Infrared light therapy led facial light therapy mask-Derma plus, Photons Facial Skin Care Machine LED Beauty Mask Therapy Beauty Machine
K200983 · Ningbo Hesi Electric Co., Ltd. · OHS · Jul 14, 2021 · General, Plastic Surgery
Device Facts
Record ID
K200983
Device Name
Infrared light therapy led facial light therapy mask-Derma plus, Photons Facial Skin Care Machine LED Beauty Mask Therapy Beauty Machine
Applicant
Ningbo Hesi Electric Co., Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Jul 14, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LED FACIAL LIGHT THERAPY MASK (Model: HK207) is intended to: - The device emitting energy in the blue is intended to reduce the mild to moderate inflammatory acne vulgaris. - The device emitting energy in the red and infrared spectrum is intended for the treatment of full-face wrinkles. The FLEXIBLE LED LIGHT THERAPY (Model: HK209) is intended to: - The device emitting energy in the blue is intended to reduce the mild to moderate inflammatory acne vulgaris. - The device emitting energy in the red and infrared spectrum is intended for the treatment of full-face wrinkles - The device is intended to deliver heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
Device Story
Wearable LED phototherapy devices (mask HK207; flexible panel HK209) for home use; intended for aesthetic and therapeutic applications. Devices emit specific light wavelengths: blue (465nm) for acne; red (640nm) and infrared (880nm) for wrinkle reduction; infrared for topical heating. User-operated by adults; devices fit anatomical contours using biocompatible materials. Output consists of light energy and heat; intended to elevate tissue temperature, promote muscle relaxation, and increase local blood circulation. Clinical decision-making is patient-driven based on aesthetic or minor pain relief needs. Benefits include non-invasive management of acne, wrinkle appearance, and temporary pain relief.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility (ISO 10993-5, ISO 10993-10) and electrical safety/EMC (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57).
Technological Characteristics
Wearable LED phototherapy device; biocompatible materials; wavelengths 465nm, 640nm, 880nm; irradiance 6.5 mW/cm²; dose 11.7 J/cm²; power input 100–240Vac, 50/60Hz. Compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 60601-2-57.
Indications for Use
Indicated for adults with mild to moderate inflammatory acne vulgaris (blue light), full-face wrinkles (red/infrared light), and for topical heating to relieve minor muscle/joint pain, arthritis, muscle spasm, stiffness, and to increase local blood circulation (infrared light).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 14, 2021
Ningbo Hesi Electric Co., Ltd Ms. Cassie Lee Guangzhou GLOOMED Biological Technology Co., Ltd. 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou, Guangdong 511495 China
Re: K200983
Trade/Device Name: LED facial light therapy mask (Model: HK207), Flexible LED light therapy (Model: HK209) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS, OLP, ILY Dated: April 2, 2020 Received: April 14, 2020
Dear Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200983
#### Device Name
LED FACIAL LIGHT THERAPY MASK (Model: HK207), FLEXIBLE LED LIGHT THERAPY (Model: HK209)
Indications for Use (Describe)
The LED FACIAL LIGHT THERAPY MASK (Model: HK207) is intended to:
- The device emitting energy in the blue is intended to reduce the mild to moderate inflammatory acne vulgaris.
- The device emitting energy in the red and infrared spectrum is intended for the treatment of full-face writtles.
The FLEXIBLE LED LIGHT THERAPY (Model: HK209) is intended to:
- The device emitting energy in the blue is intended to reduce the mild to moderate inflammatory acne vulgaris.
- The device emitting energy in the red and infrared spectrum is intended for the treatment of full-face wrinkles
- The device is intended to deliver heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stifficess; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Sponsor: | NINGBO HESI ELECTRIC CO., LTD |
|-----------------|--------------------------------------------------------------------------------------------|
| Subject Device: | LED FACIAL LIGHT THERAPY MASK (Model: HK207), FLEXIBLE LED LIGHT<br>THERAPY (Model: HK209) |
| Document Name: | 510(k) Summary |
| K number: | K200983 |
# 510(k) Summary of K200983
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
# 1. Submitter's Information
Sponsor Name: NINGBO HESI ELECTRIC CO., LTD Establishment Registration Number: 3009784099 Address: NO.818-23-156 Qiming Rd, Yinzhou, Ningbo City Zhejiang, CHINA Postal code: 315000 Contact name: MA LULU (General Manager) Tel: 15825567078 Fax: 0574-88300553 E-mail: 463415782@gq.com
## 2. Date of the summary prepared: July 13, 2021
### 3. Application Correspondent:
Contact Person: Ms. Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou, Guangdong, China Tel: +86 20 8266 2446 Email: regulatory@glomed-info.com
# 4. Subject Device Information
Type of 510(k): Traditional Classification Name: Light Based Over-The Counter Wrinkle Reduction (OHS); Over-The-Counter Powered Light Based Laser For Acne (OLP); Lamp, Infrared, Therapeutic Heating (ILY) Trade Name: LED FACIAL LIGHT THERAPY MASK (Model: HK207), FLEXIBLE LED LIGHT THERAPY (Model: HK209) Review Panel: General & Plastic Surgery Product Code: OHS, OLP, ILY
{4}------------------------------------------------
| Sponsor: | NINGBO HESI ELECTRIC CO., LTD |
|-----------------|--------------------------------------------------------------------------------------------|
| Subject Device: | LED FACIAL LIGHT THERAPY MASK (Model: HK207), FLEXIBLE LED LIGHT<br>THERAPY (Model: HK209) |
| Document Name: | 510(k) Summary |
| K number: | K200983 |
Regulation Number: 890.5500, 878.4810 Regulatory Class: II
### 5. Predicate Device Information
| Sponsor | Biophotas, Inc | Biophotas, Inc |
|-----------------------|--------------------|--------------------|
| Device Name and Model | BioPhotas Celluma³ | BioPhotas Celluma³ |
| 510(k) Number | K152280 | K171323 |
| Product Code | ILY, OHS, OLP | OHS |
| Regulation Number | 890.5500, 878.4810 | 878.4810 |
| Regulation Class | II | II |
# 6. Device Description
LED FACIAL LIGHT THERAPY MASK (Model: HK207) is a home use wearable LED phototherapy device designed to fit the contours of the target areas of the anatomy, covered with a biocompatible material, which uses specific wavelengths of light to manage aesthetic conditions. This device produces light in the blue light (465nm) is intended to help reduce the appearance of mild to moderate inflammatory acne. The red light (640nm) in combination with infrared light (880nm) is intended to improve the appearance of wrinkles.
FLEXIBLE LED LIGHT THERAPY (Model: HK209) is a highly shapeable LED panel designed to fit the contours of the target areas of the anatomy, covered with a biocompatible material, which uses specific wavelengths of light to manage aesthetic conditions. This device produces light in the near-infrared region of the spectrum (880nm) intended to provide topical heating to temporary pain relief. Blue light (465nm) is intended to help reduce the appearance of mild to moderate inflammatory acne. Red light (640nm) in combination with infrared light (880nm) is intended to improve the appearance of wrinkles.
### 7. Intended Use / Indications for Use
The LED FACIAL LIGHT THERAPY MASK (Model: HK207) is intended to:
- The device emitting energy in the blue is intended to reduce the mild to moderate inflammatory acne vulgaris.
- -The device emitting enerqy in the red and infrared spectrum is intended for the treatment of full-face wrinkles.
{5}------------------------------------------------
| Sponsor: | NINGBO HESI ELECTRIC CO., LTD |
|-----------------|--------------------------------------------------------------------------------------------|
| Subject Device: | LED FACIAL LIGHT THERAPY MASK (Model: HK207), FLEXIBLE LED LIGHT<br>THERAPY (Model: HK209) |
| Document Name: | 510(k) Summary |
| K number: | K200983 |
The FLEXIBLE LED LIGHT THERAPY (Model: HK209) is intended to:
- -The device emitting energy in the blue is intended to reduce the mild to moderate inflammatory acne vulgaris.
- The device emitting energy in the red and infrared spectrum is intended for the treatment of full-face wrinkles
- The device is intended to deliver heat in the IR spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
### 8. Comparison to predicate device
Compare with predicate devices, the subject device is very similar in design principle, intended use, indications for use, functions, material, and the applicable standards. The differences between subject devices and predicate devices do not raise new questions of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Verdict |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------------|
| Company | Ningbo Hesi Electronic<br>Co., Ltd | Biophotas, Inc | Biophotas, Inc | -- |
| Trade Name | LED FACIAL LIGHT<br>THERAPY MASK<br>(Model: HK207),<br>FLEXIBLE LED LIGHT<br>THERAPY (Model:<br>HK209) | BioPhotas Celluma3 | BioPhotas Celluma3 | -- |
| 510(k) Number | K200983 | K152280 | K171323 | -- |
| Product Code | OHS, OLP, ILY | ILY, OHS, OLP | OHS | SE |
| Intended Use /<br>Indications for Use | The LED FACIAL<br>LIGHT THERAPY<br>MASK (Model: HK207)<br>is intended to:<br>- The device<br>emitting energy in | The BioPhotas Celluma3<br>is intended to deliver<br>heat in the IR spectrum<br>to provide topical<br>heating for the purpose<br>of elevating tissue | The BioPhotas<br>Celluma3<br>is intended to emit<br>energy in the visible<br>and<br>infrared region of | SE<br>Note 1 |
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Verdict |
| | the blue is<br>intended to reduce<br>the mild to<br>moderate<br>inflammatory acne<br>vulgaris. | temperature; for the<br>temporary relief of minor<br>muscle and joint pain,<br>arthritis and muscle<br>spasm; relieving<br>stiffness; promoting the<br>relaxation of muscle<br>tissue; and to<br>temporarily increase<br>local blood circulation. | the<br>spectrum for use in<br>the<br>treatment of full face<br>wrinkles. | |
| | - The device<br>emitting energy in<br>the red and<br>infrared spectrum<br>is intended for the<br>treatment of full-<br>face wrinkles.<br>The FLEXIBLE LED<br>LIGHT THERAPY<br>(Model: HK209) is<br>intended to: | The blue spectrum light<br>is intended to<br>reduce mild to moderate<br>inflammatory acne<br>vulgaris. The Celluma³<br>is intended to emit<br>energy in the red and<br>infrared spectrum for<br>use in dermatology for<br>the treatment of<br>periorbital wrinkles. | | |
| | - The device<br>emitting energy in<br>the blue is<br>intended to reduce<br>the mild to<br>moderate<br>inflammatory acne<br>vulgaris. | | | |
| | - The device<br>emitting energy in<br>the red and<br>infrared spectrum<br>is intended for the<br>treatment of full- | | | |
{6}------------------------------------------------
Sponsor:
Subject Device:
Document Name: K number:
NINGBO HESI ELECTRIC CO., LTD LED FACIAL LIGHT THERAPY MASK (Model: HK207), FLEXIBLE LED LIGHT THERAPY (Model: HK209) 510(k) Summary K200983
{7}------------------------------------------------
| Sponsor: | NINGBO HESI ELECTRIC CO., LTD |
|-----------------|--------------------------------------------------------------------------------------------|
| Subject Device: | LED FACIAL LIGHT THERAPY MASK (Model: HK207), FLEXIBLE LED LIGHT<br>THERAPY (Model: HK209) |
| Document Name: | 510(k) Summary |
| K number: | K200983 |
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Verdict |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|---------|
| | face wrinkles<br>- The device is<br>intended to deliver<br>heat in the IR<br>spectrum to<br>provide topical<br>heating for the<br>purpose of<br>elevating tissue<br>temperature; for<br>the temporary<br>relief of minor<br>muscle and joint<br>pain, arthritis and<br>muscle spasm;<br>relieving stiffness;<br>promoting the<br>relaxation of<br>muscle tissue; and<br>to temporarily<br>increase local<br>blood circulation. | | | |
| Regulation<br>Number | 890.5500, 878.4810 | 890.5500, 878.4810 | 878.4810 | SE |
| Regulation Name | Light Based Over-The<br>Counter Wrinkle<br>Reduction (OHS); Over<br>-The-Counter Powered<br>Light Based Laser For<br>Acne (OLP); Lamp,<br>Infrared, Therapeutic | Light Based Over-The<br>Counter Wrinkle<br>Reduction (OHS); Over-<br>The-Counter Powered<br>Light Based Laser For<br>Acne (OLP); Lamp,<br>Infrared, Therapeutic | Light Based Over –<br>The Counter<br>Wrinkle Reduction<br>(OHS) | SE |
{8}------------------------------------------------
| Sponsor: | NINGBO HESI ELECTRIC CO., LTD |
|-----------------|--------------------------------------------------------------------------------------------|
| Subject Device: | LED FACIAL LIGHT THERAPY MASK (Model: HK207), FLEXIBLE LED LIGHT<br>THERAPY (Model: HK209) |
| Document Name: | 510(k) Summary |
| K number: | K200983 |
| Elements of<br>Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Verdict |
|-------------------------------------------------------------|-----------------------------------------------------------|------------------------------------------|-------------------------------------------------------|--------------|
| | Heating (ILY) | Heating (ILY) | Heating (ILY) | |
| Regulatory Class | Class II | Class II | Class II | SE |
| Treatment areas | Entire Face and body | Entire Face and body | Whole Face | SE<br>Note 3 |
| Intended<br>population | Adult | Adult | Adult | SE |
| Power Source(s) | Input: 100 – 240Vac,<br>2.0 A, 50/60Hz | 110 – 120V | 110-120V | SE<br>Note 2 |
| Wavelength(s)(nm) | 465nm, 640nm, 880nm | 465nm, 640nm, 880nm | Red: 640nm±25nm<br>NIR: 880nm±50nm | SE |
| Irradiances<br>(mW/cm²) | 6.5 mW/cm² | 6.5 mW/cm² | 6.5 mW/cm² | SE |
| Treatment Dose<br>(J/cm²) | 11.7 J/cm² | 11.7 J/cm² | 11.7 J/cm² | SE |
| Treatment time | 3 times a week for 30<br>min.<br>4 weeks | 3 times a week for 30<br>min.<br>4 weeks | 3 treatments per<br>week<br>(1800 seconds)<br>4 weeks | SE |
| The distance<br>between the LEDs<br>to treatment<br>surface | For model: HK207 0.5-<br>3cm<br>For model: HK209 10<br>cm | As closed to the skin | As closed to the skin | SE |
| Irradiation area | HK207: 510 cm²<br>HK209: 890 cm² | 820 cm² | 15" x 8" | SE<br>Note 3 |
| Electrical safety | IEC 60601-1<br>IEC 60601-1-2 | IEC 60601-1<br>IEC 60601-1-2 | IEC 60601-1<br>IEC 60601-1-2 | SE |
# Comparison in Detail(s):
Note 1:
{9}------------------------------------------------
| Sponsor: | NINGBO HESI ELECTRIC CO., LTD |
|-----------------|--------------------------------------------------------------------------------------------|
| Subject Device: | LED FACIAL LIGHT THERAPY MASK (Model: HK207), FLEXIBLE LED LIGHT<br>THERAPY (Model: HK209) |
| Document Name: | 510(k) Summary |
| K number: | K200983 |
Although the Intended Use is slightly different from the predicate devices in the description, we can find that the subject device emits the same wavelengths and the same intended purpose as the predicate device. So, the slight differences will not raise any safety or effectiveness issues.
### Note 2:
Although the "Power Source(s)" of the subject device is a little different from the predicate devices, they are all compliant with IEC 60601-2-57 requirements and the range of power source can be safely used in the USA area. So, the differences will not raise any safety or effectiveness issues.
### Note 3:
Although the "Treatment area" is a little different from the predicate devices, it's just the factor depends on the shape of the product which will not affect the safety and effectiveness, and they all compliant with IEC 60601-2-57 requirements. So, the differences will not raise any safety or effectiveness issues.
### 9. Test Summary
1) Bench test:
The LED FACIAL LIGHT THERAPY MASK (Model: HK207) and FLEXIBLE LED LIGHT THERAPY (Model: HK209) have been evaluated the safety and performance by lab bench testing as following:
- � ISO 10993-5:2009, biological evaluation of medical devices -- part 5: tests for in vitro cytotoxicity. (Biocompatibility).
- � ISO 10993-10 2010, biological evaluation of medical devices - part 10: tests for irritation and skin sensitization. (Biocompatibility).
- � IEC 60601-1: 2005+A1:2012, Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance .
- � IEC 60601-1-2: 2014-02, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - collateral standard: Electromagnetic Compatibility.
- � IEC 60601-1-11 (Edition2.0): 2015-01, Medical electrical equipment -- part 1-11: General requirements for basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- � IEC 60601-2-57 (First Edition): 2011 for use in conjunction with IEC 60601-1:2005, Medical electrical equipment - Part 2: Particular requirements for the basic safety and essential performance of non-
{10}------------------------------------------------
| Sponsor: | NINGBO HESI ELECTRIC CO., LTD |
|-----------------|--------------------------------------------------------------------------------------------|
| Subject Device: | LED FACIAL LIGHT THERAPY MASK (Model: HK207), FLEXIBLE LED LIGHT<br>THERAPY (Model: HK209) |
| Document Name: | 510(k) Summary |
| K number: | K200983 |
laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
2) Clinic test
N/A
# 10. Conclusion:
The subject device K200983 "LED FACIAL LIGHT THERAPY MASK (Model: HK207), FLEXIBLE LED LIGHT THERAPY (Model: HK209)" is Substantial Equivalent to the predicate devices K152280 and K171323.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.