EchoTip Ultra Endoscopic Ultrasound Needle, EchoTip ProCore HD Ultrasound Biopsy Needle, EchoTip Ultra Endobronchial High Definition Ultrasound Needle, EchoTip ProCore Endobronchial High Definition Ultrasound Biopsy Needle
Applicant
Cook Ireland, Ltd.
Product Code
FCG · Gastroenterology, Urology
Decision Date
May 20, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The EchoTip® Ultra Endoscopic Ultrasound Needle is used with an ultrasound endoscope for fine needle aspiration (FNA) of submucosal lesions, mediastinal masses, lymph nodes and intraperitoneal masses within or adjacent to the gastrointestinal tract. The EchoTip ProCore® HD Ultrasound Biopsy Needle is used with an ultrasound endoscope for fine needle biopsy (FNB), of submucosal lesions, mediastinal masses, lymph nodes and intraperitoneal masses within or adjacent to the gastrointestinal tract. Echotip® Ultra Endobronchial High Definition Ultrasound Needle for Olympus scopes: This device is used to sample targeted submucosal and extramural lesions within or adjacent to the tracheobronchial tree or gastrointestinal tract through the accessory channel of an ultrasound endoscope for Fine Needle Aspiration (FNA). Echotip® Ultra Endobronchial High Definition Ultrasound Needle for Pentax scopes: This device is used to sample targeted submucosal and extramural lesions within or adjacent to the tracheobronchial tree through the accessory channel of an ultrasound endoscope for Fine Needle Aspiration (FNA). Echotip Procore® Endobronchial High Definition Ultrasound Biopsy Needle for Olympus scopes: This device is used with an ultrasound endoscope for fine needle biopsy, (FNB), of submucosal and extramural lesions within or adjacent to the tracheobronchial tree or gastrointestinal tract. Echotip Procore® Endobronchial High Definition Ultrasound Biopsy Needle for Pentax scopes: This device is used with an ultrasound endoscope for fine needle biopsy, (FNB), of submucosal and extramural lesions within or adjacent to the tracheobronchial tree.
Device Story
Endoscopic/endobronchial ultrasound needles; used for tissue sampling (FNA/FNB). Input: physical tissue samples via needle puncture. Operation: stainless steel cannula with dimpled tip for ultrasound visualization; nitinol stylet; handle for length adjustment; syringe for aspiration. Used in clinical settings by physicians during endoscopic procedures. Output: tissue specimen for diagnostic analysis. Benefit: enables minimally invasive biopsy of deep-seated lesions/masses.
Clinical Evidence
Bench testing only. No clinical data provided. Performance validated via tensile, crumple, joint strength testing, finite element analysis, and simulated use/drop testing.
Technological Characteristics
Stainless steel cannula with dimpling pattern; nitinol stylet; sterile, single-use. Dimensions: 19, 20, 22, 25 Ga. Sterilization: Ethylene oxide (EO). Manual operation via handle for needle/sheath adjustment. No software or electronic components.
Indications for Use
Indicated for patients requiring fine needle aspiration (FNA) or fine needle biopsy (FNB) of submucosal, extramural, or intraperitoneal lesions, mediastinal masses, or lymph nodes within or adjacent to the gastrointestinal tract or tracheobronchial tree.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
EchoTip ProCore HD Ultrasound Biopsy Needle (K142688)
EchoTip Ultra / ProCore Endobronchial High Definition Ultrasound Needle (K160229)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 20, 2021
Cook Ireland Ltd. Jane Kennedy Senior Regulatory Affairs Specialist O'Halloran Road. National Technology Park Limerick, Ireland
Re: K210476
Trade/Device Name: EchoTip Ultra Endoscopic Ultrasound Needle, EchoTip ProCore HD Ultrasound Biopsy Needle, EchoTip Ultra Endobronchial High Definition Ultrasound Needle, EchoTip ProCore Endobronchial High Definition Ultrasound Biopsy Needle Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: FCG Dated: February 12, 2021 Received: February 19, 2021
Dear Jane Kennedy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210476
Device Name
Echotip® Ultra Endoscopic Ultrasound Needle, Echotip Procore® HD Ultrasound Biopsy Needle Echotip® Ultra Endobronchial High Definition Ultrasound Needle Echotip Procore® Endobronchial High Definition Ultrasound Biopsy Needle
### Indications for Use (Describe)
The EchoTip® Ultra Endoscopic Ultrasound Needle is used with an ultrasound endoscope for fine needle aspiration (FNA) of submucosal lesions, mediastinal masses, lymph nodes and intraperitoneal masses within or adjacent to the gastrointestinal tract.
The EchoTip ProCore® HD Ultrasound Biopsy Needle is used with an ultrasound endoscope for fine needle biopsy (FNB), of submucosal lesions, mediastinal masses, lymph nodes and intraperitoneal masses within or adjacent to the gastrointestinal tract.
Echotip® Ultra Endobronchial High Definition Ultrasound Needle for Olympus scopes: This device is used to sample targeted submucosal and extramural lesions within or adjacent to the tracheobronchial tree or gastrointestinal tract through the accessory channel of an ultrasound endoscope for Fine Needle Aspiration (FNA).
Echotip® Ultra Endobronchial High Definition Ultrasound Needle for Pentax scopes: This device is used to sample targeted submucosal and extramural lesions within or adjacent to the tracheobronchial tree through the accessory channel of an ultrasound endoscope for Fine Needle Aspiration (FNA).
Echotip Procore® Endobronchial High Definition Ultrasound Biopsy Needle for Olympus scopes: This device is used with an ultrasound endoscope for fine needle biopsy, (FNB), of submucosal and extramural lesions within or adjacent to the tracheobronchial tree or gastrointestinal tract.
Echotip Procore® Endobronchial High Definition Ultrasound Biopsy Needle for Pentax scopes: This device is used with an ultrasound endoscope for fine needle biopsy, (FNB), of submucosal and extramural lesions within or adjacent to the tracheobronchial tree.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
### I. SUBMITTER
Cook Ireland Ltd. O' Halloran Road National Technology Park Limerick, V94 N8X2 Ireland
Phone: +353 61 334440 Fax: +353 61 334441
Contact: Jane Kennedy, Senior Regulatory Affairs Specialist
Phone: +353 61 334440 Fax: +353 61 334441
Date Prepared: February 12th, 2021
{4}------------------------------------------------
### II. DEVICE
Trade Name of Device: EchoTip Ultra Endoscopic Ultrasound Needle
The model numbers are ECHO-19, ECHO-1-22, and ECHO-25.
Common or Usual Name: Endoscopic Ultrasound Needle (Biopsy Needle Kit)
Classification Name: Gastroenterology/Urology (21 CFR 876.1075)
Regulatory Class: II
Product Code: FCG
Trade Name of Device: EchoTip ProCore HD Ultrasound Biopsy Needle The model numbers are ECHO-HD-19-C. ECHO-HD-3-20-C. ECHO-HD-22-C. and ECHO-HD-25-C. Common or Usual Name: Endoscopic Ultrasound Needle (Biopsy Needle Kit) Classification Name: Gastroenterology/Urology (21 CFR 876.1075) Regulatory Class: II Product Code: FCG
Trade Name of Device: EchoTip Ultra Endobronchial High Definition Ultrasound Needle The model numbers are ECHO-HD-22-EBUS-O, ECHO-HD-22-EBUS-P, ECHO-HD-25-EBUS-O. and ECHO-HD-25-EBUS-P. (O = for use with Olympus scope, P = for use with Pentax scope) Common or Usual Name: Endoscopic Ultrasound Needle (Biopsy Needle Kit) Classification Name: Gastroenterology/Urology (21 CFR 876.1075) Regulatory Class: II
Product Code: FCG
Trade Name of Device: EchoTip ProCore Endobronchial High Definition Ultrasound Biopsy Needle
The model numbers are ECHO-HD-22-EBUS-O-C. ECHO-HD-22-EBUS-P-C, ECHO-HD-25-EBUS-O-C, and ECHO-HD-25-EBUS-P-C.
(O = for use with Olympus scope, P = for use with Pentax scope)
Common or Usual Name: Endoscopic Ultrasound Needle (Biopsy Needle Kit)
Classification Name: Gastroenterology/Urology (21 CFR 876.1075)
Regulatory Class: II
Product Code: FCG
{5}------------------------------------------------
#### III. PREDICATE DEVICE
Endoscopic Ultrasound Needle, K083330 cleared on February 6th, 2009. EchoTip ProCore HD Ultrasound Biopsy Needle, K142688 cleared on December 194, 2014. EchoTip Ultra / ProCore Endobronchial High Definition Ultrasound Needle, K160229 cleared on March 21st, 2016.
The predicate devices detailed above have never been subject to a design related recall.
#### IV. DEVICE DESCRIPTION
### EchoTip Ultra Endoscopic Ultrasound Needle
This Needle is used in conjunction with an ultrasound endoscope and is available in needle gauge sizes of 19, 22 and 25 Ga. The device comprises of a needle assembly and a syringe. The needle assembly consists of the needle cannula, stylet, sheath and handle. The stainless steel needle cannula has a dimpling pattern on the distal end to allow visualization of the needle tip under endoscopic ultrasound. The purpose of the needle cannula is for puncturing of the target site. The preloaded nitinol stylet is withdrawn from the needle when obtaining sample from the target site. The sheath covers the needle when it is not in use. The device handle allows for needle and sheath length adjustment. The syringe can aid in sample aspiration. The device is supplied sterile and is intended for single use. The device is for Rx use only.
### EchoTip ProCore HD Ultrasound Biopsy Needle
This Needle is used in conjunction with an ultrasound endoscope and is available in needle gauge sizes of 19, 20, 22 and 25 ga. The device comprises of a needle assembly and a syringe. The needle assembly consists of the needle cannula, stylet, sheath and handle. The stainless steel needle cannula has a dimpling pattern on the distal end to allow visualization of the needle tip under endoscopic ultrasound. The needle cannula has a bevelled tip design and a notch. The purpose of the needle cannula is for puncturing/biopsy of the target site. The preloaded nitinol stylet is withdrawn from the needle for biopsy. Some variants have a stylet that coils when not in the needle. The sheath covers the needle when it is not in use. The device handle allows for needle and sheath length adjustment. The syringe can aid in tissue aspiration. The device is supplied sterile and is intended for single use. The device is for Rx use only.
{6}------------------------------------------------
## EchoTip Ultra / ProCore Endobronchial High Definition Ultrasound Needle
These Needles are used in conjunction with an endobronchial ultrasound endoscope and is available with needle gauge sizes of 22 and 25 Ga. The device is composed of a needle assembly and a syringe. An adapter can also be supplied for use with endobronchial ultrasound endoscopes with metal non-Luer hubs. The needle assembly consists of the needle cannula, stylet, sheath and handle. The stainless steel needle cannula has a dimpling pattern on the distal end to allow visualization of the needle tip under endoscopic ultrasound. The needle cannula has a bevelled tip design and comes either with or without a notch. The purpose of the needle cannula is for puncturing/sampling of the target site. The needle is provided with a preloaded stylet which remains in place during advancement of the needle. The sheath covers the needle when the needle is retracted and not in use. The device handle allows for needle and sheath length adjustment. The stylet/syringe can aid in specimen retrieval. The device is supplied sterile and intended for single use only. The device is for Rx use only.
| Subject Device | Predicate Device |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The EchoTip Ultra Endoscopic Ultrasound<br>Needle is used with an ultrasound endoscope for<br>fine needle aspiration (FNA) of submucosal<br>lesions, mediastinal masses, lymph nodes and<br>intraperitoneal masses within or adjacent to the<br>gastrointestinal tract. | EchoTip Ultra Ultrasound Needle<br>(K083330): This device is intended to be<br>used with an ultrasound endoscope for<br>delivery of injectable materials into<br>tissues during endoscopic procedures<br>and fine needle aspiration (FNA) of<br>submucosal lesions, mediastinal masses,<br>lymph nodes and intraperitoneal masses<br>within or adjacent to the gastrointestinal<br>tract.<br>There were no new risks introduced as<br>the intended use of the subject device<br>falls within the broader intended use of<br>the predicate device cleared under<br>K083330. |
| The EchoTip ProCore HD Ultrasound Biopsy<br>Needle is used with an ultrasound endoscope for<br>fine needle biopsy (FNB), of submucosal lesions,<br>mediastinal masses, lymph nodes and<br>intraperitoneal masses within or adjacent to the<br>gastrointestinal tract. | Unchanged from EchoTip ProCore<br>HD Ultrasound Biopsy Needle,<br>K142688. |
#### INDICATIONS FOR USE V.
{7}------------------------------------------------
| Subject Device | Predicate Device |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Echotip® Ultra Endobronchial High<br>Definition Ultrasound Needle for Olympus<br>scopes: This device is used to sample targeted<br>submucosal and extramural lesions within or<br>adjacent to the tracheobronchial tree or<br>gastrointestinal tract through the accessory<br>channel of an ultrasound endoscope for Fine<br>Needle Aspiration (FNA). | Unchanged from EchoTip Ultra<br>Endobronchial High Definition<br>Ultrasound Needle, K160229. |
| Echotip® Ultra Endobronchial High<br>Definition Ultrasound Needle for Pentax<br>scopes: This device is used to sample targeted<br>submucosal and extramural lesions within or<br>adjacent to the tracheobronchial tree through the<br>accessory channel of an ultrasound endoscope for<br>Fine Needle Aspiration (FNA). | Unchanged from EchoTip Ultra<br>Endobronchial High Definition<br>Ultrasound Needle, K160229. |
| Echotip Procore® Endobronchial High<br>Definition Ultrasound Biopsy Needle for<br>Olympus scopes: This device is used with an<br>ultrasound endoscope for fine needle biopsy,<br>(FNB), of submucosal and extramural lesions<br>within or adjacent to the tracheobronchial tree or<br>gastrointestinal tract. | Unchanged from EchoTip ProCore<br>Endobronchial High Definition<br>Ultrasound Needle, K160229. |
| Echotip Procore® Endobronchial High<br>Definition Ultrasound Biopsy Needle for<br>Pentax scopes: This device is used with an<br>ultrasound endoscope for fine needle biopsy,<br>(FNB), of submucosal and extramural lesions<br>within or adjacent to the tracheobronchial tree. | Unchanged from EchoTip ProCore<br>Endobronchial High Definition<br>Ultrasound Needle, K160229. |
{8}------------------------------------------------
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH A PREDICATE DEVICE
The modified devices are substantially equivalent to the currently marketed predicate devices, the Endoscopic Ultrasound Needle, K083330 cleared on February 6th, 2009, the EchoTip ProCore HD Ultrasound Biopsy Needle, K142688 cleared on December 19th, 2014 and the EchoTip Ultra / ProCore Endobronchial High Definition Ultrasound Needle, K160229 cleared on March 21st, 2016.
In brief, the modified devices are identical to their respective predicate devices with regard to the following:
- | Needle gauge size
- । Needle material
- | Needle length extension range
- Stylet wire material —
- Sheath length extension range
- -Sheath material
- | Handle (method of needle and sheath adjustment)
- Endoscope compatibility |
- Syringe —
- -Shelf life
- -Sterility (Ethylene oxide, EO)
- | For single use
- For professional use
- Principle of operation —
The following technological differences exist between the modified devices and the currently marketed predicate devices:
- । Dimensional changes to inner components of the handle.
- EchoTip Ultra / ProCore Endobronchial High Definition Ultrasound Needle: । Dimensional change to the Female Luer Lock Adaptor (FLLA) connector on the Olympus Adaptor.
- । EchoTip Ultra Endoscopic Ultrasound Needle: The stylet distal tip for the 22 Ga needle is changed from a bevelled tip to a ball tip and the stylet wire for the 19 Ga and 22 Ga needles are slightly reduced.
These differences in technological characteristics do not raise different questions of safety and effectiveness.
{9}------------------------------------------------
#### VII. PERFORMANCE DATA
A biocompatibility evaluation for the proposed modifications was conducted in accordance with the FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process" issued on September 4, 2020 and the International Standard ISO 10993-1: 2018 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," as recognised by FDA.
The device specific guidance document was consulted in preparing this premarket submission, Guidance for the Content of Premarket Notifications for Biopsy Devices Used in Gastroenterology and Urology.
Performance testing was performed as per Cook Ireland's design control system; the following tests were conducted:
- Tensile testing -
- -Crumple testing
- -Joint strength testing
- Finite element analysis -
- Simulated use and drop testing -
#### VIII. CONCLUSIONS
The non-clinical data supports the safety of the modified devices and demonstrates that the EchoTip Ultra Endoscopic Ultrasound Needle, the EchoTip ProCore HD Ultrasound Biopsy Needle and the EchoTip Ultra / ProCore Endobronchial High Definition Ultrasound Needle is safe and effective and should perform as intended in the specified use conditions. This nonclinical data supports the substantial equivalence of the EchoTip Ultra Endoscopic Ultrasound Needle, the EchoTip ProCore HD Ultrasound Biopsy Needle and the EchoTip Ultra / ProCore Endobronchial High Definition Ultrasound Needle to their respective predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.