K983111 · Apteryx, Inc. · LLZ · Nov 16, 1998 · Radiology
Device Facts
Record ID
K983111
Device Name
XRAY VISION
Applicant
Apteryx, Inc.
Product Code
LLZ · Radiology
Decision Date
Nov 16, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Uses for XrayVision in the dental industry are as follows: - Permits sending of data by email, client/server, ftp, phone and floppy disk - Organize and archive x-rays and patient related files - Facilitate sharing and transmission of information between doctors, other applications and other computers - Ability to place markers and annotations on x-rays for diagnostic purposes - Allows user to take calibrated measurements - Provide image-processing tools such as sharpening, colorizing, resizing, etc. (none of which physically alter images). - Electronic transmission of images to insurance companies or specialist When using XrayVision for diagnostic purposes the general public will be the doctor/dentist population. This software is primarily used to store a patient's imaging data (i.e. x-rays) in a form compatible with today's networked online practice management software. This software may also be used to transmit radiographs to specialists who will be treating the patient or the insurance company for approval of procedure and etc.
Device Story
XrayVision is a PC-based software system for dental practices; it acquires digital images from scanners, digital cameras, and intra-oral cameras via TWAIN interface. The software organizes, archives, and displays radiographs; it provides tools for image enhancement (sharpening, colorizing, resizing) and calibrated measurements. Users can add markers and annotations to images to aid diagnosis. The system supports electronic transmission of images via email, FTP, client/server, or physical media for specialist review or insurance claims. Security features include digital signatures, encryption, and password protection. The software does not control x-ray hardware or generate images; it processes existing digital files. Clinicians use the output for diagnostic review and treatment planning; the system ensures data integrity through lossless compression and prevents unauthorized image modification.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
PC-based software for Windows 95/higher. TWAIN interface for image acquisition. Standard image processing algorithms for contrast/sharpness. Lossless compression for primary storage; JPEG for import/export. Connectivity via TAPI (phone), FTP, and client/server. Industry-standard encryption for data integrity. Supports markers/annotations (ellipses, arrows, text).
Indications for Use
Indicated for use by dentists and doctors to organize, archive, enhance, measure, and electronically transmit dental radiographic images and patient files for diagnostic review and insurance claims processing.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K983111
Image /page/0/Picture/2 description: The image shows the logo for Apteryx, a software development company. The logo features the company name in a stylized, cursive font, with a horizontal line underlining the name. Above the name is a black feather, possibly representing a quill, which adds a creative and symbolic element to the logo.
Apteryx, Inc. . 554 White Pond Dr. Suite A . Akron, OH 44320 . (330) 867-607
apteryx@apteryxware.com · www.apteryxware.col
### Safety and Effectiveness Informatio
| Submitter: | Apteryx, Inc.<br>554 White Pond Dr. Suite A<br>Akron, OH 44320<br>(330) 867-6077 |
|--------------------|----------------------------------------------------------------------------------|
| Contact Person: | Kevin Crucs |
| Date Prepared: | September 2, 1998 |
| Device Name: | XrayVision |
| Equivalent Device: | TigerView, FDA registration number 955237 |
## Device Description -----------------------------------------------------------------------------------------------------------------------------------------------------------
XrayVision is a software package designed to work with standard PC compatible computers capable of running Windows 95 or higher (32-bit operating systems). This software is designed to be a fully functional digital radiograph scanning, archiving, electronic transmission and diagnostic review system. XrayVision permits the acquisition of images directly from commercially available scanners, digital cameras, intra-oral cameras, etc. Once an image has been converted into digital form, Xray Vision permits the user to easily organize and archive digital radiographs, images, and other patient related files to various archiving media (e.g. removable media, storage servers, hard drives, etc.). XrayVision provides tools that permit the user to enhance images and place markers and annotations on radiographs to aid in the diagnostic process. The image enhancements provided by this software package utilize industry standard algorithms that do not result in the alteration of an image's or a radiograph's content. Unlike a direct digital system, XrayVision does not control the x-ray taking system and does not generate x-ray images directly from the physical world (XrayVision processes images and radiographs produced from other devices and physical media). XrayVision only permits enhancing radiographs for diagnostic purposes and does not allow the introduction of false data or the modification of the images and radiographs in any way. To facilitate better correlation of information with images, XrayVision permits direct association of custom information with an image. Using these associations the user is able to relate doctor information, a patient's information, procedure, etc. with an image. When an image is transferred to another location, these associations guarantee that the correct information is always associated with an image (it is stored along with the image). Electronic transmission
{1}------------------------------------------------
capabilities are provided so that images can be sent to specialists and insurance companies for claims processing and diagnostic review. To ensure the originality and security of information, XrayVision contains security features such as digital signatures, encrypted transfers, image tracking and password protection to keep radiographs and other information safe from others and to prevent image modifications.
## Intended Use -----------------------------------------------------------------------------------------------------------------------------------------------------------------
Uses for XrayVision in the dental industry are as follows:
- Permits sending of data by email, client/server, ftp, phone and floppy disk �
- Organize and archive x-rays and patient related files ◆
- Facilitate sharing and transmission of information between doctors, other � applications and other computers
- Ability to place markers and annotations on x-rays for diagnostic purposes �
- Allows user to take calibrated measurements �
- Provide image-processing tools such as sharpening, colorizing, resizing, � etc. (none of which physically alter images).
- Electronic transmission of images to insurance companies or specialist �
When using XrayVision for diagnostic purposes the general public will be the doctor/dentist population. This software is primarily used to store a patient's imaging data (i.e. x-rays) in a form compatible with today's networked online practice management software. This software may also be used to transmit radiographs to specialists who will be treating the patient or the insurance company for approval of procedure and etc.
# Technological Characteristics
- Utilization of the TWAIN interface standard to acquire images from imaging 1. devices (scanner, digital cameras and interoral cameras, etc.).
- Utilizes standard image processing algorithms to improve image contrast, 2. sharpness and quality.
- 3. Utilizes standard image file formats for the storage and retrieval of images. Primary file format uses lossless date compression (ensures data integrity). Supports JPEG (lossy) only as an import/export format.
- 4. Telephone: common J/O and TAPI interface supported for transferring images over conventional phone lines.
FTP: File transfer protocol.
- C/S: Standard Internet transfer method for sending files via Internet client-server connection.
- Email: MIMI attachments utilized to transfer files via standard Internet email.
{2}------------------------------------------------
Disk: Standard PC I/O
- 5. Archiving utilizes a knowledge base for recording the location used to store files related to a patient.
- 6. Utilizes standard markers and annotations to draw attention to problem areas in an image (ellipse, arrow, text, freehand line, dots, rectangles, etc).
- 7. Industry standard encryption algorithms used for ensuring data integrity.
XrayVision has the same technological characteristics as the predicate device, TigerView and TigerScan [510(k) K95-5237]. A summary of the technologies used by these devices is as follows:
## End of 510(k) SUMMARY
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized eagle. The eagle is depicted with three lines forming its body and wings, giving it a modern and abstract appearance. The logo is in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 6 1998
Kevin Crucs President Apteryx, Inc. 554 White Pond Dr. Suite A Akron, OH 44320
Re:
K983111
X-Ray Vision (PACS) Dated: September 4, 1998 Received: September 4, 1998 Regulatory class: II 21 CFR 892.2050/Procode: 90 LLZ
Dear Mr. Crucs:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitto diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdr/dsmadsmamain.html".
Sincerely yours,
Lillian Yin, Ph.D.
Lillian Yin, Ph.D. Director, Division of Reprodu Abdominal, Ear, Nose and Thro and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 Of 1
510 (k) Number (if known): K983111
Device Name: Xray Vision
Indications For Use:
Uses for XrayVision in the dental industry are as follows:
- Permits sending of data by email, client/server, flp, phone and floppy disk .
- Organize and archive x-rays and patient related files .
- Facilitate sharing and transmission of information between doctors, other applications and � other computers
- Ability to place markers and annotations on x-rays for diagnostic purposes �
- Allows user to take calibrated mcasurements �
- Anows asor to and canbretted as sharpening, colorizing, resizing, ctc. (none of which � physically alter images).
- Electronic transmission of images to insurance companies or specialist �
When using XrayVision for diagnostic purposes the general public will be the doctor/dentist which using xiny is tof the case as patient's imaging data (i.e. x-rays) in a form compatible with today's networked online practice management software. This software may also be used to transmit radiographs to specialists who will be treating the patient or the insurance company for approval of procedure and etc.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Usc (Per 21 CFR 801.109) OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
2
(Optional Format 1-2-96)
David A. Segner
---
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devi
510(k) Number K983111
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.