This device is used with an ultrasound endoscope for fine needle biopsy, (FNB), of submucosal lesions, mediastinal masses, lymph nodes and intraperitoneal masses within or adjacent to the gastrointestinal tract.
Device Story
EchoTip Procore HD Ultrasound Biopsy Needle is a manual surgical instrument used with ultrasound endoscopes for tissue sampling. Device consists of a stainless steel needle cannula with a bevelled tip and notch, a nitinol stylet, a sheath, and a handle. Operated by clinicians in clinical settings to perform fine needle biopsies of GI-adjacent masses. The needle tip features a dimpling pattern for enhanced visualization under ultrasound guidance. The clinician inserts the needle into the target site; the stylet is withdrawn to facilitate tissue collection, optionally aided by an attached syringe for aspiration. The handle allows for adjustment of needle and sheath length. The device is supplied sterile for single use. By enabling precise tissue acquisition under ultrasound visualization, the device assists in the diagnosis of lesions and masses, supporting clinical decision-making regarding patient pathology.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (cytotoxicity, sensitization, intracutaneous reactivity), visual and dimensional inspection, simulated use, stylet removal force, and joint strength testing.
Technological Characteristics
Stainless steel needle cannula with dimpled tip and notch; nitinol stylet; handle for length adjustment. Single-use, sterile (EO). Manual operation. No electronic components or software.
Indications for Use
Indicated for fine needle biopsy (FNB) of submucosal lesions, mediastinal masses, lymph nodes, and intraperitoneal masses within or adjacent to the gastrointestinal tract using an ultrasound endoscope.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 19, 2014
Cook Ireland Ltd Nora O'Connor Regulatory Affairs Specialist O'halloran Road National Technology Park Limerick Ireland
Re: K142688
> Trade/Device Name: Echotip Procore® Hd Ultrasound Biopsy Needle Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: FCG Dated: September 18, 2014 Received: September 22, 2014
Dear Nora O'Connor,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Cook Ireland EchoTip Procore® HD Ultrasound Biopsy Needle
510(k) Number (if known):
Device Name: EchoTip Procore® HD Ultrasound Biopsy Needle
Indications for Use:
This device is used with an ultrasound endoscope for fine needle biopsy, (FNB), of submucosal lesions, mediastinal masses, lymph nodes and intraperitoneal masses within or adjacent to the gastrointestinal tract.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | <div style="text-align:center;">✓</div> | AND/OR |
|-------------------------------------------------|-----------------------------------------|--------|
|-------------------------------------------------|-----------------------------------------|--------|
Over-The-Counter Use ____________ (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1_of 1
{3}------------------------------------------------
## 510(k) Summary
#### I. SUBMITTER
Cook Ireland Ltd. O'Halloran Road National Technology Park Limerick, Ireland
Phone: +353 61 334440 Fax: +353 61 239293
Contact Persons: Nora O'Connor, Regulatory Affairs Specialist Jacinta Kilmartin, Senior Regulatory Affairs Specialist
Phone: +353 61 334440 Fax: +353 61 239293
Date Prepared: October 31, 2014
#### II. DEVICE
Trade Name of Device: EchoTip Procore® HD Ultrasound Biopsy Needle. The model number is ECHO-HD-(3)-XX-C where XX denotes the needle gauge size.
Common or Usual Name: Endoscopic Ultrasound Needle (Biopsy Needle Kit)
Classification Name: Gastroenterology-urology biopsy instrument (21 CFR 876.1075)
Regulatory Class: II
Product Code: FCG
{4}------------------------------------------------
#### III. PREDICATE DEVICE
Cook Endoscopic Ultrasound Needle, K083330. Biopsy Sciences, LLC Maxi-Cell Biopsy Needle, K021847
#### IV. DEVICE DESCRIPTION
The EchoTip Procore® HD Ultrasound Biopsy Needle is used in conjunction with an ultrasound endoscope and is available with needle gauge sizes of 19, 20, 22 and 25 ga. The device comprises of a needle assembly and a syringe. The needle assembly consists of the needle cannula, stylet, sheath and handle. The stainless steel needle cannula has a dimpling pattern on the distal end to allow visualization of the needle tip under endoscopic ultrasound. The needle cannula has a bevelled tip design and a notch. The purpose of the needle cannula is for puncturing/biopsy of the target site. The preloaded nitinol stylet is withdrawn from the needle for biopsy. Some variants have a stylet that coils when not in the needle. The sheath covers the needle when it's not in use. The device handle allows for needle and sheath length adjustment. The syringe can aid in tissue aspiration. The device is supplied sterile and is intended for single use. The device is for Rx used only.
#### INDICATIONS FOR USE V.
This device is used with an ultrasound endoscope for fine needle biopsy, (FNB), of submucosal lesions, mediastinal masses, lymph nodes and intraperitoneal masses within or adjacent to the gastrointestinal tract.
### COMPARISON OF TECHNOLOGICAL CHARACETERISTICS WITH A VI. PREDICTE DEVICE
The modified device is substantially equivalent to the currently marketed device, the Cook Endoscopic Ultrasound Needle, as cleared by K083330. Both devices operate in the same manner to obtain a tissue biopsy using an ultrasound endoscope.
The modified device is substantially equivalent to the currently marketed device, Biopsy Sciences, LLC Maxi-Cell Biopsy Needle, as cleared by K021847. Both the modified and predicate device operate in a similar manner to obtain a sample.
{5}------------------------------------------------
At a high level, the modified device is identical to/ within the range of the Cook Endoscopic Ultrasound Needle predicate device with respect to the following:
- . Needle gauge range and material,
- . Needle length adjustment range and sheath extension adjustment range,
- Stylet diameter range and material,
- Stylet tip options, ●
- Compatibility with an ultrasound endoscope, ●
- . Use with a syringe,
- Mechanism/principle of operation (manual, sampling using cutting and aspiration),
- Method of needle and sheath adjustment (handle), .
- For single use, and
- Sterility (using EO).
The following technological differences exist between the modified device and the Cook Endoscopic Ultrasound Needle predicate device:
- . Needle cannula - addition of notch,
- Stylet shape upon removal (coiled), stylet distal tip location and stylet cap, ●
- . Sheath - material (addition of a covering, change in colour), outer diameter change,
- . Handle - colour change (non-patient contacting).
The modified device is identical to/ within the range of the Biopsy Sciences, LLC Maxi-Cell Biopsy Needle with respect to the following:
- Needle material , .
- Presence of a notch on the needle, .
- Used with a syringe, .
- Mechanism/principle of operation (manual, sampling using cutting and aspiration),
- For single use, and
- Sterility (using EO).
The EchoTip Procore® HD Ultrasound Biopsy Needle has the following differences to the features of the Biopsy Sciences, LLC Maxi-Cell biopsy needle:
- . Needle gauge range.
- Stylet material, shape upon removal, .
- The addition of a sheath, and ●
- The handle type.
{6}------------------------------------------------
#### VII. PERFORMANCE DATA
The biocompatibility evaluation for EchoTip Procore® HD Ultrasound Biopsy Needle was conducted in accordance with ISO 10993-1: 2009 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process" and FDA's biocompatibility guidance, G95-1 Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" (May 1, 1995). The following tests were completed: Cytotoxicity, Sensitization and Intracutaneous reactivity.
The device specific guidance document was consulted in preparing this premarket submission, Guidance for the Content of Premarket Notifications for Biopsy Devices Used in Gastroenterology and Urology. Performance testing such as visual inspection, dimensional inspection, simulated use, stylet removal force testing and joint strength testing were performed as per Cook Ireland's design control system.
#### CONCLUSIONS VIII.
Non-clinical testing carried out on the device supports the substantial equivalence of the EchoTip Procore® HD Ultrasound Biopsy Needle to the predicate devices and provides a reasonable assurance of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.