The Bausch + Lomb Stellaris Elite vision enhancement system is intended for the emulsification and removal of cataracts, anterior and posterior segment vitrectomy. The system is designed for use in both anterior and posterior segment surgeries. It provides capabilities for phacoemulsification, coaxial and bimanual irrigation/aspiration, bipolar coagulation, vitrectomy, viscous fluid injection/removal, and air/fluid exchange operations. The Stellaris Elite Vision Enhancement System configured with the laser module is additionally intended for retinal photocoagulation and laser trabeculoplasty.
Device Story
Integrated ophthalmic microsurgical system for anterior/posterior segment surgery; performs phacoemulsification, vitrectomy, irrigation/aspiration, bipolar coagulation, viscous fluid injection/removal, and air/fluid exchange. Optional 532 nm laser module for photocoagulation and trabeculoplasty. Operated by surgeons in clinical settings via LCD touch screen and footswitch. System incorporates Vitesse vitrectomy feature; supports adaptive fluidics. Provides real-time surgical control; output includes mechanical tissue removal and laser energy delivery. Benefits include enhanced surgical flexibility and precision for cataract and vitreo-retinal procedures.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-2, IEC 60601-2-22, and IEC 80601-2-58. Successful functional, simulated use, environmental, and transport testing performed. Software verified and validated per EN ISO IEC 62304.
Technological Characteristics
Ophthalmic microsurgical system; LCD touch screen interface; footswitch control. Electrical: 90–130 VAC or 200–240 VAC, 50/60 Hz. Features phacoemulsification, vitrectomy (including Vitesse), and optional 532 nm laser. Standards: IEC 60601-1, 60601-1-2, 60601-1-6, 60601-2-2, 60601-2-22, 80601-2-58. Software lifecycle per EN ISO IEC 62304.
Indications for Use
Indicated for patients undergoing cataract removal, anterior or posterior segment vitrectomy, retinal photocoagulation, or laser trabeculoplasty.
Regulatory Classification
Identification
A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 19, 2017
Bausch + Lomb Mr. Dan Regan Director, Regulatory Affairs 3365 Tree Ct. Industrial Blvd St. Louis, MO 63122
Re: K170052
Trade/Device Name: Stellaris Elite Vision Enhancement System Regulation Number: 21 CFR 886.4670 Regulation Name: Phacofragmentation System Regulatory Class: Class II Product Code: HOC, HOE, HOF Dated: March 27, 2017 Received: March 28, 2017
Dear Mr. Dan Regan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Denise L. Hampton -S
Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K170052
Device Name Stellaris Elite vision enhancement system
#### Indications for Use (Describe)
The Bausch + Lomb Stellaris Elite vision enhancement system is intended for the emulsification and removal of cataracts, anterior and posterior segment vitrectomy. The system is designed for use in both anterior segment surgeries. It provides capabilities for phacoemulsification, coaxial and bimanual irrigation, bipolar coagulation, vitrectomy, viscous fluid injection/removal, and air/fluid exchange operations. The Stellaris Elite vision enhancement system configured with the laser module is additionally intended for retinal photocoagulation and laser trabeculoplasty.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| <span style="display:inline-block;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="display:inline-block;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
#### General Information 1.
| <b>Submitter:</b> | <b>Contact Person:</b> |
|-----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Bausch + Lomb Inc.<br>3365 Tree Court Industrial Blvd.<br>St. Louis MO 63122<br>General Telephone: 636-226-3017 | Dan Regan<br>3845 Corporate Centre Drive<br>O'Fallon, MO 63368<br>(636) 794-5013<br>dan.regan@valeant.com |
Preparation Date: November 11, 2016
#### 2. Names
| Device Name: | Stellaris Elite vision enhancement system |
|-----------------------|------------------------------------------------------------------------|
| Classification Names: | Phacofragmentation Unit, Vitreous Aspiration and Cutting<br>Instrument |
| Common Name: | Ophthalmic surgical system for cataract and vitreo-retinal<br>surgery |
| CFR References: | 21 CFR 886.4670, 21 CFR 886.4150, 21 CFR 886.4390 |
| Product Codes: | HQC, HQE, HQF |
#### Predicate Devices 3.
K133486 - Stellaris PC Vision Enhancement System, Bausch + Lomb K133242 - Stellaris Vision Enhancement System, Bausch + Lomb
#### Product Description 4.
The Bausch + Lomb Stellaris Elite Vision Enhancement System is comprised of an integrated ophthalmic microsurgical system designed for use in anterior and posterior segment surgery including phacofragmentation and vitreous aspirating and cutting as well as endoillumination. Additionally, the Stellaris Elite Vision Enhancement System may be configured with a 532 nm laser module for photocoagulation.
The Stellaris Elite Vision Enhancement System is a rebranding of the currently cleared Stellaris (K133242) Vision Enhancement System and Stellaris PC (K133486) Vision Enhancement System. The system is based on the technology and the performance of the
{4}------------------------------------------------
existing Stellaris and Stellaris PC Vision Enhancement Systems, and this traditional 510(k) incorporates software and hardware revisions to support the introduction of the new Vitesse vitrectomy feature (including 2 new Vitesse handpieces and supporting accessories). The new Vitesse vitrectomy feature is available on the Stellaris Elite Vision Enhancement System based on the configuration matrix listed in Table 1. The various available system configurations allow for system feature flexibility and are made available for physician convenience.
| | Configuration by SKU | | |
|------------------------------------------------------|----------------------|-----------|-----------|
| Feature | BL11145 | BL14455 | BL15455 |
| Adaptive Fluidics<br>Feature cleared under K162342 | Available | Available | Available |
| 2500 cpm vitrectomy<br>Feature cleared under K162342 | Available | -- | -- |
| 7500 cpm vitrectomy<br>Feature cleared under K162342 | -- | Available | Available |
| Laser module<br>Feature cleared under K133486 | -- | -- | Available |
| Vitesse vitrectomy<br>[new feature] | -- | Available | Available |
Table 1: Stellaris Elite Vision Enhancement System Configuration Table
Abbreviations: cpm = cuts per minute
--: Not available
BL11145: formerly the Stellaris Anterior Vision Enhancement System
BL14455: formerly the Stellaris PC Vision Enhancement System
BL15455: formerly the Stellaris PC Vision Enhancement System with laser
#### Indications for Use 5.
The Bausch + Lomb Stellaris Elite Vision Enhancement System is intended for the emulsification and removal of cataracts, anterior and posterior segment vitrectomy. The system is designed for use in both anterior and posterior segment surgeries. It provides capabilities for phacoemulsification, coaxial and bimanual irrigation/aspiration, bipolar coagulation, vitrectomy, viscous fluid injection/removal, and air/fluid exchange operations. The Stellaris Elite Vision Enhancement System configured with the laser module is additionally intended for retinal photocoagulation and laser trabeculoplasty.
The above indication is a slightly reworded version of the already cleared indications for the Stellaris (K133242) and Stellaris PC (K133486) Vision Enhancement Systems. The rewording includes language regarding the rebranding of the product to Stellaris Elite Vision Enhancement System as well as clarification in regard to the laser module in the last sentence of the indications for use. The addition of the Vitesse vitrectomy feature did not further change the above indication for use statement.
{5}------------------------------------------------
#### 6. Summary of Technological Characteristics
The technological characteristics of the Stellaris Elite Vision Enhancement System are substantially equivalent to those of the predicate devices.
| Characteristic | Subject Device:<br>Stellaris Elite Vision<br>Enhancement System | K133486<br>Stellaris PC Vision<br>Enhancement System | K133242<br>Stellaris Vision<br>Enhancement System |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Anterior/Posterior<br>ophthalmic surgery | Same as subject device | Use in ophthalmic<br>procedures |
| Indications for Use | The Bausch + Lomb Stellaris<br>Elite Vision Enhancement<br>System is intended for the<br>emulsification and removal<br>of cataracts, anterior and<br>posterior segment vitrectomy.<br>The system is designed for<br>use in both anterior and<br>posterior segment surgeries.<br>It provides capabilities for<br>phacoemulsification, coaxial<br>and bimanual<br>irrigation/aspiration, bipolar<br>coagulation, vitrectomy,<br>viscous fluid<br>injection/removal and<br>air/fluid exchange operations.<br>The Stellaris Elite Vision<br>Enhancement System<br>configured with the laser<br>module is additionally<br>intended for retinal<br>photocoagulation and laser<br>trabeculoplasty. | The Bausch + Lomb<br>Stellaris PC Vision<br>Enhancement System is<br>intended for the<br>emulsification and removal<br>of cataracts, anterior and<br>posterior segment<br>vitrectomy. The system is<br>designed for use in both<br>anterior and posterior<br>segment surgeries. It<br>provides capabilities for<br>phacoemulsification,<br>coaxial and bimanual<br>irrigation/aspiration,<br>bipolar coagulation,<br>vitrectomy, viscous fluid<br>injection/removal and<br>air/fluid exchange<br>operations.<br>The laser modes are<br>intended for retinal<br>photocoagulation and laser<br>trabeculoplasty. | The Bausch + Lomb<br>Stellaris Vision<br>Enhancement System<br>device is intended for the<br>emulsification and removal<br>of cataracts, anterior and<br>posterior segment<br>vitrectomy. The system is<br>designed for use in both<br>anterior and posterior<br>segment surgeries. It<br>provides capabilities for<br>phacoemulsification,<br>irrigation/aspiration,<br>bipolar, coaxial, and<br>bipolar coagulation,<br>vitrectomy, viscous fluid<br>injection/removal and<br>air/fluid exchange<br>operations. Use only<br>Bausch + Lomb disposable<br>packs and handpieces<br>designated for use with the<br>system. |
| Laser capabilities | Yes | Same as subject device | No |
| User interface | LCD touch screen | Same as subject device | Same as subject device |
| Footswitch | Yes | Same as subject device | Same as subject device |
| Electrical<br>Characteristics | 90 – 130 VAC, 50/60 Hz<br>200 - 240 VAC, 50/60 Hz | Same as subject device | Same as subject device |
#### 7. Safety and Effectiveness Information
The review of the indications for use and technical characteristics provided demonstrates that the Stellaris Elite Vision Enhancement System is substantially equivalent to the predicate devices and is safe and effective for the emulsification and removal of cataracts, anterior and posterior segment vitrectomy.
{6}------------------------------------------------
#### 8. Brief Summary of Nonclinical Tests
Safety tests of the Stellaris Elite Vision Enhancement System have demonstrated its compliance with applicable requirements of the following electrical standards:
| IEC 60601-1:2005 + C1(2006) +<br>C2(2007) + AM1(2012)<br>or IEC 60601-1:2012 | Medical electrical equipment - Part 1: General requirements<br>for basic safety and essential performance |
|------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2 ed3.0 (2007) | Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential performance -<br>Collateral standard: Electromagnetic compatibility -<br>Requirements and tests |
| IEC 60601-1-6:2010 | Medical electrical equipment. Part 1-6: General<br>requirements for basic safety and essential performance.<br>Collateral standard: Usability |
| IEC 60601-2-2:2009 | Medical electrical equipment - Part 2-2: Particular<br>requirements for the basic safety and essential performance<br>of high frequency surgical equipment and high frequency<br>surgical accessories |
| IEC 60601-2-22:2007 | Medical electrical equipment - Part 2-22: Particular<br>requirements for the safety of diagnostic and therapeutic<br>laser equipment |
| IEC 80601-2-58:2008 | Medical electrical equipment - Part 2-58: Particular<br>requirements for the basic safety and essential performance<br>of lens removal devices and vitrectomy devices for<br>ophthalmic surgery |
Substantial equivalence is supported the successful tests for functional, simulated use, environmental and transport testing performed on representative units.
Software changes were verified and validated in accordance with the Bausch + Lomb software quality procedures which comply with EN ISO IEC 62304:2006 Medical device software -- Software life cycle processes.
The Stellaris Elite Vision Enhancement System passed all of the above referenced testing. This testing demonstrates that the functional requirements have been met and that the modified device is equivalent to the predicate devices.
{7}------------------------------------------------
#### 9. Conclusion
The Stellaris Elite Vision Enhancement System shares the same indications for use, design features, and functional features with, and thus is substantially equivalent to, the predicate devices. Non-clinical test results demonstrate that the Stellaris Elite Vision Enhancement System is substantially equivalent to the predicate devices and no new issues of safety or effectiveness have been raised.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.