The Strados Remote Electronic Stethoscope Platform (RESP) is a non-invasive battery-operated device, including a wearable component, intended to longitudinally acquire, record, and store lung sounds from adult patients in a healthcare or outpatient setting including transition from healthcare setting to outpatient care without interruption. The device stores the data for later playback, review, and analysis by a clinician and comparison with earlier data from the same patient.
Device Story
Non-invasive, battery-operated wearable device; records lung sounds (cough, wheeze, rhonchi, rales, crackles, bronchial sounds, quiet breathing, chest wall movement). Device adheres to patient chest wall; records to internal memory. Controlled via Bluetooth by Strados Mobile Application (SMA) on smartphone. Used in healthcare and outpatient/home settings; clinician configures settings; patient may place/replace device in outpatient setting. Recordings transmitted to SMA for clinician playback, review, and analysis. Enables longitudinal monitoring and comparison of lung sounds; assists clinicians in respiratory assessment.
Clinical Evidence
Clinical performance testing validated the quality of recorded lung sounds when self-placed and recorded by patients in an outpatient setting. Bench testing included frequency response, device cleaning, and usability testing per IEC 62366-1.
Technological Characteristics
Battery-operated wearable; Bluetooth connectivity; internal memory storage. Complies with IEC 60601-1, IEC 60601-1-11 (home healthcare), IEC 62366-1 (usability), IEC 62304 (software lifecycle), and UL 62368-1. Reusable device with disposable adhesive components.
Indications for Use
Indicated for adult patients requiring longitudinal acquisition, recording, and storage of lung sounds in healthcare or outpatient settings.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
April 27, 2022
Strados Labs, Inc. % Prithul Bom Most Resonsible Person Regulatory Technology Services, LLC 1000 Westgate Drive. Suite 510k Saint Paul, Minnesota 55114
Re: K220893
Trade/Device Name: Strados Remote Electronic Stethoscope Platform (RESP) Regulation Number: 21 CFR 870.2800 Regulation Name: Medical magnetic tape recorder Regulatory Class: Class II Product Code: DSH Dated: March 25, 2022 Received: March 28, 2022
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Brandon Blakely, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220893
Device Name
Strados Remote Electronic Stethoscope Platform (RESP)
Indications for Use (Describe)
The Strados Remote Electronic Stethoscope Platform (RESP) is a non-invasive battery-operated device, including a wearable component, intended to longitudinally acquire, record, and store lung sounds from adult patients in a healthcare or outpatient setting including transition from healthcare setting to outpatient care without interruption. The device stores the data for later playback, review, and analysis by a clinician and comparison with earlier data from the same patient.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"><b> X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;"><b> </b></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image contains the word "strados" in lowercase letters. The word is written in a simple, sans-serif font and is gray. Above the word is a graphic of several blue lines that are curved and parallel to each other.
## Section 5 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Strados RESP Traditional 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.
| Sponsor: | Strados Labs, Inc.<br>1315 Walnut Street, Suite 1101<br>Philadelphia, PA 19107<br>Tel: (888) STRADOS |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission Contact: | Grace Powers, MS, MBA, RAC<br>Founder/Principal Consultant<br>Powers Regulatory Consulting<br>grace@powersregulatory.com |
| Submission Date: | March 1, 2022 |
| Subject Device: | Trade Name: Strados Remote Electronic Stethoscope Platform (RESP)<br>Common Name: Medical magnetic tape recorder<br>Classification Name: Recorder, Magnetic Tape, Medical<br>Regulation: 21 CFR §870.2800<br>Regulatory Classification: Class 2<br>Product Code: DSH |
| Predicate Device: | Trade Name: Strados Remote Electronic Stethoscope Platform (RESP)<br>Common Name: Medical magnetic tape recorder<br>Classification Name: Recorder, Magnetic Tape, Medical<br>Regulation: 21 CFR §870.2800<br>Regulatory Classification: Class 2<br>Product Code: DSH<br>510(k): K201077 |
{4}------------------------------------------------
### Device Description
The Strados Remote Electronic Stethoscope Platform (RESP) is a non-invasive battery-operated device, including a wearable component, intended to longitudinally acquire, record, and store lung sounds from adult patients. The device stores the data for later playback, review, and analysis by a clinician and comparison with earlier data from the same patient. The Strados Remote Electronic Stethoscope Platform (RESP) is comprised of the Strados Wearable Device (SWD), Strados Charging Station (SCS) and external power supply, and Strados Patient Adhesive (SPA) used to adhere the SWD to the patient. The SWD is controlled by a mobile device running the Strados Mobile Application (SMA) via Bluetooth connection. The Strados Wearable Device (SWD) sits on the chest wall and passively records the patient's airway sounds, including but not limited to cough, wheeze, rhonchi, rales, crackles, coarse (bronchial) breath sounds, quiet breathing, and chest wall movement onto internal memory. The Strados Mobile App (SMA) on a smartphone allows playback of lung sounds from the wearable device in order for clinicians to listen to the patient's lung sounds. The device is used in a healthcare and outpatient setting including the home.
| Product Offering | Catalog Number | Contents |
|----------------------------------------------------------|----------------|--------------------------------------------------------------------------------------------------------------------------------|
| Strados Remote Electronic<br>Stethoscope Platform (RESP) | RESP1 | One system contains:<br>• Two (2) Strados Wearable Device<br>• Adhesives<br>• Charging Station with wall adaptor<br>• Labeling |
| Strados Wearable Device | SWD1 | Sold individually |
| Strados Patient Adhesives | SPA1 | Sold in a pack |
| Strados Charging Station | SCS1 | Sold individually. Contains Charging<br>Station and wall adapter |
| Strados Mobile Application | 540-00001 | Downloaded on a mobile device |
| Table 5-1: Strados Labs, Inc. Strados Remote Electronic Stethoscope Platform (RESP) |
|-------------------------------------------------------------------------------------|
| Product Offerings |
## Intended Use/Indications for Use
The Strados Remote Electronic Stethoscope Platform (RESP) is a non-invasive battery-operated device, including a wearable component, intended to longitudinally acquire, record, and store lung sounds from adult patients in a healthcare or outpatient setting including transition from healthcare setting to outpatient care without interruption. The device stores the data for later playback, review, and analysis by a clinician and comparison with earlier data from the same patient.
#### Technological Characteristics
An overview comparison of the Strados Remote Electronic Stethoscope Platform (RESP) (subject device) to the predicate device is presented in the table below.
{5}------------------------------------------------
| Table 5-2: Device Comparison | | |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device:<br>Strados RESP | Predicate Device:<br>Strados RESP K201077 |
| Name | Strados Remote Electronic Stethoscope Platform (RESP) | Strados Remote Electronic Stethoscope Platform (RESP) |
| Manufacturer | Strados Labs, Inc. | Strados Labs, Inc. |
| Product Code | DSH | DSH |
| Regulation Number | 21 CFR 870.2800 | 21 CFR 870.2800 |
| Device Classification Name | Recorder, Magnetic Tape, Medical | Recorder, Magnetic Tape, Medical |
| Device Classification | Class 2 | Class 2 |
| Indication for Use | The Strados Remote Electronic Stethoscope Platform (RESP) is a non-invasive battery-operated device, including a wearable component, intended to longitudinally acquire, record, and store lung sounds from adult patients in a healthcare or outpatient setting including transition from healthcare setting to outpatient care without interruption. The device stores the data for later playback, review, and analysis by a clinician and comparison with earlier data from the same patient. | The RESP is a non-invasive battery-operated device, including a wearable component, intended to longitudinally acquire, record, and store lung sounds from adult patients in a healthcare setting. The device stores the data for later playback, review, and analysis by a clinician on a mobile app. |
| Age of Device Use | Adults. | Adults. |
| Principle of Operation | <b>Placement</b> - The Strados RESP is adhered in an area of interest by a clinician to the skin (Healthcare setting). Device placement is determined by the clinician. The patient places and replaces the device on themselves (Outpatient setting).<br><b>Recording</b> - Lung sounds are recorded by wearable and stored in internal storage. A clinician configures the options.<br><b>Transmission</b> - Recordings are transmitted to the Strados Mobile App (SMA) from internal storage on the SWD.<br><b>Playback</b> - Recordings can be played back via the SMA or Clinical Portal. | <b>Placement</b> - The Strados RESP is adhered in an area of interest by a clinician to the skin.<br><b>Recording</b> - Lung sounds are recorded by wearable and stored in internal storage. A clinician configures the options.<br><b>Transmission</b> - Recordings are transmitted to the Strados Mobile App (SMA) from internal storage on the SWD.<br><b>Playback</b> - Recordings can be played back via the SMA or Clinical Portal. |
| Condition of Use | Reusable | Reusable |
| Rx or OTC | Prescription Only | Prescription Only |
| Wearable | Yes | Yes |
# Table 5-2: Device Comnarison
{6}------------------------------------------------
Section 5 510(k) Summary
### Non-Clinical Performance Data
The subject device was subject to non-clinical performance testing. There are no device-specific guidance documents, special controls document, and/or requirements in a device-specific regulation that is applicable to the subject device. A list of the standards, guidance and additional testing for all the devices is listed below:
- . IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1:2012- Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- . AAMI/IEC 60601-1:2005 + AMD 1:2012- Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-11:2015- Medical electrical equipment — Part 1-11: General requirements for basic safety and essential performance — Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- . IEC 62366-1:2015- Medical devices — Part 1: Application of usability engineering to medical devices
- IEC 62304:2006-Medical device software — Software life cycle processes
- . UL 62368-1:2012- Audio/video, information and communication technology equipment - Part 1: Safety requirements
IEC 60601-1-6:2010- Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- Additional Non-Clinical performance testing conducted for this submission includes:
- O Frequency response equivalence testing
- Device Cleaning Testing O
- Usability Testing O
#### Clinical Performance Data
Clinical performance testing was performed to validate the quality of the subject device's recorded lung sounds when self-placed and recorded in an outpatient setting.
#### Conclusion
The Strados Remote Electronic Stethoscope Platform (RESP) is substantially equivalent to the legally marketed predicate device as demonstrated by the similar indication for use, similar technologies and performance data, and does not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.