K231132 · J & P Dental Technologies · NHA · Mar 26, 2024 · Dental
Device Facts
Record ID
K231132
Device Name
J & P Click Attachments
Applicant
J & P Dental Technologies
Product Code
NHA · Dental
Decision Date
Mar 26, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
J & P Click Attachments are designed to support fixed, partial or full arch restorations on endosseous dental implants in the mandible or maxilla for the purpose of restoring masticatory function. They are used in fixed hybrid restorations that can be attached with a click in system. Click attachments are indicated for use with vertical implant placements.
Device Story
J & P Click Attachments provide rigid connection for fixed, partial, or full arch hybrid restorations to endosseous dental implants. System components include abutments, attachment housings, and inserts. Abutments are manufactured from Ti-6AL-4V ELI. Device functions by providing a mechanical 'click-in' interface between the dental implant and the prosthetic restoration. Used in clinical dental settings by practitioners to restore masticatory function. Output is a stable, retrievable attachment for hybrid dentures. Benefits include improved prosthetic stability and ease of maintenance for fixed hybrid restorations.
Clinical Evidence
Bench testing only. Evidence includes reverse engineering tolerance analyses covering OEM implant bodies, abutments, and screws. Cytotoxicity testing performed per ISO 10993-5. Steam sterilization validated per ISO 17665-1. Non-clinical MRI safety assessment conducted using scientific rationale and literature to evaluate magnetically induced displacement force and torque.
Technological Characteristics
Materials: Ti-6AL-4V ELI (ASTM F136). Form factor: Abutments, attachment housings, and inserts with collar heights 1-6mm. Connectivity: Mechanical interface for dental implants. Sterilization: Non-sterile, user-sterilized via steam (ISO 17665-1).
Indications for Use
Indicated for patients requiring fixed, partial, or full arch restorations on endosseous dental implants in the mandible or maxilla to restore masticatory function. Compatible with specific implant systems from Biohorizons, Nobel Biocare, Implant Direct, Surgikor, Neodent, MIS, Zimmer, Hiossen, and SIN. Indicated for use with vertical implant placements.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Zest Anchors High Retention Attachment System (K220252)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
J & P Dental Technologies c/o Angela Blackwell Senior Consultant Blackwell Device Consulting P.O. Box 718 Gresham, Oregon 97030-0172
March 26, 2024
Re: K231132
Trade/Device Name: J & P Click Attachments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: February 27, 2024 Received: February 28, 2024
Dear Angela Blackwell:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
3
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231132
Device Name J & P Click Attachments
Indications for Use (Describe) J & P Click Attachments are designed to support fixed, partial or full arch restorations on endosseous dental implants in the mandible or maxilla for the purpose of restoring masticatory function. They are used in fixed hybrid restorations that can be attached with a click in system. Click attachments are indicated for use with vertical implant placements. They are indicated for the following implant systems: Biohorizons Tapered Tissue Level implants in diameters 3.8, 4.2, 4.6, 5.2 mm Nobel Biocare NobelActive including 3.5. 4.3, 5.0 mm diameter NobelActive, 3.75, 4.3, 5.0mm diameter NobelParallel and 3.5. 4.3. 5.0mm diameter NobelReplace Conical Connection Implants Implant Direct Legacy 3 for 3.7, 4.2, 4.7, 5.2mm diameter implants Surgikor Versatile for 3.5, 3.75, 4.2, 4.5, 5.0, 6.0mm diameter implants Surgikor Fixation for 3.5. 3.9. 4.3. 5.0mm diameter implants Surgikor Solution for 3.5, 4.0, 4.5, 5.0, 5.5, 6.0mm diameter implants Neodent Grand Morse for 3.5, 3.75, 4.0, 4.3 and 5.0 mm diameter implants MIS Seven for implant diameters 3.75, 4.2, 5, and 6mm Zimmer for Tapered Screw-Vent in 3.7, 4.1 and 4.7mm implant diameter Hiossen ETIII for 3.5, 4.0, 4.5, 5.0, 5.5, 6.0, and 7.0mm implant diameters
SIN Cone Morse 11.5° and 16° implant lines
11.5° Strong SW/SW Plus implant diameters 3.5. 3.8.4.5. 5.0mm Unitite implant diameters 3.5. 4.0.4.3. 5.0. 6.0 Tryon CM Conical implant diameters 3.5, 4.5, 5.0 Tryon CM Cylindrical implant diameters 3.5, 3.75, 4.0. 5.0 Epikut CM/CM Plus implant diameters 3.5, 3.8, 4.0. 4.5, 5.0
16 ° Strong SW CM Plus implant diameters 3.5, 3.8, 4.5 and 5.0mm Epikut S/S Plus implant diameters 3.5, 3.8. 4.0. 4.5. 5.0mm
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510k Summary K231132 March 25, 2024 J & P Click Attachments
- Name and address: J & P Dental Technologies 1527 W State Hwy 114 Ste 500-280 Grapevine, TX 76051 Contact Person: Dr. Aaron Paz Phone Number: (310) 877-9858 Name of device: J & P Click Attachments Classification Name: Endosseous dental implant abutments CFR: 21 CFR 872.3630 Primary Product Code: NHA Primary Predicate Device: Zest Anchors K220252 High Retention Attachment System Reference Predicates: Nobel Active K142260 Allied Dental Solutions K152845 Implant Direct K192221 Legacy 3 and Interactive SIN K221453 Epikut CM and K222231 Epikut S MIS K040807 Seven Biohorizons K071638 Tapered Tissue Level implant Neodent K203382 Neodent GM Implants Zimmer K160398 Tapered Screw Vent Hiossen K140934 ET III Surgikor K182615 Surgikor Dental Implant System
#### Submission Contact:
Angela Blackwell Blackwell Device Consulting P.O. Box 718 Gresham, OR 97030-0172 (704)450-9934 angela@blackwelldevice.com
Device Description: J & P Click Attachments provide a rigid connection of fixed, partial and full arch restorations (fixed/detachable hybrid dentures) to endosseous dental implants. They consist of abutments, attachment housings, and inserts. The abutments are provided in various OEM implant and abutment connections. The abutments are made from Ti-6AL-4V ELI which meets ASTM F136. All varieties of click attachments come in collar heights of 1, 2, 3, 4, 5 and 6mm. Abutment platform diameters include:
Biohorizons 3.5, and 4.5mm
{5}------------------------------------------------
Nobel Biocare Nobel Active 3.5 and 3.9mm (NobelParallel and NobelReplace Conical Connection are the same) Implant Direct Legacy 3.5 and 4.5mm Surgikor Versatile 3.5 and 4.5mm Surgikor Fixation and Solution 3.5 and 3.9mm Neodent Grand Morse 3.0mm MIS Seven 3.5 and 4.5mm Zimmer for Tapered ScrewVent 3.5 and 4.5mm Hiossen ETIII 3.35mm SIN 2.5mm for 11.5° cone morse and 2.72 mm for 16° cone morse
#### Indications for Use:
J & P Click Attachments are designed to support fixed, partial or full arch restorations on endosseous dental implants in the mandible or maxilla for the purpose of restoring masticatory function. They are used in fixed hybrid restorations that can be attached with a click in system. Click attachments are indicated for use with vertical implant placements.
They are indicated for the following implant systems:
Biohorizons Tapered Tissue Level implants in diameters 3.8, 4.2, 4.6, 5.2 mm
Nobel Biocare NobelActive including 3.5, 4.3, 5.0 mm diameter NobelActive,3.75, 4.3, 5.0mm diameter NobelParallel and 3.5, 4.3, 5.0mm diameter NobelReplace Conical Connection Implants
lmplant Direct Legacy 3 for 3.7, 4.2, 4.7, 5.2mm diameter implants
Surgikor Versatile for 3.5, 3.75, 4.2, 4.5, 5.0, 6.0mm diameter implants
Surgikor Fixation for 3.5, 3.9, 4.3, 5.0mm diameter implants
Surgikor Solution for 3.5, 4.0, 4.5, 5.0, 5.5, 6.0mm diameter implants
Neodent Grand Morse for 3.5, 3.75, 4.0, 4.3 and 5.0 mm diameter implants
MIS Seven for implant diameters 3.75, 4.2, 5, and 6mm
Zimmer for Tapered Screw-Vent in 3.7, 4.1 and 4.7mm implant diameter
Hiossen ETIII for 3.5, 4.0, 4.5, 5.0, 5.5, 6.0, and 7.0 mm implant diameters
SIN Cone Morse 11.5° and 16° implant lines
11.5° Strong SW/SW Plus implant diameters 3.5, 3.8, 4.5, 5.0mm Unitite implant diameters 3.5, 4.0,4.3, 5.0, 6.0 Tryon CM Conical implant diameters 3.5, 4.5, 5.0 Tryon CM Cylindrical implant diameters 3.5, 3.75, 4.0, 5.0 Epikut CM/CM Plus implant diameters 3.5, 3.8, 4.0, 4.5, 5.0
{6}------------------------------------------------
16 ° Strong SW CM/CM Plus implant diameters 3.5, 3.8, 4.5 and 5.0mm Epikut S/S Plus implant diameters 3.5, 3.8, 4.0, 4.5, 5.0mm
Testing Summary: Abutment steam sterilization was done according to ISO 17665-1. Cytotoxicity testing was conducted according to ISO 10993-5. Reverse engineering tolerance analyses were conducted for all OEM implant systems in the indications for use. These analyses covered the entire OEM implant system and included OEM implant body models, OEM abutment models and OEM abutment screw models.
### MR Environment Condition
Non-clinical worst-case MRI review was performed to evaluate the metallic J & P Click Attachment devices in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system including all variations (all compatible implant bodies, dental abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
# Substantial Equivalence:
| Click attachments are substantially equivalent to Zest High Retention Attachments abutments in | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| indications for use, materials, design, and implant system compatibility. | | |
| Abutment Comparison | L&P Click Attachments | Zest High Retention Attachments |
| Abutment Comparison<br>Table | J & P Click Attachments | Zest High Retention Attachments<br>K220252 |
| Sterilization of<br>abutments | Provided non-sterile with<br>instructions for user to sterilize<br>them | Provided non-sterile with instructions<br>for user to sterilize them |
| Abutments available<br>for these implant<br>systems | Nobel Active implant diameters<br>3.5, 4.3, and 5.00mm<br>NobelParallel implant diameters<br>3.75, 4.3, and 5.0mm<br>NobelReplace implant diameters<br>3.5, 4.3 and 5.0mm<br>Biohorizons Tapered Tissue Level<br>implant diameters 3.8, 4.6mm<br>Implant Direct Legacy 3 implant<br>diameters 3.7, 4.2, 4.7, 5.2mm<br>Surgikor Versatile for 3.5, 3.75, 4.2,<br>4.5, 5.0, 6.0mm diameter implants | Nobel Active implant diameters 3.5,<br>4.3, 5.5mm<br>Biohorizons Tapered Tissue Level<br>implant diameters 3.8, 4.6mm<br>Implant Direct Legacy 3 implant<br>diameters 3.7, 4.2, 4.7, 5.2, 5.7mm<br>Neodent Grand Morse implant<br>diameters 3.5, 3.75, 4.0, 4.3 and 5.0<br>mm<br>MIS Seven implant diameters 3.75,<br>4.2, 5.0, 6.0mm |
| Surgikor Fixation for 3.5, 3.9, 4.3, 5.0mm diameter implants<br>Surgikor Solution for 3.5, 4.0, 4.5, 5.0, 5.5, 6.0mm diameter implants<br>Neodent Grand Morse implant diameters 3.5, 3.75, 4.0, 4.3 and 5.0 mm<br>MIS Seven implant diameters 3.75, 4.2, 5, and 6mm<br>Zimmer Tapered Screw-Vent implant diameters 3.7, 4.1 and 4.7mm<br>Hiossen ETIII 3.5, 4.0, 4.5, 5.0, and 5.5mm implant diameters | Zimmer Tapered Screw-vent implant diameters 3.7, 4.1, 4.7 and 6.00mm<br>Hiossen ET III implant diameters 3.5, 4.0, 4.5, 5.0, 6.0, 7.0mm | |
| SIN Cone Morse 11.5° and 16° implant lines<br>11.5° Strong SW/SW Plus implant diameters 3.5, 3.8, 4.5, 5.0mm<br>Unitite implant diameters 3.5, 4.0,4.3, 5.0, 6.0 Tryon CM Conical implant diameters 3.5, 4.5, 5.0<br>Tryon CM Cylindrical implant diameters 3.5, 3.75, 4.0, 5.0 Epikut CM/CM Plus implant diameters 3.5, 3.8, 4.0, 4.5, 5.0<br>16 ° Strong SW CM/CM Plus implant diameters 3.5, 3.8, 4.5 and 5.0mm Epikut S/S Plus implant diameters 3.5, 3.8, 4.0, 4.5, 5.0mm | Neodent GM Interface with gingival heights of 1,2,3,4,5,6mm<br>Hiosen ETIII Mini and Regular Interfaces with gingival heights of 1,2,3,4,5,6mm<br>Biohorizons Tapered Pro 3.5 Interface/MIS Seven Standard Interface/Zimmer 3.5 Interface/Implant Direct Legacy 3.5 | Neodent GM Interface with gingival heights of 1,2,3,4,5,6mm<br>Hiosen ETIII Mini and Regular Interfaces with gingival heights of 1,2,3,4,5,6mm<br>Biohorizons Tapered Pro 3.5 Interface/MIS Seven Standard Interface/Zimmer 3.5 Interface/Implant Direct Legacy 3.5 |
| Abutments are available in the following platform and gingival height descriptions | | |
| | | |
| | Interface/Surgikor Hex Interface<br>with gingival heights of<br>1,2,3,4,5,6mm | Interface/Surgikor Hex Interface with<br>gingival heights of 0, 1, 2.5, 3.5, 4.5,<br>5.5, 6.5mm |
| | Biohorizons Tapered Pro Wide<br>interface/MIS Seven Wide<br>Interface/Zimmer 4.5<br>interface/Implant Direct Legacy<br>Wide Interface with gingival<br>heights of 1,2,3,4,5,6mm | Biohorizons Tapered Pro Wide<br>interface/MIS Seven Wide<br>Interface/Zimmer 4.5<br>interface/Implant Direct Legacy Wide<br>Interface with gingival heights of<br>0,1,2,3,4,5,6mm |
| | Nobel Active NP Interface/Surgikor<br>RC Interface with gingival heights<br>of 1,2,3,4,5,6mm | Nobel Active NP Interface/Surgikor<br>RC Interface with gingival heights of<br>1,2,3,4,5,6mm |
| | Nobel Active RP Interface/Surgikor<br>WC Interface with gingival heights<br>of 1,2,3,4,5,6mm | Nobel Active RP Interface/Surgikor<br>WC Interface with gingival heights of<br>1,2,3,4,5,6mm |
| | SIN 11.5° Cone Morse and 16°<br>Cone Morse with gingival heights<br>of 1,2,3,4,5,6mm | |
| Indications for Use | J & P Click Attachments are<br>designed to support fixed, partial<br>or full arch restorations on<br>endosseous dental implants in the<br>mandible or maxilla for the<br>purpose of restoring masticatory<br>function. They are used in fixed<br>hybrid restorations that can be<br>attached with a click in system.<br>Click attachments are indicated for<br>use with vertical implant<br>placements. | The High Retention Attachment<br>System is designed to support fixed,<br>partial or full arch restorations on<br>endosseous dental implants in<br>the mandible or maxilla for the<br>purpose of restoring masticatory<br>function. It is used in fixed hybrid<br>restorations that can be attached<br>with a snap-in system. |
| Material | Ti6Al4V | Ti6Al4V |
| Mechanism of action | Used with various dental implant<br>systems to make fixed hybrid<br>restorations | Used with various dental implant<br>systems to make fixed hybrid<br>restorations |
{7}------------------------------------------------
{8}------------------------------------------------
# Conclusion:
J & P Click Attachments are substantially equivalent to Zest High Retention Attachments. They both have the same indications for use (other than the vertical implant placement restriction), same mechanism of action, are of the same material, have the similar implant system compatibilities, and similar platform interfaces and gingival heights. Vertically placed implants are included in the indications for the predicate device so the indications for the subject device are a subset of those of the predicate. The
{9}------------------------------------------------
indications only being a subset of the predicate indications do not change the substantial equivalence because the restriction is there to make the devices safer and it does not change the other parts of the indications for use. Those interfaces not found in the predicate device are found in reference devices. The attachment designs are very similar between the predicate device and the current submission. Any differences in attachment design are slight differences in the dimensions which do not change the substantial equivalence of the two devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.