Legacy2, simplyLegacy2, Legacy3, simplyLegacy3, and Legacy4 dental implants are two-piece implants for one-stage or two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework. Implants can be indicated for immediate loading when good primary stability has been achieved and with appropriate occlusal loading. Narrow (3.2mmD) implants: Indicated for single-tooth replacement (mandibular central and lateral incisors; maxillary lateral incisors), multiple-tooth replacements or denture stabilization. Short (<10mm) 3.7mm implants: Indicated for single-tooth (mandibular and maxillary central and lateral incisors), multiple tooth replacements or denture stabilization. The Legacy 2, Legacy3, and Legacy4 fixture-mounts are intended for use with the corresponding dental implants (Legacy2, Legacy3, and Legacy4, respectively). The fixture-mounts can function as an abutment. As an abutment, fixture-mounts are intended for use with dental implants in the maxillary and/or mandibular arches to provide support for crowns or bridges for edentulous or partially edentulous patients. Fixture-mounts as an abutment for narrow (3.2mmD) implants: Indicated for single-tooth replacement of mandibular central and lateral incisors and maxillary lateral incisors. Fixture-mounts as an abutment for short (8mm) 3.7mmD implants: Indicated for tooth replacement of mandibular and maxillary central and lateral incisors.
Device Story
Threaded, root-form dental implants and fixture-mounts; used in one-stage or two-stage surgical procedures for edentulous/partially edentulous patients. Implants serve as anchors for single/multiple-unit restorations or fixed bridgework; fixture-mounts function as carriers for placement, impression transfers, and abutments. Operated by dental surgeons/clinicians in clinical settings. Mechanical load-bearing principle; implants provide stability for prosthetic restorations. Benefits include restoration of chewing function and aesthetics. Subject devices are titanium alloy; supplied sterile (implants) or require sterilization (fixture-mounts used as abutments).
Clinical Evidence
No clinical or animal studies performed. Bench testing included fatigue testing (ISO 14801:2016), biocompatibility (ISO 10993 series), surface area/bone-to-implant contact analysis, pull-out testing, insertion torque testing, sterilization validation (ISO 11137-2, ISO 17665), endotoxin testing (ANSI/AAMI ST72), shelf-life validation, and distribution testing (ASTM D4169). Results demonstrate safety and effectiveness equivalent to predicate.
Technological Characteristics
Titanium 6AL4V ELI alloy; threaded root-form design; internal hex connection. Surface modifications: Soluble Blast Media (SBM) or Hydroxyapatite (HA) per ASTM F1185. Dimensions: 3.2mm-7.0mm diameter; 6mm-16mm length. Single-use. Sterilization: Gamma irradiation (implants) or steam sterilization (fixture-mounts/abutments).
Indications for Use
Indicated for partially/fully edentulous patients requiring single/multiple-unit restorations or fixed bridgework. Narrow (3.2mmD) implants/abutments: single-tooth replacement (mandibular central/lateral incisors, maxillary lateral incisors), multiple-tooth replacement, or denture stabilization. Short (<10mm) 3.7mm implants/abutments: single-tooth replacement (mandibular/maxillary central/lateral incisors), multiple-tooth replacement, or denture stabilization. Suitable for immediate loading with sufficient primary stability.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Custom Legacy and Custom InterActive Titanium Abutments (K192218)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right, the FDA logo is displayed in blue, with the words "U.S. FOOD & DRUG" above the word "ADMINISTRATION".
June 25, 2020
Implant Direct Sybron Manufacturing, LLC % Kelliann Payne Partner Hogan & Lovells US LPP 1735 Market Street Floor 23 Philadelphia, Pennsylvania 19103
Re: K192221
Trade/Device Name: Legacy2, Legacy4, simplyLegacy2, simplyLegacy3 dental implants; Legacy2, Legacy3, Legacy4 fixture-mounts Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: May 27, 2020 Received: May 27, 2020
Dear Kelliann Payne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
542 of the Act); 21 CFR 1000-1050.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K192221
### Device Name
Legacy2, Legacy4, simplyLeaacy2, simplyLegacy3 dental implants; Legacy2, Legacy4 fixture-mounts Indications for Use (Describe)
Legacy2, simplyLegacy2, Legacy3, simplyLegacy3, and Legacy4 dental implants for one-stage of two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework.
Implants can be indicated for immediate loading when stability has been achieved and with appropriate occlusal loading.
- Narrow (3.2mmD) implants: Indicated for single-tooth replacement (mandibular central and lateral incisors), multiple-tooth replacements or denture stabilization.
- . Short (<10mm) 3.7mm implants: Indicated for single-tooth (mandibular and maxillary central and lateral incisors), multiple tooth replacements or denture stabilization.
The Legacy 2. Legacy3, and Legacy4 fixture-mounts are intended for use with the corresponding dental mplants (Legacy3. and Legacy4. respectively). The fixture-mounts can furction as an abutment. fixture-mounts are intended for use with dental implants in the maxillary and/or mandbular arches to provide support for edentulous or partially edentulous patients.
- Fixture-mounts as an abutment for narrow (3.2mmD) implants: Indicated for single-tooth replacement of mandibular central and lateral incisors and maxillary lateral incisors.
- Fixture-mounts as an abutment for short (8mm) 3.7mmD implants: Indicated for tooth replacement of mandibular and ● maxillary central and lateral incisors.
Legacy2, simplyLegacy2, Legacy3, simplyLegacy3, and Leqacy4 implants are compatible with the following abutments.
| Manufacturer | Abutment Line | Platform Diameter |
|----------------|---------------|----------------------------|
| Implant Direct | Legacy | 3.0mm, 3.5mm, 4.5mm, 5.7mm |
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
## Implant Direct LLC's
## Legacy2, Legacy3, Legacy4, simplyLegacy2, and simplyLegacy3 Dental Implants; Legacy2, Legacy3, and Leqacy4 Fixture-mounts
| Submitter:<br>Address: | Implant Direct Sybron Manufacturing LLC<br>3050 East Hillcrest Drive<br>Thousand Oaks, CA 91362 | | | |
|------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| Phone:<br>Contact Person: | 818-444-3306<br>Reina Choi, Regulatory Affairs Manager | | | |
| Date Prepared: | June 25, 2020 | | | |
| Name of Device: | Legacy2, Legacy3, Legacy4, simplyLegacy2, simplyLegacy3 dental<br>implants; Legacy2, Legacy3, Legacy4 fixture-mounts | | | |
| Common or Usual Name:<br>Classification Name:<br>Regulatory Class:<br>Primary Product Code:<br>Secondary Product Code: | Endosseous dental implants<br>21 C.F.R. § 872.3640 (Endosseous dental implants);<br>Class II<br>DZE<br>NHA | | | |
| Predicate Device: | Implant Direct LLC's Spectra-System Dental Implants 2008 (now<br>Legacy3 Dental Implant) (K090234) | | | |
| Reference Devices: | lmplant Direct LLC's Legacy3 Implants (6mm Length) (K131097);<br>Implant Direct LLC's Legacy Abutment System (K060063);<br>Implant Direct LLC's Custom Legacy and Custom InterActive<br>Titanium Abutments (K192218) | | | |
## Device Description
The Legacy2, Legacy4, and simplyLegacy2 implants have implant body and are supplied in similar dimensions. This 510(k) notice includes the Legacy3 dimensions previously cleared in K090234 and K131097; the Legacy3 (7.0mmD) is the same as the previously cleared Legacy3, but in a wider diameter.
The top approximate one-third of the implant body is straight, and the lower approximate two-thirds is tapered with progressively deeper buttress-threads. The Legacy3 implant body also features progressively deeper buttress threads, with quadruple-lead micro-threads at the coronal aspect; this implant body is evenly tapered. Each implant features a color-coded internal hex with a lead-in bevel. The simplyLegacy2 and simplyLegacy3 implants are identical to the Legacy3 implants, respectively, except that they are supplied without the corresponding fixturemount/abutment.
The table below outlines the body diameters, platform diameters, and lengths in which each Legacy implant model is available.
{4}------------------------------------------------
# Dimensions of Subject Implants
| Subject Implants | Body<br>Diameter | Platform<br>Diameter | Implant Length |
|---------------------------------------------------------------------|------------------|----------------------|----------------------------------------------|
| Legacy2,<br>simplyLegacy2,<br>Legacy3,<br>simplyLegacy3,<br>Legacy4 | 3.2mmD | 3.0mmD | 8, 10, 11.5, 13, 16mm |
| | 3.7mmD | 3.5mmD | 8, 10, 11.5, 13, 16mm |
| | 4.2mmD | 3.5mmD | 8, 10, 11.5, 13, 16mm |
| | 4.7mmD | 4.5mmD | 6 (only simplyLegacy), 8, 10, 11.5, 13, 16mm |
| | 5.2mmD | 4.5mmD | 6 (only simplyLegacy), 8, 10, 11.5, 13, 16mm |
| Legacy2,<br>simplyLegacy2,<br>Legacy3,<br>simplyLegacy3,<br>Legacy4 | 5.7mmD | 5.7mmD | 6 (only simplyLegacy), 8, 10, 11.5, 13, 16mm |
| Legacy2,<br>simplyLegacy2,<br>Legacy3,<br>simplyLegacy3,<br>Legacy4 | 7.0mmD | 5.7mmD | 6 (only simplyLegacy), 8, 10, 11.5, 13mm |
The Legacy2 and Legacy3 implants have a one-piece fixture-mount, and the Legacy4 implant has a two-piece fixture-mount. The Legacy fixture-mounts serve (1) as a transfer, to carry the implant to the ostectory and help with placement if desired; (2) to take an impression after implant placement, to guide fabrication of the patient-specific dental restoration; and (3) as an abutment.
All of the subject devices are made of Titanium alloy, are single-use, and come into contact with patient tissue/bone for a long-term or permanent duration. The implants are supplied with one of two surface modifications/coatings:
- . Soluble Blast Media (SBM);
- Hydroxyapatite (HA) per ASTM F1185. ●
Legacy2, Legacy3, and Legacy4 implants are available with either surface modification/coating; simplyLegacy2 and simplyLegacy3 are only available with SBM.
The devices are supplied sterile; the fixture-mounts, if modified for use as abutments, must also be sterilized by the end user prior to intra-oral use.
# Intended Use / Indications for Use
Legacy2, simplyLegacy2, Legacy3, simplyLegacy3, and Legacy4 dental implants are two-piece implants for one-stage or two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework.
Implants can be indicated for immediate loading when good primary stability has been achieved and with appropriate occlusal loading.
- . Narrow (3.2mmD) implants: Indicated for single-tooth replacement (mandibular central and lateral incisors; maxillary lateral incisors), multiple-tooth replacements or denture stabilization.
- Short (<10mm) 3.7mm implants: Indicated for single-tooth (mandibular and maxillary central and lateral incisors), multiple tooth replacements or denture stabilization.
The Legacy 2, Legacy3, and Legacy4 fixture-mounts are intended for use with the corresponding dental implants (Legacy2, Legacy3, and Legacy4, respectively). The fixture-mounts can function as an abutment. As an abutment, fixture-mounts are intended for use with dental implants in the maxillary and/or mandibular arches to provide support for crowns or bridges for edentulous or partially edentulous patients.
- . Fixture-mounts as an abutment for narrow (3.2mmD) Implants: Indicated for single-tooth
{5}------------------------------------------------
replacement of mandibular central and lateral incisors and maxillary lateral incisors.
- Fixture-mounts as an abutment for short (8mm) 3.7mmD implants: Indicated for tooth . replacement of mandibular and maxillary central and lateral incisors.
Legacy2, simplyLegacy2, Legacy3, simplyLegacy3, and Legacy4 implants are compatible with the following abutments.
| Manufacturer | Abutment Line | Platform Diameter |
|----------------|---------------|----------------------------|
| Implant Direct | Legacy | 3.0mm, 3.5mm, 4.5mm, 5.7mm |
## Summary of Comparison to Predicate Device
The minor differences in indications between the subject and predicate devices are not critical to the intended therapeutic use of the products and do not raise different questions of safety or effectiveness when the products are used as labeled, because they merely limit the scope of use and clarify the appropriate conditions of use. Relatedly, this submission updates the indications for use statement for the previously cleared Legacy3 implant models (K090234 and K131097) to align with the indications for the other subject Legacy models.
Both the subject and predicate implants function in support of single- and multi-unit restorations to help rehabilitate partially and fully edentulous patients' chewing function. Both implants are indicated for immediate loading when good primary stability has been achieved and with appropriate occlusal loading. The underlying technological principle is mechanical load-bearing of the forces imparted in the patient's mouth in order to provide stability. While the devices differ slightly in specific features of the implant body, the core design and how it functions are the same. Similarly, the subject fixturemounts and the predicate Legacy3 straight abutment are made of the same materials and function in the same way to support dental restorations in conjunction with a corresponding implant. The minor differences in certain dimensions of the subject fixture-mounts and to incorporate the additional class l, exempt functionalities (implant carrier and transfer) do not alter the principles of operation of the system or the related safety and efficacy considerations.
A table comparing the key features of the subject and predicate devices is provided below.
## Performance Data
The company performed the following bench testing in support of the subject devices' safety and effectiveness:
- . Fatigue Testing
The subject devices were subjected to cyclic fatique testing in accordance with ISO 14801:2016 and FDA's Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments, and also leveraged fatigue testing conducted on representative devices (submitted in K192218). Results were favorable compared to the predicate and reference device and demonstrate that the subject devices can withstand foreseeable mastication forces in accordance with their specifications when used as intended.
#### . Biocompatibility Testing
Representative, worst-case subject devices were successfully tested for biocompatibility in accordance with ISO 10993-1, ISO 10993-3, ISO 10993-5, ISO 10993-6, ISO 10993-10, ISO 10993-11 and FDA's corresponding June 2016 quidance. The following endpoints were tested: cytotoxicity; sensitization; systemic toxicity; implantation (considered to also
{6}------------------------------------------------
cover sub-acute/sub-chronic and chronic toxicity); and genotoxicity/mutagenicity.
- . Surface Area and Bone-to-Implant Contact Analysis
Both surface area and bone-to-implant contact comparisons were made between the subject implants and reference device (K131097) 6mm length implants. Results show that the subject devices are substantially equivalent with respect to surface area and bone-to-implant contact.
- . Pull-out Testing
Osteotomies were made in simulated soft bone. The subiect and reference device 6mmL implants were inserted to a depth of 3mm below the bone level. Results show that the force required to pull out the 6mmL subject implants was substantially equivalent to that for the reference device.
#### . Insertion Torque Testina
Osteotomies were made in simulated soft and dense bone; subject and reference device 6mmL implants were inserted. Results show that the subject 6mmL devices have comparable insertion torque as the reference 6mmL device.
#### . Sterilization Validation Testing
Sterilization validation testing was performed on worst-case representative subject devices in accordance with ISO 11137-2. The validation successfully demonstrated a sterility assurance level (SAL) of 10th for the subject devices.
#### . Endotoxin Testina
Bacterial endotoxin (LAL) testing was performed per ANSI/AAMI ST72 on worst-case representative subject devices to verify that the subject implants, upon terminal sterilization, meet established endotoxin release limits. Results were well below the maximum thresholds set by FDA and USP, confirming that the devices are acceptably non-pyrogenic.
#### . Shelf Life Validation
The subject devices after 5 years of real time aging were subjected to microbial aerosol challenge and sterility testing in accordance with ISO 11737-2. Results were passing in support of the labeled shelf life - the device label remained attached and legible, there was no damage to the device or packaging due to aging, and the sterile barrier remained intact.
#### . Distribution Validation
To determine any impact distribution would have on the subject device, worst-case implants underwent simulated shipping following ASTM D4169, DC 13, Assurance Level II. Tests include handling, stacking, loose load vibration, and concentrated impact. All tested implants passed post-simulation quality and sterile barrier inspection.
#### . Sterilization Validation
Sterilization validation performed for K192218 on devices representing worst-case compared to the subject fixture-mount/abutments in accordance with ANSI/AAMI/ISO_14937 and ISO 17665 was relied upon. This testing confirmed the adequacy of the steam sterilization process (performed by the end user on the fixture-mount/abutment per instructions/parameters provided in the device's labeling) to achieve a sterility assurance level (SAL) of 10 ° for the subject fixture-mount/abutments.
All results confirmed the devices' adequacy for their intended use and equivalent safety and effectiveness to the predicate device.
{7}------------------------------------------------
No animal or clinical studies were performed in support of this 510(k) notice.
# Conclusions
The Legacy2, Legacy3, Legacy4, simplyLegacy2, and simplyLegacy3 dental implants and the Legacy2, Legacy3, and Legacy4 fixture-mounts are as safe and effective as the predicate Legacy3 implant system (K090234). The subject devices have intended uses and similar indications, technological characteristics, and principles of operation as the predicate device. The minor differences in indications do not alter the intended therapeutic use of the devices and do not affect their safety and effectiveness when used as labeled. In addition, the minor technological differences between the subject devices and their predicate raise no new issues of safety or effectiveness. Performance data demonstrate that the subject devices are as safe and effective as the predicate. Thus, the subject Legacy dental implants and fixture-mount/abutments are substantially equivalent.
{8}------------------------------------------------
| Characteristic | Legacy2, Legacy4, simplyLegacy2,<br>Legacy3, and simplyLegacy3 Implants;<br>Legacy2, Legacy3, Legacy4 Fixture-<br>mounts | Predicate (K090234):<br>Spectra-System Dental<br>Implants 2008 | Reference (K131097):<br>Legacy3 Implants<br>(6mmL) | Reference (K060063):<br>Legacy Abutment<br>System | Reference (K192218):<br>Custom Legacy and<br>Custom InterActive<br>Titanium Abutments |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | 21 C.F.R. § 872.3640; DZE<br>21 C.F.R. § 872.3630; NHA | | 21 C.F.R. § 872.3640; DZE | 21 C.F.R. § 872.3630;<br>NHA | 21 C.F.R. § 872.3630; NHA |
| Intended Use /<br>Indications for<br>Use | Legacy2, simplyLegacy2, Legacy3,<br>simplyLegacy3, and Legacy4 dental implants<br>are two-piece implants for one-stage or two-<br>stage surgical procedures. These implants<br>are intended for use in partially and fully<br>edentulous upper and lower jaws in support<br>of single or multiple-unit restorations and<br>terminal or intermediate abutment support for<br>fixed bridgework.<br>Implants can be indicated for immediate<br>loading when good primary stability has<br>been achieved and with appropriate<br>occlusal loading.<br>• Narrow (3.2mmD) implants: Indicated for<br>single-tooth replacement (mandibular<br>central and lateral incisors; maxillary<br>lateral incisors), multiple-tooth<br>replacements or denture stabilization.<br>• Short (<10mm) 3.7mm implants:<br>Indicated for single-tooth (mandibular<br>and maxillary central and lateral incisors),<br>multiple tooth replacements or denture<br>stabilization.<br>The Legacy 2, Legacy3, and Legacy4 fixture-<br>mounts are intended for use with the<br>corresponding dental implants (Legacy2,<br>Legacy3, and Legacy4, respectively). The<br>fixture-mounts can function as an abutment.<br>As an abutment, fixture-mounts are intended<br>for use with dental implants in the maxillary<br>and/or mandibular arches to provide support<br>for crowns or bridges for edentulous or<br>partially edentulous patients.<br>• Fixture-mounts as an abutment for<br>narrow (3.2mmD) implants: Indicated for<br>single-tooth replacement of mandibular<br>central and lateral incisors and maxillary<br>lateral incisors.<br>• Fixture-mounts as an abutment for short<br>(8mm) 3.7mmD implants: Not intended<br>for tooth replacement of canines, pre- | The Spectra-System<br>Dental Implant 2008<br>system is comprised of<br>dental implant fixtures<br>and prosthetic devices<br>that compose a two-<br>piece implant system.<br>The Dental Implants are<br>intended for use in the<br>mandible and maxilla, in<br>support of single unit or<br>multiple unit cement or<br>screw-receiving<br>restorations and for the<br>retention and support of<br>overdentures. The<br>implants are intended<br>for immediate<br>placement and function<br>for the support of single-<br>tooth or multiple-tooth<br>restorations,<br>recognizing bone<br>stability and appropriate<br>occlusal load<br>requirements. | Legacy3 6mm Length<br>implants consist of two-<br>piece implants for one-<br>stage or two-stage<br>surgical procedures.<br>These implants are<br>intended for use in<br>partially and fully<br>edentulous upper and<br>lower jaws in support of<br>single or multiple-unit<br>restorations and<br>terminal or intermediate<br>abutment support for<br>fixed bridgework.<br>Implants can be<br>indicated for immediate<br>loading when good<br>primary stability has<br>been achieved and with<br>appropriate occlusal<br>loading. | The Legacy Abutment<br>System is intended for<br>use with dental implants<br>in the maxillary and/or<br>mandibular arches to<br>provide support for<br>crowns, bridges, or<br>overdentures for<br>edentulous or partially<br>edentulous patients.<br>The Legacy Abutment<br>System is compatible<br>with implants that have<br>mating diameters, lead-in<br>bevels, internal hex sizes,<br>and 1-72UNF internal<br>threads, as shown in the<br>Zimmer Dental Tapered<br>Screw-Vent Surgical<br>Manual.<br>Implant Direct LLC will<br>monitor the compatible<br>implants for modifications<br>to ensure future<br>compatibility. In the event<br>of any modification,<br>Implant Direct LLC will<br>either modify the Legacy<br>abutment to ensure<br>compatibility, or cease<br>claiming compatibility to<br>the modified Zimmer<br>Dental Screw-Vent<br>implants. | Custom Titanium Abutments<br>are customizable devices<br>intended for use with dental<br>implants in the maxillary<br>and/or mandibular arches to<br>provide support for crowns or<br>bridges for edentulous or<br>partially edentulous patients.<br>• Custom Titanium<br>Abutment for narrow<br>(3.2mmD, 3.3mmD)<br>implants: Indicated for<br>single-tooth replacement<br>of mandibular central and<br>lateral incisors and<br>maxillary lateral incisors.<br>Also indicated for multiple<br>tooth replacements.<br>• Custom Titanium<br>Abutment for short (8mm)<br>3.7mmD implants:<br>Indicated for tooth<br>replacement of<br>mandibular and maxillary<br>central and lateral<br>incisors.<br>Custom Legacy Titanium<br>Abutments are compatible at<br>the implant-level with<br>Legacy1, Legacy2, Legacy3,<br>Legacy4, simplyLegacy2<br>and simplyLegacy3<br>implants, excluding 6mm<br>length implants.<br>Custom InterActive Titanium<br>Abutments are compatible at<br>the implant-level with<br>InterActive,<br>simplyInterActive and |
| Characteristic | Legacy2, Legacy4, simplyLegacy2,<br>Legacy3, and simplyLegacy3 Implants;<br>Legacy2, Legacy3, Legacy4 Fixture-<br>mounts | Predicate (K090234):<br>Spectra-System Dental<br>Implants 2008 | Reference (K131097):<br>Legacy3 Implants<br>(6mmL) | Reference (K060063):<br>Legacy Abutment<br>System | Reference (K192218):<br>Custom Legacy and<br>Custom InterActive<br>Titanium Abutments |
| | molars or molars. | | | | SwishActive implants, |
| | Legacy2, simplyLegacy2, Legacy3,<br>simplyLegacy3, and Legacy4 implants are<br>compatible with Legacy abutments (3.0mm,<br>3.5mm, 4.5mm, 5.7mm platform). | | | | excluding 6mm length<br>implants. |
| General Design | Threaded Root form implant | Threaded Root form implant | Threaded Root form implant | Screw- or cement-retained<br>root form abutment | Cement-retained root form<br>abutment |
| Body Diameters | 3.2mm, 3.7mm, 4.2mm, 4.7mm, 5.2mm,<br>5.7mm, 7.0mm | 3.2mm, 3.7mm, 4.2mm,<br>4.7mm, 5.2mm, 5.7mm | 3.7mm, 4.2mm, 4.7mm,<br>5.2mm, 5.7mm, 7.0mm | N/A | N/A |
| Implant<br>Lengths | 6mm (only simplyLegacy), 8mm, 10mm,<br>11.5mm, 13mm, 16mm | 8mm, 10mm, 11.5mm,<br>13mm, 16mm | 6mm | N/A | N/A |
| Cutting Flutes | 3 (Legacy2/4); 2 (Legacy3) | N/A | 2 | N/A | N/A |
| lmplant<br>Platform | 3.0mm, 3.5mm, 4.5mm, 5.7mm | 3.0mm, 3.5mm, 4.5mm,<br>5.7mm | 3.5mm, 4.5mm, 5.7mm | Mates with 3.5mm,<br>4.5mm, 5.7mm interface<br>diameter | Mates with 3.0mm, 3.4mm.<br>3.5mm, 4.5mm, 5.7mm<br>interface diameter…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.