General Total Knee Arthroplasty (TKR) Indications: - Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis. - Post-traumatic loss of knee joint configuration and function. - Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability. - Revision of previous unsuccessful knee replacement or other procedure. - Fracture of the distal femur and/or proximal tibia that cannot be standard fracture-management techniques. Additional Indications for Posterior Stabilized (PS) Components: - Ligamentous instability requiring implant bearing surface geometries with increased constraint. - Absent or non-functioning posterior cruciate ligament. - Severe anteroposterior instability of the knee joint. The Triathlon® Pro PS Femoral Components are intended for cemented use only.
Device Story
Triathlon® Pro PS Femoral Component is a line extension to the Triathlon® Total Knee System; used in total knee arthroplasty to replace damaged knee joint surfaces. Device is a sterile, single-use Cobalt-Chromium alloy femoral component; implanted by orthopedic surgeons in a clinical/OR setting. Functions as a semi-constrained cemented prosthesis to restore joint stability and function in patients with degenerative or traumatic knee disease. Provides mechanical articulation with tibial components; addresses ligamentous instability and posterior cruciate ligament deficiency. Benefits include pain relief and restoration of knee joint configuration. Substantial equivalence established via non-clinical bench testing, including fatigue, wear, and contact stress analysis.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing including material characterization, fatigue testing, wear testing (ISO 14243-3:2014), contact area/stress characterization, range of motion/constraint analysis, biocompatibility (ISO 10993-1:2018), shelf-life validation, and MR compatibility testing per FDA guidance.
Technological Characteristics
Cobalt-Chromium alloy femoral component manufactured via low-pressure metal injection molding (ASTM F2886-17). Semi-constrained cemented prosthesis. Available in sizes 1-8, left/right configurations. Sterile, single-use. Evaluated for MR compatibility (ASTM F2052-15, F2213-17, F2119-07, F2182-19e2).
Indications for Use
Indicated for patients with painful, disabling knee joint disease (osteoarthritis, traumatic arthritis, avascular necrosis, rheumatoid arthritis, post-traumatic arthritis), post-traumatic loss of function, moderate varus/valgus/flexion deformity, revision of failed knee replacement, or distal femur/proximal tibia fractures. PS components specifically indicated for ligamentous instability, absent/non-functioning posterior cruciate ligament, or severe anteroposterior instability. Intended for cemented use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is in blue and consists of the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
June 2, 2023
Howmedica Osteonics Corp. dba Stryker Orthopaedics Jane Guo Senior Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430
Re: K230952
Trade/Device Name: Triathlon® Total Knee System - Triathlon® Pro Posterior Stabilized (PS) Femoral Component Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: April 3, 2023 Received: April 4, 2023
Dear Jane Guo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jesse Muir-S
Jesse Muir, Ph.D. Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K230952
Device Name
Triathlon® Total Knee System - Triathlon® Pro Posterior Stabilized (PS) Femoral Component
Indications for Use (Describe)
General Total Knee Arthroplasty (TKR) Indications:
- · Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
- · Post-traumatic loss of knee joint configuration and function.
· Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
· Revision of previous unsuccessful knee replacement or other procedure.
· Fracture of the distal femur and/or proximal tibia that cannot be standard fracture-management techniques.
Additional Indications for Posterior Stabilized (PS) Components:
- · Ligamentous instability requiring implant bearing surface geometries with increased constraint.
- · Absent or non-functioning posterior cruciate ligament.
- · Severe anteroposterior instability of the knee joint.
The Triathlon® Pro PS Femoral Components are intended for cemented use only.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <b> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
| Sponsor | Stryker Orthopaedics<br>325 Corporate Drive<br>Mahwah, NJ 07430 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Jane Guo<br>Senior Regulatory Affairs Specialist<br>Howmedica Osteonics Corp<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Phone: +1 269-800-2081<br>Email: jane.guo@stryker.com |
| Alternate Contact | Shing Jen Tai<br>Principal Regulatory Affairs Specialist<br>Howmedica Osteonics Corp<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Phone: +1 269-800-2088<br>Email: shingjen.tai@stryker.com |
| Date Prepared: | April 3rd, 2023 |
| Proprietary Name: | Triathlon® Total Knee System – Triathlon® Pro Posterior Stabilized<br>(PS) Femoral Component |
| Common Name: | Total Knee Joint Replacement |
| Classification Name: | Knee joint patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented prosthesis (21 CFR 888.3560) |
| Product Codes: | JWH |
# Legally Marketed Primary Predicate Devices to Which Substantial Equivalence is Claimed:
o The components of the Triathlon® Total Knee System:
- Triathlon® Posterior Stabilized (PS) Femoral Component K190402 ■
# Legally Marketed Additional Predicate Devices Used to Support Substantial Equivalence:
- The components of the Triathlon® Total Knee System: O
- . Triathlon® Posterior Stabilized (PS) Femoral Component - K042993, K141056, K172326
# Legally Marketed Reference Devices Used to Support Substantial Equivalence:
- -PreludeTM PF Resurfacing Knee System - K123907
{4}------------------------------------------------
- 트 Triathlon® Cruciate Retaining (CR) Femoral Component - K040267, K141056, K172326
# Reason for 510(k) Submission:
The purpose of this submission is to introduce a new Triathlon® Pro Posterior Stabilized (PS) Femoral Component to the Triathlon® Total Knee System.
# Device Description:
The subject Triathlon® Pro PS Femoral Component is a line extension to the existing Triathlon® Total Knee System and will be a modified version of the predicate Triathlon® PS Femoral Components. The subject component is sterile, single use, and available in sizes 1-8 and in right and left configurations. The femoral component is manufactured from Cobalt-Chromium alloy, which meets ASTM material standard F2886-17 as well as F75-18 standard requirements in its final finished form. It is intended for cemented use only.
# Intended Use:
The subject PS femoral component has the same intended use as the predicate devices as specified in the 510(k) submissions for the predicate devices – the subject device is intended to be used in total knee arthroplasty and for cemented use only.
# Indications:
The subject PS femoral component has the same indications for use as the predicate devices:
General Total Knee Arthroplasty (TKR) Indications:
- · Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
- · Post-traumatic loss of knee joint configuration and function.
- · Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- · Revision of previous unsuccessful knee replacement or other procedure.
- Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture-management techniques.
Additional Indications for Posterior Stabilized (PS) Components:
- · Ligamentous instability requiring implant bearing surface geometries with increased constraint.
- · Absent or non-functioning posterior cruciate ligament.
- · Severe anteroposterior instability of the knee joint.
The Triathlon® Pro PS Femoral Components are intended for cemented use only.
# Summary of Technological Characteristics:
The subject device will be manufactured using a low-pressure metal injection molding process (Cobalt-Chromium alloy conforming to ASTM F2886-17) as opposed to the casting process (CoCr conforming to ASTM F75-18) utilized by the predicate devices. Reference device, K123907, utilizes this same manufacturing process. The device comparisons and performance testing show that the subject Triathlon® Pro PS Femoral Component is substantially equivalent to the cited predicate Triathlon® PS Femoral Components based on
{5}------------------------------------------------
intended use, indications for use, design, material, sterilization method, biocompatibility, MR compatibility, technical and performance characteristics, and operational principles.
# Non-Clinical Testing:
The following non-clinical laboratory testing and engineering analysis were performed to determine substantial equivalence:
- Material Characterization CoCr per ASTM F2886-17 and ASTM F75-18 ●
- Femoral Component Fatigue Testing
- o Wear Testing per ISO 14243-3:2014
- o Characterization of Tibiofemoral Contact Area/Contact Stress
- Characterization of Patellofemoral Contact Area/Contact Stress ●
- Characterization of Articulating/Bearing Surface Finish
- Characterization of Range of Motion and Constraint ●
- Biocompatibility evaluated per ISO 10993-1:2018 ●
- Shelf-life validated per the following standards: ●
- ISO 11607-1:2019 O
- O ISO 11607-2:2019
- ASTM F1980-21 O
- Bacterial endotoxin testing (BET) as specified in ANSVAAMI ST72:2019 was used for ● pyrogenicity testing to achieve an endotoxin limit of <20EU/Device.
- The subject Triathlon® Pro Posterior Stabilized (PS) Femoral Component has been evaluated according to the FDA guidance document entitled "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" issued May 20, 2021 and to the standards listed below:
- Magnetically Induced Displacement Force performed per ASTM F2052-15 O
- Magnetically Induced Torque performed per ASTM F2213-17 O
- Image Artifact per ASTM F2119-07(2013) O
- Heating by Radio Frequency Fields performed per ASTM F2182-19e2 O
# Clinical Testing:
Clinical testing was not required as a basis for substantial equivalence.
# Conclusion:
Based upon a comparison of the intended use, indications for use, design, material, sterilization method, biocompatibility, MR compatibility, technical and performance characteristics, and operational principles, the subject Triathlon® Pro Posterior Stabilized (PS) Femoral Component is substantially equivalent to the cited predicate Triathlon® PS Femoral Components.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.