MRI Compatibility for Existing Zimmer Dental Implant Systems
K160398 · Zimmer Dental, Inc. · DZE · May 17, 2017 · Dental
Device Facts
Record ID
K160398
Device Name
MRI Compatibility for Existing Zimmer Dental Implant Systems
Applicant
Zimmer Dental, Inc.
Product Code
DZE · Dental
Decision Date
May 17, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Indications for Use
Tapered Screw-Vent Zimmer Dental Implant: Zimmer dental implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. Zimmer One-Piece (ZOP) Dental Implant: Zimmer One-Piece Implants are indicated for the support and retention of fixed single tooth and fixed partial denture restorations in the mandibular central incisor, and maxillary lateral incisor regions of partially edentulous jaws. The Zimmer One-Piece Implant must be splinted if two or more are used adjacent to each other. The Zimmer One-Piece Implant may be immediately restored with a temporary prosthesis that is not in functional occlusion. Screw-Vent Zimmer Dental Implant: Zimmer Dental implant systems are designed for use in endentulous mandibles or maxillae for attachment of complete denture prostheses for immediate or conventional loading, or as a terminal or intermediary abutment for fixed or removable bridgework, or as a free standing single tooth replacement. Advent Zimmer Dental Implant: Zimmer dental implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. SwissPlus Dental Implant: Zimmer Dental implant systems are designed for use in endentulous mandibles or maxillae for attachment of complete denture prostheses for immediate or conventional loading, or as a terminal or intermediary abutment for fixed or removable bridgework, or as a free standing single tooth replacement. Spline Twist Dental Implant: Zimmer Dental implant systems are designed for use in edentulous mandibles or maxillae for attachment of complete denture prostheses for immediate or conventional loading, or as a terminal or intermediary abutment for fixed or removable bridgework, or as a free standing single tooth replacement. Trabecular Metal Implant™ (Tapered Screw Vent® X Implant System): Zimmer Dental implant systems are (Tapered Screw Vent® X Implant System) designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. The 3.7mmD Trabecular Metal Implants should be splinted to additional implants when used in the pre- molar region and should not be used in the molar region. The 4.1mmD Trabecular Metal Implants should be splinted to additional implants when used in the molar region. Zimmer 3.1mmD Dental Implants: The 3.1mmD Eztetic Dental Implants are designed for use in the anterior maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. The 3.1mmD Eztetic Dental Implants may be placed immediately following an extraction or loss of natural teeth provided there is sufficient volume of alveolar bone to minimally support the implant (minimum 1mm circumferential and 2mm apical). The 3.1mmD Eztetic Dental Implants should be splinted to additional implants when used in the pre-molar region and should not be used in the molar region. Zimmer Dental Tapered Screw-Vent® T Implant, HA Coated & Zimmer Dental Tapered Screw-Vent® M Implant, HA Coated: Zimmer dental implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. Zimmer Dental Tapered Screw-Vent® T Implant & Zimmer Dental Tapered Screw-Vent® M Implant: Zimmer Dental implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. Zimmer Contour Hex-Lock Abutment, Straight: Abutment is used as a terminal or intermediate abutment for a cemented prosthesis. Abutment can be used for a single or multiple-unit restoration. Zimmer Contour Hex-Lock Abutment, 17°: Abutment is used as a terminal or intermediate abutment for a cemented prosthesis where the angle needs to be offset by 17°. Abutment can be used for a single or multiple-unit restoration. Zimmer Dental Inc. Hex-Lock Abutment: Abutment is used as a terminal or intermediate abutment for a cemented prosthesis. Abutment can be used for a single or multiple-unit restoration. Hex-Lock Short Abutments: Abutment is used as a terminal or intermediate abutment for a cemented prosthesis. Abutment can be used for a single or multiple-unit restoration. Zimmer Dental Inc. 20° Angled Abutment: Zimmer Dental Implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. "Cast-To" Gold Abutments, Engaging: Zimmer Dental Implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. Zimmer Cast-To Gold Abutment for the 3.1mmD Dental Implant System, Engaging: The Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System, Engaging is intended for use in the creation of a customized abutment for a cement-retained crown or bridge, or a customized screw-retained restoration. "Cast-To" Gold Abutments, Non-Engaging: Zimmer Dental implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. Zimmer Cast-To Gold Abutment for the 3.1mmD Dental Implant System. Non-Engaging: The Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System. Non-Engaging, is intended for use in the creation of customized multiple-unit restorations on implants (e.g., bars and bridges), when anti-rotation of the abutment is not necessary. 2.9mm Angled Abutment and the 2.9mm Angled Abutment, Straight Hex: The 2.9mm Angled Abutment and the 2.9mm Angled Abutment, Straight Hex are used for attachment of restorations requiring off-axis correction. The 2.9mm Angled Abutment and the 2.9mm Angled Abutment, Straight Hex are designed to be used in edentulous or partially edentulous mandibles or maxillae for attachment of complete denture prostheses, or as a terminal or intermediary abutment for fixed or removable bridgework, or as a freestanding single tooth replacement. The 2.9mm Contour Abutment and the 2.9mm Contour Abutment, Straight Hex are used as a terminal or intermediate abutment for a cemented prosthesis. Abutment can be used for a singleor multiple-unit restoration. The 2.9mm Angled Contour Abutment and the 2.9mm Angled Contour Abutment, Straight Hex are designed to be used as a terminal or intermediate abutment for a cemented prosthesis where the angle needs to be offset by 17°. Abutment can be used for a single- or multiple-unit restoration. The 2.9mm Temporary Abutment is intended to be used to fabricate and support provisional restorations that aid in creating an esthetic emergence through the gingiva during the healing period and prior to final restoration. The 2.9mm Temporary Abutment can be used for cementretained or screw-retained provisional restorations. The abutments can be used for single-unit and multi-unit restorations. Angled Tapered Abutments, 15°: The Angled Tapered Abutment is used for a terminal or intermediate abutment for screwretained multiple-unit restorations. The 30° Angled Tapered Abutment must be used within 45° of parallelism for a splinted restoration. Angled Tapered Abutments, 30°: The Angled Tapered Abutment is used for a terminal or intermediate abutment for screwretained multiple-unit restorations. The 30° Angled Tapered Abutment must be used within 45°of parallelism for a splinted restoration. The 15° Angled Tapered Abutment must be used within 30° of parallelism for a splinted restoration. Zimmer Dental Inc. Screw Vent Ball Abutment: Zimmer Dental implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. Spline Temporary Abutment: Temporary Abutment, Engaging: Spline engaging titanium alloy cylinder with retentive parallel walls and a flared cuff. Attaches to the Spline Implant with a separate screw. For use when a temporary prosthesis for up to six weeks is desired. Single or multiple units. Do not use as a permanent abutment. Not to exceed six weeks of placement. Temporary Abutment, Non-Engaging: Adjustable titanium alloy cylinder with retentive parallel walls and a cuff. Attaches to the Spline Implant with a separate screw. Does not engage the tines. For use when a temporary prosthesis for up to six weeks is desired. Single or multiple units. Spline Fixed Abutment: For use as a terminal or intermediate abutment for cemented prosthesis. Abutments must be parallel to within 7.5° for Spline Implant or be prepped to be parallel. The Spline Abutments can be used for a single tooth or splinted to other abutments. Single use. Spline Fixed Abutment, 17°: For use as a terminal or intermediate abutment for cemented prosthetics when long axis of abutment needs to be changed by 15°, 17° or 25°. Abutment can be used for single tooth or splinted restorations. Single use. Zimmer Zfx Titanium Abutment for NobelActive Implant System: The Zimmer Zfx Titanium Abutment for NobelActive Implant System is designed for use as a terminal or intermediate abutment for cement retained prostheses. The abutment can be used with NobelActive and NobelReplace Conical Connection implants with a Narrow Platform (NP) Ø 3.5mm, or Regular Platform (RP) Ø 3.9mm. Zimmer Zfx Abutment for NobelReplace Implant System: The Zimmer Zfx Abutment for NobelReplace Implant System is designed for use as a terminal or intermediate abutment for cement retained prostheses. The abutment can be used with NobelReplace, Replace Select and NobelSpeedyReplace implants with a Narrow Platform (NP) Ø3.5 mm, Regular Platform (RP) Ø 4.3 mm, WidePlatform (WP) Ø 5.0 mm or 6.0 Platform (6.0) Ø 6.0mm. Zimmer Zfx Titanium Abutment for Straumann Bone Level Implant System: The Zimmer Zfx Titanium Abutment for Straumann Bone Level Implant System is designed for use as a terminal or intermediate abutment for cement retained prostheses. The abutment can be used with Straumann Bone Level implants with a Narrow CrossFit Platform Connection (NC) Ø 3.3mm or Regular CrossFit Platform Connection (RC) Ø 4.1mm or 4.8mm. Zimmer Zfx Titanium Base Abutment for Tapered Screw Vent Implant System: The Zimmer Zfx Titanium Base Abutment is a combination of a pre-manufactured (stock) abutment and Zimmer Zfx Abutment Coping or Crown as part of a straight or angled two piece abutment. The combination of the titanium base stock abutment and the abutment coping is designed for use as a terminal or intermediate abutment for cement or screw retained prostheses. The two-piece abutment is used for a single -unit or multi-unit(bridge) restoration. The Zimmer Zfx Abutment Coping shall be manufactured by an approved Zimmer Dental milling facility. Zimmer Zfx Abutment for Biomet 3i Certain Implant System: Titanium Abutment for Biomet 3i Certain Implant System is designed for use as a terminal or intermediate abutment for cement retained prostheses. The abutment can be used with Biomet 3i internal connection implants with a 3.4mm, 4.1mm, 5.0mm, or 6.0mm Platform. Zimmer® Patient Specific Abutment, Internal Hex, Titanium: The Zimmer Patient Specific Abutment, Internal Hex, Titanium is used as a terminal or intermediate abutment for a cemented prosthesis. The abutment can be used for a single or multiple-unit restoration. Tapered Screw-Vent Healing Collar: The Healing Collar is used to assist in the forming of the soft tissue during healing before a final restoration is placed. The Healing Collar is for single use only. 2.9mm Healing Collar: The Healing Collar is used to assist in the forming of the soft tissue during healing before a final restoration is placed. The Healing Collar is for single use only. Zimmer Dental Surgical Cover Screw: The Healing Screw is used to seal the implant internal connection and separate it from the soft tissue which is sutured over the implant during healing. The Healing Screw and Surgical Cover Screw are for single use only. 2.9mm Healing Screw: The Healing Screw is used to seal the implant internal connection and separate it from the soft tissue which is sutured over the implant during healing. The Healing Screw is for single use only. Retaining Screw: The Retaining Screws are intended to be used for securing the temporary abutments, final abutments, and impression transfers to the implant analog. The long Retaining Screw is intended to be used with Temporary Abutments for fabrication of screw-retained provisional restorations and with Impression Transfers for direct impressions.
Device Story
Device consists of existing Zimmer Dental implant systems and associated abutments/components; updated to define MRI compatibility. Implants/abutments are placed in maxilla or mandible by dental surgeons/clinicians to support dental prostheses. Device functions as structural support for teeth. MRI compatibility labeling informs clinicians of safety during MRI procedures. Output is clinical information regarding MRI safety; affects decision-making for patients requiring MRI scans. Benefits include clear safety guidance for patients with existing dental implants.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Dental implants and abutments constructed from titanium/titanium alloys. Components are mechanical, non-powered, and intended for surgical implantation. MRI compatibility determined via bench testing.
Indications for Use
Indicated for patients requiring dental restoration in the maxilla or mandible, including edentulous or partially edentulous cases. Used for single tooth replacement, fixed partial dentures, or complete denture prostheses. Supports immediate or conventional loading protocols. Contraindications include insufficient bone volume or stability for immediate loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, layered on top of each other, creating a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 17, 2017
Zimmer Dental, Inc. Christina Boydston Regulatory Affairs Manager 1900 Aston Ave. Carlsbad, California 92008
Re: K160398
Trade/Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: April 14, 2017 Received: April 17, 2017
Dear Christina Boydston:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/7 description: The image shows the name "Michael J. Ryan -S" in a large, sans-serif font. The text is black and appears to be on a white background. The name is written in a clear and legible manner.
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a light blue circle with a stylized letter 'Z' inside. Below the circle, the word 'zimmer' is written in a similar light blue color, with the word 'dental' underneath in a smaller font.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Tapered Screw-Vent Zimmer Dental Implant
Zimmer dental implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a stylized letter 'Z' in a light blue color, positioned above the words 'zimmer' and 'dental'. The word 'zimmer' is in a similar light blue color, while 'dental' is in a smaller font size and placed directly below 'zimmer'.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer One-Piece (ZOP) Dental Implant
Zimmer One-Piece Implants are indicated for the support and retention of fixed single tooth and fixed partial denture restorations in the mandibular central incisor, and maxillary lateral incisor regions of partially edentulous jaws. The Zimmer One-Piece Implant must be splinted if two or more are used adjacent to each other. The Zimmer One-Piece Implant may be immediately restored with a temporary prosthesis that is not in functional occlusion
× Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a stylized letter "Z" in a light blue color, positioned above the word "zimmer" in a smaller, sans-serif font, also in light blue. Below "zimmer" is the word "dental" in a smaller, italicized font, underlined with a thin line.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Screw-Vent Zimmer Dental Implant
Zimmer Dental implant systems are designed for use in endentulous mandibles or maxillae for attachment of complete denture prostheses for immediate or conventional loading, or as a terminal or intermediary abutment for fixed or removable bridgework, or as a free standing single tooth replacement.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a light blue circle with a stylized letter 'Z' inside. Below the circle, the word 'zimmer' is written in a similar light blue color, with the word 'dental' underneath in a smaller font.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Advent Zimmer Dental Implant
Zimmer dental implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a light blue circle with a stylized "Z" inside. Below the circle, the word "zimmer" is written in a sans-serif font, also in light blue. Underneath "zimmer", the word "dental" is written in a smaller font size and is underlined.
510(k) Number (if known): _ K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
SwissPlus Dental Implant
Zimmer Dental implant systems are designed for use in endentulous mandibles or maxillae for attachment of complete denture prostheses for immediate or conventional loading, or as a terminal or intermediary abutment for fixed or removable bridgework, or as a free standing single tooth replacement.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{7}------------------------------------------------
Image /page/7/Picture/2 description: The image contains the logo for Zimmer Dental. The logo features a light blue circle with a stylized letter 'Z' inside. Below the circle, the word 'zimmer' is written in a similar light blue color, with the word 'dental' appearing in a smaller font size directly underneath.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Spline Twist Dental Implant
Zimmer Dental implant systems are designed for use in edentulous mandibles or maxillae for attachment of complete denture prostheses for immediate or conventional loading, or as a terminal or intermediary abutment for fixed or removable bridgework, or as a free standing single tooth replacement.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{8}------------------------------------------------
Image /page/8/Picture/2 description: The image contains the logo for Zimmer Dental. The logo features a large, stylized letter 'Z' in a light blue color, positioned above the words 'zimmer' and 'dental'. The word 'zimmer' is in a similar light blue color and is placed above the word 'dental', which is in a smaller font size and underlined with a thin line.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Trabecular Metal Implant™ (Tapered Screw Vent® X Implant System)
Zimmer Dental implant systems are (Tapered Screw Vent® X Implant System) designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.
The 3.7mmD Trabecular Metal Implants should be splinted to additional implants when used in the pre- molar region and should not be used in the molar region.
The 4.1mmD Trabecular Metal Implants should be splinted to additional implants when used in the molar region.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{9}------------------------------------------------
Image /page/9/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a blue letter Z inside of a circle. Below the circle are the words "zimmer" and "dental" in a smaller font size.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer 3.1mmD Dental Implants
The 3.1mmD Eztetic Dental Implants are designed for use in the anterior maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. The 3.1mmD Eztetic Dental Implants may be placed immediately following an extraction or loss of natural teeth provided there is sufficient volume of alveolar bone to minimally support the implant (minimum 1mm circumferential and 2mm apical).
The 3.1mmD Eztetic Dental Implants should be splinted to additional implants when used in the pre-molar region and should not be used in the molar region.
× AND/OR Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{10}------------------------------------------------
Image /page/10/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a light blue circle with a stylized letter 'Z' inside. Below the circle, the word 'zimmer' is written in a similar light blue color, with the word 'dental' underneath in a smaller font.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer Dental Tapered Screw-Vent® T Implant, HA Coated & Zimmer Dental Tapered Screw-Vent® M Implant, HA Coated
Zimmer dental implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{11}------------------------------------------------
Image /page/11/Picture/2 description: The image contains the logo for Zimmer Dental. The logo features a stylized letter "Z" in a light blue color, positioned above the words "zimmer dental." The word "zimmer" is in a sans-serif font, and the word "dental" is in a smaller font size and placed directly below "zimmer."
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer Dental Tapered Screw-Vent® T Implant & Zimmer Dental Tapered Screw-Vent® M Implant
Zimmer Dental implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{12}------------------------------------------------
Image /page/12/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a stylized letter 'Z' in a light blue circle. Below the circle, the word 'zimmer' is written in lowercase, also in light blue, and below that, the word 'dental' is written in a smaller font size and gray color.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer Contour Hex-Lock Abutment, Straight
Abutment is used as a terminal or intermediate abutment for a cemented prosthesis. Abutment can be used for a single or multiple-unit restoration.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{13}------------------------------------------------
Image /page/13/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a stylized letter "Z" in a light blue color, positioned above the words "zimmer" and "dental," also in light blue. The word "zimmer" is in a larger, bolder font compared to "dental," which is placed directly underneath it.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer Contour Hex-Lock Abutment, 17°
Abutment is used as a terminal or intermediate abutment for a cemented prosthesis where the angle needs to be offset by 17°. Abutment can be used for a single or multiple-unit restoration.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{14}------------------------------------------------
Image /page/14/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a stylized letter "Z" in a light blue color, positioned above the words "zimmer" and "dental," also in light blue. The word "zimmer" is in a larger, bolder font compared to "dental," which is placed directly underneath it.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer Dental Inc. Hex-Lock Abutment
Abutment is used as a terminal or intermediate abutment for a cemented prosthesis. Abutment can be used for a single or multiple-unit restoration.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{15}------------------------------------------------
Image /page/15/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a stylized letter "Z" in a light blue color, positioned above the words "zimmer" and "dental," also in light blue. The word "zimmer" is in a larger, bolder font compared to "dental," which is placed directly underneath it.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Hex-Lock Short Abutments
Abutment is used as a terminal or intermediate abutment for a cemented prosthesis. Abutment can be used for a single or multiple-unit restoration.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{16}------------------------------------------------
Image /page/16/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a stylized letter 'Z' in a light blue color, positioned above the words 'zimmer' and 'dental'. The word 'zimmer' is in a similar light blue color, while 'dental' is in a smaller font size and placed directly below 'zimmer'.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer Dental Inc. 20° Angled Abutment
Zimmer Dental Implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.
X Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{17}------------------------------------------------
Image /page/17/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a stylized letter 'Z' in a light blue color, positioned above the words 'zimmer' and 'dental'. The word 'zimmer' is in a similar light blue color, while 'dental' is in a smaller font size and placed directly below 'zimmer'.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
"Cast-To" Gold Abutments, Engaging
Zimmer Dental Implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{18}------------------------------------------------
Image /page/18/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a light blue circle with a stylized letter 'Z' inside. Below the circle, the word 'zimmer' is written in a bold, sans-serif font, and below that, the word 'dental' is written in a smaller, lighter font.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer Cast-To Gold Abutment for the 3.1mmD Dental Implant System, Engaging
The Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System, Engaging is intended for use in the creation of a customized abutment for a cement-retained crown or bridge, or a customized screw-retained restoration.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use_ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{19}------------------------------------------------
Image /page/19/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a light blue circle with a stylized "Z" inside. Below the circle, the word "zimmer" is written in a lowercase, sans-serif font, also in light blue. Underneath "zimmer", the word "dental" is written in a smaller font size and is underlined.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
"Cast-To" Gold Abutments, Non-Engaging
Zimmer Dental implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{20}------------------------------------------------
Image /page/20/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a large, light blue letter Z above the words "zimmer" and "dental", which are stacked on top of each other. The word "zimmer" is in a larger, bolder font than the word "dental".
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer Cast-To Gold Abutment for the 3.1mmD Dental Implant System. Non-Engaging
The Zimmer Cast-to Gold Abutments for the 3.1mmD Dental Implant System. Non-Engaging, is intended for use in the creation of customized multiple-unit restorations on implants (e.g., bars and bridges), when anti-rotation of the abutment is not necessary.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{21}------------------------------------------------
Image /page/21/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a large, stylized letter "Z" in a light blue color. Below the "Z", the word "zimmer" is written in a smaller, lowercase font, also in light blue. Underneath "zimmer", the word "dental" is written in an even smaller font, with a line above it.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
2.9mm Angled Abutment and the 2.9mm Angled Abutment, Straight Hex
The 2.9mm Angled Abutment and the 2.9mm Angled Abutment, Straight Hex are used for attachment of restorations requiring off-axis correction. The 2.9mm Angled Abutment and the 2.9mm Angled Abutment, Straight Hex are designed to be used in edentulous or partially edentulous mandibles or maxillae for attachment of complete denture prostheses, or as a terminal or intermediary abutment for fixed or removable bridgework, or as a freestanding single tooth replacement.
The 2.9mm Contour Abutment and the 2.9mm Contour Abutment, Straight Hex are used as a terminal or intermediate abutment for a cemented prosthesis. Abutment can be used for a singleor multiple-unit restoration. The 2.9mm Angled Contour Abutment and the 2.9mm Angled Contour Abutment, Straight Hex are designed to be used as a terminal or intermediate abutment for a cemented prosthesis where the angle needs to be offset by 17°. Abutment can be used for a single- or multiple-unit restoration.
The 2.9mm Temporary Abutment is intended to be used to fabricate and support provisional restorations that aid in creating an esthetic emergence through the gingiva during the healing period and prior to final restoration. The 2.9mm Temporary Abutment can be used for cementretained or screw-retained provisional restorations. The abutments can be used for single-unit and multi-unit restorations.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{22}------------------------------------------------
Image /page/22/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a stylized letter 'Z' in a light blue circle. Below the circle, the word 'zimmer' is written in lowercase, also in light blue, and below that, the word 'dental' is written in a smaller font size and gray color.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Angled Tapered Abutments, 15°
The Angled Tapered Abutment is used for a terminal or intermediate abutment for screwretained multiple-unit restorations. The 30° Angled Tapered Abutment must be used within 45° of parallelism for a splinted restoration.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{23}------------------------------------------------
Image /page/23/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a stylized letter 'Z' in a light blue color, positioned above the words 'zimmer' and 'dental'. The word 'zimmer' is in a similar light blue color, while 'dental' is in a smaller font size and placed directly below 'zimmer'.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Angled Tapered Abutments, 30°
The Angled Tapered Abutment is used for a terminal or intermediate abutment for screwretained multiple-unit restorations. The 30° Angled Tapered Abutment must be used within 45°of parallelism for a splinted restoration. The 15° Angled Tapered Abutment must be used within 30° of parallelism for a splinted restoration.
× Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{24}------------------------------------------------
Image /page/24/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a stylized letter "Z" in a light blue color, positioned above the words "zimmer" and "dental." The word "zimmer" is in a smaller, sans-serif font, and the word "dental" is in an even smaller font and is underlined.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer Dental Inc. Screw Vent Ball Abutment
Zimmer Dental implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.
× Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{25}------------------------------------------------
Image /page/25/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a blue circle with a stylized letter "Z" inside. Below the circle, the word "zimmer" is written in a sans-serif font, and below that, the word "dental" is written in a smaller, thinner font.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Spline Temporary Abutment
Temporary Abutment, Engaging: Spline engaging titanium alloy cylinder with retentive parallel walls and a flared cuff. Attaches to the Spline Implant with a separate screw. For use when a temporary prosthesis for up to six weeks is desired. Single or multiple units. Do not use as a permanent abutment. Not to exceed six weeks of placement.
Temporary Abutment, Non-Engaging: Adjustable titanium alloy cylinder with retentive parallel walls and a cuff. Attaches to the Spline Implant with a separate screw. Does not engage the tines. For use when a temporary prosthesis for up to six weeks is desired. Single or multiple units.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{26}------------------------------------------------
Image /page/26/Picture/2 description: The image contains the logo for Zimmer Dental. The logo features a light blue circle with a stylized letter 'Z' inside. Below the circle, the word 'zimmer' is written in a similar light blue color, with the word 'dental' appearing in a smaller font size directly underneath.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Spline Fixed Abutment
For use as a terminal or intermediate abutment for cemented prosthesis. Abutments must be parallel to within 7.5° for Spline Implant or be prepped to be parallel. The Spline Abutments can be used for a single tooth or splinted to other abutments. Single use.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{27}------------------------------------------------
Image /page/27/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a light blue circle with a stylized "Z" inside. Below the circle, the word "zimmer" is written in a sans-serif font, also in light blue. Underneath "zimmer", the word "dental" is written in a smaller font size and is underlined.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Spline Fixed Abutment, 17°
For use as a terminal or intermediate abutment for cemented prosthetics when long axis of abutment needs to be changed by 15°, 17° or 25°. Abutment can be used for single tooth or splinted restorations. Single use.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{28}------------------------------------------------
Image /page/28/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a stylized letter Z inside of a circle. Below the circle, the word "zimmer" is written in a lowercase sans-serif font, and below that, the word "dental" is written in a smaller font.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer Zfx Titanium Abutment for NobelActive Implant System
The Zimmer Zfx Titanium Abutment for NobelActive Implant System is designed for use as a terminal or intermediate abutment for cement retained prostheses. The abutment can be used with NobelActive and NobelReplace Conical Connection implants with a Narrow Platform (NP) Ø 3.5mm, or Regular Platform (RP) Ø 3.9mm.
X Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use_ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{29}------------------------------------------------
Image /page/29/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a light blue circle with a stylized letter 'Z' inside. Below the circle, the word 'zimmer' is written in a similar light blue color, with the word 'dental' underneath in a smaller font.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer Zfx Abutment for NobelReplace Implant System
The Zimmer Zfx Abutment for NobelReplace Implant System is designed for use as a terminal or intermediate abutment for cement retained prostheses. The abutment can be used with NobelReplace, Replace Select and NobelSpeedyReplace implants with a Narrow Platform (NP) Ø3.5 mm, Regular Platform (RP) Ø 4.3 mm, WidePlatform (WP) Ø 5.0 mm or 6.0 Platform (6.0) Ø 6.0mm.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{30}------------------------------------------------
Image /page/30/Picture/2 description: The image contains the logo for Zimmer Dental. The logo features a light blue circle with a stylized "Z" inside. Below the circle, the word "zimmer" is written in a sans-serif font, with the word "dental" written in a smaller font size directly underneath and separated by a horizontal line.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer Zfx Titanium Abutment for Straumann Bone Level Implant System
The Zimmer Zfx Titanium Abutment for Straumann Bone Level Implant System is designed for use as a terminal or intermediate abutment for cement retained prostheses. The abutment can be used with Straumann Bone Level implants with a Narrow CrossFit Platform Connection (NC) Ø 3.3mm or Regular CrossFit Platform Connection (RC) Ø 4.1mm or 4.8mm.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{31}------------------------------------------------
Image /page/31/Picture/2 description: The image contains the logo for Zimmer Dental. The logo features a light blue circle with a stylized letter 'Z' inside. Below the circle, the word 'zimmer' is written in a lowercase, sans-serif font, and below that, the word 'dental' is written in a smaller, sans-serif font with a line above it.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer Zfx Titanium Base Abutment for Tapered Screw Vent Implant System
The Zimmer Zfx Titanium Base Abutment is a combination of a pre-manufactured (stock) abutment and Zimmer Zfx Abutment Coping or Crown as part of a straight or angled two piece abutment. The combination of the titanium base stock abutment and the abutment coping is designed for use as a terminal or intermediate abutment for cement or screw retained prostheses. The two-piece abutment is used for a single -unit or multi-unit(bridge) restoration. The Zimmer Zfx Abutment Coping shall be manufactured by an approved Zimmer Dental milling facility.
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{32}------------------------------------------------
Image /page/32/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a blue circle with a white "Z" inside. Below the circle, the word "zimmer" is written in blue, with the word "dental" written in a smaller font size below it. The logo is simple and clean, with a focus on the company name.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer Zfx Abutment for Biomet 3i Certain Implant System
Titanium Abutment for Biomet 3i Certain Implant System is designed for use as a terminal or intermediate abutment for cement retained prostheses. The abutment can be used with Biomet 3i internal connection implants with a 3.4mm, 4.1mm, 5.0mm, or 6.0mm Platform.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{33}------------------------------------------------
Image /page/33/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a blue circle with a stylized letter "Z" inside. Below the circle, the word "zimmer" is written in a sans-serif font, and below that, the word "dental" is written in a smaller, thinner font.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer® Patient Specific Abutment, Internal Hex, Titanium
The Zimmer Patient Specific Abutment, Internal Hex, Titanium is used as a terminal or intermediate abutment for a cemented prosthesis. The abutment can be used for a single or multiple-unit restoration.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{34}------------------------------------------------
Image /page/34/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a stylized letter 'Z' in a light blue circle. Below the circle, the word 'zimmer' is written in lowercase, also in light blue, and below that, the word 'dental' is written in a smaller font size and gray color.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Tapered Screw-Vent Healing Collar
The Healing Collar is used to assist in the forming of the soft tissue during healing before a final restoration is placed. The Healing Collar is for single use only.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{35}------------------------------------------------
Image /page/35/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a light blue circle with a stylized letter "Z" inside. Below the circle, the word "zimmer" is written in a lowercase, sans-serif font, and below that, the word "dental" is written in a smaller, sans-serif font.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
2.9mm Healing Collar
The Healing Collar is used to assist in the forming of the soft tissue during healing before a final restoration is placed. The Healing Collar is for single use only.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{36}------------------------------------------------
Image /page/36/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a stylized letter "Z" in a light blue color, positioned above the word "zimmer" in a smaller, sans-serif font, also in light blue. Below "zimmer" is the word "dental" in a smaller, italicized font, underlined with a thin line.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Zimmer Dental Surgical Cover Screw
The Healing Screw is used to seal the implant internal connection and separate it from the soft tissue which is sutured over the implant during healing. The Healing Screw and Surgical Cover Screw are for single use only.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{37}------------------------------------------------
Image /page/37/Picture/2 description: The image shows the logo for Zimmer Dental. The logo features a stylized letter "Z" in a light blue color, positioned above the words "zimmer" and "dental," also in light blue. The word "zimmer" is larger and bolder than the word "dental," which is placed directly below it.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
2.9mm Healing Screw
The Healing Screw is used to seal the implant internal connection and separate it from the soft tissue which is sutured over the implant during healing. The Healing Screw is for single use only.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{38}------------------------------------------------
Image /page/38/Picture/2 description: The image contains the logo for Zimmer Dental. The logo features a light blue circle with a stylized letter 'Z' inside. Below the circle, the word 'zimmer' is written in a similar light blue color, with the word 'dental' appearing in a smaller font size directly underneath.
510(k) Number (if known): K160398
Device Name: MRI Compatibility for Existing Zimmer Dental Implant Systems
Indications For Use:
Retaining Screw
The Retaining Screws are intended to be used for securing the temporary abutments, final abutments, and impression transfers to the implant analog. The long Retaining Screw is intended to be used with Temporary Abutments for fabrication of screw-retained provisional restorations and with Impression Transfers for direct impressions.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.