K110266 · Ge Medical Systems Information Technologies, Inc. · DPS · Apr 29, 2011 · Cardiovascular
Device Facts
Record ID
K110266
Device Name
MAC 5500 HD, MAC 3500
Applicant
Ge Medical Systems Information Technologies, Inc.
Product Code
DPS · Cardiovascular
Decision Date
Apr 29, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric
Indications for Use
MAC 5500 HD: The MAC 5500 HD ECG Analysis System is intended to acquire, analyze, display, and record electrocardiographic information from adult and pediatric populations. Basic systems deliver 3, 6, 12, or 15 lead ECG's, interpretive analysis, vector loops; and can be upgraded to provide software analysis options such as high resolutions signal averaging of QRS and P wave portions of the electrocardiogram. Transmission and reception of ECG data to and from a central ECG cardiovascular information system is optional. The MAC 5500 HD is intended to be used under the direct supervision of a licensed healthcare practitioner, by trained operators in a hospital or medical professional's facility. MAC 3500: The MAC 3500 ECG Analysis System is intended to acquire, analyze, display, and record resting electrocardiographic information from adult and pediatric populations. Basic systems deliver 3.6. 12, or 15 lead ECG's, including interpretive andlysis. Transmission and reception of ECG data to and from a central ECG cardiovascular information system is optional. The MAC 3500 is intended to be used under the direct supervision of a licensed healthcare practitioner, by trained operators in a hospital or medical professional's facility.
Device Story
MAC 5500 HD and MAC 3500 are mobile electrocardiograph systems; acquire, analyze, display, and record ECG signals from surface electrodes. MAC 5500 HD uses external CAM HD acquisition module for improved signal fidelity; MAC 3500 has integrated acquisition module. Systems provide 3, 6, 12, or 15 lead ECGs; interpretive analysis; vector loops (MAC 5500 HD); and high-resolution signal averaging (optional). Devices feature alphanumeric keyboard, VGA display, thermal writer, and removable storage. Operated by trained personnel in clinical settings; optional connectivity to central cardiovascular information systems. Output assists clinicians in cardiac assessment and diagnosis.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Electrocardiograph; mains or rechargeable battery powered. Components: processing unit, patient acquisition module (CAM HD for 5500 HD, integrated for 3500). Connectivity: optional transmission/reception to central information systems. Form factor: mobile (trolley) or desktop. Software analysis: interpretive analysis, vector loops, high-resolution signal averaging.
Indications for Use
Indicated for adult and pediatric patients requiring acquisition, analysis, display, and recording of electrocardiographic information. Used in hospital or medical professional facilities under supervision of a licensed healthcare practitioner.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
K110266 P1/3
APR 2 9 2011
510(k) Premarket Notification Submission
: :
## 510(k) Summary
| In accordance with 21 CFR 807.92 the following summary of information is provided: | | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Date: | 24-January-2011 | |
| Submitter: | GE Medical Systems Information Technologies, Inc.<br>9900 Innovation Drive<br>Wauwatosa, WI 53226 | |
| Primary Contact Person: | Patricia Taige<br>Regulatory Affairs Leader<br>GE Medical Systems Information Technologies, Inc.<br>Phone:(414) 721-3222<br>Fax: (414) 721-3863 | |
| Secondary Contact Person: | Joe Lucas<br>Regulatory Affairs Leader<br>GE Medical Systems Information Technologies, Inc.<br>Phone:(414) 721-2593<br>Fax: (414) 721-3863 | |
| Device: | Trade Name: MAC 5500 HD ECG Analysis-System.<br>MAC 3500 ECG Analysis System | |
| Common/Usual Name: | Electrocardiograph | |
| Classification Names: | 21 CFR 870.2340 Electrocardiograph | DPS |
| Product Code: | 21 CFR 870.1425 Programmable Diagnostic Computer<br>21 CFR 870.2920 Telephone Electrocardiograph<br>Transmitter and Receiver | DQK<br>DXH |
| Predicate Device(s): | K073625 MAC 5500 ECG Analysis System | |
| Device Description: | The MAC 5500 HD and MAC 3500 ECG Analysis Systems are<br>designed to acquire, analyze, display, and record ECG signals from<br>surface ECG electrodes. The devices consist of two basic<br>components: the processing unit and the patient acquisition module.<br>The MAC 5500 HD and MAC 3500 can deliver 3, 6, 12, or 15 lead<br>ECG's, including interpretive analysis and 12 or 15 lead ECG's on full-<br>size reports. In addition MAC 5500 HD can deliver vector loops and<br>can be upgraded to provide software analysis options such as high<br>resolution signal averaging of QRS and P wave portions of the<br>electrocardiogram.<br>The MAC 5500 HD acquires ECG data using the external CAM HD<br>patient data acquisition module. By placing the data acquisition<br>device closer to the patient, signal fidelity is improved and noise is<br>reduced: MAC 3500 has the patient data acquisition module<br>integrated into the main device.<br>The MAC 5500 HD and MAC 3500 incorporate an alphanumeric<br>keyboard for patient demographics and other data entry, a full VGA<br>graphics and waveform display, an integrated thermal writer and | |
Image /page/0/Picture/5 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circle. The letters are stylized and appear to be handwritten. The logo is black and white.
::
:
{1}------------------------------------------------
KI10266 PZ
GE Healthcare
Image /page/1/Picture/1 description: I am sorry, but there is no discernible information in the image to provide a description.
### 510(k) Premarket Notification Submission
removable data storage.
Models provide mains or, for customer convenience, rechargeable battery operation as well as optional transmission and reception of ECG data to and from a central ECG cardiovascular information system via communication links.
MAC 5500 HD and MAC 3500 are intended as mobile devices by offering an optional trolley for transporting the equipment but the main units can be separated from the trolley and used as desktop units.
Intended Use:
MAC 5500 HD: The MAC 5500 HD ECG Analysis System is intended to acquire, analyze, display, and record electrocardiographic information from adult and pediatric populations. Basic systems deliver 3, 6, 12, or 15 lead ECG's, interpretive analysis, vector loops; and can be upgraded to provide software analysis options such as high resolutions signal averaging of QRS and P wave portions of the electrocardiogram. Transmission and reception of ECG data to and from a central ECG cardiovascular information system is optional. The MAC 5500 HD is intended to be used under the direct supervision of a licensed healthcare practitioner, by trained. operators in a hospital or medical professional's facility.
MAC 3500: The MAC 3500 ECG Analysis System is intended to acquire, analyze, display, and record resting electrocardiographic information from adult and pediatric populations. Basic systems deliver 3.6. 12, or 15 lead ECG's, including interpretive andlysis. Transmission and reception of ECG data to and from a central ECG cardiovascular information system is optional. The MAC 3500 is intended to be used under the direct supervision of a licensed healthcare practitioner, by trained operators in a hospital or medical professional's facility.
The MAC 5500 HD and MAC 3500 employ.the same fundamental. Technology: scientific technology as their predicate device MAC 5500 ECG Analysis System (K073625).
Determination of Substantial Equivalence:
Summary of Non-Clinical Tests:
The MAC 5500 HD and MAC 3500 and their applications comply with voluntary standards as detailed in Section 9, 11, and 17 of this premarket submission. The following quality assurance measures were applied to the development of the systems:
- Requirement Definition
- Risk Analysis .
- . Software Safety Classification
- Technical Review ●
- Formal Design Review ●
- Code Inspection
ﺗﻮ
{2}------------------------------------------------
K110266 P3/3
# GE Healthcare
ి
:
#### 510(k) Premarket Notification Submission
- Integration Testing (Module and System verification) .
- Final Acceptance Testing .
- Performance Testing .
- Safety testing .
#### Summary of Clinical Tests:
The subject of this premarket submission, MAC 5500 HD and MAC 3500, did not require clinical studies to support substantial equivalence.
GE Medical Systems Information Technologies, Inc. considers the Conclusion: MAC 5500 HD / MAC 3500 to be as safe, as effective, and performance is substantially equivalent to the predicate device.
Image /page/2/Picture/10 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are enclosed within a circular border, creating a recognizable and iconic symbol.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, enclosed within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the upper portion of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
APR 2 9 2011
GE Medical Systems Information Technologies, Inc. c/o Ms. Patricia Taige Regulatory Affairs Leader 9900 Innovation Drive Wauwatosa, WI 53226
Re: K110266
Trade/Device Name: MAC 5500 HD and MAC 3500 ECG Analysis Systems Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (two) Product Codes: DPS, DQK, DXH Dated: January 24, 2011 Received: February 1, 2011
Dear Ms. Taige:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The rou may, ateresters, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Patricia Taige
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must of any : each all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the quality of events (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized font, enclosed within a circle. There are decorative swirls or flourishes around the letters, adding to the logo's distinctive appearance. The logo is black and white.
## GE Healthcare 510(k). Premarket. Notification. Submission.
510(k) Number (if known): K110266
Device Name: MAC 5500 HD
Indications for Use:
The MAC 5500 HD ECG Analysis System is intended to acquire, analyze, display, and record electrocardiographic information from adult and pediations. Basic systems deliver 3,6, 12, or 15 lead ECG's, interpretive analysis, vector loops, and can be upgraded to provide software analysis options such as high resolutions signal averaging of QRS and P wave portions of the electrocardiogram. Transmission and reception of ECG data to and from a central ECG cardiovascular information system is optional.
The MAC 5500 HD is intended to be used under the direct supervision of a licensed healthcare practitioner, by trained operators in a hospital or medical professional's facility.
AND/OR Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
IPLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
.
Risten Sign-Off sion of Cardiovascular Devices
510(k) Number_k110266
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for GE Healthcare. The text is in a simple, sans-serif font. The words "GE Healthcare" are on the first line. The image is a document that is likely a premarket notification submission.
510(k) Premarket Notification Submission
510(k) Number (if known): K I (0766
Device Name: MAC 3500
Indications for Use:
The MAC 3500 ECG Analysis System is intended to acquire, analyze, display, and record resting electrocardiographic information from adult and pediatric populations. Basic systems deliver 3, 6, 12, or 15 lead ECG's, including interpretive analysis. Transmission and reception of ECG data to and from a central ECG cardiovascular information system is optional.
The MAC 3500 is intended to be used under the direct supervision of a licensed healthcare proctitioner, by trained operators in a hospital or medical professional's facility.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE).
K110266
(Division Sign-Off)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K1102616
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.