K231010 · AliveCor, Inc. · MHX · Jun 7, 2024 · Cardiovascular
Device Facts
Record ID
K231010
Device Name
Corvair
Applicant
AliveCor, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Jun 7, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, PCCP
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
ECG rhythm and morphology determinations
Deep neural networks
—
—
Approximately 1 million 12-Lead ECGs from 400K patients (1985-2010).
>1 (physicians)
—
—
ECG interval estimation
Deep neural networks
—
—
Approximately 1 million 12-Lead ECGs from 400K patients (1985-2010).
>1 (physicians)
Common Standards for Quantitative Electrocardiography Standard Database (CSEDB) and Mayo Clinic Genetic Heart Rhythm Clinic dataset.
—
Indications for Use
AliveCor's Corvair ECG analysis system assists the healthcare professional (HCP) in measuring and interpreting resting diagnostic ECGs for rhythm and morphological information by providing an initial automated interpretation. The interpretation by the analysis program may then be confirmed, edited, or deleted by the HCP. The analysis program is intended for use in the general population ranging from healthy subjects to patients with cardiac and/or non-cardiac abnormalities. Corvair is intended for use by healthcare professionals, or trained personnel in healthcare facilities (e.g. the doctor's office or hospital) and in acute settings. Corvair analyses should be used only as an adjunct to clinical history, symptoms, and the results of other non-invasive and/or invasive tests. Corvair analyses are considered unconfirmed and must be reviewed by a qualified physician. The provisional automated ECG analysis should not be used for clinical action if it has not been reviewed by a qualified healthcare professional capable of independently interpreting the ECG signal.
Device Story
Corvair is a Software as a Medical Device (SaMD) providing automated analysis of 10-second, 4-lead resting ECGs (Leads I, II, V2/V1, V4). It functions as an API library integrated into target devices (e.g., GE MAC series, AliveCor Impala). The device uses deep neural networks (DNNs) to generate 35 determinations (14 rhythm, 21 morphology) and global interval measurements (PR, QRS, QT, QTc, Heart Rate). Operated by healthcare professionals in clinical/acute settings, it offers 'Symptomatic' and 'Asymptomatic' modes to balance sensitivity and specificity. Outputs are provisional and require physician review/confirmation before clinical action. Benefits include rapid, automated ECG interpretation support, aiding in the detection of arrhythmias, conduction defects, hypertrophy, and myocardial infarction.
Clinical Evidence
No clinical testing was conducted. Performance was established via bench testing using a large dataset of ~1 million 12-lead ECGs (400K patients) and validated against the Common Standards for Quantitative Electrocardiography Standard Database (CSEDB) and a proprietary Mayo Clinic dataset. Metrics included sensitivity, specificity, PPV, mean error/SD for intervals, and mean absolute error for heart rate. Performance was found to be substantially equivalent to the predicate.
Technological Characteristics
SaMD (API library); C++ interface. Analyzes 4-lead ECGs (wet Ag/AgCl electrodes). Uses deep neural networks (DNNs) for analysis. Operates in Symptomatic or Asymptomatic modes. Validated per IEC 60601-2-25. Includes a Predetermined Change Control Plan (PCCP) for model retraining.
Indications for Use
Indicated for use in the general population, including healthy subjects and patients with cardiac or non-cardiac abnormalities, to assist healthcare professionals in measuring and interpreting resting diagnostic ECGs for rhythm and morphological information.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
12SL ECG Analysis Program by GE Medical Systems (K141963)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 7, 2024
AliveCor, Inc. % Prabhu Raghavan Principal Consultant Mdar, LLC. 1790 Montemar Way San Jose. California 95125
Re: K231010
Trade/Device Name: Corvair Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II Product Code: MHX Dated: April 6, 2023 Received: April 7, 2023
Dear Prabhu Raghavan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not
{1}------------------------------------------------
required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Kimberly Crowley
For: Jennifer Kozen Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231010
Device Name Corvair
#### Indications for Use (Describe)
AliveCor's Corvair ECG analysis system assists the healthcare professional (HCP) in measuring and interpreting resting diagnostic ECGs for rhythm and morphological information by providing an initial automated interpretation. The interpretation by the analysis program may then be confirmed, edited, or deleted by the HCP. The analysis program is intended for use in the general population ranging from healthy subjects to patients with cardiac abnormalities. Corvair is intended for use by healthcare professionals, or trained personnel in healthcare facilities (e.g. the doctor's office or hospital) and in acute settings.
Corvair analyses should be used only as an adjunct to clinical history, symptoms, and the results of other non-invasive and/or invasive tests. Corvair analyses are considered unconfirmed and must be reviewed by a qualified physician. The provisional automated ECG analysis should not be used for clinical action if it has not been reviewed by a qualified healthcare professional capable of independently interpreting the ECG signal.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary for K231010
Prepared in accordance with the requirements of 21 CFR 807.92
#### Submitter Information [807.92(a)(1)]
| Submitter/Applicant | AliveCor Inc.<br>189 N Bernardo, Suite 100<br>Mountain View, CA 94043<br>Establishment Registration: 3009715978<br>Submitter Contact: Samip Shah<br>Phone: (650) 396-8557<br>Email: sshah323@alivecor.com |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person | Prabhu Raghavan<br>Regulatory Consultant for AliveCor, Inc.<br>Principal Consultant, MDQR, LLC<br>Phone: 408-316-5707<br>Email: prabhu@mdqr.solutions |
| Date Prepared | June 06, 2024 |
| Device Information [807.92(a)(2)] | |
| Trade Name | Corvair |
|--------------|-------------------------------------------------------------------------|
| Common Name | Monitor, Physiological, Patient(With Arrhythmia Detection<br>Or Alarms) |
| Regulation | 21 CFR§ 870.1025 |
| Device Class | II |
| Product Code | MHX |
#### Predicate Information [807.92(a)(3)]
K141963. 12SL ECG Analysis Program by GE Medical Predicate(s) Systems®
## Device Description [807.92(a)(4)]
Corvair is Software as a Medical Device (SaMD) intended for use by healthcare professionals to analyze a diagnostic-bandwidth ECG. Corvair analyzes a 10-second ECG and provides rhythm analysis, morphological analysis, and ECG interval estimation. Corvair provides 35 separate determinations with 14 rhythm and 21 morphology determinations. Rhythm determinations include Normal Sinus Rhythm, Atrial fibrillation, Atrial flutter, Paced Rhythm, Junctional Rhythm, and Bigeminy, with the modifiers of 1* Degree AV Block, Higher Degree AV Block (including 2nd and 3rd degree AV blocks), Sinus Arrhythmia, Marked Sinus Arrhythmia, Marked Bradycardia, Sinus Tachycardia, and PVCs. Morphology determinations include Intraventricular block (RBBB, LBBB, and Other Intraventricular Block), Hypertrophy (LVH, and RVH), Atrial Enlargement (LAE and RAE), Acute Myocardial Infarction (Anterior MI, Inferior MI, Lateral MI), Old/Previous Myocardial Infarction (Anterior Old MI, Inferior Old MI,
{5}------------------------------------------------
Lateral Old MI), Ischemia (Anterior, Inferior, Lateral), Prolonged QT, Paced ECG, Other Morphological Defects (Early Repolarization, Wolff-Parkinson-White Syndrome (WPW)), and Normal or Otherwise Normal. Rhythm and morphology determinations are overlapping. i.e., an ECG could receive multiple rhythm and morphology determinations (e.g., Sinus Rhythm, Acute MI). The device also provides global ECG measurements (PR, ORS, OT, OTcB, OTcF, and Heart Rate). No beat-level analysis is provided by the device. Corvair may fail to detect or misidentify conduction system pacing and demand pacing. Corvair does not detect sinus pause. While Corvair provides PR interval estimation and does detect WPW, it does not have a separate determination of abnormally short PR intervals.
This SaMD provides these capabilities in the form of an Application Program Interface (API) library. Any software or device ("target device") can incorporate the Corvair API library into its device software to provide users with resting ECG analytics. The input ECG is provided by the target device to Corvair, to which the various Corvair algorithms are applied, and outputs generated accordingly. Corvair has a C++ interface and a distributed binary (library), which is used by the target device to statically link to Corvair. Viewing of Corvair's ECG analysis is handled by the target device.
Corvair is intended to be used with standard diagnostic-bandwidth, resting ECG recordings collected using 'wet' Ag/AgCl electrodes with conductive gel/paste. Corvair only requires 4 ECG leads for analysis, specifically, either Leads {I, II, V2, and V4}, or Leads {I, II, V1, and V4}. Compatible devices include resting ECGs from GE Medical Systems® (e.g., K081437, MAC 1600, K110266, MAC 5500, K173830, MAC VU360, etc.), and AliveCor's Impala (K232035). Regardless of the lead configuration, Corvair provides the same set of rhythm, morphological, and interval determinations. Corvair has two modes of operation, Symptomatic Mode, which is used when the pre-test probability for a specific rhythm or morphology is high, and Asymptomatic Mode, which optimizes the PPV, by optimizing the specificity, to detect the various rhythms and morphologies. The target device can choose which lead set and which mode of determinations to utilize based on the target clinical application and the patient's clinical presentation.
Corvair utilizes several deep neural networks (DNNs) for its analysis. These DNNs were trained on a dataset of approximately 1 million 12-Lead ECGs acquired from about 400K clinical patients at the Emory University Hospital over several decades between 1985 and 2010. Each ECG has a physician overread confirmed diagnosis with multiple diagnostic codes. The dataset had a 52%/48% ratio of ECGs from male and female patients, respectively. The average age of the patient was 61.3 ± 16. The dataset included 56% white, 33% African American, 2.2% Asian, 9% other races/ethnicities.
## Indications for use [807.92(a)(5)]
AliveCor's Corvair ECG analysis system assists the healthcare professional (HCP) in measuring and interpreting resting diagnostic ECGs for rhythm and morphological information by providing an initial automated interpretation. The interpretation by the analysis program may then be confirmed, edited, or deleted by the HCP. The analysis program is intended for use in the general population ranging from healthy subjects to patients with cardiac and/or non-cardiac
{6}------------------------------------------------
abnormalities. Corvair is intended for use by healthcare professionals, or trained personnel in healthcare facilities (e.g. the doctor's office or hospital) and in acute settings.
Corvair analyses should be used only as an adjunct to clinical history, symptoms, and the results of other non-invasive and/or invasive tests. Corvair analyses are considered unconfirmed and must be reviewed by a qualified physician. The provisional automated ECG analysis should not be used for clinical action if it has not been reviewed by a qualified healthcare professional capable of independently interpreting the ECG signal.
#### Substantial Equivalence
#### Comparison of intended use and indications for use
Both Corvair and the K141963 predicate have identical intended uses, i.e., intended for measuring and interpreting diagnostic-bandwidth, resting ECG. For both devices, the ECG analysis outputs represent potential findings for review and interpretation by a qualified healthcare professional and do not represent complete diagnoses.
The Indications for Use (IFU) statement for Corvair is largely equivalent to the predicate K141963 device IFU. The subject and predicate devices are both intended to provide analysis of resting ECG recordings. Specifically, both devices assists the healthcare professional (HCP) in measuring and interpreting diagnostic-bandwidth ECGs for rhythm and morphological information. In both devices, the provisional automated ECG analysis should not be used for clinical action if it has not been reviewed by a qualified healthcare professional capable of independently interpreting the ECG signal.
Note that while the intended use population of both devices include adults 18 years old and over, the predicate is also intended for use with a pediatric patient population. As such, the subject device's intended use population is a subset of the predicate device's intended use population.
## Comparison of technological characteristics
While there are many similarities between the two devices, there are three main differences:
- a) Number of leads analyzed: While the predicate requires and analyzes all 12 ECG leads, the subject device is intended to analyze a reduced set of 4 standard leads.
- b) Number of determinations: The subject device provides determinations that are a subset of all the determinations provided by the predicate device.
- c) Machine Learning Models: While both devices utilize machine learning models, the architecture of these models are different. The predicate device utilizes traditional signal processing and regression based learning models, and the subject device utilizes signal processing as well as deep neural networks.
Both devices use signal processing and machine learning algorithms to detect arrhythmia, perform analysis of ECG morphology, and provide interval estimates. Both the predicate and subject devices focus on providing high specificity to improve the positive value (PPV) of the final determinations, especially since many of the determinations have low prevalence in the intended use population. The subject device additionally offers a mode where the sensitivity is improved at the cost of some reduction in specificity. This is intended to assist
{7}------------------------------------------------
HCPs who have a sense of the pretest probability and can choose the results of this mode in case they are able to tolerate more false positives and a reduced PPV.
The differences noted above do not by themselves raise different questions of safety and effectiveness. The differences were addressed through an extensive performance validation using a large ECG dataset to demonstrate safety and effectiveness and substantial equivalence. The proposed methods to validate the subject device and to demonstrate ECG analysis performance utilize well understood scientific methods.
In conclusion, Corvair has the same intended use as the predicate device, and any differences in technological characteristics do not raise different questions of safety or effectiveness.
#### Performance Data [807.92(b)]
All necessary testing was conducted on Corvair to support a determination of substantial equivalence to the predicate device.
#### Nonclinical Testing Summary [807.92(b)(1)]
Nonclinical performance testing was conducted to demonstrate that Corvair satisfies its design and software requirements as well as demonstrate that the ECG analysis performance is substantially equivalent to the predicate device (K141963, 12SL ECG Analysis Program).
AliveCor evaluated Corvair against a large set of ECGs and compare its analysis output against a known reference. The comparison uses standard ECG performance metrics including, sensitivity, specificity, and PPV for the interpretive outputs, mean error and standard deviation of error for the interval outputs, and mean absolute error for heart rate accuracy. These outputs were evaluated against a clinically relevant acceptance criteria to demonstrate Corvair's effectiveness. PR, QRS, QT interval estimation performance was demonstrated using the Common Standards for Quantitative Electrocardiography Standard Database (CSEDB) as referenced in the IEC 60601-2-25 and against an AliveCor proprietary dataset developed from ECGs that were collected in a clinical study at the Mayo Clinic's Genetic Heart Rhythm Clinic. Heart rate and OTc were also validated using ECGs from the latter database. Corvair performance of was also compared against the predicate and the subject device demonstrated substantially equivalent performance.
#### Clinical Testing Summarv [807.92(b)(2)]
No clinical testing was required or conducted to support a determination of substantial equivalence.
#### Predetermined Change Control Plan (PCCP)
As part of K231010, Corvair includes a PCCP to improve algorithm performance by retraining with additional data without modifying the architecture:
- Such improvements will be made by acquiring additional high quality, diverse training . data from major clinical institutions similar to the data used to train the models within the 510(k).
- The performance of the retrained models will be evaluated using the same datasets used ● within the 510(k). Additional large validation datasets will also be created from sites
{8}------------------------------------------------
independent of the training data to ensure model generalization. The performance will be acceptable when the overall performance is noninferior to the performance of the models used in the 510(k). Some minor individual determination performance variation would be acceptable.
- When such improvements are made and found to be acceptable, the Corvair device labeling will be updated to incorporate the updated performance specifications. The changes would be also communicated to Corvair API software integrators so that they can also communicate the changes to the end users.
#### Conclusions [807.92(b)(3)]
The results of these testing therefore demonstrate that the device performs as intended and confirm that the technological differences between the subject device and the predicate devices do not raise different questions of safety and effectiveness, and that the device is as safe and as effective for its intended use as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.