K232084 · Bausch and Lomb · HQC · Feb 26, 2024 · Ophthalmic
Device Facts
Record ID
K232084
Device Name
Stellaris Elite vision enhancement system
Applicant
Bausch and Lomb
Product Code
HQC · Ophthalmic
Decision Date
Feb 26, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4670
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Bausch + Lomb Stellaris Elite™ vision enhancement system is intended for the emulsification and removal of cataracts, anterior and posterior segment vitrectomy. The system is designed for use in both anterior and posterior segment surgeries. It provides capabilities for phacoemulsification, coaxial and bimanual irrigation/ aspiration, bipolar coagulation, viscous fluid injection/removal and air/fluid exchange operations. The Bausch + Lomb Stellaris Elite™ vision enhancement system configured with the laser module is additionally intended for retinal photocoagulation and laser trabeculoplasty.
Device Story
Integrated ophthalmic microsurgical system for anterior/posterior segment surgery; performs phacoemulsification, vitrectomy, irrigation/aspiration, bipolar coagulation, viscous fluid injection/removal, air/fluid exchange, and laser procedures (photocoagulation/trabeculoplasty). Inputs include user commands via touchscreen GUI, foot controller, or remote; system monitors fluidics, vacuum, and IV pole position. Operates via electromechanical base with modules for phaco, illumination, fluidics, and optional laser. Outputs include ultrasonic energy for lens disruption, vacuum for tissue aspiration, and laser energy for retinal/trabecular treatment. Used in OR by ophthalmologists. System updates include new Congatec System on Module (SOM) and Windows 10 OS migration. Provides real-time surgical parameter control and microscope overlay (DMS option). Benefits include precise tissue removal and integrated multi-modality surgical support.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical performance testing, including software verification/validation, electrical safety, and electromagnetic compatibility testing.
Technological Characteristics
Electromechanical base with modular architecture (phaco, illumination, fluidics, laser). Materials: standard medical-grade components. Energy: ultrasonic (phaco), pneumatic (fluidics/vitrectomy), 532 nm diode-pumped solid-state laser. Connectivity: touchscreen GUI, foot controller, remote, optional digital media system (DMS). Software: Windows 10 platform, Congatec SOM. Sterilization: autoclave for surgical tray. Standards: IEC 60601-1, 60601-1-2, 60601-1-6, 60601-2-2, 60601-2-22, 80601-2-58.
Indications for Use
Indicated for patients undergoing anterior or posterior segment ophthalmic surgery, including cataract removal, vitrectomy, retinal photocoagulation, and laser trabeculoplasty.
Regulatory Classification
Identification
A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.
Predicate Devices
Stellaris Elite™ vision enhancement system (K170052)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 26, 2024
Bausch + Lomb Ken Nehmer Director, Regulatory Affairs 3365 Tree Court Industrial Blvd. St. Louis, Missouri 63122
Re: K232084
Trade/Device Name: Stellaris Elite™ vision enhancement system Regulation Number: 21 CFR 886.4670 Regulation Name: Phacofragmentation System Regulatory Class: Class II Product Code: HQC, HQE, HQF Dated: January 26, 2024 Received: January 29, 2024
Dear Ken Nehmer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Claudine H. Krawczyk -S
Claudine Krawczyk Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K232084
Device Name
Bausch + Lomb Stellaris Elite™ vision enhancement system
Indications for Use (Describe)
The Bausch + Lomb Stellaris Elite™ vision enhancement system is intended for the emulsification and removal of cataracts, anterior and posterior segment vitrectomy. The system is designed for use in both anterior and posterior segment surgeries. It provides capabilities for phacoemulsification, coaxial and bimanual irrigation/ aspiration, bipolar coagulation, viscous fluid injection/removal and air/fluid exchange operations. The Bausch + Lomb Stellaris Elite™ vision enhancement system configured with the laser module is additionally intended for retinal photocoagulation and laser trabeculoplasty.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------|---------------------------------------------------------------------|
| <div><span></span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span></span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image displays the logo for Bausch + Lomb, a company specializing in eye health products. The logo features the company name in a bold, sans-serif font, with a plus sign in a lighter shade of teal connecting the two words. Below the logo, the tagline "See better. Live better." is written in a smaller, softer font, emphasizing the company's mission to improve people's vision and quality of life.
## K232084-510(k) Summary
#### 8. 510(K) SUMMARY
#### 8.1 General Information
| Submitter: | Contact Person: |
|-----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Bausch + Lomb Inc.<br>3365 Tree Court Industrial Blvd.<br>St. Louis MO 63122<br>General Telephone: 636-226-3017 | Ken Nehmer<br>3365 Tree Court Industrial Blvd.<br>St. Louis MO 63122<br>Telephone: 415-297-0408<br>ken.nehmer@bausch.com |
| Preparation Date: | February 26, 2024 |
| <b>8.2 Names</b> | |
| Device Name: | Stellaris EliteTM vision enhancement system |
| Classification Names: | Phacofragmentation Unit, Vitreous Aspiration and Cutting Instrument |
| Common Name: | Ophthalmic surgical system for cataract and vitreo-retinal surgery |
| CFR References: | 21 CFR 886.4670, 21 CFR 886.4150, 21 CFR 886.4390 |
| Product Codes: | HQC, HQE, HQF |
#### 8.3 Predicate Device
K170052 - Stellaris Elite™ vision enhancement system, Bausch + Lomb
#### 8.4 Product Description
The Bausch + Lomb Stellaris Elite™ vision enhancement system is comprised of an integrated ophthalmic microsurgical system designed for use in anterior and posterior segment surgery including phacofragmentation and vitreous aspirating and cutting as well as endoillumination. Additionally, the Stellaris Elite™ vision enhancement system may be configured with a 532 nm laser module for photocoagulation. The Stellaris Elite™ vision enhancement system was initially cleared under K 162342 which included the introduction of an Adaptive Fluidics feature as well as increase in vitrectomy cut rate. The Vitesse vitrectomy handpiece was introduced and cleared for use on the Stellaris Elite™ vision enhancement system under K170052.
A selection of disposable single-use procedure packs is available for use with this system. These packs contain the necessary tubing to facilitate delivery and removal of air
{5}------------------------------------------------
# BAUSCH + LOMB
See better. Live better.
## K232084-510(k) Summary
and fluids to/from the patient as well as a selection of components (cannulas, cutters, probes, etc.) that facilitate the surgical procedure. The items are arranged for physician convenience and may be presented as a group intended to support all the needs of a procedure or packaged singularly to allow the physician greatest flexibility. Additional Stellaris Elite™ vision enhancement system procedure packs have been made available and these packs contain components that have been used in previously available procedure packs. Some of these Stellaris Elite™ vision enhancement system procedure packs now include the BiBlade vitrectomy cutter produced by Medical Instruments Development Lab (separately cleared via K153168).
The previously cleared Stellaris Elite™ vision enhancement system introduced new feature sets that are not available on the legacy Stellaris PC systems previously cleared under K133242/K133486 respectively. A summary of these additional feature sets for the Stellaris Elite™ vision enhancement system are listed in Table 8-1. The various available system configurations allow for system feature flexibility and are made available for physician convenience.
The most recent clearance to the Stellaris Elite™ vision enhancement system introduced the Vitesse vitrectomy feature and was cleared via K170052 on April 19, 2017.
This traditional 510(k) incorporates updates to User Interface Computer System on Module (SOM) with an introduction of a new replacement Congatec SOM. In addition. the Stellaris Software platform is updated to Windows 10, since the current Stellaris software platform Windows XP is obsolete/incompatible with the new Congatec SOM.
The features remain the same as the previously approved Stellaris Elite™ vision enhancement system. The configuration matrix is listed in Table 8-1. The various available system configurations allow for system feature flexibility and are made available for physician convenience.
{6}------------------------------------------------
## BAUSCH + LOMB
See better. Live better.
## K232084–510(k) Summary
| | | | Table 8-1: Stellaris Elite™ vision enhancement system Configuration Table |
|-------------------------------------------------------------------------------------------------------------------------|----------------------|---------------|----------------------------------------------------------------------------|
| | | | |
| | Configuration by SKU | | |
| Feature | BL11145 | BL14455 | BL15455 |
| 2500 cpm vitrectomy<br>Initially cleared via K162342 – no substantial<br>changes to the feature since initial clearance | Available | Not Available | Not Available |
| 7500 cpm vitrectomy<br>Initially cleared via K162342 – no substantial<br>changes to the feature since initial clearance | Not Available | Available | Available |
| Vitesse vitrectomy<br>Initially cleared via K170052 | Not Available | Available | Available |
| Adaptive Fluidics<br>Initially cleared via K162342 for use in the anterior<br>segment of the eye | Available | Available | Available |
| Laser module<br>Initially cleared via K133486 – no substantial<br>changes to the feature since initial clearance | Not Available | Not Available | Available |
Abbreviations: cpm = cuts per minute
BL11145: formerly the Stellaris Anterior Vision Enhancement System
BL14455: formerly the Stellaris PC Vision Enhancement System
BL15455: formerly the Stellaris PC Vision Enhancement System with laser
| The Stellaris Elite™ vision enhancement system is composed of the following: | | | |
|------------------------------------------------------------------------------|--|--|--|
| | | | |
| Module/Assembly | Description |
|----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electromechanical<br>Base (EMB) | The EMB is intended to act as the central frame<br>including the outer skin through which all system<br>modules attach or receive power. The EMB also serves<br>as the attachment point for all external electrical cords<br>and fluid lines. |
| Computer Unit | Consists of a standard personal computer (PC)<br>customized to include the computer user interface with<br>the Stellaris Elite vision enhancement system. |
| User Interface<br>Module | Provides user control of operation of the Stellaris Elite<br>vision enhancement system. It provides a touch screen<br>display with user input controls to command the system<br>into its various modes and controls activation of<br>functions within each mode. |
| Power Supply<br>Module | Provides electrical power to all parts of the Stellaris Elite<br>vision enhancement system integral to the EMB. |
| Device<br>Relationship<br>Management<br>(DRM) | Designed to interact with an external management<br>system to monitor specific activities of the Stellaris Elite<br>vision enhancement system and to provide Bausch +<br>Lomb with a tool to monitor certain system metadata (no<br>patient identifiable data is collected) |
| Anterior Module<br>(Phaco) | The Anterior Module contains all anterior segment<br>connections, circuits, and operating hardware and<br>firmware necessary for anterior segment procedures. It<br>enables ultrasound for continuous, burst, and modulated<br>modes, bimanual phacoemulsification, and bipolar<br>coagulation capability. |
| Posterior<br>Illumination<br>Module<br>(Configurations<br>BL14455 and<br>BL15455 only) | The Stellaris Elite vision enhancement system<br>Illumination Module has two light outputs from two<br>separate light bulbs. The module is designed to support<br>the use of Xenon or Mercury light bulbs. One of the<br>light outputs has a built-in switchable filter to produce<br>tinted Green, Amber or Yellow lights. Lights are<br>transmitted through fiber optics of varying gauge sizes. |
| Compressor<br>Module | The Compressor Module provides the necessary pressure<br>to drive the 2 fluidics modes available for operation of<br>the system vacuum pumps and accessories that require<br>external assistance as drivers for fluid delivery and<br>exchange. This module is the source of bottle infusion<br>and also provides Adaptive Fluidics functionality. It<br>provides pressure to vitrectomy and pinch valves. |
| Fluidics Module | This module is designed for use with anterior and<br>posterior surgical procedures. The module provides<br>vacuum control. The vacuum control is done by the<br>software via a transducer located inside the Fluidics<br>Module. The Fluidics Module provides the electronics,<br>pneumatic actuators, and mechanisms necessary to<br>interface and control the fluidics of the cartridge. The<br>Fluidics Module also monitors the fluid level. A<br>pneumatic vitrectomy cutter control and interface is<br>included. |
| Automated IV<br>Pole | The unit contains electrical sensors to place the pole at<br>pre-ordained positions via screen generated touch control<br>commands for height requirements in delivery of fluids<br>at certain pressures. |
| Surgical Instrument<br>Tray | The tray is ergonomically located within the base unit<br>and set at a height useful to the surgeon and the<br>assistant. The tray is capable of tilting, and is removable<br>for autoclave sterilization. The tray accommodates a<br>surgical drape. The tray is retractable within the unit<br>when not in use. |
| Foot Controller | The foot controller allows the user to control various<br>features of the system. |
| Remote Control | The remote control unit mimics the GUI functions. The<br>device can control operation of IV pole height, vacuum<br>pressure, ultrasound power, and bipolar power. |
| Digital Media<br>System (option) | Provides a microscope overlay capability. The DMS is<br>used to overlay the surgical parameters output from the<br>Stellaris Elite to the video image of the surgical site<br>captured by the operating microscope camera. The<br>combined image is output to a video monitor and/or a<br>video recorder. |
| Phaco Handpiece | Electrically powered hand-held instrument for anterior<br>segment surgery. There are three<br>capabilities: Generation of ultrasonic energy for<br>disruption of lens tissue; provision for an aspiration path<br>for disrupted tissue from the eye, and the delivery of<br>saline irrigation fluid to the eye for standard, burst, or<br>pulsed phacoemulsification. A phacofragmentation<br>(posterior) handpiece is also available and has two<br>functions: generation of ultrasonic energy for disruption<br>of lens tissue and aspiration path for disrupted tissue<br>from the eye for standard, burst, or pulsed<br>phacofragmentation. |
| Laser Module<br>(Configuration<br>BL15455 only) | The laser module uses an off the shelf 2 Watt 532 nm<br>diode pumped frequency doubled solid state laser<br>coupled coaxially with an off the shelf diode sourced 0.8<br>mW 635 nm red aiming beam. These are enclosed in a<br>separate box located inside the system. The laser module |
| can be driven in either single shot, continuous or pulsed | |
| mode. The user may control treatment power, duty | |
| cycle, repetition rate, and aiming beam power. | |
| | |
{7}------------------------------------------------
## BAUSCH+LOMB See better. Live better.
## K232084–510(k) Summary
{8}------------------------------------------------
## BAUSCH+LOMB See better. Live better.
## K232084–510(k) Summary
{9}------------------------------------------------
# BAUSCH + LOMB
See better. Live better.
## K232084-510(k) Summary
#### 8.5 Indications for Use
The Bausch + Lomb Stellaris Elite™ vision enhancement system is intended for the emulsification and removal of cataracts, anterior and posterior segment vitrectomy. The system is designed for use in both anterior and posterior segment surgeries. It provides capabilities for phacoemulsification, coaxial and bimanual irrigation/ aspiration, bipolar coagulation, vitrectomy, viscous fluid injection/removal and air/fluid exchange operations. The Bausch + Lomb Stellaris Elite™ vision enhancement system configured with the laser module is additionally intended for retinal photocoagulation and laser trabeculoplasty.
#### 8.6 Summary of Technological Characteristics
The technological characteristics of the Stellaris Elite™ vision enhancement system are substantially equivalent to those of the predicate devices.
| Characteristic | Subject Device:<br>Stellaris EliteTM vision<br>enhancement system | K170052<br>Stellaris PC Vision Enhancement System |
|----------------|-------------------------------------------------------------------|---------------------------------------------------|
| Intended Use | Anterior/Posterior ophthalmic<br>surgery | Same as subject device |
{10}------------------------------------------------
## BAUSCH+LOMB See better. Live better.
## K232084–510(k) Summary
| Indications for<br>Use | The Bausch + Lomb Stellaris Elite™<br>vision enhancement system is<br>intended for the emulsification and<br>removal of cataracts, anterior and<br>posterior segment vitrectomy. The<br>system is designed for use in both<br>anterior and posterior segment<br>surgeries. It provides capabilities for<br>phacoemulsification, coaxial and<br>bimanual irrigation/ aspiration,<br>bipolar coagulation, vitrectomy,<br>viscous fluid injection/removal and<br>air/fluid exchange operations. The<br>Bausch + Lomb Stellaris Elite™<br>vision enhancement system<br>configured with the laser module is<br>additionally intended for retinal<br>photocoagulation and laser<br>trabeculoplasty. | The Bausch + Lomb Stellaris PC Vision<br>Enhancement System is intended for the<br>emulsification and removal of cataracts,<br>anterior and posterior segment vitrectomy.<br>The system is designed for use in both<br>anterior and posterior segment surgeries. It<br>provides capabilities for<br>phacoemulsification, coaxial and bimanual<br>irrigation/aspiration, bipolar coagulation,<br>vitrectomy, viscous fluid injection/removal<br>and air/fluid exchange operations.<br>The laser modes are intended for retinal<br>photocoagulation and laser trabeculoplasty. |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Laser capabilities | Yes | Same as subject device |
| User interface | LCD touch screen | Same as subject device |
| Footswitch | Yes | Same as subject device |
| Electrical<br>Characteristics | 90 - 130 VAC, 50/60 Hz<br>200 - 240 VAC, 50/60 Hz | Same as subject device |
| User Interface<br>SOM (System on<br>Module) | Congatec | Kontron |
| Operating System | Windows 10 | Windows XP |
| Maximum vacuum | 660 mmHg* | 600 mmHg |
| eyeTELLIGENCE | Available* | Available |
{11}------------------------------------------------
### BAUSC TT KOMB See better. Live better.
## K232084-510(k) Summary
| Vitesse revisions including: | Available* | Available |
|------------------------------------------------------|------------|-----------|
| a. Introduction of Vitesse procedure pack BL5631TD | | |
| b. Vitesse handpiece manufacturing process revisions | | |
| c. Firmware update to introduce Vitesse auto tune | | |
* incremental improvements introduced since the predicate of K170052
## Performance Data
## Biocompatibility testing
Biocompatibility assessment is not required to support the premarket notification. The Stellaris Elite™ vision enhancement systems do not contain direct or indirect patient contacting materials.
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety tests of the Stellaris Elite™ vision enhancement system have demonstrated its compliance with applicable requirements of the following electrical standards:
| IEC 60601-1:2020 | Medical electrical equipment - Part 1: General requirements for<br>basic safety and essential performance |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2:2020 | Medical electrical equipment - Part 1-2: General requirements<br>for basic safety and essential performance - Collateral standard:<br>Electromagnetic compatibility - Requirements and tests |
| IEC 60601-1-6:2020 | Medical electrical equipment. Part 1-6: General requirements<br>for basic safety and essential performance. Collateral standard:<br>Usability |
| IEC 60601-2-2:2017 | Medical electrical equipment - Part 2-2: Particular<br>requirements for the basic safety and essential performance of<br>high frequency surgical equipment and high frequency surgical<br>accessories |
{12}------------------------------------------------
## BAUSCH + LOMB See better. Live better.
## K232084-510(k) Summary
| IEC 60601-2-22:2019 | Medical electrical equipment - Part 2-22: Particular<br>requirements for the safety of diagnostic and therapeutic laser<br>equipment |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 80601-2-58:2016 | Medical electrical equipment - Part 2-58: Particular<br>requirements for the basic safety and essential performance of<br>lens removal devices and vitrectomy devices for ophthalmic<br>surgery |
Details of the electromagnetic compatibility and safety standard testing can be found in Section 17.
## Software Verification and Validation Testing
Functional, simulated use, environmental and transport testing were also performed on representative units. Software changes were verified and validated in accordance with the Bausch + Lomb software quality procedures which comply with EN ISO IEC 62304:2006 Medical device software -- Software life cycle processes. Refer to Section 16 for additional details on the specifications and reports.
The Stellaris Elite™ vision enhancement system passed all of the above referenced testing. This testing demonstrates that the functional requirements have been met and that the modified device is substantially equivalent to the predicate devices.
### Mechanical and acoustic testing
There was no mechanical or acoustical testing performed to support substantial equivalence of this premarket notification.
## Animal Study
There was no animal study performed to support substantial equivalence of this premarket notification.
## Non-Clinical Performance Data
Non-clinical performance testing was conducted to establish substantial equivalence between the predicate device and the revisions as identified within this premarket notification.
## Clinical Studies
There was no clinical study performed to support substantial equivalence of this premarket notification.
#### 8.7 Conclusion
The Stellaris Elite™ vision enhancement system shares the same indications for use, design features, and functional features with, and thus is substantially equivalent to, the predicate devices. Non-clinical test results demonstrate that the Stellaris Elite™ vision enhancement system is substantially equivalent to the predicate devices and no new issues of safety or effectiveness have been raised.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.