product Code: | MUH |
#### 4. Description:
This Intraoral Digital Imaging Sensor employs CMOS (Complementary Metal-Oxide-Semiconductor), protective optical fiber and scintillator. This sensor was developed to obtain a high-quality x-ray image from the human mouth and its structures. The acquisition process is made by positioning the sensor inside the mouth, behind the structure you want to perform the exam. The structure must be exposed to an x-ray dose using an external source. Once exposed, the sensor performs a conversion of the x-ray photons into a digital signal and transfers it to a computer through USB connection (Universal Serial Bus). The x-ray generator (an integral part of a complete dental x-ray system) is not part of the device. Device sensor sizes:
Size 1: 24.1 x 36.2 x 5.9mm Size 2: 30.5 x 42.8 x 5.7mm
The I/O sensor is compatible with the following cleared software: Xray Vision / XV Capture (XV Web) / Curve / Sota. This software must be purchased separately by the end user.
## 5. Indications for use:
Dental Sensors, models Tuxdeluxe, 6100B-Size 1, and 6101B-Size 2 are intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, viewed, and manipulated for diagnostic use by dentists. This device must only be used in hospital environments, clinics or dental offices by trained and qualified dental personnel, and not used in the oxygen rich environment. This device is suitable for providing dental radiography imaging for both adult and pediatric patients
{4}------------------------------------------------
## 6. Comparison with predicate device:
The new devices Tuxdeluxe Sensors (and its predicate) are small digital imaging receptors that may be used in place of dental x-ray film.
Predicate sensor sizes: Size 1: 25 x 38.5 mm , Size 2: 31 x 40 mm Subject device sensor sizes: Size 1: 24.1 x 36.2 mm Size 2: 30.5 x 42.8. mm
The images are displayed on a computer workstation. The technologies employed by the predicate and our new device are almost identical. The proposed sensors have greater resolution than the predicate via smaller pixel size and greater numbers of total pixels. The same sensor sizes are available, and the computer interface is the same (USB). Both sensors use CMOS technology. Similarities: Indications for Use, available sizes, USB interface, CMOS technology employed. Differences: New device has greater resolution via smaller pixel size and greater number of pixels. The sensor sizes are nearly identical. A comparison table follows:
| PRODUCT | K220422 Trade/Device Name: Dental<br>Sensors NanoPix1, NanoPix2 | Tuxdeluxe Intraoral Dental Digital<br>Imaging Sensor |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for use | Dental Sensors, models NanoPix1 and<br>NanoPix2 are intended to collect dental xray<br>photons and convert them into electronic<br>impulses that may be stored, viewed, and<br>manipulated for diagnostic use by dentists.<br>This device must only be used in hospital<br>environments, clinics or dental offices by<br>trained and qualified dental personnel, and<br>not used in the oxygenrich environment.<br>This device is suitable for providing dental<br>radiography imaging for both adult<br>and pediatric. | Dental Sensors, models Tuxdeluxe, 6100B-Size<br>1, and 6101B-Size 2 are intended to collect<br>dental x-ray photons and convert them into<br>electronic impulses that may be stored,<br>viewed, and manipulated for diagnostic use by<br>dentists. This device must only be used in<br>hospital environments, clinics or dental offices<br>by trained and qualified dental personnel, and<br>not used in the oxygenrich environment. This<br>device is suitable for providing dental<br>radiography imaging for both adult and<br>pediatric patients. SAME |
| Where Used | Clinics, hospitals, dental offices | SAME |
| Operating<br>Temperature<br>Range | 10°C to 30°C | +5°C to +35°C<br>Greater operating temperature range. |
| Supply Voltage | +5 Vdc (USB) | +5 Vdc (USB) |
| Technology | CMOS | CMOS |
| Contrast | 12 bits | 12 bits |
| Gray Level | 4096 | 4096 |
| Pixel Size | 20 μm | 14 μm |
| Number of pixels | Size 1 1000 x 1500<br>Size 2 1300 x 1800 | Size 1: 1404 x 2104 pixels<br>Size 2: 1852 x 2574 pixels |
| Line pairs/mm | 16 Line pairs/mm | 35 Line pairs/mm |
| MTF | 0.1 at 12.5lp/mm | 0.095 at 12.5lp/mm (Essentially the same) |
| PRODUCT | K220422 Trade/Device Name: Dental<br>Sensors NanoPix1, NanoPix2 | Tuxdeluxe Intraoral Dental Digital<br>Imaging Sensor |
| DQE | DQE @ RQA5: >61.3% @ 0 lp/mm | DQE @ RQA5: >65% @ 0 lp/mm |
| Active Sensor<br>Area | Size 1: 25 x 38.5 mm ,<br>Size 2: 31 x 40 mm | Size 1: 24.1 x 36.2 mm<br>Size 2: 30.5 x 42.8. mm<br>SIMILAR |
| Imaging Software | NanoPix included | Not included. Compatibility verified with<br>these previously cleared software packages:<br>XrayVision K983111<br>XVCapture/XVWeb K983111<br>Curve K110139<br>Sota K210682 |
| Target Computer<br>System Type | Windows with USB | SAME |
| Connection type | USB 2 or 3 | SAME |
| Cable Length | 10 ft. | 2 or 3 m (6 ft or 9 ft) SIMILAR |
| Patient<br>Protection | Single Use Patient Protective Barrier, FDA<br>ckeared | Single Use Patient Protective Barrier, FDA<br>ckeared SAME |
| Photo | Image: NanoPix 1 and NanoPix 2 | Image: Tuxdeluxe Intraoral Dental Digital Imaging Sensor |
{5}------------------------------------------------
- 7. Non-clinical Testing Performed: Safety, EMC, Biocompatibility and Performance Data: Safety testing: Electrical, mechanical, environmental safety and performance testing according to standards:
SAFETY TEST ACCORDING TO THE STANDARDS IEC 60601-1: 2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012 (or IEC 60601-1: 2012 reprint) and EN 60601-1:2006+ A1:2013 + A12:2014, Recognition #19-49
ELECTROMAGNETIC COMPATIBILITY TESTS according to the standards: IEC 60601-1-2 Ed4.0 (2014) EN 60601-1-2 Ed4.0 (2015) Recognition 19-36 with a declaration of conformity to IEC TR 60601-4-2 Medical electrical equipment – Part 4-2: Guidance and interpretation – Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems Recognition 19-19. USABILITY ACCORDING TO THE STANDARDS IEC 60601-1-6:2010 (Third Edition) + A1:2013 and EN 60601-1-6:2010 +A1:2015 Recognition 5-89
{6}------------------------------------------------
Verification of degrees of protection IP68 provided by the enclosure according to the standards IEC 60529: 2013 and NF EN 60529: 1992 + A1: 2000 + A2: 2014
The main patient contact material is the FDA cleared barrier sheath (K160232, Disposable Barrier Sleeves and Covers), not supplied by us.
Successful operation has been demonstrated with these previously cleared dental imaging software packages: XrayVision K983111 XVCapture/XVWeb K983111 Curve K110139 Sota K210682 The user must purchase one of these software programs separately.
Risk analysis was conducted. All test results were satisfactory. Cybersecurity concerns were addressed Reference: Content of Premarket Submissions for Management of Cybersecurity in via labeling. Medical Devices. We also had an image quality analysis performed by a USA Board Certified Dentist who reviewed the proposed device images. He concluded that the images are of good quality, clinically acceptable, and suitable for the intended use.
## 8. Clinical Testing Performed: Clinical testing is not required for a finding of substantial equivalence.
- 9. Conclusions: Our conclusions drawn from the nonclinical tests (discussed above) are that we have demonstrated that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device identified according to 807.92(b)(3), 57 FR 18066, Apr. 28, 1992, as amended at 59 FR 64295, Dec. 14, 1994, According to the Federal Food, Drug and Cosmetic Law, 21 CFR Part 872. Based on the information provided in this pre-marketing notification, Tuxedo Imaging, LLC concludes that Intraoral Dental Digital Imaging Sensor is safe and effective and substantially equivalent to predicated devices, as described in this document.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.