The Triathlon™ CR Total Knee System consists of a femoral component, tibial insert, and all-polyethylene patellar components that are intended to be used with the previously cleared Triathlon™ Primary Cemented Tibial Tray in primary or revision total knee arthroplasty. The Triathlon™ All Polyethylene Patellar components are intended to be used with the femoral components of the previously released Duracon® Total Knee System, as well as the previously released Triathlon™ PS Femoral component in situations where replacement of the articular surface of the patella is required. The Triathlon™ CR Total Knee System is intended to accommodate the posterior cruciate ligament (PCL) if it is present. Specific indications and contraindications are listed below: Indications: Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or, rheumatoid arthritis Post-traumatic loss of knee joint configuration and function. Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability. Revision of previous unsuccessful knee replacement or other procedure. Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture management techniques.
Device Story
Total knee arthroplasty system; replaces articulating surfaces of distal femur and patella; preserves posterior cruciate ligament (PCL). Components: cobalt-chromium-molybdenum femoral component; ultra-high molecular weight polyethylene (UHMWPE) tibial insert and patellar components. Femoral component features grit-blasted interior for PMMA cement interdigitation and cast-in pegs for rotational stability. Tibial insert uses wire-tab locking mechanism to secure to previously cleared tibial baseplate. Used in OR by orthopedic surgeons for primary or revision knee replacement. Benefits: restores joint function, stability, and configuration in patients with degenerative or traumatic knee disease.
Indicated for patients with painful, disabling knee joint disease (osteoarthritis, traumatic arthritis, avascular necrosis, rheumatoid arthritis), post-traumatic loss of function, moderate varus/valgus/flexion deformity, revision of failed knee replacement, or distal femur/proximal tibia fractures. Contraindicated in patients with active/latent knee infection, distant infection foci, mental/neuromuscular disorders risking instability, compromised bone stock, skeletal immaturity, severe instability due to collateral ligament loss, or obesity.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Triathlon™ PS Total Knee System
Duracon® CR Femoral Component
Scorpio® CR Femoral Component
Duracon® A/P Lipped Tibial Insert
Duracon® Symmetric Patellar Component
Duracon® Asymmetric Patellar Component
Kinemax® All Polyethylene Patella
Submission Summary (Full Text)
{0}------------------------------------------------
# K040267 (P9 1 of 5)
Triathlon™ Cruciate Retaining (CR) Total Knee System 510(k) Premarket Notification ________________________________________________________________________________________________________________________________________________
Confidential
| MAY - 5 2004 | 510(k) Summary |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission Information | |
| Name and Address of Sponsor: | Howmedica Osteonics Corp.<br>DBA (doing business as) Stryker Orthopaedics<br>325 Corporate Drive<br>Mahwah, New Jersey 07430 |
| For Information contact: | Margaret F. Crowe<br>Regulatory Affairs Consultant<br>Howmedica Osteonics Corp.<br>(DBA Stryker Orthopaedics)<br>325 Corporate Drive<br>Mahwah, New Jersey 07430 |
| Device Identification | |
| Proprietary Name: | Triathlon™ Cruciate Retaining (CR) Total Knee<br>System |
| Common Name: | Cruciate Retaining Total Knee Replacement |
| Classification Name and Reference: | Knee Joint Patellofemorotibial<br>Polymer/Metal/Polymer Semi-Constrained<br>Cemented Prosthesis<br>21 CFR §888.3560 |
| Proposed Regulatory Class: | Class II |
| Device Product Code: | OR (87) JWH<br>Prosthesis, Knee Patellofemorotibial, Semi-<br>Constrained, Cemented, Polymer/Metal/ Polymer |
The Triathlon™ CR Total Knee System consists of a femoral component, tibial insert, and all-polyethylene patellar components that are intended to be used with the previously cleared Triathlon™ Primary Cemented Tibial Tray in primary or revision total knee arthroplasty. The Triathlon™ All Polyethylenc Patcllar components are intended to be used with the femoral components of the previously released Duracon® Total Knce System, as well as the previously released Triathlon™ PS Femoral component in
{1}------------------------------------------------
# K040267 ( Pop 2 of 5
## Triathlon™ Cruciate Retaining (CR) Total Knee System
## 510(k) Premarket Notification
Confidential
situations where replacement of the articular surface of the patella is required. The Triathlon™ CR Total Knee System is intended to accommodate the posterior cruciate ligament (PCL) if it is present. Specific indications and contraindications are listed below:
Indications:
- · Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or, rheumatoid arthritis
- · Post-traumatic loss of knee joint configuration and function.
- · Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- · Revision of previous unsuccessful knee replacement or other procedure.
- Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture management techniques.
## Contraindications
- · Any active or suspected latent infection in or about the knee joint.
- Distant foci of infection which may cause hematogenous spread to the implant site
- · Any mental or neuromuscular disorder which would create an unacceptable risk of prosthesis instability, prosthesis fixation failure, or complications in postopcrative care.
- · Bone stock compromised by disease, infection or prior implantation that cannot provide adequate support and/or fixation to the prosthesis.
- Skeletal immaturity. .
- · Severe instability of the knec joint secondary to the absence of collateral ligament integrity and function.
- · Obesity. An overweight or obesc patient can produce loads on the prosthesis that can lead to failure of the fixation of the device or to failure of the device itself.
## Device Description
{2}------------------------------------------------
## K040267 (89 3 of 5)
### Triathlon™ Cruciate Retaining (CR) Total Knee System Confidential 510(k) Premarket Notification ________________________________________________________________________________________________________________________________________________
The Triathlon™ Cruciate Retaining (CR) Total Knee System consists of three primary components: Triathlon™ Cruciate Retaining (CR) Femoral Component, Triathlon™ Cruciate Retaining (CR) Tibial Insert, and Triathlon™ Patellar components (available in two styles – symmetric and asymmetric). These components are intended to be used with the previously cleared Triathlon™ Primary Cemented Tibial Tray in primary or revision total knee arthroplasty. The Triathlon™ Patellar components are intended to be used with both the Triathlon™ CR Femoral component, the previously cleared Triathlon™ PS Femoral component, or the femoral components of the Duracon® Total Knee System. Duracon® patellar components may be used with both the Triathlon™ PS and Triathlon™ CR femoral components. The use of a patellar component is optional. The Triathlon™ CR components are described below:
The Triathlon" Total Knec Cruciate Retaining (CR) Femoral Component is fabricated from cast cobalt-chromium-molybdenum alloy, and is intended for cemented application to replace the articulating surface of the distal femur. This cruciate retaining femoral component is utilized when total knee replacement is indicated, and accommodates the posterior cruciate ligament if it is present.
The Triathlon"" Total Knee Cruciate Retaining (CR) Femoral Component is available in right and left configurations, and eight proportional sizes (sizes 1 to 8) to accommodate differences in patient anatomy. The interior surface of the component is grit-blasted to increase surface roughness - this is intended to promote interdigitation of the polymethylmethacrylate (PMMA) bone cement with the surface texture and the apposing bone. This femoral component features cast-in pegs to help in femoral component placement, and to provide rotational stability.
The Triathlon" Cruciate Retaining (CR) Tibial Insert is neutral in configuration, and is available in seven proportional sizes (sizes 2 to 8) and varying thicknesses (9mm, 11mm, 13mm, 16mm, and 19mm). The insert is fabricated from ultra high molecular weight polyethylene and cobalt-chromium alloy. The minimum thickness of the tibial insert on the bearing surface is 6mm. The tibial inscrt is designed to accommodate the posterior
{3}------------------------------------------------
#### 4 of 5) K040267 ( pg
## Triathlon™ Cruciate Retaining (CR) Total Knee System 510(k) Premarket Notification ________________________________________________________________________________________________________________________________________________
Confidential
cruciate ligament if it is present. There is a relief on the anterior aspect of the tibial insert to accommodate the patellar tendon and patellar fat pad.
The Triathlon™ CR Tibial Inscrt incorporates a locking wire feature on the anterior aspect of the insert. This locking wire is fabricated from cobalt-chromium alloy, and engages under tabs on the anterior rim of the Triathlon" " Primary Tibial Baseplate. This wire-tab locking mechanism secures the insert into the baseplate. This insert-baseplate locking mechanism is identical to the locking mechanism utilized on the Triathlon™ PS Total Knee System.
Triathlon™ Patellar components are available in two styles: symmetric and asymmetric. Both are fabricated from ultra high molecular weight polyethylene. The symmetric design is available in six diameters (27mm, 31mm, 31mm, 36mm, 36mm, and 39mm) and four thicknesses (8mm, 9mm, 10mm, and 11mm). The symmetric design features a central cement recess, and three pegs on the bone interface surface. The Triathlon™ Symmetric patellar component is intended to be implanted via a resurfacing or inset surgical technique. The asymmetric design is available in five superior/inferior dimensions (29mm, 32mm, 35mm, 38mm, and 40mm) and three thicknesses (9mm, 10mm, and 11mm). The asymmetric design features a flare on the latcral aspect of the component. The asymmetric patcllar component also has a central recess with three pegs for cement fixation. The Triathlon™ Asymmetric patellar component is intended to be implanted using a resurfacing surgical technique.
Equivalent products include:
- 1. Triathlon™ PS Total Knec System
- 2. Duracon® CR Femoral Component
- 3. Scorpio® CR Femoral Component
- 4. Duracon® A/P Lipped Tibial Insert
- 5. Duracon® Symmetric Patellar Component
- 6. Duracon® Asymmetric Patellar Component
- Kinemax® All Polyethylene Patella 7.
{4}------------------------------------------------
K040267 (pg 5 of 5)
Triathlon™ Cruciate Retaining (CR) Total Knce System
510(k) Premarket Notification_________________________________________________________________________________________________________________________________________________
Confidential
Testing was presented to support a claim of substantial equivalence to the predicate devices.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized wing-like shapes.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY - 5 2004
Ms. Margaret F. Crowe Regulatory Affairs Consultant Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, New Jersey 07430
Re: K040267
KU40267
Trade/Device Name: Triathlon™ Cruciate Retaining (CR) Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: 21 CF X 866.5500
Regulation Name: Knee joint, patellofemorotibial, polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: II Product Code: JWH Dated: February 4, 2004 Received: February 5, 2004
Dear Ms. Crowc:
We have reviewed your Section 510(k) premarket notification of intent to market the device we have teviewed your becamed the device is substantially equivalent (for the indications felerenced above and nave actoring ally marketed predicate devices marketed in interstate for use stated in the encreative to regars) as tegally the Medical Device Amendments, or to conninered prior to May 20, 2017 11:11 accordance with the provisions of the Federal Food, Drug, devices mat have been recuired in avere approval of a premarket approval application (PMA). and Cosmetic rear (11ct) that as nevice, subject to the general controls provisions of the Act. The r ou may, merelevely mains of the Act include requirements for annual registration, listing of general controls profile proctice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can may or sabyee to does friederal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalcnce determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or uns I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Ms. Margaret F. Crowe
This letter will allow you to begin marketing your device as described in your Section 510(k) rms retect notification. The FDA finding of substantial equivalence of your device to a legally premaince device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark A. Wilkerson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
K040267 (pg 1 of 2)
Indications for U
510(k) Number (if known): K040267
Device Name: Triathlon™ CR Total Knee System
Indications For Use:
Division of General, Restorative and Neurological Devices
**510(k) Number** K040267
The Triathlon" CR Total Knec System consists of a femoral component, tibial insert, and allpolyethylene patellar components that are intended to be used with the previously cleared Triathlon™ Primary Cemented Tibial Tray in primary or revision total knee arthroplasty. The Triathlon™ All Polycthylene Patellar components are intended to be used with the femoral components of the previously released Duracon® Total Knee System, as well as the previously released Triathlon™ PS Femoral component in situations where replacement of the articular surface of the patella is required. The Triathlon™ CR Total Knee System is intended to accommodate the posterior cruciate ligament (PCL) if it is present. Specific indications and contraindications are listed below:
Indications:
- · Painful, disabling joint disease of the knee resulting from: nonintlammatory degenerative joint discass: (including osteoarthritis, traumatic arthritis or avascular necrosis) or, theumatoid arthrius
- · Post-traumatic loss of knee joint configuration and function.
- · Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- · Revision of previous unsuccessful knee replacement or other procedure.
- · Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture management techniques.
### Contraindications
- · Any active or suspected latent infection in or about the knee joint.
- · Distant foci of infection which may cause hematogenous spread to the implant site
- · Any mental or neuromuscular disorder which would create an unacceptable risk of prostlyesis instability, prosthesis fixation failure, or complications in postoperative care,
- · Bone stock compromised by disease, infection or prior implantation that cannot provide adequate support and/or fixation to the prosthesis.
- Skeletal immaturity. .
- Severe instability of the knec joint secondary to the absence of collateral ligament integrity and . function.
- Obesity. An overweight or obese patient can produce loads on the prosthesis that can lead to failure of the fixation of the device or to failure of the device itself.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Uso (21 CFR 807 Subpart C)
ge lof
{8}------------------------------------------------
K040267 (pg 2 of 2)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark A. Milleson
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number K040267
Page of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.