The RED Light Device is an OTC device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles.
Device Story
Battery-operated, handheld LED device for home-use treatment of periorbital wrinkles. Emits red (633 ±5nm) and infrared (830 ±5nm) light via 25 LEDs over 17 cm² area. Constant waveform; power density 125 mW/cm² (70 mW/cm² red; 55 mW/cm² IR). Device includes handpiece, charging base, and AC adapter. User operates device by applying light energy to treatment zones for 5-7 minutes per zone, 3 times weekly for the first month. No clinical data provided; usability study confirmed user ability to self-select, set up, perform light sensitivity test, operate, and clean device.
Clinical Evidence
No clinical data. Bench testing only. Usability/Label Comprehension Study with 35 participants demonstrated 100% success in self-selection, setup, light sensitivity testing, operation, and cleaning; 95% success in questionnaire comprehension.
Technological Characteristics
Materials: ABS and stainless steel. Light source: 25 LEDs (633nm/830nm). Power: 125 mW/cm² total density. Energy: 2x3.7V Lithium battery or 100-240V AC adapter. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57, IEC 60601-1-11, IEC 62471, ISO 10993-5, ISO 10993-10. Software: Moderate level of concern.
Indications for Use
Indicated for individuals with periorbital wrinkles for use in dermatology.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines suggesting hair or movement. The profiles are arranged in a way that they appear to be connected. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 13, 2017
Zhongshan Bisen Plastic Electronic Products Co., Ltd. c/o Mr. Ray Wang Beijing Believe Technology Service Co., Ltd. 5-1206, Build 332, DaFangJu, No. 25 BanBiDian Rd. LiYuan Town, TongZhou District, Beijing 101121 China
Re: K162489
Trade/Device Name: RED Light Device Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHS Dated: December 2, 2016 Received: December 7, 2016
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
Page 2 - Mr. Ray Wang
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K162489
Device Name
RED Light Device
Indications for Use (Describe)
The RED Light Device is an OTC device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
### Tab #7 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: __
- 1. Date of Preparation
08/23/2016
- Sponsor 2.
### Zhongshan Bisen Plastic Electronic Products Co.,Ltd.
2/F,Building1,No.81,Dongzhen North Road,Zhangjiabian 1st village,Torch Development Zone,Zho ngShan City,GuangDong,China Establishment Registration Number: Not yet registered or the Number Contact Person: FengPing Li Position: General Manager Tel: 86-760-88297142 Fax:86- 760-88297142 Email: binzim@163.com
- Submission Correspondent 3.
Mr. Ray Wang
### Beijing Believe Technology Service Co., Ltd.
5-1206, Build 332, DaFangJu, No.25 BanBiDian Rd., LiYuan Town, TongZhou District, Beijing, 101121, China Tel: +86-21-50313932 Fax: +86-21-68093116 Email: ray.wang@believe-med.com
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: RED Light Device Common Name: Light Emitting Diode (LED) Device Model(s): BZ-0606
#### Regulatory Information:
Classification Name: Light Based Over The Counter Wrinkle Reduction Classification: II; Product Code: OHS; Regulation Number: 21 CFR 878.4810; Review Panel: General & Plastic Surgery;
### Indication for Use Statement:
The RED Light Device is an OTC device indicated to emit energy in the red and IR region of the spectrum for use in dermatology for the treatment of periorbital wrinkles.
#### Device Description 5.
The proposed device, RED Light Device is a battery operated device that uses low power light spectrum at red and infrared LED, at wavelength of 633 ±5nm, 830 ±5nm emitting optical power in a uniform distribution with no hot spots.
The device is composed of a handpiece for delivery of light energy, base unit for charging and storage when not in use, and A.C. charging adapter.
It is a hand held light emitting diode (LED) device for the treatment of periorbital wrinkles designed for home-use.
The proposed device is NOT for life-supporting or life-sustaining, not for implant. The proposed device is NOT provided sterile and is NOT a reprocessed single-use device.
{5}------------------------------------------------
#### 6. Identification of Predicate Device
Predicate 510(k) Number: K152332 Product Name: Perfectio LED Infrared Device Manufacturer: OMM IMPORTS INC DBA ZERO GRAVITY
#### 7. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
IEC 60601-1:2012, Medical electrical equipment-Part 1: General requirements for basic safety, and essential performance.
IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
IEC 60601-2-57:2011, Medical electrical equipment - Part 2-57: Particular requirements for basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use.
IEC 60601-1-11:2010, Medical electrical equipment - Part 1-11: General requirements for medical device equipment and medical electrical systems used in the home healthcare environment.
IEC 62471:2006 , Photobiological Safety of lamps and lamp systems.
ISO 10993-5: 2009, Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity.
ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
Performance Test: Temperature Test
A Usability/Label Comprehension Study was conducted with 35 participants.
The results of the study showed that (100%) were able to:
-Correctly self-select as being an appropriate user of the device.
-Correctly demonstrate how to set up the device, perform the Light Sensitivity Test, operate the device (apply Tx), and clean the device.
And that (95%) of participants were able to correctly answer each question for the Questionnaire portion of the Study.
- Clinical Test Conclusion 8.
Clinical data was not included in this submission.
{6}------------------------------------------------
# 9. Substantially Equivalent (SE) Comparison
| ITEM | Proposed Device | Predicate Device | Remark |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Product Code | OHS | OHS | SE |
| Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 | SE |
| Class | 2 | 2 | SE |
| OTC use | Yes | Yes | SE |
| Indication for<br>Use | The RED Light Device is an OTC device<br>indicated to emit energy in the red and IR<br>region of the spectrum for use in<br>dermatology for the treatment of<br>periorbital wrinkles. | Perfectio LED infrared device is an<br>over the counter device indicated to<br>emit energy in the red and IR region<br>of the spectrum for use in<br>dermatology for the treatment of<br>periorbital wrinkles. | SE |
# Table 1 General Comparison
{7}------------------------------------------------
| ITEM | Proposed Device | Predicate Device | Remark |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Handheld | Yes | Yes | SE |
| Contacting<br>Materials | ABS & Stainless Steel | ABS & Stainless Steel | SE |
| Measurement<br>wavelength | Red: 633 ±5nm<br>Infrared: 830 ±5nm | Red: 633 ±5nm<br>Infrared: 830 ±5nm | SE |
| Light Source | Light Emitting Diode (LED) | Light Emitting Diode (LED) | SE |
| Waveform | Constant | Constant | SE |
| Energy Source | 25 LEDs over 17 cm² | 25 LEDs over 17 cm² | SE |
| Power Density | 125 mW/cm2<br>70 mW/cm2 (633 nm); 55 mW/cm2 (830 nm) | 125 mW/cm2<br>70 mW/cm2 (633 nm); 55 mW/cm2 (830 nm) | SE |
| Power Supply | Adaptor: 100~240V AC 50/60Hz<br>Lithium battery: 2x3.7V | Adaptor: 100~240V AC 50/60Hz<br>Lithium battery: 2x3.7V | SE |
| Initial Treatment<br>Course | For the first month (4 weeks), treatment<br>should be performed 3 times a week for<br>15-20 minutes each time.(5-7 minutes on<br>each treatment zone). | For the first month (4 weeks), treatment<br>should be performed 3 times a week for<br>15-20 minutes each time.(5-7 minutes on<br>each treatment zone). | SE |
| Maintenance<br>Regime | Individuals with periorbital lines and<br>wrinkles | Individuals with periorbital lines and<br>wrinkles | SE |
Table 2 Performance Comparison
{8}------------------------------------------------
| ITEM | Proposed Device | Predicate Device | Remark |
|-------------------|----------------------------|----------------------------|--------|
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 | SE |
| | Comply with IEC 60601-1-11 | Comply with IEC 60601-1-11 | |
| | Comply with IEC 62471 | Comply with IEC 62471 | |
| | Comply with IEC 60601-2-57 | / | |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | SE |
| Biocompatibility | Comply with ISO 10993-1 | Comply with ISO 10993-1 | SE |
| Label and | Conforms to FDA Regulatory | Conforms to FDA Regulatory | SE |
| Labeling | Requirements | Requirements | |
| Level of Concern | Moderate | Moderate | SE |
| of the Software | | | |
# Table 3 Safety Comparison
# 10. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.