K232035 · AliveCor, Inc. · DPS · Jun 7, 2024 · Cardiovascular
Device Facts
Record ID
K232035
Device Name
Impala
Applicant
AliveCor, Inc.
Product Code
DPS · Cardiovascular
Decision Date
Jun 7, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
AI/ML
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
12-lead resting electrocardiogram (ECG)
Machine learning model for lead synthesis
—
—
Approximately 110,000 12-Lead ECGs acquired from clinical patients at the Mayo Clinic (1985-2010).
—
3,000 12-Lead ECGs acquired from clinical patients at the Mayo Clinic (1985-2010), independent from the training dataset.
—
ECG rhythm and morphological interpretation
—
—
—
—
—
—
—
Indications for Use
Impala is intended to record, store, and transfer a 12-lead resting electrocardiogram (ECG). Impala acquires four standard diagnostic-bandwidth leads (Leads I, II, V2, V4, or Leads I, II, V1, V4). The device derives four standard diagnostic-bandwidth, Lead-III and unipolar limb leads aVR, aVF and aVL. The device also synthesizes Leads V1 or V2, V3, V5, V6, which are similar to but not identical to the same leads of a standard diagnostic 12-lead. The 4 synthesized chest leads are not intended for diagnostic use and may fail to show important findings limited to those leads. This device is not a substitute for a diagnostic 12-lead ECG and is contraindicated for use in ruling out any condition (including but not limited to certain ischemia/infarcts, Brugada syndrome) for which the diagnosis may be solely dependent on the synthesized leads. The device also provides ECG measurements and ECG analysis (rhythm and morphological interpretation) using the acquired leads. The automated ECG analysis results are provisional and should not be used for clinical action if it has not been reviewed by a qualified physician capable of interpreting the ECG signal in the context of the patient's condition. The automated analysis may then be confirmed, edited, or qualified physician. ECG analysis should be used only as an adjunct to clinical history, symptoms, and the results of other non-invasive tests. Impala is intended for use with patients aged 18 years and older. Impala is intended for use by healthcare professionals, or trained personnel in healthcare facilities (e.g. the doctor's office or hospital) and in acute settings.
Device Story
Portable 12-lead resting ECG system; consists of ECG module, patient lead wire with 5 snap-on electrodes, and mobile app (Impala App) on smartphone. User places electrodes on patient; device acquires 4 leads (I, II, V2/V1, V4); derives limb leads; synthesizes chest leads (V1/V2, V3, V5, V6) via machine learning model. Used by healthcare professionals in clinics/hospitals/acute settings. Integrates AliveCor Corvair software for rhythm/morphological analysis and measurements. Output displayed on mobile app for physician review; automated results are provisional and require physician confirmation. Benefits include diagnostic ECG capability in settings where traditional 10-electrode systems are impractical.
Clinical Evidence
No clinical testing conducted. Bench testing only. Performance validated using proprietary Mayo Clinic dataset (3,000 ECGs) and public databases (Physionet PTB-XL, CSEDB) to compare synthesized leads against standard 10-electrode 12-lead ECGs. Electrical safety/EMC testing per IEC 60601-1, 60601-1-2, 60601-2-25, 60601-1-11, 60601-1-12. Human factors usability testing confirmed user ability to collect data and interpret app output.
Technological Characteristics
Portable ECG module with 5-electrode lead wire. Uses standard gel electrodes. Connectivity via mobile app on smartphone. Machine learning model for lead synthesis trained on 110,000 ECGs. Complies with IEC 60601-1, 60601-1-2, 60601-2-25, 60601-1-11, 60601-1-12; patient lead wire conforms to ANSI EC53.
Indications for Use
Indicated for patients aged 18+ to record, store, and transfer 12-lead resting ECGs. Contraindicated for ruling out conditions where diagnosis depends solely on synthesized leads (e.g., certain ischemia/infarcts, Brugada syndrome).
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 7, 2024
AliveCor, Inc. % Prabhu Raghavan Principal Consultant Mdar, LLC. 1790 Montemar Way San Jose. California 95125
Re: K232035
Trade/Device Name: Impala Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: July 7, 2023 Received: July 7, 2023
Dear Prabhu Raghavan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Kimberly C
For: Jennifer Kozen Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics
{2}------------------------------------------------
and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K232035
Device Name Impala
#### Indications for Use (Describe)
Impala is intended to record, store, and transfer a 12-lead resting electrocardiogram (ECG). Impala acquires four standard diagnostic-bandwidth leads (Leads I, II, V2, V4, or Leads I, II, V1, V4). The device derives four standard diagnosticbandwidth, Lead-III and unipolar limb leads aVR, aVF and aVL. The device also synthesizes Leads V1 or V2, V3, V5, V6, which are similar to but not identical to the same leads of a standard diagnostic 12-lead. The 4 synthesized chest leads are not intended for diagnostic use and may fail to show important findings limited to those leads. This device is not a substitute for a diagnostic 12-lead ECG and is contraindicated for use in ruling out any condition (including but not limited to certain ischemia/infarcts, Brugada syndrome) for which the diagnosis may be solely dependent on the synthesized leads.
The device also provides ECG measurements and ECG analysis (rhythm and morphological interpretation) using the acquired leads. The automated ECG analysis results are provisional and should not be used for clinical action if it has not been reviewed by a qualified physician capable of interpreting the ECG signal in the context of the patient's condition. The automated analysis may then be confirmed, edited, or qualified physician. ECG analysis should be used only as an adjunct to clinical history, symptoms, and the results of other non-invasive tests.
Impala is intended for use with patients aged 18 years and older. Impala is intended for use by healthcare professionals, or trained personnel in healthcare facilities (e.g. the doctor's office or hospital) and in acute settings.
| Type of Use (Select one or both, as applicable) | | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|---------------------------------------------|--|--|
| | X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | | | | |
| *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete | | | | |
and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "ALIVECOR" in a sans-serif font. The letters are a dark teal color. A small registered trademark symbol is located to the right of the "R".
### 510(k) Summary for K232025
Prepared in accordance with the requirements of 21 CFR 807.92
#### Submitter Information [807.92(a)(1)]
| Submitter/Applicant | AliveCor Inc.<br>189 N Bernardo, Suite 100<br>Mountain View, CA 94043<br>Establishment Registration: 3009715978 |
|----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Submitter Contact: Samip Shah<br>Phone: (650) 396-8557<br>Email: sshah323@alivecor.com | |
| Primary Contact Person | Prabhu Raghavan<br>Regulatory Consultant for AliveCor, Inc.<br>Principal Consultant, MDQR, LLC |
| Phone: 408-316-5707<br>Email: prabhu@mdqr.solutions | |
| Date Prepared | June 06, 2024 |
#### Device Information [807.92(a)(2)]
| Trade Name | Impala |
|--------------|--------------------|
| Common Name | Electrocardiograph |
| Regulation | 21 CFR§870.2340 |
| Device Class | II |
| Product Code | DPS |
#### Predicate Information [807.92(a)(3)]
K173952, Universal SmartECG by VectraCor, Inc Predicate(s)
#### Device Description [807.92(a)(4)]
Impala is a portable 12-Lead resting electrocardiograph (ECG) device that acquires 4 standard diagnostic-bandwidth ECG leads from a patient (I, II, V1 or V2, V4), derives 4 standard diagnostic-bandwidth leads (aVL, aVR, aVF, III), and using software generates the remaining leads (V2 or V1, V3, V5, V6) to create a 12-Lead ECG recording. The device can be used by healthcare professionals (HCPs) to record a reduced lead, diagnostic-bandwidth, resting ECG, where traditional 10 electrode, 12 lead ECG recorders are not practical to administer due to size, time, or need for specialized clinicians to administer. Examples may include physician offices, and remote and field locations.
The Impala hardware consists of the Impala ECG Module that connects to the Patient Lead Wire. The Patient Lead Wire is a single cable that includes five snap-on electrodes. Impala also consists of a mobile software application, the Impala App that executes on a mobile computing
{5}------------------------------------------------
platform (MCP), such as an Apple iPhone smartphone. To record an ECG, the user positions standard off-the-shelf (OTS) ECG gel electrodes on the patient and snaps on the connectors in the Patient Lead Wire on to the electrodes. Impala allows for two options for which set of leads are acquired:
- 1. Lead Set 1: Leads {I, II, V2, and V4}, with electrodes on RA, LA, LL, V2, V4; and
- 2. Lead Set 2: Leads {I, II, V1, and V4} with electrodes on RA, LA, LL, V1, and V4.
Impala utilizes a machine learning model for lead synthesis. This model was trained on a dataset of approximately 110.000 12-Lead ECGs acquired from clinical patients at the Mayo Clinic over several decades between 1985 and 2010. Each ECG has a physician overread confirmed diagnosis with multiple diagnostic codes. This dataset included various types of morphology including intraventricular blocks, hypertrophy, atrial enlargement, myocardial infarction, ischemia, and prolonged OT. The dataset had a 56%/44% ratio of ECGs from male and female patients, respectively. The dataset included patients with a wide distribution of age groups from 18-99 years.
Impala incorporates AliveCor's Corvair (K231010), a software as a medical device, to provide rhythm analysis, morphological analysis, and ECG measurements. Corvair has two modes of operation, Symptomatic Mode, which is used when the pre-test probability for a specific rhythm is high, and Asymptomatic Mode, which optimizes the PPV, by optimizing the specificity, to detect the various rhythms and morphologies. The user can choose which Impala Lead Set and which Corvair mode of determinations to utilize based on the target clinical application.
#### Indications for use [807.92(a)(5)]
Impala is intended to record, store, and transfer a 12-lead resting electrocardiogram (ECG). Impala acquires four standard diagnostic-bandwidth leads (Leads I, II, V2, V4, or Leads I, II, V1, V4). The device derives four standard diagnostic-bandwidth. Lead-III and unipolar limb leads aVR, aVF and aVL. The device also synthesizes Leads V1 or V2, V3, V5, V6, which are similar to but not identical to the same leads of a standard diagnostic 12-lead. The 4 synthesized chest leads are not intended for diagnostic use and may fail to show important findings limited to those leads. This device is not a substitute for a diagnostic 12-lead ECG and is contraindicated for in ruling out any condition (including but not limited to certain ischemia/infarcts, Brugada syndrome) for which the diagnosis may be solely dependent on the synthesized leads.
The device also provides ECG measurements and ECG analysis (rhythm and morphological interpretation) using the acquired leads. The automated ECG analysis results are provisional and should not be used for clinical action if it has not been reviewed by a qualified physician capable of independently interpreting the ECG signal in the context of the patient's condition. The automated analysis may then be confirmed, edited, or deleted by a qualified physician. ECG analysis should be used only as an adjunct to clinical history, symptoms, and the results of other non-invasive and/or invasive tests.
Impala is intended for use with patients aged 18 years and older. Impala is intended for use by healthcare professionals, or trained personnel in healthcare facilities (e.g. the doctor's office or hospital) and in acute settings.
{6}------------------------------------------------
### Substantial Equivalence
#### Comparison of intended use and indications for use
Both Impala and the predicate are intended to record a reduced lead, diagnostic-bandwidth, resting ECG. Both Impala and the predicate are intended for Adult patients. Note that similar to other AliveCor devices, the proposed "Adult" population for Impala would include Transitional Adolescents, i.e., Adults 18 and over. Both Impala and the predicate are intended to be used by healthcare professionals, or trained personnel.
Like the predicate, the subject device is also intended to record a reduced lead set but utilizes a standard lead position instead of the Frank X, Y, Z leads. The subject device does not include the capability to directly record a 12-lead ECG using the standard 10 electrodes. As such, the subject device provides a subset of the predicate's feature set, and this difference does not raise different questions of safety or effectiveness.
#### Comparison of technological characteristics
Both Impala and the predicate allow for recording reduced set of ECG leads. The K173952 predicate uses 5-electrodes to acquire a 3-lead recording. Like the Impala subject device, the K173952 predicate derives or produces a full 12-lead ECG output from these reduced leads. This derivation of the full 12-lead utilizes VectraCor's K102378, VectraplexECG that can derive a 12 to 15-lead ECG report with the recorded 5-electrodes. Both Impala and the K173952 predicate use standard gel-electrodes with both devices including lead wires that clip onto these electrodes. While both Impala and the K173952 predicate can record reduced leads, the predicate records 3 leads instead of the 4 leads needed by Impala. The K173952 predicate requires the electrodes to be placed in RA, LA, V2, RL, LL, from which it records 3 leads (L1, L2, and V2 based on available public information). Impala records Leads I, II, V1 or V2, and V4. However both devices are utilizing the same location on the patient, i.e., arms, legs, and torso.
The K173952 predicate device utilizes the K102378, VectraplexECG device to provide the ECG Lead Generation. In Impala, this Lead Generation has been incorporated within the device software. Like the predicate, the synthesized leads are not intended for diagnostic use and may fail to show important findings limited to those leads.
In conclusion, Impala has the same intended use as the predicate device, and any differences in technological characteristics do not raise different questions of safety or effectiveness.
#### Performance Data [807.92(b)]
All necessary testing was conducted on Impala to support a determination of substantial equivalence to the predicate device.
#### Nonclinical Testing Summary [807.92(b)(1)]
The subject device has equivalent safety and technological characteristics and the differences have been addressed through the following performance testing:
- Electrical safety and electromagnetic compatibility testing verified that the subject device . complies with the requirements of IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-25, IEC 60601-1-11, and IEC 60601-1-12.
{7}------------------------------------------------
- The patient lead wire was verified and conforms to ANSI EC53.
- . Bench performance testing was conducted to demonstrate that Impala satisfies its design, hardware and software requirements.
- Bench performance testing was also conducted to evaluate the 4 synthesized chest leads ● generated by Impala from the 4 recorded leads in comparison with the corresponding chest leads of a standard 10-electrode resting 12-lead ECG. This validation utilized a proprietary AliveCor database as well as two public databases (Physionet PTB-XL and the Common Standards for Quantitative Electrocardiography Standard Database (CSEDB) as referenced in the IEC 60601-2-25). The proprietary dataset consisted of ECGs recorded 3,000 12-Lead ECGs acquired from clinical patients at the Mayo Clinic over several decades between 1985 and 2010. Each ECG has a physician overread confirmed diagnosis with multiple diagnostic codes. This dataset included various types of morphology including intraventricular blocks, hypertrophy, atrial enlargement, myocardial infarction, ischemia, and prolonged OT. The dataset had a 56%/44% ratio of ECGs from male and female patients, respectively. The dataset included patients with a wide distribution of age groups from 18-99 years. This dataset was independent from the training dataset. Results suggest that the synthesized leads tend to be similar to the corresponding 12-leads, so the display of these leads is not likely to confuse human interpreter. However significant differences exist in many cases. Thus the 4 synthesized chest leads are not intended for diagnostic use and may fail to show important findings limited to those leads. Further, this device is contraindicated for the diagnosis of any condition (including but not limited to certain ischemia/infarcts, Brugada syndrome) for which the diagnosis may be solely dependent on the synthesized leads.
- Human factors usability testing of Impala demonstrated that representative users could ● use the device as intended, including collecting data using Impala hardware and the associated Impala App, and understand the data presented on the App.
As such, the subject device has substantially equivalent safety and technological characteristics and any differences, based on the performance testing, do not raise different questions of safety and effectiveness.
#### Clinical Testing Summary [807.92(b)(2)]
No clinical testing was required or conducted to support a determination of substantial equivalence.
#### Conclusions [807.92(b)(3)]
The results of these testing therefore demonstrate that the device performs as intended and confirm that the technological differences between the subject device and the predicate devices do not raise different questions of safety and effectiveness, and that the device is as safe and as effective for its intended use as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.