The AeviceMD is a non-invasive battery-operated device, including a wearable component, intended to longitudinally acquire, record and store lung sounds from adult patients in a clinical or non-clinical setting. The device stores the data for later playback, review, and analysis by a clinician and comparison with earlier data from the same patient.
Device Story
AeviceMD is a wearable electronic stethoscope system for adult patients. It consists of a sensor, silicone patch, docking station, mobile/web apps, and cloud platform. The sensor detects lung sounds and transmits data via Bluetooth to a gateway (docking station), which uploads to the cloud. Healthcare professionals (HCP) use the AeviceMD HCP Web App to review, playback, and analyze recorded lung sounds to support clinical decision-making. Patients use the mobile app to view recorded sounds for sharing with HCPs. The device is intended for longitudinal monitoring; it does not provide alarms, is not for emergency use, is not for sleep apnea, and is not for self-diagnosis. It facilitates remote or in-clinic auscultation, allowing clinicians to compare current lung sounds with historical data to track patient status.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included frequency response, stethoscope performance against a cleared reference device, human factors usability, shipping validation (ASTM D4169-16), cleaning validation, and biocompatibility (ISO 10993-5, ISO 10993-10). Electrical safety and electromagnetic compatibility verified per IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-11.
Technological Characteristics
Wearable electronic stethoscope. Materials: Silicone patch (biocompatible). Connectivity: Bluetooth (sensor to gateway), Cloud-based platform. Software: Embedded firmware and mobile/web applications. Standards: ISO 10993-5/10, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-1-6, ANSI AAMI IEC 62366-1, ANSI AAMI IEC 62304, ISO 14971. Power: Battery-operated.
Indications for Use
Indicated for longitudinal acquisition, recording, and storage of lung sounds in adult patients in clinical or non-clinical settings for clinician review and comparison.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 7, 2023
Aevice Health Pte. Ltd. Adrian Ang CEO 18 Howard Road, #06-11 Novelty Bizcentre Singapore, Singapore 369585 Singapore
Re: K223382
Trade/Device Name: AeviceMD Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: DSH, DOD Dated: November 7, 2022 Received: November 7, 2022
Dear Adrian Ang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ethan L. Nyberg -S
Ethan Nyberg, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223382
Device Name AeviceMD
Indications for Use (Describe)
The AeviceMD is a non-invasive battery-operated device, including a wearable component, intended to longitudinally acquire, record and store lung sounds from adult patients in a clinical setting. The device stores the data for later playback, review, and analysis by a clinician and comparison with earlier data from the same patient.
Type of Use (Select one or both, as applicable)
| <span></span> | <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|--------------------------------------------------------------------------------------|
| <span></span> | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
## I. SUBMITTER
Aevice Health Pte. Ltd. 18 Howard Road, #06-11 Novelty Bizcentre Singapore 369585 Tel: +65 90058825 Fax: N/A
Company Contact: Adrian Ang Chief Executive Officer
- Contact Person: Adrian Ang Chief Executive Officer
- Contact Email: adrian@aevice.com
- Date Prepared: 9 June 2023
| II. DEVICE | |
|----------------------|--------------------------------|
| Name of Device: | AeviceMD |
| Classification Name: | Medical Magnetic Tape Recorder |
| Regulation: | 21 CFR §870.2800 |
| Regulatory Class: | Class II |
| Product Code: | DSH; DQD |
## III. PREDICATE AND REFERENCE DEVICES
Predicate Manufacturer: Strados Labs, Inc. Predicate Trade Name: Strados Remote Electronic Stethoscope Platform (RESP) Predicate 510(k): K220893
Reference Device Manufacturer: Eko Devices, Inc. Reference Device Trade Name: Eko CORE Reference Device 510(k): K200776
{4}------------------------------------------------
## IV. DEVICE DESCRIPTION
The AeviceMD is designed as an electronic stethoscope to acquire and record lung sounds from users for healthcare professionals (HCP) to playback and interpret the sounds recorded. AeviceMD does not contain any alarm feature and it is not intended for emergency use. It is also not a sleep apnea device. The device is not intended for self-diagnosis.
The AeviceMD consists of hardware and embedded software. It is a five-part system that includes the following components:
1. AeviceMD Sensor – an embedded electronic wearable device that detects and records lung sounds and transmits data to an electronic gateway via Bluetooth.
2. AeviceMD Silicone Patch - silicone patch that houses and attaches the Sensor to the user's body (i.e., chest). This silicone patch undergoes biocompatibility testings which allow AeviceMD Sensor to be worn on the skin.
3. AeviceMD Docking Station - gateway device that serves as a computational hub and linkage from the Sensor to the Cloud Platform, and as a charqer for the Sensor.
4. AeviceMD App (for patients) / AeviceMD HCP Web App (for healthcare professionals in a clinical setting) - The AeviceMD App is a mobile app that downloads the post-processed information from the Cloud Platform and presents users with their recorded lung sounds at the auscultation locations which they can share with their HCP during their next consultation. A separate app, AeviceMD HCP Web App is tailored for HCP to review their patient's data in a clinical setting.
5. AeviceMD Cloud Platform – secure cloud server that receives data from gateway units and analyzes user's data using meaningful output information.
## V. INTENDED USE/ INDICATIONS FOR USE
The AeviceMD is a non-invasive battery-operated device, including a wearable component, intended to longitudinally acquire, record and store lung sounds from adult patients in a clinical or non-clinical setting. The device stores the data for later playback, review, and analysis by a clinician and comparison with earlier data from the same patient.
# VI. COMPARISON OF DEVICE CLASSIFICATION CHARACTERISTICS WITH PREDICATE AND REFERENCE DEVICES
The table below includes a comparison of the product code, regulation number, device classification name, intended user (including which data or measurements), and intended use environment among the subject device, the predicate device and the reference device.
{5}------------------------------------------------
-
| Parameter: | Subject Device:<br>AeviceMD | Predicate Device:<br>Strados RESP<br>(K220893) | Reference Device:<br>Eko CORE<br>(K200776) |
|------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | DSH; DQD | DSH | DQD |
| Regulation Number | 21 CFR 870.2800; | 21 CFR 870.2800 | 21 CFR 870.1875 |
| Device<br>Classification Name | Recorder, Magnetic<br>Tape, Medical;<br>Stethoscope,<br>Electronic | Recorder, Magnetic<br>Tape, Medical | Stethoscope, Electronic |
| Indications for Use | The AeviceMD is a<br>non-invasive<br>battery-operated<br>device, including a<br>wearable component,<br>intended to<br>longitudinally acquire,<br>record and store lung<br>sound from adult<br>patients in a clinical or<br>non-clinical setting.<br>The device stores the<br>data for<br>later playback, review,<br>and analysis by a<br>clinician and<br>comparison with<br>earlier data from the<br>same patient. | The Strados Remote<br>Electronic<br>Stethoscope<br>Platform (RESP) is a<br>non-invasive<br>battery-operated<br>device, including a<br>wearable<br>component, intended<br>to longitudinally<br>acquire, record, and<br>store lung sounds<br>from adult patients in<br>a healthcare or<br>outpatient setting<br>including transition<br>from healthcare<br>setting to outpatient<br>care without<br>interruption. The<br>device stores the<br>data for later<br>playback, review,<br>and analysis by a<br>clinician and<br>comparison with<br>earlier data from the<br>same patient. | The Eko CORE is an<br>electronic stethoscope<br>that enables<br>amplification, filtering,<br>and transmission of<br>auscultation sound data<br>(heart, lungs, bowel,<br>arteries, and veins),<br>whereby a clinician at<br>one location on network<br>can listen to the<br>auscultation sounds of<br>a patient on site or at a<br>different location on the<br>network. Eko CORE is<br>intended for use on<br>pediatric and adult<br>patients. The Eko<br>CORE is intended to be<br>used by professional<br>users in a clinical<br>environment or by lay<br>users in a nonclinical<br>environment. The<br>device is not intended<br>for self-diagnosis. |
| Intended User | Adults only | Adults only | Adults and pediatrics |
| Intended Use<br>Environment | Clinical and<br>Non-clinical Setting | Clinical and<br>Non-clinical Setting | Clinical and Non-clinical<br>Setting |
| User Interface | Mobile App<br>Web Portal (For HCP) | Mobile App<br>Web Portal (For<br>HCP) | Mobile App<br>Web Portal (Mentioned<br>on their website for |
| | | | HCP) |
| Record and<br>Playback Sounds | Yes | Yes | Yes |
| Data Transfer to<br>Compatible<br>Computing platform | Yes | Yes | Yes |
| Condition of Use | Reusable | Reusable | Reusable |
| Rx or OTC | Prescription Only | Prescription Only | Over-the-Counter |
| Wearable | Yes | Yes | No |
{6}------------------------------------------------
Table 1. Comparison of Device Classification Characteristics with Predicate and Reference Devices
#### Summary of Subject Device Comparison to predicate and reference device
The subject device as it longitudinally acquires sounds and allows recording and playback. The subject device identifies as an electronic stethoscope as the auscultation locations consist of the anterior body where a manual stethoscope would be placed during a consultation. These locations are supported by the cleared reference device.
#### Summary of Technological Characteristics
The AeviceMD has very similar technological characteristics compared to predicate and reference devices. All three devices have the same frequency range and can connect to mobile applications for recording and sharing data with HCP.
#### Non-Clinical Performance Data
The subject device was subject to non-clinical performance testing. There are no device-specific special control documents/ regulations that apply to the subject device. A list of the standards, guidance and additional testings for the device is listed below:
- . ISO 10993-5:2009 Biological evaluation of medical devices — Part 5: Tests for in vitro cvtotoxicity
- ISO 10993-10:2010 Biological evaluation of medical devices — Part 10: Tests for irritation and skin sensitization
- IEC 60601-1:2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007 +A1:2012 (or IEC . 60601-1: 2012 reprint)
- EN 60601-1-2:2015 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- . IEC 60601-1-11: 2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
{7}------------------------------------------------
- . IEC 60601-1-6:2010- Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- ANSI AAMI IEC 62366- 1:2015+AMD1:2020(Consolidated Text) Medical devices Part ● 1: Application of usability engineering to medical devices
- . ANSI AAMI IEC 62304:2006/A1:2016 Medical device software - Software life cycle processes [Including Amendment 1 (2016)] ISO 14971:2019 Medical devices -Applications of risk management to medical devices
- Additional testing: - Non-clinical Frequency Response Test
- Stethoscope Performance Test against a 510(k) cleared reference stethoscope
- Human Factors Usability
- Shipping Validation Test according to ASTM D4169-16
- Cleaning Validation Testing
# VII. CONCLUSIONS
The conclusions drawn from the non-clinical tests demonstrate that the proposed subject device performs as well as the leqally marketed predicate device and is substantially equivalent. The minor differences in the indications for use do not introduce a new intended use and do not raise any new issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.